Last week we measured the bandwidth of cavity. By using this data, we also did the extrapolation and got the losses.
We considered all the losses come from the increase of end mirror transmissivity. Then we did like this:
Velocity | Bandwidth | Finesse | Losses | r2 |
200Hz/s | 119Hz | 4191 | 134 | 0.999931 |
400Hz/s | 114Hz | 4355 | 78 | 0.999959 |
80Hz/s | 115Hz | 4312 | 92 | 0.999952 |