LOG-IN
Displaying reports 2101-2120 of 3201.Go to page Start 102 103 104 105 106 107 108 109 110 End
R&D (FilterCavity)
Print this report.
Pierre Prat - 10:49, Thursday 18 April 2019 (1308)Get code to link to this report
Modification of the CC-1 Servo-filter (Green Phase Coherent Control)
The following settings and modifications were done for the CC-1 (Green Phase Coherent Control) Servo-filter to the original Servo-filter which electronic schematics and bill of material are saved on the wiki.

0- Current configuration:

Notch filter 1, Notch filter 2 ans LP filter are disable.
The Servo-filter must be set only on 1/f integrator.
The gain is set to minimum (position 0).
An attenuator of 20dB with a 50 Ohm load is set on the ERROR IN input.
The unity gain frequency was measured to 85Hz.


1-Setting of switches on the front panel:

* The differentiator shall be disabled on the front panel in setting the switch on "OFF".

* The switch INV/NON INV on the front panel, shall be set on INV.


2-Setting of the 8 straps on the board:

Notch filter 1, notch filter 2 and Low-pass filter are disabled in setting strap on connectors P7, P8 and P9 between pins 2 and 3.

* The transmission signal is ont used.
The strap on connector P4 (3 pins) is set between pin 2 and 3.

* Strap is set on connector P11 (3 pins), between pins 2 and 3, in order to activate the sample-and-hold on the triangular signal, on the locking.
* Strap is set on connector P3 (2 pins) to connect the triangular signal to the output stage.

* Strap is set on connector P2 (3 pins), between pins 1 and 2, for test purpose.
To check notch 1 and notch 2 filters (in scan mode) between TEST IN and TEST OUT. For this test the differentiator, shall be set on "ON" (not intuitive but important). After this test, the differentiator shall be disabled the front panel in setting the switch on "OFF".

* Strap is set on connector P1 (2 pins), in order to be able to tune the offset.


3-Modification of components:

* Integrator 1/f: corner frequency changed to 22 kHz
Capacitor CMS 1206: C38 = 3.3nF

* Integrator 1/f2: corner frequency changed to 22.5 Hz
Capacitor CMS 1206: C26 = C33 = 2200nF
Capacitor CMS 1206: C25 = C32 = 1000nF

* Low-pass filter: unchanged

* Notch filter 1: notch frequency changed to 11.8 kHz / quality factor changed to 0.9 (measured)
[Capacitor CMS 0805 1% : C49 ; C50 ; C51 ; C53 = unchanged (560 pF)]
Resistor CMS 1206 : R65 ; R66 ; R67 ; R68 = 24k
Resistor CMS 1206 : R73 = 13k

* Notch filter 2: notch frequency changed to 14.2 kHz / quality factor changed to 4.85 (measured)
[Capacitor CMS 0805 1% : C60 ; C61 ; C62 ; C63 = unchanged (560 pF)]
Resistor CMS 1206 : R79 ; R80 ; R81 ; R82 = 16k
Resistor CMS 1206: R89 = 1.3k

* Gain adjustment (G): Gmin = 0.4 / Gmax = 16.5 / Gtyp = 6

* Input impedance
Resistor CMS 1206 : R145 and R146 removed

R&D (FilterCavity)
Print this report.
NaokiAritomi - 10:05, Thursday 18 April 2019 (1307)Get code to link to this report
Re-alignment of ML PLL fiber and homodyne
 
[Aritomi, Yuhang]
 
This is work on April 16th.
A mirror for ML PLL moved again... We replaced the folk and fixed it. Then we re-aligned ML PLL fiber.
 
Current fiber coupling:
ML AUX1: 3.3 mW → 0.5 mW, coupling: 0.5*2/3.3 = 30%
ML AUX2: 4 mW → 0.5 mW, coupling: 0.5*2/4 = 25%
 
We checked space for telescope for a filter cavity and moved some folks and HWP in squeezing path to save space.
Then we re-aligned homodyne, but we couldn't see any squeezing at that time.
Images attached to this report
1307_20190418030715_mirrorfiber.png
R&D (FilterCavity)
Print this report.
NaokiAritomi - 07:51, Thursday 18 April 2019 (1306)Get code to link to this report
Comment to IR injection and reflection telescopes new scheme (Click here to view original report: 1305)

-200 lens for reflection telescope is on OPO transmission path and it changes mode matching of OPO transmission. So this configuration is not feasible.

