LOG-IN
Displaying reports 2501-2520 of 3274.Go to page Start 122 123 124 125 126 127 128 129 130 End
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 12:14, Tuesday 17 July 2018 (888)Get code to link to this report
Circuit box to separate AC and DC from PD

I upload the circuit diagram of the box used in the original setup from the SPTS company. (they made it)
I'm still using it in the current setup, but since the InGaAs PD has a higher Current/Power responsivity, It could be that I should reduce the load resistance and send more power on the PD. If it helps to reduce the noise, I will make a new box for the InGaAs PD.

Images attached to this report
888_20180717051321_acdcboxdiagram.jpg 888_20180717051326_acdcboxphoto.jpg
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 11:48, Tuesday 17 July 2018 (887)Get code to link to this report
Correlation between the two PDs

Using the Yokogawa2400 spectrum analyzer I measured the correlation between the signals coming from the two photodiodes:

- PM100D (incident power 85mW)       DC 470mV, analog out
- DET10N (incident power 15microW) DC 104mW (load resistance 6kOhm) 

The signals don't look coherent. It could be that they are really not coherent, or that the spikes at high frequencies invalidate the spectrum measurement.

Images attached to this report
887_20180717044738_20180713coherence.png
R&D (FilterCavity)
Print this report.
Eleonora, Matteo B, Yuhang - 19:07, Friday 13 July 2018 (886)Get code to link to this report
Further rampeauto noise reduction

After having disconnected the potentiometer of the ramp in the actuator section of the rampeauto (see previous entry # 875) we have further modified the rampeauto to reduce its noise. Following Pierre's instructions, we have soldered a 2.2K resistor in parallel to the resistor R24, in order to reduce the gain of the PA83 amplifier and then the effect of its input noise. In practice, since the resistor R24 was difficult to access, we have put the resistor in parallel with C23. 

- The attached PDF shows the electronic scheme of the rampeauto (actuator part). 

- Figure 1 and 2 show the rampeauto noise taken at the PZT output without any input signal. We see that the rampeauto noise is reduced by a factor ~ 20 in the kHz region, as expected. 

- Figure 3 shows the error signal of the filter cavity (green) with this new configuration, compared with the old ones. Note that, since now the gain of the P83 is reduced, we should increase the gain of the loop increasing the signal at the input of the P83 with the potentiometer "PZT gain". With respect to before, the error signal at high frequency is reduced by a factor ~ 8. 

- Figure 4 shows the error signal of the filter cavity (green) compared with the noise projection of the laser frequency noise (7.5 kHz/f [Hz/sqrt(Hz]) and rampeauto noise. Both the laser frequency noise and the rampeauto noise have been filtered by the pole of the cavity (2.4 kHz) and multiplied by the open loop transfer function of the loop, which assumes a 1/f behaviour at f> 3 kHz and a unity gain frequency of 10 kHz. We see that with this new configuration the error signal of the filter cavity (green) is explained very well by the laser frequency noise. 

Note that after the modification of the rampeauto it was not possible to obtain a ugf larger than 10 kHz. This should be understood better. 

Images attached to this report
886_20180717065008_pztres.png 886_20180717081438_nn.png 886_20180717081456_nb1.png 886_20180717081504_figure1.png
Non-image files attached to this report
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 16:03, Friday 13 July 2018 (885)Get code to link to this report
Comment to Control loop investigation (Click here to view original report: 880)

I repeated the same measurements with another spectrum analyzer, the Agilent35670a

This instrument spectra are much cleaner, I did only 32 averages instead of 256 with the Yokogawa sa2400

First plot: gain 1000 in-loop pd: PM100D

Second plot: gain 1000 in-loop pd: DET10N

Third plot: gain 5000 in-loop pd: DET10N

Images attached to this comment
885_20180713091810_inlooppm100doutofloopdet10ngain1000agilent35670a.png 885_20180713091818_inloopdet10noutoflooppm100dgain1000agilent35670a.png 885_20180713091831_inloopdet10noutoflooppm100dgain5000agilent35670a.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 11:57, Friday 13 July 2018 (884)Get code to link to this report
Monitor the reflection of green mode cleaner

Participiant: Matteo.B, Yuefan and Yuhang.

Yesterday, we installed a photo dector(PD) to monitor the reflection of green mode cleaner. This PD was put in the path which is the pick off of green reflection. And the monitor point has almost all the reflected power.

