LOG-IN
Displaying reports 2461-2480 of 3274.Go to page Start 120 121 122 123 124 125 126 127 128 End
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 11:20, Friday 10 August 2018 (941)Get code to link to this report
Comment to Simulation: change pump size (Click here to view original report: 905)

I reduced the probe size as well, from 180um to 120um, to be 3 times larger than the pump (which is 40um), but the signal doesn't change much. 

Images attached to this comment
941_20180810042043_bulkandsapphireprobe120and180.png
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 10:48, Thursday 09 August 2018 (939)Get code to link to this report
Preliminar design of the pump size reduction

I order to reduce the size of the pump and keep the waist on the same position we should add a diverging lens before the converging one.

We designed a configuration on Jammt using a -100 lens and a +100 lens. See the attached image.

The new pump size will be 37.6um (radius).

A more precise design will be after a better characterization of the pump beam profile before the lenses

Images attached to this report
939_20180809034739_reducepumpsize.png
Comments related to this report
ManuelMarchio - 15:21, Friday 10 August 2018 (942)

I made a measurement of the pump beam profile to make sure of the parameters without the last converging lens.
Result:
waist = 500um
waist position = -1m

the axis origin is the converging lens mount.

Then I updated the Jammt design using these parameters for the initial beam.
The current converging lens is f=150mm, so Jammt gives a resulting waist of 67um at 165mm from the lens. See figure. 

I also made a measurement and fit after putting back the lens. the fit gives a waist of 70um at 171mm. Quite comparable with the jammt result.

then I updated the configuration to have a pump size of 70 - 75um (diameter).
See the diagram.

I need a diverging lens f=-125mm and a converging lens f=100mm

KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 10:20, Thursday 09 August 2018 (938)Get code to link to this report
Comment to Simulation: change pump size (Click here to view original report: 905)

Simulating the absorption of the surface reference, I optimized the Imagin Unit distances to have the maximum signal in the two cases, pump waist 40um and pump waist 80um. See the first plot, it shows the signal as a function of the distance d2 from the lens and the small sphere.

Using the optimum value of d2 in the two cases, I repeated the simulation of elog entry 905.

In the case of pump size 40micron, the absorption is 14.7ppm/cm, which, compared with the 60ppm/cm gives a material correction factor of 4.09
In the case of pump size 80micron, the absorption is 19.7ppm/cm, which, compared with the 60ppm/cm gives a material correction factor of 3.03

the probe size is still 180um in both cases, next step is to reduce it as well to be 3 times larger than the pump

Images attached to this comment
938_20180809031937_surfiuoptimization40um80um.png 938_20180809031950_pumpsize40and80iuoptim.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 13:31, Wednesday 08 August 2018 (937)Get code to link to this report
Comment to Green beam reflected by the FC (Click here to view original report: 935)

I  used ABCD matrix to calculate the beam parameter before the lens(the lens used to perform better beam measurement). The result is as following:

beam waist position: -3.2m (relative to Faraday Isolator)

beam waist size: 877.20um

I also attached the python code, if you are interested, please have a look.

Non-image files attached to this comment
R&D (FilterCavity)
Print this report.
MarcEisenmann - 20:09, Tuesday 07 August 2018 (936)Get code to link to this report
CC and P-Pol beam recombination

Today we work on the recombination of the 2 auxiliary laser beams.

  1. We put the rail after the EOM so that the lens before the EOM is on an usual post. The purpose of this change was to be able to match the CC and P-Pol beam sizes. After trying a lot of lenses before the EOM, we couldn't see any drastic changes on the beam shape so we guessed this EOM doesn't affect the beam shape as much as the green one. We now have 85% transmission of the EOM. The missing few % can come from dust inside the EOM making the alignment a bit more tricky.
  2. We put 2 steering mirrors before the lens in order to save some space on the bench.
  3. The second lens is set such as the beam is collimated with diameters 2150 and 2050 um close to the CC beam size (2100 um)
  4. we put the lambda/2 on the rail in order to save place
  5. We put a 2' PBS cube and checked that the 2 beams are superposed. It is still the case after the last steering mirror.

Today the 2 beams seem to be overlapped.

One issue is that we have some power loss on the p-pol beam (EOM path) :

before EOM After EOM After 1st steering mirror before lambda/2 After PBS transmission PBS reflection
13.5 mW 11.5 mW 10.1 mW 9.5 mW 9.36 mW 8.07 mW 1.12 mW

This might be linked with some polarization troubles. We will look this hypothesis tomorrow.

