LOG-IN
Displaying reports 2401-2420 of 3201.Go to page Start 117 118 119 120 121 122 123 124 125 End
R&D (FilterCavity)
Print this report.
YuhangZhao - 10:21, Thursday 02 August 2018 (928)Get code to link to this report
The measurement of Finesse of green mode cleaner(p-pol and s-pol)

We measured the Finesse of green mode cleaner (while p-pol and s-pol)

For s-pol, the measurement is fine. But for p-pol, the main peak overlaps with sidebands, as you can see in the attached figure 2. I take only the central part of data to do fit to avoid the influence of sideband. The result is listed as following.

For s-pol, the Finesse is 300

For p-pol, the Finesse is 36.7

Images attached to this report
928_20180802031737_spol.png 928_20180802031810_ppol1.png 928_20180802031815_ppol2.png
R&D (FilterCavity)
Print this report.
MarcEisenmann - 16:49, Wednesday 01 August 2018 (926)Get code to link to this report
New sections on the NAOJ wiki

I opened 2 new sections on NAOJ wiki :

 

"Pictures" which can be useful for external people to know what space is available on the bench for example.

It isn't yet totally working but it has already 1 picture on it.

 

"Available optics and datasheets" could also be useful (e.g for future telescope design)

R&D (FilterCavity)
Print this report.
YuhangZhao - 20:09, Monday 30 July 2018 (924)Get code to link to this report
The installation of telescope for p-pol path

During the last week and today, we did a lot of investigation about how to put lens to achieve many limitations. Today, we finally found it really diffcult to design telescope with all the condition we have now. We talked with Matteo, we got an important conclusion. It is we can decrease laser low to 30mW.

However, after measured the beam dimension, we found the beam is not like the simulation result of Jammt. We guess this comes from that the inital beam is not very collimated. We decided to decrease power further more. So we decrease power to 18mW. Then we characterize beam again. The beam is really similar with before. See Fig.1

Then we put the EOM, the position is decided as shown in attached Fig.2. After put EOM, we characterized beam again. The Fig.2 is the characterization result.

According to this result. We tried to put another lens. We can recover the beam to beam waist as 940um. See attached Fig.3. This is a little different from what we need. By putting more lens, we can have good dimension of 1000um of waist. However, more lenses cost too much space.

Images attached to this report
924_20180730130924_powercomp.png 924_20180730130929_untitled.png 924_20180730130945_lensdesign.png
KAGRA MIR (Absorption)
Print this report.
RaffaeleFlaminio - 15:17, Sunday 29 July 2018 (923)Get code to link to this report
Comment to Separate setup for noise investigation on 1310nm laser (Click here to view original report: 921)
- There is a large peak at 35 kHz when the loop is closed. Is the loop gain too high?
- The dark noise measured above 1 kHZ is not as flat as one would expect. Is this real or is it a problem with the measurement?
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 23:13, Thursday 26 July 2018 (921)Get code to link to this report
Separate setup for noise investigation on 1310nm laser
I moved another optical table in the cleanroom because I didn't have enough space to properly test the laser noise reduction.
 
I removed the laser fiber from its collimator without moving the collimator, so that, when I will put it back, I will not loose the alignment.
 
I set the laser at 200mA which is half of the typical operation, this current gives 60mW in output. I put an OD2 filter after the collimator and a beam dumper to collect its reflection. Then a non-polarizing cube beam splitter 50:50 and a 50mm lens for each beam to focus the beam on 2 identical photodiodes (DET10N and DET10N/M). On one beam I measure the beam size at different positions. To do that, I put a blade moved by the old translation stage and controlled by the old software that makes scans of the DC. On the DC channel, I sent the analog output of the PM100D power meter .
 
I found a waist of 20micron at 54mm from the lens and I put the PD at about that position. I didn't repeat the waist measurement for the other lens but I put the other PD at the same distance from the lens. Each PD is mounted on a XY micrometric translation stage, so, by watching the signal at the oscilloscope I centered the beams in the PDs (moving the PDs instead of the beam).
 
