I remeasured the transfer function whose resonant frequency is set around 0.5Hz.
I measured it from 2Hz to 0.1Hz.
The peak of resonant looks like 0.4Hz.
This time, peak of resonant is crushed by range of photo sensor.
When I measured it around resonant frequency, displacement is bigger than linear range of photo sensor.