R&D (FilterCavity)
Print this report.
EleonoraPolini - 14:48, Wednesday 17 April 2019 (1305)Get code to link to this report
IR injection and reflection telescopes new scheme

I found a new solution, better than the previous one, considering a larger database of lenses.

The robustness is good, moving one lens in a range of 1cm, the total mismatch is lower than 20%.

Images attached to this report
1305_20190417074618_injectionbetter.png 1305_20190417074628_reflectionbetter.png 1305_20190417074645_newscheme.png 1305_20190417074713_mismatchreflection1stlensbetter.png 1305_20190417074726_mismatchreflection2ndlensbetter.png 1305_20190417074747_mismatchinjection1stlensbetter.png 1305_20190417074800_mismatchinjection2ndlensbetter.png
Comments related to this report
NaokiAritomi - 07:51, Thursday 18 April 2019 (1306)

-200 lens for reflection telescope is on OPO transmission path and it changes mode matching of OPO transmission. So this configuration is not feasible.

R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 15:40, Tuesday 16 April 2019 (1303)Get code to link to this report
More boards for GALVO control

I have found (between BS adn NM1 chamber) a rack with 5 more boards for the galvo control. See attached picture.

Maybe some of them are the "new version" Yuefan was talking about?

On two of them there is also a label specifing if the QPD has big or small range. 

Images attached to this report
1303_20190416083944_qpdcentering.jpeg
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 15:19, Tuesday 16 April 2019 (1302)Get code to link to this report
PR yaw loop closed with new DGS

After solving the DGS issue with the filter loading, I could test the simulink model on the control of YAW of PR.

The mechanical TF and the closed loop TF are shown in pic 1 and 2.  The comparison between the open and closed loop spectrum is shown in pic 3. The control seems to work fine.

UGF is at ~ 6 Hz and phase margin is ~ 50 deg.  A first, rather basic version of medm screen developped for the control is shown in pic 4.

Error and correction signals are currenty in counts and needs to be calibrated.

Images attached to this report
1302_20190416082226_prtf160419.png 1302_20190416082235_olpryaw160419.png 1302_20190416082244_spectrapryaw150419.png 1302_20190416082251_medmpr.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 22:46, Monday 15 April 2019 (1300)Get code to link to this report
200Hz noise from our cleaning room fan

Today we used the sound spectrum analyzer characterized the sound environment. We found a clear frequency from our cleanroom fan. It is 200Hz.

Images attached to this report
1300_20190415154634_wechatimg383.jpeg 1300_20190415154640_wechatimg384.jpeg
R&D (FilterCavity)
Print this report.
YuhangZhao - 22:42, Monday 15 April 2019 (1299)Get code to link to this report
Broken connector for Alignment-Mode-Cleaner PZT

Yuhang, Pierre, Aritomi

In the beginning, we tried to check the alignment of homodyne. But we cannot see a meaningful signal from AMC transmission.

We found the PZT of AMC was broken. I guess it is related with the strong force we(mainly it's me) enforced on this connector or its wire. I am sorry that I made very ugly soldering.

Anyway, we repaired it and we heard the sound of PZT by sending a 4kHz signal.

So in the future, let's be kind for our wireswink!

Images attached to this report
1299_20190415154016_wechatimg382.jpeg 1299_20190415154021_wechatimg381.jpeg
R&D (FilterCavity)
Print this report.
EleonoraPolini - 19:32, Monday 15 April 2019 (1298)Get code to link to this report
Robustness of injection telescope and reflection telescope

I report the robustness of the injection and reflection telescopes described in entry #1296.

The two telescopes can be consiedered robust enough (till 20% of mismath) only in a range of +/- 2 mm.

Images attached to this report
1298_20190416114435_mismatchinjection1stlensnew.png 1298_20190416114541_mismatchinjection2ndlensnew.png 1298_20190416114553_mismatchreflection1stlens.png 1298_20190416114602_mismatchreflection2ndlens.png
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 22:35, Friday 12 April 2019 (1297)Get code to link to this report
DGS: Filter loading issue fixed going back to old version

[T.Yamamoto, Y.Fujii, Eleonora]

Here the report from Yamamoto-san about today's work:

- Real-time models was not able to read filter files.
Models detected the modification of filter files.
But “COEFF LOAD” button did not work well.