I put a T with a load resistance of 7.5kOhm for both the PDs. This gives 2.5V DC signal for one PD and 2.8V for the other. The two signals, coupled in AC, overlap on the oscilloscope and show the "jumps" that are probably the noise reason.
 
I closed the loop with a lowpass at 3Hz and gain 1000. See the spectra.
 
The noise in the out-of-loop PD reduces by a factor of 10 (20dB)
 
I also connected the PD at the lockin and after demodulation at 380Hz (with the internal oscillator) the noise significantly reduces (by about 10 times by eye).
 
Next thing to do is to place the PDs in the absorption setup and check if the absorption signal noise reduces in the same way
Images attached to this report
921_20180726151806_z00mm.png 921_20180726151813_z05mm.png 921_20180726151817_z10mm.png 921_20180726151822_z15mm.png 921_20180726151827_z20mm.png 921_20180726151832_z25mm.png 921_20180726151837_z30mm.png 921_20180726151840_z35mm.png 921_20180726151844_z40mm.png 921_20180726151848_z45mm.png 921_20180726151852_z50mm.png 921_20180726151906_after50mmlens.png 921_20180726161329_011.jpg 921_20180726161334_01.jpg 921_20180726161340_001.jpg 921_20180726161345_00.jpg 921_20180726161442_20180726inloopdet10noutofloopdet10nm.png
Comments related to this report
RaffaeleFlaminio - 15:17, Sunday 29 July 2018 (923)
- There is a large peak at 35 kHz when the loop is closed. Is the loop gain too high?
- The dark noise measured above 1 kHZ is not as flat as one would expect. Is this real or is it a problem with the measurement?
R&D (FilterCavity)
Print this report.
YuhangZhao - 21:26, Thursday 26 July 2018 (919)Get code to link to this report
Comment to Beam characterizations and EOM telescope design (Click here to view original report: 909)

At the begining, we used the wrong beam dimmsion, the initial beam DIAMETER is 2000um.

BUT, all the telescopes we designed are using RADIUS as 2000um.

Today, we realized this problem. I designed the telescope again. The EOM doesn't make a large difference. This design can be a fine reference.

Lesson: actually we have many chances to realized this problem, we checked every time after putting the lens. But everytime, we checked only the beam waist position. We never checked the beam waist size. So we didn't realize this problem. So next time we should check both of them carefully.

Images attached to this comment
919_20180727024829_opoeomt2.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 19:29, Wednesday 25 July 2018 (915)Get code to link to this report
Astigmatism after put EOM(for opo)

Participiant: Marc and Yuhang

We want to know how the beam will look like after putting the EOM. So we put EOM yesterday and checked the output of it today.

Before putting EOM: we found for the far field(more than 40cm), we can see clearly the astigmatism even with the card. See attached Fig.1

After putting EOM: we found the astigmatism disappear for the far field(the beam looks very perfect by card now, unfortunatly I didn't take picture). But we found it in near field. See attached Fig.2. You can see the waist position is really different. But for the far field, it looks fine.

Images attached to this report
915_20180725122912_2449420449222702925520180725192602.jpg 915_20180725122925_fiteomopo.png
R&D (FilterCavity)
Print this report.
MarcEisenmann - 18:43, Wednesday 25 July 2018 (913)Get code to link to this report
Comment to Beam characterizations and EOM telescope design (Click here to view original report: 909)

As Eleonora pointed out we used a wrong datasheet for the EOM (and also did some wrong calculations for the max beam size inside the EOM...)

Here is the good size range : between 300 and 80 um

We designed a new telescope (Fig1) as the following : f=125mm lens and 10cm after f=-25mm lens.

This should allow to have a beam size around 200um inside the EOM.

 

 

Question : For the green EOM, the astigmatism depended a lot on the lens position.

