Tama -size Sapphire Sample2. I plot together the data taken:
- with the original setup 2 years ago (red line);
- with the current setup, which should have all the same parameters as the original setup, according to the company instructions (blue line)
- at LMA (black dots)
It looks like we are back to the factor of ~2 we had at the beginning.
(note: the phase is not calibrated, so it is different in the 2 plots mainly because of the different position of the chopper)