Pierre, Aritomi and Yuhang
We found that when we just connect the source port of network analyzer to servo, the error signal presents a very very large 50Hz noise.
When I connect it with the same power with rack, the 50Hz noise becomes even higher!
Another thing to be noticed is that when I use SR560 as servo, and inject noise from network analyzer to SR560, there is no 50Hz noise!
When I connect it with the same power with rack, the 50Hz noise becomes even higher!
Another thing to be noticed is that when I use SR560 as servo, and inject noise from network analyzer to SR560, there is no 50Hz noise!