I installed the another AOM as attached picture as following procedure.
Current diffraction efficiency is about 80% and still there is a room for improvement.
In addition, I worked on the double-pass AOM alignment previously installed.
The power of double-passed beam is 5.76mW in contrast with 8.86mW input to AOM.
Hence the double-pass diffraction efficiency is about 65% which is reasonable value.
The next step is (quantitatively) measure the beam jitter with scanning frequency and minimize it.