- We unified the RCG version as v3.1.1.
At first master and slave model run as v2.8.8 and v3.1.1, respectively.
But the problem was not solved by unifying the version.

- System clock of STDA on BIOS was fixed.
System clock should be set as UTC. But it was set as JST.
So system time showed 9 hours future and date of file modification was wrong.
We fixed the time-stamp of filter file, but problem was not solved.

- Filter files re-generated after fixing system clock.
We moved filter files and re-generated them by rebuilding models.
But the problem was not solved.

- The version of real-time system returned back form v3.1.1 to v2.8.8.
The problem was solved by using v2.8.8.

#####################

Thanks a lot to Yamanoto-san and Fujii-san for the precious help and for all the time spent!

R&D (FilterCavity)
Print this report.
EleonoraPolini - 19:34, Friday 12 April 2019 (1296)Get code to link to this report
IR injection e reflection telescope

I made new simulations taking into account also the telescope for the beam reflected from the cavity into the homodyne. 

Injection telescope:

focal lenght 1 = -101.65 mm

focal lenght 2= 204.6 mm

Reflection telescope:

focal lenght 1= 203.3 mm

focal lenght 2 = -101.65 mm

In fig 3 you can find the scheme of the two telescopes on the bench.

Next step:

- Test the two telescopes

Images attached to this report
1296_20190412123340_injnew2.png 1296_20190412123350_ref1.png 1296_20190412123410_injrefpaint2.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 19:21, Friday 12 April 2019 (1295)Get code to link to this report
Comment to System performances with CC power reduced (Click here to view original report: 1252)

I just measured there are 3uW of p-pol is going also into homodyne.

R&D (FilterCavity)
Print this report.
Pierre Prat - 13:39, Friday 12 April 2019 (1294)Get code to link to this report
Modification of the OPO Servo-filter
The following settings and modifications were done for the OPO Servo-filter to the original Servo-filter which electronic schematics and bill of material are saved on the wiki.


1-Setting of switches on the front panel:

* The differentiator shall be disabled on the front panel in setting the switch on "OFF".

* The switch INV/NON INV on the front panel, shall be set on INV.


2-Setting of the 8 straps on the board:

Low-pass filter, Notch filter 1 and notch filter 2 are activated on the board in setting strap on connectors P7, P8 and P9 (3 pins) between pins 1 and 2

* The transmission signal is positive with a peak at 1.26V.
It shall be inverted: the strap on connector P4 (3 pins) is set between pin 2 and 3.
The threshold level must normally be tuned to a negative level of 600mV (THRESHOLD OUT).
We had also to increase the sample-hold capacitor (C89). See below.

* Strap is set on connector P11 (3 pins), between pins 2 and 3, in order to activate the sample-and-hold on the triangular signal, on the locking.
* Strap is set on connector P3 (2 pins) to connect the triangular signal to the output stage.

* Strap is set on connector P2 (3 pins), between pins 1 and 2, for test purpose.
To check low-pass filter, notch 1 and notch 2 filters (in scan mode) between TEST IN and TEST OUT. For this test the differentiator, shall be set on "ON" (not intuitive but important). After this test, the differentiator shall be disabled the front panel in setting the switch on "OFF".

* Strap is set on connector P1 (2 pins), in order to be able to tune the offset.


3-Modification of components:

* Integrator 1/f: corner frequency changed to 3.3 kHz
Capacitor CMS 1206: C38 = 22nF

* Integrator 1/f2: corner frequency changed to 220 Hz
Capacitor CMS 1206: C26 = C33 = 330nF

* Low-pass filter: cut-off frequency changed to 3.3 kHz
Capacitor CMS 0805 : C45 = 2.2nF (0805)
Resistor CMS 1206 : R59 = 22k

* Notch filter 1: notch frequency changed to 10.5 kHz / quality factor changed to 6 (measured)
[Capacitor CMS 0805 1% : C49 ; C50 ; C51 ; C53 = unchanged (560 pF)]
Resistor CMS 1206 : R65 ; R66 ; R67 ; R68 = 27k
Resistor CMS 1206 : R73 = 820

* Notch filter 2: notch frequency changed to 14.2 kHz / quality factor changed to 6 (measured)
[Capacitor CMS 0805 1% : C60 ; C61 ; C62 ; C63 = unchanged (560 pF)]
Resistor CMS 1206 : R79 ; R80 ; R81 ; R82 = 20k
Resistor CMS 1206: R89 = 820

* Gain adjustment (G): Gmin = 0.125 / Gmax = 05 / Gtyp = 1
No modification.

