Although the locking accuracy is 1.8Hz with CCFC, the FDS detuning fluctuation is ~10Hz even with CCFC/Z correction/pointing/AA. The possible reasons are parameter estimation error and IR misalignment change and FC macroscopic length change.
sqz_dB = 10.5; % produced SQZ (dB)
L_rt = 120e-6; % FC losses
L_inj = 0.31; % Injection losses
L_ro = 0.24; % Readout losses
A0 = 0.05; % Squeezed field/filter cavity mode mismatch
C0 = 0.05; % Squeezed field/local oscillator mode mismatch
ERR_L = 1e-12; % Lock accuracy (m)
ERR_csi = 30e-3; % Phase noise (rad)
Note: In this calculation, shot noise level is assumed to be -133.2 dBVrms/rtHz, but actually it may not be true. So the result in this log could be wrong.