Then we scan the mode cleaner by sending a ramp signal to the mode cleaner. Actually here I made a mistake which was mentioned by Matteo.L before, but I didn't pay attention. It is I use a very large ramp amplitude, which is 700mV p-p. In that case, the scan is very fast. So I cannot see the reflection peak. See attached Figure one.

Then we decrease the ramp amplitude to 100mV. And we can see the reflection peak.

We can see from the figure 2, when we are in the resonance, the reflection is around 55%.

Images attached to this report
884_20180713045311_inkedfigure1li.jpg 884_20180713045410_figure11.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 11:37, Friday 13 July 2018 (883)Get code to link to this report
Comment to Rampeauto noise reduction (Click here to view original report: 875)

We compare the green error signal spectrum before and after the disconnection of Ramp injection point in the circuit. We can see the noise level is reduced by a factor up to 3 from 2kHz.

Images attached to this comment
883_20180713043710_spectrumcomp.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 19:10, Thursday 12 July 2018 (882)Get code to link to this report
Comment to Power budget for green mode cleaner(MC) (Click here to view original report: 879)

I mwasured the power loss due to detection point is not exactly the point of reflection and transmission.

For reflection, we lose 1%(0.12mW) of power. For transmission we lose 0.8%(0.097mW) of power.

So the loss of 2.56mW cannot be explained by this.

KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 20:13, Wednesday 11 July 2018 (880)Get code to link to this report
Control loop investigation

Manuel, Matteo B, Eleonora.

I recorded a first video of the signal of the photodiodes on the oscilloscope without control loop .
Channel1 (yellow) is the out-of-loop PD (DET10N) with a load of 6kOhm that gives 104mV of DC.
Channel2 (blue) is the in-loop PD (PM100D) analog output. DC = 470mV
The noise is mostly at high frequencies. The PM100D shows some spikes, while the DET10N doesn't.

The loop consists of a subtraction of an offset from a waveform generator, then a low pass (order one) at 100Hz and a Gain of 1000 (using the sr560)

I recorded a second video around the moment of closing the loop. On the oscilloscope there are the signal of the in-loop photodiode (blue channel), and the correction signal (yellow channel).

It shows that the in loop PD increases the noise when closing the loop. This is because the noise is mostly at high frequencies and the noise reduction at low frequencies doesn't show clearly in the time domain.

Plot1:  I made the spectrum of the 2 PDs in 3 conditions: dark, loop open, loop closed
It shows that the loop works for the inloop PD, but it doesn't for the out of loop PD.
 

Plot2: I exchanged the two PDs, so in this case the out-of-loop PD is the PM100D and the in-loop PD is the DET10N.
I copied the dark and openloop curves from the plot1 and I updated the closed loop curves. In this case I could increase the gain to 5000.

Images attached to this report
880_20180711131216_inlooppm100doutofloopdet10n.png 880_20180711131315_inloopdet10noutoflooppm100d.png
Non-image files attached to this report
Comments related to this report
ManuelMarchio - 16:03, Friday 13 July 2018 (885)

I repeated the same measurements with another spectrum analyzer, the Agilent35670a

This instrument spectra are much cleaner, I did only 32 averages instead of 256 with the Yokogawa sa2400

First plot: gain 1000 in-loop pd: PM100D

Second plot: gain 1000 in-loop pd: DET10N

Third plot: gain 5000 in-loop pd: DET10N

R&D (FilterCavity)
Print this report.
YuhangZhao - 18:32, Wednesday 11 July 2018 (879)Get code to link to this report
Power budget for green mode cleaner(MC)

Participiant: Yuefan and Yuhang

Today we locked the MC and Mach-Zehnder(MZ). We found some problems while locking.

1. The lock of MC sometimes has drifting. It brings the locking point to a lower point.

2. The error signal for MZ locking has a fixed frequency oscillation.

3. Sometimes we can see the MC transmission goes away very large.

We tried to make sure the MC transmission has the same level and then make the measurement.

input(unlock) 27.51mW 0.002
reflec(lock) 11.96mW 0.002
trans(lock) 12.17mW 0.007
end mirror loss 0.814mW Nan

The second cloumn is the averange power. The third cloumn is the standard deviation normalized by the averange power.

Note: the end mirror loss is diffcult to measure, so I take only one measurement.