* Modification of he sample-hold capacitance on the triangular signal:
Capacitor 1206 of 4.7microFarad added on the 1microFarad capacitor (C89).
R&D (FilterCavity)
Print this report.
Pierre Prat - 13:23, Friday 12 April 2019 (1293)Get code to link to this report
Modification of the IRMC Servo-filter
The following settings and modifications were done for the IRMC Servo-filter to the original Servo-filter which electronic schematics and bill of material are saved on the wiki.


1-Setting of switches on the front panel:

* The differentiator shall be disabled on the front panel in setting the switch on "OFF".

* The switch INV/NON INV on the front panel, shall be set on INV.


2-Setting of the 8 straps on the board:

Low-pass filter, Notch filter 1 and notch filter 2 are activated on the board in setting strap on connectors P7, P8 and P9 (3 pins) between pins 1 and 2

* The reflection signal is negative with a base at about -600mV (after the change of the trans-impedance resistor of the photodetector from 51 Ohm to 510 Ohm ).
It shall be inverted: the strap on connector P4 (3 pins) is set between pin 2 and 3.
The threshold level must normally be tuned to a positive level of +300mV (THRESHOLD OUT).
In practice, we had best result with a threshold level of about +500mV.
We had also to increase the sample-hold capacitor (C89). See below.

* Strap is set on connector P11 (3 pins), between pins 2 and 3, in order to activate the sample-and-hold on the triangular signal, on the locking.
* Strap is set on connector P3 (2 pins) to connect the triangular signal to the output stage.

* Strap is set on connector P2 (3 pins), between pins 1 and 2, for test purpose.
To check low-pass filter, notch 1 and notch 2 filters (in scan mode) between TEST IN and TEST OUT. For this test the differentiator, shall be set on "ON" (not intuitive but important). After this test, the differentiator shall be disabled the front panel in setting the switch on "OFF".

* Strap is set on connector P1 (2 pins), in order to be able to tune the offset.


3-Modification of components:

* Integrator 1/f: corner frequency changed to 154 Hz
Capacitor CMS 1206: C38 = 470nF

* Integrator 1/f2: corner frequency changed to 220 Hz
Capacitor CMS 1206: C26 = C33 = 330nF

* Low-pass filter: cut-off frequency changed to 154 Hz
Capacitor CMS 0805 : C45 = 2.2nF (0805) + 270nF (through capacitor)
Resistor CMS 1206 : R59 = 3.9k

* Notch filter 1: notch frequency changed to 11.85 kHz / quality factor changed to 3 (measured)
[Capacitor CMS 0805 1% : C49 ; C50 ; C51 ; C53 = unchanged (560 pF)]
Resistor CMS 1206 : R65 ; R66 ; R67 ; R68 = 24k
Resistor CMS 1206 : R73 = 2.7k

* Notch filter 2: notch frequency changed to 17.75 kHz / quality factor changed to 0.9 (measured)
[Capacitor CMS 0805 1% : C60 ; C61 ; C62 ; C63 = unchanged (560 pF)]
Resistor CMS 1206 : R79 ; R80 ; R81 ; R82 = 16k
Resistor CMS 1206: R89 = 13k

* Gain adjustment (G): Gmin = 0.0125 / Gmax = 0.5 / Gtyp = 0.1 
Resistor CMS 1206 : R33 = 2.2k
Resistor CMS 1206 : R5 = 10k
Resistor CMS 1206 : R7 = 10k

* Modification of threshold circuit in order to have a better tuning.
Resistor CMS 1206 : R108 = 1M (hysteresis resistor).
Resistor 2k added between pin 1 of the J14 connector and the wire (red) going to potentiometer 2k on front panel
Resistor 2k added between pin 3 of the J14 connector and the wire (blue) going to potentiometer 2k on front panel

* Modification of he sample-hold capacitance on the triangular signal:
Capacitor 1206 of 4.7microFarad added on the 1microFarad capacitor (C89).
R&D (FilterCavity)
Print this report.
YuhangZhao - 23:55, Thursday 11 April 2019 (1292)Get code to link to this report
Loop characterization of coherent control loop between pump beam and coherent control beam

In the beginning, we measured the optomechanical transfer function of the CC loop with Stanford Research 560 at the beginning. However, since this lock is not so stable, the measurement was very noisy.

