I measured the round trip losses with new beam spot which has more stable IR transmission. The round trip losses are 139 ppm.
P_res = 245; % BAB reflection power on resonance (uW)
P_in = 362; % BAB reflection power off resonance (uW)
R_gamma = P_res/P_in;
gamma = 0.05; % uncoupled power to the cavity: 5% from mode mismatch
R = (R_gamma-gamma)/(1-gamma);
T = 0.00136; % input mirror transmissivity
L = (T/2)*(1-R)/(1+R) = 139 ppm