The new optical path is attached to this entry and the optical scheme will soon be updated.

After some rough simulations, it seems that one f = 125mm lens 5 holes before the OPO can provide a beam waist of 40 um. This is close to the nominal 36 um.

Anyway, we will use this lens tomorrow to make sure the 2 beam parameters are similar.

Images attached to this report
936_20180807130643_beamrecombinations.jpg
R&D (FilterCavity)
Print this report.
MarcEisenmann - 19:52, Tuesday 07 August 2018 (935)Get code to link to this report
Green beam reflected by the FC

In previous entry ( #934 ) we reported that the BS pitch was saturating causing a lot of residual motion of the beam.

Yesterday, we found out that the BS pitch correction was not saturating anymore (~ -6V) and the motion was at the order of 10 urad both for pitch and yaw.

This is still yet ~5 times higher than other mirror motions at the exception of EM pitch which has similar motion.

 

To do the characterization we installed a f = 100 mm lens

In order to be sure that we had a good beam shape we installed a CCD after the periscope.

When the cavity is locked, the beam is shaking a lot preventing to take much more points.

 

Good news is that the beam doesn't seems astigmatic.

Images attached to this report
935_20180807125132_reflgbeamfcwithlens.jpg
Comments related to this report
YuhangZhao - 13:31, Wednesday 08 August 2018 (937)

I  used ABCD matrix to calculate the beam parameter before the lens(the lens used to perform better beam measurement). The result is as following:

beam waist position: -3.2m (relative to Faraday Isolator)

beam waist size: 877.20um

I also attached the python code, if you are interested, please have a look.

MarcEisenmann - 22:38, Monday 27 August 2018 (958)

There were few mistakes made on this entry corrected in this one.

Attached to this entry is the proper fit of beam after the lens (previously a wrong wavelength was used for the plot).

The mean profile was used (w0 = 113.37 um 0.6981 m after the f = 100mm lens used for the characterization [lens is 40 cm after the Faraday Isolator])

 

The beam parameter is the following : w0 = 18.921 um @ 0.1168 m before the lens ie roughly 0.28320m after the faraday isolator.

YuhangZhao - 09:45, Wednesday 29 August 2018 (962)

Since my result is different from Marc's result, I did calculation again. I found a mistake in my calculation of ABCD matrixs.

From the calculation point of view, Marc's result is correct.

I will check in actual case to see if the calculation aggres with calculation or not. As I have already mentioned in the meeting, we can see the reflected beam is shaking while the filter cavity is locking. So if they don't agree with the actual case, I think the discrepancy comes from the beam shaking.

R&D (FilterCavity)
Print this report.
MarcEisenmann - 11:46, Monday 06 August 2018 (934)Get code to link to this report
Comment to Green beam reflected of FC characterization (Click here to view original report: 929)

Last Friday we wanted to do this characterization again because it seems that our previous points were still inside the Rayleigh range of the beam (which should be few meters)

The solution is to add a lens and recover the beam parameters without the lens.

We faced few issues during this measurement :

Faraday Isolator

  1. The Faraday Isolator output (facing the FC) is clipping probably higher order mode which are quite large w.r.t. the FI aperture
  2. The FI PBS which reflects the beam toward the locking photodiode is a bit broken ( it is the cause of the "strange" behavior of the beam we could observe. However, with proper alignment it is possible to avoid hitting the broken part of the PBS and have a good beam ( meaning we can avoid to have black line on the beam)

For now it seems we can keep going on with this FI at least for the beam characterization.

It may be useful in the future to change this FI for a larger aperture one or at least change the broken PBS.

BS

  1. We found out that the beam was shaking a lot ( ~40 urad rms ) while performing this measurement. This is due to the fact that the BS pitch correction is close to saturation (-10V) and some peaks lead to correction saturation.

 

The solution is to move the BS pitch picomotor. Matteo also recommends to follow the GEO procedure : Every ~2,4 weeks, use picomotors to put mirrors on good positions in order to avoid this kind of saturation.

R&D (FilterCavity)
Print this report.
YuhangZhao - 19:49, Thursday 02 August 2018 (933)Get code to link to this report
Comment to The characterization of beam going out from Green Mode-cleaner (Click here to view original report: 911)

Last time I used the wrong set up of the beam profiler. This time we used a correct one. The result is attached. The beam is not a round shape in this measurement. And I found the beam is shaking  while measurement. 

