I started some analysis for finesse of the folded cavity. The finesse was obtained by fitting the transmitted signal.
The finesse at room temp. was about 1.678*1e4(+/- 11) though that at 8 K was 1.579*1e4(+/- 26).
Further analysis is ongoing...