Members [Manuel, Matteo L.]
I measured the divergence of the 1310nm laser. Since the beam profiler doesn't work over 1100nm, I used the DET10N-InGaAs Detector to measure the time-dependent intensity while the beam is cut by the chopper. On the oscilloscope, I measure the rising time from 10% to 90% of the gaussian integral and calculate the waist and the divergence. I repeated for several chopper positions along the beam axis.
divergence = 1.0 ± 0.1 mrad.
waist = 420±40 um