Images attached to this comment
933_20180802124946_figure1.png
R&D (FilterCavity)
Print this report.
MarcEisenmann - 18:39, Thursday 02 August 2018 (932)Get code to link to this report
Comment to Green beam reflected of FC characterization (Click here to view original report: 929)

Here are the data :

 

z = [ 0 1 2 3 4 5 6 15]*2.5+40 [number of holes]+position of first measurement

wW = [2170 2192 2181 2137 2239 2183 2214 2380]/2
wV = [ 2169 2207 2216 2211 2207 2202 2112 2114]/2

R&D (FilterCavity)
Print this report.
MarcEisenmann - 18:35, Thursday 02 August 2018 (931)Get code to link to this report
leakage in Tama (?)

Since few days (weeks?) it is quite hot and humid in the 2nd floor of Tama.

If I remember correctly it wasn't the same last summer.

We also found few mosquitoes inside Tama and even a spider web in the "clean" booth.

This seems to indiquate that there are some ways from the outside air to enter Tama quite freely..

R&D (FilterCavity)
Print this report.
MarcEisenmann - 18:31, Thursday 02 August 2018 (930)Get code to link to this report
Few beams characterizations

Today we did some progress on the EOM telescope.

 

As shown in Fig1, by using a f = 100mm lens at the output of the EOM, we can obtain a collimated beam roughly the size of the CC beam (Fig. 2).

We our many attempts to design the EOM telescope, we have placed at least 3 times a f = 100 mm lens before the EOM without noticing any huge difference on the astigmatism of the beam at the output of the EOM.

Therefore, we think that it can be useful to put the lens before the EOM of a "simple" post (its position is on a hole) and place the lens after the EOM on the rail.

This might be useful to match more precisely the beams of the CC and p-pol laser.

 

The next steps are :

install the 2 steerings mirrors and the lambda/2 on the p-pol path.

install the steering mirror and the pbs on the cc path to recombine the 2 beams.

Install a converging lens after the pbs to finely match the 2 beams.

Start the design of the OPO telescope with CC beam as reference. For this telescope, we plan to use a 30 cm long rail to be able to finely match the beams with the OPO

 

Images attached to this report
930_20180802112612_after2ndlensofeom.png 930_20180802112623_cclaser.png
R&D (FilterCavity)
Print this report.
MarcEisenmann - 11:26, Thursday 02 August 2018 (929)Get code to link to this report
Green beam reflected of FC characterization

Participants : Eleonora, Marc, Yuefan, Yuhang

 

Since last week there were few attempts to characterize the green beam reflected by the FC.

One first task was to design a periscope able to extract the beam going inside the photodiode used to lock the FC.

As Matteo Tacca requested, this periscope was installed before the lens of the photodiode.

We couldn't find any green BS so we used a IR BS as the bottom mirror of the periscope.

We still need to use 2 optical density in order to reach power values similar to the previous configuration.

This let around 200 uW going inside the photodiode which is enough to lock

[ By the way one attempt let only 18 uW inside the photodiode and lock wasn't possible.]

 

Another difficulty was to avoid the clipping of the reflected beam at the FI output (face toward the FC) which needed to change a bit the alignment condition of the FC.

There are also some strange big features on this beam which might be investigate a bit.

 

We then installed a steering mirror at the output of the periscope in order to have enough space to characterized the reflected green beam.

The characterization is Fig.1 attached to this entry.

 

The first points seems to suggest that the beam is collimated around 2.2 mm.

However the last points are quite differents in size.

This might come from the fact that the FC unlocked several times during this measurement.

We started from a quite good alignment (1.2V transmission) to end with a poor one (0.8 V).

Images attached to this report
929_20180802042127_gbeamreflectedofthefc.png
Comments related to this report
MarcEisenmann - 18:39, Thursday 02 August 2018 (932)

Here are the data :

 

z = [ 0 1 2 3 4 5 6 15]*2.5+40 [number of holes]+position of first measurement

wW = [2170 2192 2181 2137 2239 2183 2214 2380]/2
wV = [ 2169 2207 2216 2211 2207 2202 2112 2114]/2

MarcEisenmann - 11:46, Monday 06 August 2018 (934)

Last Friday we wanted to do this characterization again because it seems that our previous points were still inside the Rayleigh range of the beam (which should be few meters)

The solution is to add a lens and recover the beam parameters without the lens.

We faced few issues during this measurement :

Faraday Isolator

  1. The Faraday Isolator output (facing the FC) is clipping probably higher order mode which are quite large w.r.t. the FI aperture
  2. The FI PBS which reflects the beam toward the locking photodiode is a bit broken ( it is the cause of the "strange" behavior of the beam we could observe. However, with proper alignment it is possible to avoid hitting the broken part of the PBS and have a good beam ( meaning we can avoid to have black line on the beam)

For now it seems we can keep going on with this FI at least for the beam characterization.