Is the astigmatism also that problematic? We will have quite a short beam path until the OPO.

Anyway, we found 2 trails on which we can translate the lenses borrowed from Manuel's experiment.

 

 

Fig 2 : beam size before the EOM

The beam didn't seem to be too much astigmatic after the EOM (posted soon)

Images attached to this comment
913_20180725113502_eomtelescope.png 913_20180725114206_beforeeom.jpg
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 14:09, Tuesday 24 July 2018 (912)Get code to link to this report
Comment to Beam characterizations and EOM telescope design (Click here to view original report: 909)

You can find attached the complete data sheet for the 88 MHz EOM which I get from Quibig. Specs may be a bit different from that reported in the entry. 

Non-image files attached to this comment
R&D (FilterCavity)
Print this report.
YuhangZhao - 10:14, Tuesday 24 July 2018 (911)Get code to link to this report
The characterization of beam going out from Green Mode-cleaner

Yesterday, we measured the Green Mode-cleaner output beam. While measurement, I found the lock of green mode cleaner is more stable for half fringe. However, the lock we did last week is only for fringe higher than half fringe. If we lock it on half fringe we will have less power transmitted but more stable. However, I should say that even with half-fringe lock, the lock can be destoried by vibration. So maybe we should consider to put some rubber under the MZ to isolate the vibration.

Anyway, by locking the beam many times. I must say here that the measurement is performed with different locking condition. Although I tried to make the lock the same, I cannot make sure they are exactly the same. But the result is fine. I attached the figure here.

Images attached to this report
911_20180724031401_figure1.png
Comments related to this report
YuhangZhao - 19:49, Thursday 02 August 2018 (933)

Last time I used the wrong set up of the beam profiler. This time we used a correct one. The result is attached. The beam is not a round shape in this measurement. And I found the beam is shaking  while measurement. 

R&D (FilterCavity)
Print this report.
MatteoBarsuglia - 00:56, Tuesday 24 July 2018 (910)Get code to link to this report
Comment to Phase noise measurement after changing rampeauto (Click here to view original report: 903)

I'm wondering if the increase of the phase noise at high frequency when the cavity is locked is due to the fact that, when the cavity is locked, the frequency changes of the master laser are very large ~ MHz.

Possible tests to check this hypothesis are to damp the mirrors (sending the PZT correction signal to the mirrors, upon filtering) or to excite the mirror oscillations, to artificially increase the laser frequency changes. 

A related question: when we compute the residual RMS phase noise between the main laser and the auxiliary laser we integrate down to 100 Hz. Maybe the 1 Hz region is dominant with respect to the high frequency region, and thus we should solve in any case this problem of  the main laser frequency changes in the Hz region.

R&D (FilterCavity)
Print this report.
MarcEisenmann - 23:44, Monday 23 July 2018 (909)Get code to link to this report
Beam characterizations and EOM telescope design

Participants : Eleonora, Yuefan, Yuhang

 

Since last friday we started designing the EOM telescope.

 

EOM parameters:

Following the EOM datasheet (attached to this entry) the beam conditions inside the EOM are the following :

Max beam size defined by EOM aperture (3x3mm) : max beam radius = 425 um

Min beam size defined by max optical intensity (20W/mm^2) : min beam radius = 75 um (as the input power is around 350mW)

This requirements can be meet if we use a f=175 lens and place the EOM 10cm after it. (actually we first used the wrong value of max optical density first meaning that the beam is now more than 100um inside the EOM)

Issues :

Because of the Faraday Isolators, the beams after the two 98:2 are quite astigmatics (the datas will be added tomorrow morning).

The beam were also vertically tilted (3.1 mrad for the beam going to the EOM).

By using only 1 lens after the 98:2 we couldn't achieve better than 86% of transmission.

Possible solution and future work :

We then installed 2 steerings mirrors before the lens in order to correct the beam tilt. This means that the EOM path is now shifted 5cm away from the laser with respect to the nominal position.

