LOG-IN
Displaying reports 2041-2060 of 3201.Go to page Start 99 100 101 102 103 104 105 106 107 End
R&D (FilterCavity)
Print this report.
EleonoraPolini - 13:31, Monday 03 June 2019 (1382)Get code to link to this report
Green path FI installation

[Eleonora P, Yuhang]

We replaced the FI's stage along the green path, after a hole has been drilled on it as explained in entry #1367

In fig.1 the picture of the stage on the bench.

Images attached to this report
1382_20190603063032_20.jpeg
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 17:56, Saturday 01 June 2019 (1381)Get code to link to this report
Silicon mirror installed and beam profile measurement

I installed one silicon mirror, which was labeled "No.1", inside the cryostat and did rough alignment.
Then I measured beam profile for mode matching.

R&D (Cryogenic)
Print this report.
SatoshiTanioka - 18:45, Friday 31 May 2019 (1380)Get code to link to this report
Installation work

Today, I re-installed some optics for higher-order-modes.
I wanted to put a iris for the mark of alignment, but I could not find it...
From tomorrow, I'm going to install silicon cavity inside the cryostat.

R&D (Cryogenic)
Print this report.
SatoshiTanioka - 19:35, Thursday 30 May 2019 (1379)Get code to link to this report
Optical layout changed

I changed optical layout to achieve better alignment.
I will update the schematic figure of optical layout later...
Anyway, the alignment seems ok so far.
 

R&D (FilterCavity)
Print this report.
EleonoraPolini - 17:24, Thursday 30 May 2019 (1378)Get code to link to this report
Robustness of injection telescope considering the error on beam waist position

As shown in entry #1363, the error on the beam waist position is large ( around 8%).

I checked if the solution #3 of entry #1366 would be good also considering the border positions of beam waist in the error interval, running again the simulations (fig 1, 2, 3) and analyzing the robustness on Python.

Both focal lengths are 750 mm.

I summarize the results in the following table:

z0 (m) pos lens 1 (m) pos lens 2 (m) pos lens 1 at min mismatch (m) pos lens 2 at min mismatch (m) min mismatch (%)
0.089 0.420 0.673 0.424 0.673 0.9
0.081 0.420 0.673 0.4276 0.683 1.8
0.097 0.420 0.673 0.410 0.6755 1.6

where the positions of the lenses is referred to the PBS after the OPO.

I computed on Python the position of the two lenses corresponding to the condition of minimum total mismatch.

In fig. 4, 5 and 6 are shown the robustness of the telescope with z=0.081 (min position), 0.089 (real position), 0.097 (max position), that are still pretty good (less than 10% mismatch).

Images attached to this report
1378_20190530101050_081.png 1378_20190530101100_096.png 1378_20190530101109_097.png 1378_20190531093840_injmin081.png 1378_20190531093851_injmin089.png 1378_20190531093902_injmin097.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 00:12, Thursday 30 May 2019 (1377)Get code to link to this report
Need to have better PD for CC1 locking if we want to reduce CC laser power and phase noise

[Yuhang, Chien-Ming]

Today we tried to decrease CC laser power. But when we just decreased it to half, we couldn't lock the CC1 loop. 

The cc error signals that we can use to lock and cannot be used to lock are attached in Fig1.

The RMS dark noise for CC1 and CC2 PD is measured and attached in Fig2. We can see that the CC1 PD has RMS dark noise almost 5 times higher than CC2 PD. 

We also suspected that this noisy PD for CC1 locking brings also phase noise.

Images attached to this report
1377_20190529171218_cc1.png 1377_20190529171227_cc2.png
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 23:13, Wednesday 29 May 2019 (1376)Get code to link to this report
Operation check of FI for TEM00 and some modifications

I installed a FI for TEM00 PDH locking, and confirmed that one can pick off reflected beam with simple setup (QWP and mirror).
The attached picture shows the reflected beam picked up by FI.

Then, I continued the installation of optics.
However, I found that the height of FI was not good, which is not the one I mentioned but already installed one before.
So I modified the alignment.
Actually, the beam height is still higher than the target height, though I cannot adjust anymore.

So I am planning to install STMs in front of the EOM, which can adjust the beam height.