It may be useful in the future to change this FI for a larger aperture one or at least change the broken PBS.

BS

  1. We found out that the beam was shaking a lot ( ~40 urad rms ) while performing this measurement. This is due to the fact that the BS pitch correction is close to saturation (-10V) and some peaks lead to correction saturation.

 

The solution is to move the BS pitch picomotor. Matteo also recommends to follow the GEO procedure : Every ~2,4 weeks, use picomotors to put mirrors on good positions in order to avoid this kind of saturation.

R&D (FilterCavity)
Print this report.
MarcEisenmann - 11:07, Thursday 02 August 2018 (927)Get code to link to this report
EOM telescope

Paticipants : Yuhang

 

After some troubles to design the telescope, we finally found a convenient solution :

use 2 steerings mirrors to have the beam tranmitted by the 98:2 at the nominal height (75 mm)

Install a lambda/2 and replacing one steering mirror by a pbsw : power is now 14mW which is low enough to have the beam waist inside the EOM ( the ihgher limit is 18.1 mW)

We installed a beam dump to get rid of the spurious reflection of the (wedged) pbsw.

use one f=100mm lens to obtain a good beam size inside the EOM. It was installed on a rail in prevision for some EOM induced astigmatism.

 

The last picture shows the setup before installing the EOM 4 holes away of the lens on the rail.

 

The first figure is a fit of the beam after installing the lens.

(following is comparison of W and V profiles)

difference in waist position : 4mm

difference in waist size : 1.3 um

 

The second one is a fit after installing the EOM.

(following is comparison of W and V profiles)

difference in waist position : 2mm

difference in waist size : 2 um

 

The installation of the EOM changed the beam waists sizes by 2um and beam waists positions by 2 mm

Images attached to this report
927_20180802040140_eomtelescope.png 927_20180802040151_aftereom.png 927_20180802040229_20180801150358.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 10:21, Thursday 02 August 2018 (928)Get code to link to this report
The measurement of Finesse of green mode cleaner(p-pol and s-pol)

We measured the Finesse of green mode cleaner (while p-pol and s-pol)

For s-pol, the measurement is fine. But for p-pol, the main peak overlaps with sidebands, as you can see in the attached figure 2. I take only the central part of data to do fit to avoid the influence of sideband. The result is listed as following.

For s-pol, the Finesse is 300

For p-pol, the Finesse is 36.7

Images attached to this report
928_20180802031737_spol.png 928_20180802031810_ppol1.png 928_20180802031815_ppol2.png
R&D (FilterCavity)
Print this report.
MarcEisenmann - 16:49, Wednesday 01 August 2018 (926)Get code to link to this report
New sections on the NAOJ wiki

I opened 2 new sections on NAOJ wiki :

 

"Pictures" which can be useful for external people to know what space is available on the bench for example.

It isn't yet totally working but it has already 1 picture on it.

 

"Available optics and datasheets" could also be useful (e.g for future telescope design)

R&D (FilterCavity)
Print this report.
YuhangZhao - 20:09, Monday 30 July 2018 (924)Get code to link to this report
The installation of telescope for p-pol path

During the last week and today, we did a lot of investigation about how to put lens to achieve many limitations. Today, we finally found it really diffcult to design telescope with all the condition we have now. We talked with Matteo, we got an important conclusion. It is we can decrease laser low to 30mW.

However, after measured the beam dimension, we found the beam is not like the simulation result of Jammt. We guess this comes from that the inital beam is not very collimated. We decided to decrease power further more. So we decrease power to 18mW. Then we characterize beam again. The beam is really similar with before. See Fig.1

Then we put the EOM, the position is decided as shown in attached Fig.2. After put EOM, we characterized beam again. The Fig.2 is the characterization result.

According to this result. We tried to put another lens. We can recover the beam to beam waist as 940um. See attached Fig.3. This is a little different from what we need. By putting more lens, we can have good dimension of 1000um of waist. However, more lenses cost too much space.