It seems that there is enough space to use this solution (and to recombine the 2 beams we could then use 1 steering mirror and rotate the PBS).

Tomorrow we will installed EOM and characterized the output beam.

It should then be possible to use this solution to recombine the 2 beams.

Images attached to this report
909_20180723162331_eomdatasheet.jpg
Comments related to this report
EleonoraCapocasa - 14:09, Tuesday 24 July 2018 (912)

You can find attached the complete data sheet for the 88 MHz EOM which I get from Quibig. Specs may be a bit different from that reported in the entry. 

MarcEisenmann - 18:43, Wednesday 25 July 2018 (913)

As Eleonora pointed out we used a wrong datasheet for the EOM (and also did some wrong calculations for the max beam size inside the EOM...)

Here is the good size range : between 300 and 80 um

We designed a new telescope (Fig1) as the following : f=125mm lens and 10cm after f=-25mm lens.

This should allow to have a beam size around 200um inside the EOM.

 

 

Question : For the green EOM, the astigmatism depended a lot on the lens position.

Is the astigmatism also that problematic? We will have quite a short beam path until the OPO.

Anyway, we found 2 trails on which we can translate the lenses borrowed from Manuel's experiment.

 

 

Fig 2 : beam size before the EOM

The beam didn't seem to be too much astigmatic after the EOM (posted soon)

YuhangZhao - 21:26, Thursday 26 July 2018 (919)

At the begining, we used the wrong beam dimmsion, the initial beam DIAMETER is 2000um.

BUT, all the telescopes we designed are using RADIUS as 2000um.

Today, we realized this problem. I designed the telescope again. The EOM doesn't make a large difference. This design can be a fine reference.

Lesson: actually we have many chances to realized this problem, we checked every time after putting the lens. But everytime, we checked only the beam waist position. We never checked the beam waist size. So we didn't realize this problem. So next time we should check both of them carefully.

KAGRA MIR (Absorption)
Print this report.
RaffaeleFlaminio - 17:35, Monday 23 July 2018 (907)Get code to link to this report
Comment to Simulation: change pump size (Click here to view original report: 905)
Yes, it would be interesting to see what happens if the optimization is done in both cases.
KAGRA MIR (Absorption)
Print this report.
RaffaeleFlaminio - 17:33, Monday 23 July 2018 (906)Get code to link to this report
Comment to simultaneous power fluctuations measurement with 2 PDs (Click here to view original report: 895)
- The video shows some "jump" in the signals. Are these real i.e. due to the laser or due to some
effect in the setup?
- When the loop is closed there is an oscillation at high frequency. Is the loop gain too high?
- Having the beams well focused within the photodiodes is important for this test.
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 16:08, Monday 23 July 2018 (905)Get code to link to this report
Simulation: change pump size

Still looking for the sapphire calibration factor of ~3

I simulated the absorption signal for bulk reference and sapphire sample in two different conditions:
- pump size (waist radius) 40micron
- pump size (waist radius) 80micron 

The absorption parameter used to simulate the sapphire absorption signal is 60ppm/cm.

To calculate the sapphire absorption from the simulated signal I applied the formula Abs = AC/DC / P / R
where the calibration factor is R = (AC/DC)_ref / P_ref / Abs_ref
with P = 10W and P_ref = 30mW.
I didn't apply the material correction because that's the final estimation of this simulation.

In the case of pump size 40micron, the absorption is 19.5ppm/cm, which, compared with the 60ppm/cm gives a material correction factor of 3.08
In the case of pump size 80micron, the absorption is 23.8ppm/cm, which, compared with the 60ppm/cm gives a material correction factor of 2.52

Conclusion:
decreasing the pump size results in a better estimation of the material correction (comparing it with the value of 3.34 given by the SPTS company),
But it is still far from the factor of ~3 discrepancy of my measurements.
If the discrepancy was all due to the pump size and the simulation were exact, the material correction in the case of 80micron pump size should have been about 1, instead of 2.52.

