This measurement was done on 20210622.
I measured the CCFC locking accuracy with the new beam spot (Fig 1). CCFC calibration amplitude is 166mVpp and CCFC filter is gain of 1000 and LPF of 30Hz. CC2 mass feedback gain is 2.7.
Compared with old beam spot, the locking accuracy without CCFC is smaller. The locking accuracy with CCFC is similar to before because it is limited by high frequency noise.
Then I measured FDS with CCFC (Fig 2). The squeezing degradation parameters are as follows. Here are some points of this result.
sqz_dB = 10.8; % generated squeezing
L_rt = 120e-6; % FC losses
L_inj = 0.32; % Injection losses
L_ro = 0.25; % Readout losses (propagation loss is 49%)
A0 = 0.06; % Squeezer/filter cavity mode mismatch
C0 = 0.02; % Squeezer/local oscillator mode mismatch
ERR_L = 1e-12; % Lock accuracy (m)
ERR_csi = 30e-3; % Phase noise (rad)