I set IR beam power to 1.6mW, I checked vertical and horizontal AOI with the razor blades and got respectively -0.008 deg and -0.022 deg.
I tuned the QWP and HWP angle to minimize p polarization while injection s polarization.
I did the calibration and started measurement of #4 (was cleaned with first contact) with s polarization at the input.