Images attached to this report
924_20180730130924_powercomp.png 924_20180730130929_untitled.png 924_20180730130945_lensdesign.png
KAGRA MIR (Absorption)
Print this report.
RaffaeleFlaminio - 15:17, Sunday 29 July 2018 (923)Get code to link to this report
Comment to Separate setup for noise investigation on 1310nm laser (Click here to view original report: 921)
- There is a large peak at 35 kHz when the loop is closed. Is the loop gain too high?
- The dark noise measured above 1 kHZ is not as flat as one would expect. Is this real or is it a problem with the measurement?
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 23:13, Thursday 26 July 2018 (921)Get code to link to this report
Separate setup for noise investigation on 1310nm laser
I moved another optical table in the cleanroom because I didn't have enough space to properly test the laser noise reduction.
 
I removed the laser fiber from its collimator without moving the collimator, so that, when I will put it back, I will not loose the alignment.
 
I set the laser at 200mA which is half of the typical operation, this current gives 60mW in output. I put an OD2 filter after the collimator and a beam dumper to collect its reflection. Then a non-polarizing cube beam splitter 50:50 and a 50mm lens for each beam to focus the beam on 2 identical photodiodes (DET10N and DET10N/M). On one beam I measure the beam size at different positions. To do that, I put a blade moved by the old translation stage and controlled by the old software that makes scans of the DC. On the DC channel, I sent the analog output of the PM100D power meter .
 
I found a waist of 20micron at 54mm from the lens and I put the PD at about that position. I didn't repeat the waist measurement for the other lens but I put the other PD at the same distance from the lens. Each PD is mounted on a XY micrometric translation stage, so, by watching the signal at the oscilloscope I centered the beams in the PDs (moving the PDs instead of the beam).
 
I put a T with a load resistance of 7.5kOhm for both the PDs. This gives 2.5V DC signal for one PD and 2.8V for the other. The two signals, coupled in AC, overlap on the oscilloscope and show the "jumps" that are probably the noise reason.
 
I closed the loop with a lowpass at 3Hz and gain 1000. See the spectra.
 
The noise in the out-of-loop PD reduces by a factor of 10 (20dB)
 
I also connected the PD at the lockin and after demodulation at 380Hz (with the internal oscillator) the noise significantly reduces (by about 10 times by eye).
 
Next thing to do is to place the PDs in the absorption setup and check if the absorption signal noise reduces in the same way
Images attached to this report
921_20180726151806_z00mm.png 921_20180726151813_z05mm.png 921_20180726151817_z10mm.png 921_20180726151822_z15mm.png 921_20180726151827_z20mm.png 921_20180726151832_z25mm.png 921_20180726151837_z30mm.png 921_20180726151840_z35mm.png 921_20180726151844_z40mm.png 921_20180726151848_z45mm.png 921_20180726151852_z50mm.png 921_20180726151906_after50mmlens.png 921_20180726161329_011.jpg 921_20180726161334_01.jpg 921_20180726161340_001.jpg 921_20180726161345_00.jpg 921_20180726161442_20180726inloopdet10noutofloopdet10nm.png
Comments related to this report
RaffaeleFlaminio - 15:17, Sunday 29 July 2018 (923)
- There is a large peak at 35 kHz when the loop is closed. Is the loop gain too high?
- The dark noise measured above 1 kHZ is not as flat as one would expect. Is this real or is it a problem with the measurement?
R&D (FilterCavity)
Print this report.
YuhangZhao - 21:26, Thursday 26 July 2018 (919)Get code to link to this report
Comment to Beam characterizations and EOM telescope design (Click here to view original report: 909)

At the begining, we used the wrong beam dimmsion, the initial beam DIAMETER is 2000um.

BUT, all the telescopes we designed are using RADIUS as 2000um.

Today, we realized this problem. I designed the telescope again. The EOM doesn't make a large difference. This design can be a fine reference.

Lesson: actually we have many chances to realized this problem, we checked every time after putting the lens. But everytime, we checked only the beam waist position. We never checked the beam waist size. So we didn't realize this problem. So next time we should check both of them carefully.

Images attached to this comment
919_20180727024829_opoeomt2.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 19:29, Wednesday 25 July 2018 (915)Get code to link to this report
Astigmatism after put EOM(for opo)

Participiant: Marc and Yuhang

We want to know how the beam will look like after putting the EOM. So we put EOM yesterday and checked the output of it today.

Before putting EOM: we found for the far field(more than 40cm), we can see clearly the astigmatism even with the card. See attached Fig.1

After putting EOM: we found the astigmatism disappear for the far field(the beam looks very perfect by card now, unfortunatly I didn't take picture). But we found it in near field. See attached Fig.2. You can see the waist position is really different. But for the far field, it looks fine.

Images attached to this report
915_20180725122912_2449420449222702925520180725192602.jpg 915_20180725122925_fiteomopo.png