Images attached to this report
1376_20190529160637_fi20190529.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 22:44, Wednesday 29 May 2019 (1375)Get code to link to this report
Measurement of 5.5dB SQZ down to 50Hz after homodyne improvement

[Yuhang, Chien-Ming, EleonoraP]

After the improvement of homodyne suggested by Chien-Ming, we measured squeezing again.

This time, we could see only two peaks for homodyne shot noise limited down to 10Hz when LO incident. The 34Hz peak is related to some unknown device(maybe air-conditioner). The characterization of this 34Hz peak is done by Federico and Irene. While another peak is a 50Hz power line. (We didn't perform a low-frequency high-resolution measurement, we may see some peaks at low frequency in that case. We will do it soon.)

For SQZ, we could see the noise floor(5.5dB SQZ) down to 10Hz with several narrow peaks. But notice that now there is still 2.4mV DC signal(for SQZ path) not balanced on homodyne. We could possibly improve CMRR for SQZ path soon.

For anti-SQZ, we could see 13.25dB. And it is also very flat.

The result is shown in the attached figure. We didn't pay attention to the measurement bandwidth of anti-SQZ and LO-shot noise. But we may do it again after the refinement of BAB balance.

Images attached to this report
1375_20190529154545_shotnoise.png 1375_20190529154552_5dbsqz.png
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 16:26, Wednesday 29 May 2019 (1372)Get code to link to this report
INPUT MIRROR cotrolled with new DGS

The local controls for the input mirror have been implemented in the new DGS.

As usual,  we amplified and filtered the PSD signals for pitch and yaw with a SR560 (2nd order lowpass, cutoff 100 Hz, gain 100)

TF for yaw and pitch are reported in pic 1 and 2. (injected noise: 5000 and 7000 counts respectively)

As for PR and BS the pitch TF and coherence is not very good.

Comparison between open and closed loop spectra are shown in pic 3 (yaw) and 4 (pitch)

I attach the filters values (model and damp for both yaw and pitch):

Fig 5  yaw model

Fig 6  yaw damp

Fig 7 pitch model

Fig 8 pitch damp

I will compute the calibration soon.

Images attached to this report
1372_20190529092004_inputyawtf209519.png 1372_20190529092010_inputpitchtf290519.png 1372_20190529092115_inputyawspe290519.png 1372_20190529092120_inputpitchspe290519.png 1372_20190529092139_inputyawmodel.png 1372_20190529092147_inputyawdamp.png 1372_20190529092152_inputpitchmodel.png 1372_20190529092355_inputpitchdamp.png
R&D (FilterCavity)
Print this report.
ChienMingWu - 22:27, Tuesday 28 May 2019 (1370)Get code to link to this report
Homodyne DC balance

Chien-Ming, Yu-Hang, and Aritomi

Yesterday we adjusted the BS horizontal angle to make the LO beam reach the DC balance at homodyne detector. We measured the CMRR, it is better than 80 dB at 1 kHz.

However, in this case, if we let that BAB and LO have a good overlapping, we need to change the optical set up of BAB a lot.

So today we keep the BAB beam in its original condition and adjust the BS angle to make the BAB beam reach the DC balance at homodyne. Then we adjust the LO beam to achieve a good overlapping with BAB beam. In this case, when the BAB is DC balanced, the homodyne DC offset of the LO shows -10mV.

We repeatedly adjust the BS angle and the alignment of both beams a little bit, finally, we make the LO beam become DC balanced and the DC Offset of BAB is 2.4 mV.

R&D (FilterCavity)
Print this report.
NaokiAritomi - 19:27, Monday 27 May 2019 (1369)Get code to link to this report
Alignment method of homodyne
 
[Aritomi, Chien-Ming, Yuhang]
 
Our BS for homodyne (BSW41-1064) is R = 49.14513016% (p-pol) and T= 50.49834442% (p-pol) when the injection angle is 45 deg. That means we have 1.35% unbalance when the injection angle is 45 deg. So far we made the injection angle 45 deg and actually we had 74 mV homodyne DC offset even if the homodyne is aligned. Each PD has 5V so the unbalance is 1.48%. To reduce this unbalance, the injection angle should be below 45 deg.
 
The alignment method of homodyne will be as follows. The rough sketch is also attached.
 