Then Pierre put an integrator before 10kHz. We could lock loop with high gain. However, this is a small stable region. If we increase or decrease gain, we can both have oscillation. Anyway, we can lock in that small region. Then we measured both open loop transfer function and optomechanical transfer function.

The result is shown in the attached figure. There are several peaks in the low-frequency region.

Images attached to this report
1292_20190411165511_ccomtf.png 1292_20190411165517_ccoltf.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 23:02, Thursday 11 April 2019 (1291)Get code to link to this report
open loop transfer function of OPO

Pierre and Yuhang

Here is the open loop transfer function of OPO.

Images attached to this report
1291_20190411160224_opooltf.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 22:43, Thursday 11 April 2019 (1290)Get code to link to this report
Comment to The comparision between old and new servo on homodyne noise spectrum with only LO (Click here to view original report: 1286)

After connecting the power of homodyne and spectrum analyzer(this can remove the ground loop noise), I measured noise spectrum of homodyne again. During this measurement, there is only an infrared beam incident and has a power value of 1.3mW.

The result is shown in the attached figure.

Images attached to this comment
1290_20190411154444_figure1.png
R&D (FilterCavity)
Print this report.
EleonoraPolini - 18:46, Thursday 11 April 2019 (1289)Get code to link to this report
IR injection telescope simulation

I did a new simulation extending the optical path (of 45 cm), using 4 additional mirrors (fig 1).

The best simulation is shown in fig 2 where the first lens is 47.5 cm far from PBS and the second one is 86.6 cm far from PBS, both with the same focal of 1020 mm.

This solution is more robust than the previous one (entry #1283)  as shown in fig 3 and fig 4. Moving the lenses of +/- 1 cm we have a mismatch < 10%.

Next step:

- Simulation for the FC reflected beam (first idea in fig 5).

Images attached to this report
1289_20190411114258_benchinjection.png 1289_20190411114316_2lensesallpathbis.png 1289_20190411114324_mis1stlenslongpath2.png 1289_20190411114339_mis2ndlenslongpath2.png 1289_20190411120316_injectionrefl1.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 22:10, Wednesday 10 April 2019 (1287)Get code to link to this report
Coherent control PLL works again and stability measurement

Yuhang and Matteo

After Pierre replaced the chip of coherent control PLL, we measured the correction signal within a long period of time. This correction signal tells us the stability of CC PLL loop.

The coefficient of variation of the fast correction signal is 0.0538.

While the coefficient of variation of the slow correction signal is 0.1311.

Although we saw a clear change of fast correction signal. The slow correction signal shows much more high-frequency noise.

I also calculated the correlation coefficient between these two signals. It is -0.5155 as we can see one signal is going down while the other is going up.

Images attached to this report
1287_20190410151125_figure11.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 19:11, Wednesday 10 April 2019 (1286)Get code to link to this report
The comparision between old and new servo on homodyne noise spectrum with only LO

Since we have new IRMC servo working. I measured again the noise spectrum of homodyne when there is only LO incidence. I did the measurement after common mode noise rejection. The result is shown in the attached figure. There is three main difference.

1. New servo has more flat low-frequency noise.

2. New servo shows much more peaks in the low-frequency region. It looks like electronic noise. As pointed by Matteo, this is because of the ground loop.

3. The peaks at the high frequency of these two situations are different.

Images attached to this report
1286_20190410121058_figure1.png
Comments related to this report
YuhangZhao - 22:43, Thursday 11 April 2019 (1290)

After connecting the power of homodyne and spectrum analyzer(this can remove the ground loop noise), I measured noise spectrum of homodyne again. During this measurement, there is only an infrared beam incident and has a power value of 1.3mW.

The result is shown in the attached figure.