[Aritomi, Yuhang]
First we made LO DC balanced and aligned LO and BAB into AMC. DC balance of BAB is 10mV. We measured shot noise of LO (Pic 1). Although there are large 50Hz peak and its harmonics , homodyne is shot noise limited above 20Hz. To remove the 50Hz peak, we floated ground of homodyne, but the shot noise was same. We also measured shot noise spectrum when LO and CC are injected into homodyne. Shot noise with LO and CC was same as shot noise with LO only.
Then we locked green and IR phase with SR560 (bandwidth ~ 100Hz) and found squeezing spectrum is very bad. So we locked green and IR phase with Pierre's servo (bandwidth ~ 1kHz) and finally could recover the 5dB squeezing (Pic 2).
We thought green power is 50mW, but actually green power was smaller than 50mW due to misalignment of GRMC at that time.