1. inject BAB to homodyne PD along the hole line of optical bench
2. adjust the angle of BS so that homodyne DC of BAB is 0
3. align BAB to AMC with two  mirrors after homodyne BS
4. align LO to AMC with two mirrors before homodyne BS
 
Our BS for homodyne (BSW41-1064) is R = 49.14513016% (ppol) and T= 50.49834442% (ppol).
That means we have 1.35% unbalance when the injection angle is 45 deg. So far we made the injection angle 45 deg and actually we had 70 mV homodyne DC offset even if homodyne is aligned. Each PD has 5V so the unbalance is 1.4%. We have to make the injection angle below 45 deg to get 1:1 beam splitting ratio.
Our BS for homodyne (BSW41-1064) is R = 49.14513016% (ppol) and T= 50.49834442% (ppol).
That means we have 1.35% unbalance when the injection angle is 45 deg. So far we made the injection angle 45 deg and actually we had 70 mV homodyne DC offset even if homodyne is aligned. Each PD has 5V so the unbalance is 1.4%. We have to make the injection angle below 45 deg to get 1:1 beam splitting ratio.
Images attached to this report
1369_20190527122659_img7475.jpg
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 19:21, Monday 27 May 2019 (1368)Get code to link to this report
Cryostat 4K shield

Takahashi-san, Sato-san, Tanioka

We removed the 4K shield of cryostat in ATC in order to ship it to KEK.
Some insulator are detached from 4K shield to remove screws.

Images attached to this report
1368_20190527122128_shield4k.jpg
R&D (FilterCavity)
Print this report.
EleonoraPolini - 17:51, Monday 27 May 2019 (1367)Get code to link to this report
Green path pre-alignment and FI stage pre-installation

[Yuhang and EleonoraP]

First we aligned the green beam before PR chamber.

Then we changed the stage of Faraday Isolator with a smaller one, in order to make space for the wind shields. We put some schotch to cover the borders of the FI, to avoid damages (fig 1).

We found a problem: not enough space for the mirror under the FI (fig 2). The black plastic edge of the mirror mount is 3.4mm while the beam reflected by the FI crystal is 5 mm far from the stage edge (fig 3).

Possible solutions:

1. make a hole in the mount in order to make space for the mirror. Size of the hole: 30 mm (diameter) and 12 mm (depth) in fig 4.

2. build a different stage with other components, in fig 5.

Images attached to this report
1367_20190527104206_img20190527wa0003.jpg 1367_20190527104251_img20190527wa0001.jpg 1367_20190527104922_fcgreenrefclipping.png 1367_20190527105015_36pm.jpeg 1367_20190527105025_img20190527172411.jpg
R&D (FilterCavity)
Print this report.
EleonoraPolini - 11:55, Tuesday 21 May 2019 (1366)Get code to link to this report
Injection telescope with lenses we have in the lab

I found two good options using the lenses we already have in the lab.

Version 2:

1000 mm @ z = 0.517 m (distance from PBS)

750 mm @ z = 0.618 m

Version 3:

750 mm @ z = 0.42 m

750 mm @ z = 0.673 m

As we can see in fig 2 and fig 4, the robustness is not as good as the one of entry #1365 (fig 5), but we are still under 10% of mismatch.

In fig 6 there is the scheme on the bench of version 2 (PURPLE) and version 3 (BLUE).

Images attached to this report
1366_20190521045408_injv2.png 1366_20190521045435_injv2.png 1366_20190521045503_injv3.png 1366_20190521045519_injv3.png 1366_20190521045532_injv1.png 1366_20190615124554_versione2e3new.png
R&D (FilterCavity)
Print this report.
EleonoraPolini - 00:21, Tuesday 21 May 2019 (1365)Get code to link to this report
IR injection telescope simulation after improvement of homodyne matching

After the improvement of homodyne matching entry #1354, the parameters of the beam entering the injecion telescope have changed (results of measurement in entry #1363).

Parameters changes: 

w0

126um --> 94um

z0 (with respect to PBS)

15 cm --> 8.9 cm

Focal lenghts needed for the new injection telescope are:

600 mm (instead of -1000mm)

750 mm (instead of 500mm)

Robustness is still really good (fig 2).

Images attached to this report
1365_20190520171916_injection.png 1365_20190520171922_injnew.png 1365_20190520172030_schemenew.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 14:03, Saturday 18 May 2019 (1363)Get code to link to this report
Fit of beam going to filter cavity after the improvement of homodyne matching

As I mentioned in the entry, I moved the first lens after OPO. So the beam parameter going to filter cavity is changed. To have a better design for the telescope matching OPO transmission to filter cavity. I made the beam measurement again.

