Michael and Yuhang
Today we found a new issue about new OPO component No.28. The issue is described in the attached PDF file. The issue is that a counter bore hole is made from a wrong side of no.28 component.
We also took photos about no.28 component with wrong and correct holes, as shown in the attached Fig. 1 and 2. (The one with correct hole actually has wrong thread as reported in elog2724)
For solution, we found that a longer screw seems to be able to solve the problem. We will look for some longer screws and clean them. Then try to use them for the new OPO.