Comments:
 - There is a strong approximation on the bulk reference material, which is schottglass#21 but in the simulation is silica (because I couldn't find the thermal properties of schottglass).
 - The phase for the same material shouldn't change with the pump size. But the simulation gives different values of the phase. This may be due to a different optimal position of the Imaging Unit for different pump sizes, and I didn't optimize it for the new 40micron pump size.

Images attached to this report
905_20180723085838_pumpsize40and80.png
Comments related to this report
RaffaeleFlaminio - 17:35, Monday 23 July 2018 (907)
Yes, it would be interesting to see what happens if the optimization is done in both cases.
ManuelMarchio - 10:20, Thursday 09 August 2018 (938)

Simulating the absorption of the surface reference, I optimized the Imagin Unit distances to have the maximum signal in the two cases, pump waist 40um and pump waist 80um. See the first plot, it shows the signal as a function of the distance d2 from the lens and the small sphere.

Using the optimum value of d2 in the two cases, I repeated the simulation of elog entry 905.

In the case of pump size 40micron, the absorption is 14.7ppm/cm, which, compared with the 60ppm/cm gives a material correction factor of 4.09
In the case of pump size 80micron, the absorption is 19.7ppm/cm, which, compared with the 60ppm/cm gives a material correction factor of 3.03

the probe size is still 180um in both cases, next step is to reduce it as well to be 3 times larger than the pump

ManuelMarchio - 11:20, Friday 10 August 2018 (941)

I reduced the probe size as well, from 180um to 120um, to be 3 times larger than the pump (which is 40um), but the signal doesn't change much. 

R&D (FilterCavity)
Print this report.
YuhangZhao - 10:16, Monday 23 July 2018 (904)Get code to link to this report
Comment to The procedure to use PLL (Click here to view original report: 860)

For the 7 of step, first thing is to demodulate this signal with the frequency of beat note. Then by chaning the phase of this demodulation signal, we can make the demodulation output close to zero. This is crucial for the measurement of phase noise with DC coupling.

R&D (FilterCavity)
Print this report.
YuhangZhao - 10:11, Monday 23 July 2018 (903)Get code to link to this report
Phase noise measurement after changing rampeauto

Participaint: Eleonora and Yuhang

Following the procedure of Marco, we measured the phase noise again. The difference is now rampeauto doesn't have ramp-in port and one of the resistors is changed.

Before, we found the error noise spectrum is limited by the laser frequency noise after change.

The measurement of phase is performed in three different cases: rampeauto off, rampeauto on and lock on.

The result is shown in attached figure. However, we found the phase noise level is comparable with before while locking.

Images attached to this report
903_20180723031126_figure1.png
Comments related to this report
MatteoBarsuglia - 00:56, Tuesday 24 July 2018 (910)

I'm wondering if the increase of the phase noise at high frequency when the cavity is locked is due to the fact that, when the cavity is locked, the frequency changes of the master laser are very large ~ MHz.

Possible tests to check this hypothesis are to damp the mirrors (sending the PZT correction signal to the mirrors, upon filtering) or to excite the mirror oscillations, to artificially increase the laser frequency changes. 

A related question: when we compute the residual RMS phase noise between the main laser and the auxiliary laser we integrate down to 100 Hz. Maybe the 1 Hz region is dominant with respect to the high frequency region, and thus we should solve in any case this problem of  the main laser frequency changes in the Hz region.

KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 23:09, Sunday 22 July 2018 (902)Get code to link to this report
Comment to simultaneous power fluctuations measurement with 2 PDs (Click here to view original report: 895)

I replaced the PM100D power meter with another DET10N that I borrowed from Tanioka-kun, and I repeated the measurement.

This time the two signals at the oscilloscope look really similar. See the attached videos. (also the coherence on the spectrum analyzer is close to 1, sorry I didn't save the coherence data).

