Measured the power of the probes with and without the surface calibration sample to check the transmission of the sample at different wavelengths
633nm: 2.87 mW -> 1.89 mW Transmission= 65%
1310nm: 24.8 mW -> 16.6 mW Transmission= 67%
----------------------------------------------------------------------------------------
To check linearity of the Si PD, I measured the HeNe power and changed it putting some filters while measuring the PD DC signal
2.87 mW -> 6.58V
2.44 mW -> 6.52V
1.45 mW -> DC=4.28V
1.22mW (2 filters) -> 3.4V
328uW -> DC=0.88V
Result: the PD is saturated, see plot