The attached file shows the very preliminary results of the ringdown measurements.
As shown in the figure, the finesse at 134 K, and 165 K is smaller than that of 170 K (judged by my eyeballs).
This might imply that the molecular layer i.e., amorphous ice can desorb around between 165 K and 170 K.
It should be noted that the fitting has not yet done...
Tomorrow, I will measure the TF of PDH locking loop.
Then I will re-cooling the cavity.
At this moment, the target temperature is 10 K and monitor the behavior of finesse in order to estimate the loss induced by cryogenic molecular layer.
I think this result would provide some implications to the Einstein Telescope.