R&D (FilterCavity)
MarcEisenmann - 17:31, Thursday 03 June 2021 (2553)
Oplev and AA SNR comparison

Today I sent 15 Hz line on input and end mirrors pitch and yaw to calibrate the AA signals.

The 4 attached figures show the comparison of Oplev signal with calibrated AA signals.

This line can also be used to estimate the SNR of both Oplev and AA signals.

  INPUT PITCH INPUT YAW END PITCH END YAW
Oplev 17 3 23 3
AA 47 11 54 6
Images attached to this report
2553_20210603103107_inputpitch.jpg 2553_20210603103112_inputyaw.jpg 2553_20210603103116_endpitch.jpg 2553_20210603103119_endyaw.jpg