The way that I did measurement is illustrated in the attached figure 1. 

By measuring seven points along a long distance, I fit the beam. The result is shown in the attached figure 2.

The beam waist is even smaller than the last measurement.

Images attached to this report
1363_20190518070251_scheme.png 1363_20190518185646_opotra2.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 07:46, Saturday 18 May 2019 (1362)Get code to link to this report
Comment to Install the new Faraday Isolator (Click here to view original report: 1360)

I also matched BAB into OPO. The spectrum is shown in the attached figure.

The matching of main laser optical fiber is also changed after the cleaning of some mirrors.

The good thing is we see the main laser is very stable after the installation of this second FI. The situation is definitely much better than the situation before.

Images attached to this comment
1362_20190518004712_tek00025.png
R&D (FilterCavity)
Print this report.
ChienMingWu - 23:52, Friday 17 May 2019 (1360)Get code to link to this report
Install the new Faraday Isolator

Chien-Ming, Yu-Hang

We installed the new Faraday Isolator (FI) on the main laser beam. The optical intensity transmission efficiency we measured was >98% although the polarizer at the entrance of this FI was accidentally scratched yesterday.

The mode matching statuses before and after inserting this new FI are shown in the attached figure. It got a little worse after inserting FI.

Now the SHG can provide 218mW when input 580mW IR. We cleaned some optical component, then we can get more IR power (up to 880 mW) to the SHG if needed.

Other mode matching status are also shown.

Images attached to this report
1360_20190517165217_shg.png 1360_20190517165221_otherstatus.png
Comments related to this report
YuhangZhao - 07:46, Saturday 18 May 2019 (1362)

I also matched BAB into OPO. The spectrum is shown in the attached figure.

The matching of main laser optical fiber is also changed after the cleaning of some mirrors.

The good thing is we see the main laser is very stable after the installation of this second FI. The situation is definitely much better than the situation before.

R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 19:23, Friday 17 May 2019 (1359)Get code to link to this report
Current filters configuration for PR and BS on photon

Since I have been changing name of the channels and modified the RTmodel quit a lot I put here the screenshots of the filter/model current configuration of photon for BS and PR, to recover it in case it is lost.

For each d.o.f of the two suspensions there is the model and the contoller (called damp).  The d.o.f and suspension is indentifed by the "module"  field and the ZPK configuration is relative the filter selected in the "section" field.

Fig1: BS PITCH  DAMP

Fig2: BS PITCH  MODEL

Fig3: BS YAW  DAMP

Fig4: BS YAW  MODEL

Fig5: PR PITCH  DAMP

Fig6: PR PITCH  MODEL

Fig7: PR YAW  DAMP

Fig8: PR YAW  MODEL

Images attached to this report
1359_20190517122201_bspitchdamp.jpeg 1359_20190517122206_bspitchmodel.jpeg 1359_20190517122211_bsyawdamp.jpeg 1359_20190517122217_bsyawmodel.jpeg 1359_20190517122843_prpitchdamp.jpeg 1359_20190517122850_prpitchmodel.jpeg 1359_20190517122855_pryawdamp.jpeg 1359_20190517122901_pryawmodel.jpeg
R&D (FilterCavity)
Print this report.
Eleonora Capocasa - 18:57, Friday 17 May 2019 (1358)Get code to link to this report
Two things I coudn't make work on DGS/medm

There are two minor issues I couldn't solve so far:

1) Despite many suggestion from Oshino-san, I could not make work the medm screen botton which saves and restores the EPICS channels values (so called screenshot). It seems there is no error in the coding but it doesn't work. So I have done two simple scripts (located in /home/controls) to save and restore the snapshot from terminal. The commands are respectively

./takesnap.sh  and ./restoresnap.sh

2) I also tried to implement the "wave rotator" function to rotate the oplev signal in order to compensate a possible inclination of the PSD but I couldn't find the name of the "angle channel" to be written in the medm. I asked the help of Shoda-san and we tried to implement the function on a test model in the ATC DGS but we failed as well. We could find some EPICS channels created into the fuction on the dataviewer but when we tried to read or write them on the terminal (commands caget, caput) it said they do not exist. They cannot be shown in a "text monitor object" either. The workaround for the moment is to use a matrix and put in the already computed cos/sin values corresponding to the desired rotation.