NAOJ GW Elog Logbook 3.2

[Aritomi, Yuhang, Yaochin, Eleonora]
We measured IR error signal of filter cavity using BAB with TAMA PD which is same as CC1 PD in the reflection path from filter cavity. Pic. 1 shows IR error signal and IR transmission when filter cavity is scanned with AOM (setting is Pic. 2). Calculation method of calibration factor is same as entry 750.
Note that factor of 2 in calibration factor is because slope of error signal on resonance is as twice as slope of peak-peak of error signal.
Then we measured spectrum of IR error signal when IR is locked on resonance and calculated rms (Pic.3). IR locking accuracy is 4.2 Hz which corresponds to 4.5 pm of locking accuracy. Pic. 4 shows squeezing degradation with this locking accuracy. In addition to mode mismatch, locking accuracy is also dominating. We need IR locking.
Most of the RMS is accumualted below 10 Hz. We observed a correlation between the PDH error signal for IR (demodualtion of 15 Mhz sidebands) and the lock correction signal from RAMPEAUTO.
At the time of the measument we know the IR alignment was not good. This can be the cause of such behaviour. We should repeat this measurement in good alignment condition.