Despite 2 windows updates that turned off the computer I finally finished the measurement of the absorption map of the spare viewport (see fig 1).
In order to recover the lost time I changed the step size to 0.5 mm.
The mean absorption is at the order of 18 ppm while there is some dust/surface defects creating few spikes in the absorption.