Then I closed the loop using the PD#1 in-loop and the PD#2  out-of-loop.
Then I exchanged them and closed the loop. See the two figures.

The control loop reduces a lot the noise in-loop but it doesn't really work for he out-of-loop PD (same situation when they are exchanged).

One possible reason could be the clipping noise, because I'm not sure how precisely the beam is focused inside the area of the PD. 

Another possible reason could be the OD2 filter (that I'm putting after the laser to limit the power and avoid the PDs saturate). If I remove it, 40mW would imping on each PD. I'm not sure how safe it will be for the PD, and In order to avoid saturation, I will have to drastically reduce the load resistance.

Another way to reduce the power would be to enlarge the beams up to much more than the PD size (which is 1mm).

Images attached to this comment
902_20180722155700_20180721inloopdet10n1outofloopdet10n2.png 902_20180722155711_20180721inloopdet10n2outofloopdet10n1.png
Non-image files attached to this comment
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 20:05, Friday 20 July 2018 (901)Get code to link to this report
Preparation for the characterisation of the beam reflected by the FC

Participants : Yuefan, Eleonora

We prepare the set up to characterize the reflected beam from the FC. In the final configuration we will need to take a part of the reflected beam to send it to the quadrants for the AA.

At present the green FI reflects about 4.5 mW which are attenauted in order to send only 150 uW to the PD used for the filter cavity lock. (150 uW corresponds to a DC output of 88 mV )

Up to now we used a set of optical densities just placed on the bench to attenuate the power and today we changed them with two "mirror shaped" optical densities  ( ND 1 and ND 0.5)  which we coud directly screw on the last lens before the PD, making the setup more stable.

Since the beam height is only 3.8 cm,  we prepere a periscope to be installed before the PD. The lower mirror of the telecope should be a BS which trasmit at least 3% to the PD and send the rest to the quadrants.  Since we couldn't find any suitable mirror we could not perform the beam characterization with the FC cavity locked. We will do it as soon as we can get the mirror for the green.

Pic 1 and 2 show the periscope we assembled.

We remark that there won't be a lot of space to put quadrants and galvo in that area of the bench, so the desing has to be studied carefully.

Images attached to this report
901_20180720130134_pp2.jpg 901_20180720130141_pp3.jpg
R&D (FilterCavity)
Print this report.
MarcEisenmann - 11:19, Friday 20 July 2018 (899)Get code to link to this report
Procedure to lock the MZ and the green MC

Following Yuhang: Procedure to lock MC green (MCG) and MZ

  1. Lock SHG
  2. turn on high-voltage drivers of MZ and MCG
  3. Check alignment of MCG using a ramp  (typical value 7.05Hz, 1Vpp) now we are using s-pol but anyway it should be the same procedure with p-pol
  4. if the alignment is fine, set the gain of standford to gain = 1. Use a simple lowpass pole at 3Hz on the standford.
  5. tune MCG PZT driver offset to go as close as possible to the TEM00 peak
  6. close the loop of MCG (put MCG correction in input of MCG driver)
  7. increase the gain of standford to 5
  8. tune the MZ high voltage offset. You should see max and min. Now we are locking a little higher than mid-fringe in order to get enough tranmitted power by MCG (1V on oscilloscope which corresponds to 12.5mW)
  9. On MZ servo check EP mon (Error Point monitor)  and put it to 0 changing its offset
  10. NOTICE : WE NEED TO MAKE SURE THE GAIN IS HIGH ENOUGH (USUALLY 10) TO AVOID THE UNSTABILITY AT 600 HZ
  11. On MZ servo turn on lock and integrator

To measure MZ openloop TF we injected noise in the servo input ''add'' and take TF between  EP mon and input. For a swept sine typical noise value is 50 mVpk

Notice: if you vibrate the bench too much, you will destory the lock of MZ.