I checked data especially of the measurement with 45deg input polarization angle.
It seems there was some issue with the calibration of the DC signal (the DC with approximately 45 deg input polarization is higher by a factor 2 than the expected maximum value when we inject pure p polarization).
We will redo this measurement again today.
Again issue with the translation stage so we have to remove the mirror to reset the motors reading from Zaber..
I deactivated the windows update for 1 month
Marc, Michael
To facilitate the removal of the mirror, we used the knob controller of X direction to move the mirror.
After that, we found out that the Zaber reading got good so we started the measurement with 45 deg input polarization angle.