KAGRA MIR (Absorption)
ManuelMarchio - 01:30, Friday 08 April 2016 (239)
Comment to Checking of optical chopper SR540 (Click here to view original report: 238)
I tightly fixed the chopper to the optical table;
I monitored the frequency value for about 1.5 hours;
I didn't change any parameter or move anything;
Just took some snapshot of the spectrum analyzer every few minutes.
The attached file shows each snapshot and a summary plot
The result is:
 
- For a short time (few minutes), the frequency fluctuations are within the specifications (2%), in agreement with previous Tatsumi-san's measurements.
 
- For a long period, the frequency fluctuations can be larger than 2%
 
 
I couldn't upload the file because it's a bit more than 10MB, so here is the link https://www.dropbox.com/s/ajzduvo0ingj5an/chopper%20long%20time%20stability.pdf?dl=0