So reflection is 43.48%, transmission is 44.24%, end mirror tans is 2.96%. We loss 2.56mW (9.32%).

Comments related to this report
YuhangZhao - 19:10, Thursday 12 July 2018 (882)

I mwasured the power loss due to detection point is not exactly the point of reflection and transmission.

For reflection, we lose 1%(0.12mW) of power. For transmission we lose 0.8%(0.097mW) of power.

So the loss of 2.56mW cannot be explained by this.

R&D (FilterCavity)
Print this report.
YuhangZhao - 11:33, Wednesday 11 July 2018 (878)Get code to link to this report
Update of optical layout on the inner-air bench

Please refer to wiki page.

https://gwpo.nao.ac.jp/wiki/FilterCavity/OpticalLayout

You can download if you want. Refer the instruction from the attached figure.

Images attached to this report
878_20180711043743_wikiinstruc.png
R&D (FilterCavity)
Print this report.
RaffaeleFlaminio - 20:30, Tuesday 10 July 2018 (877)Get code to link to this report
Comment to Green mode cleaner trasmission issue (Click here to view original report: 850)
At the last weekly meeting doubts were raised on the measured powers.
Are the values given above now confirmed?
P_in = 22 mW
P_tra = 8.5 mW
P_ref = 10.5 mW
P_tra from end mirror = 0.5 mW
Missing = 2.5 mW
Are the 2.5 mW still missing?
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 18:43, Tuesday 10 July 2018 (875)Get code to link to this report
Rampeauto noise reduction

[Eleonora, Matteo B.]

We have measured the rampeuto PZT output noise by connecting the PZT output to the spectrum analyzer. It is about 2 uV/sqrt(Hz) at low frequency and it starts to decrease above 5 kHz, as can be seen from pic 1 (blue curve) 

We also observed with the oscilloscope that there is an offset of -10mV. (pic 2).

In order to reduce the rampeauto output noise, following the advice of Pierre, we have disconnected the ramp input (which normaly is not used for suspended cavities). (pic 3) 

The output noise has been reduced by a factor 3 and now  it is about 800 nV/sqrt(Hz),  below 5 kHz.  (see pic. 1 and 4)

In pic 5,  we have compared the rampeauto noise (actuation noise)  with the free running laser noise and found that it becomes dominant above 2 kHz.

For the conversion to Hz we used a piezo gain of 2 MHz/V, as measeured in entry  #859.

The next step is to re-measure the phase noise, as done in entry #863, with this new configuration.

Images attached to this report
875_20180710112527_pl1.png 875_20180710112640_offset.jpg 875_20180710113359_ramp.jpg 875_20180710114558_plot2.png 875_20180710114606_plcf3.png
Comments related to this report
YuhangZhao - 11:37, Friday 13 July 2018 (883)

We compare the green error signal spectrum before and after the disconnection of Ramp injection point in the circuit. We can see the noise level is reduced by a factor up to 3 from 2kHz.

R&D (FilterCavity)
Print this report.
YuhangZhao - 18:42, Tuesday 10 July 2018 (876)Get code to link to this report
Investigation about the tolerance of mode cleaner mirror reflectivity difference(error)

Participaint: Eleonora, Matteo Barsuglia and Yuhang.

As we found before, the transmissivity of mode cleaner is not as expected. We also found for different Finesse requirement, we can have different tolerance of reflectivity error.

According to the measurement result of entry871 and entry548, we firstly get the reflectivity of mirror (r) and then we use this result and scan the reflectivity amplitude from 0 to 0.13.

We put this difference to each mirror as delta/2, for example we use r_1=r+delta/2, r_2=r-delta/2. However, when we convert this amplitude difference to power difference, we bring back a factor of 2. So they balance.

We also plot the case for Finesse as 100. As the experience of Matteo Barsuglia, this Finesse can give us 99% of clean to have TEM00. And we can increase reflectivity difference tolerance dramatically. For details, refer to attahced figure.

We also investigate the reason of higher transmission with lower Finesse. The reason is we loss some of the power through the end mirror of mode cleaner. The higher the Finesse is, the more losses we have.

Images attached to this report
876_20180710114205_mcfinal.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 16:17, Monday 09 July 2018 (867)Get code to link to this report
The simulation of opto-mechanical transfer function and filter design

Participaint: Yuhang and Eleonora

It is necessary to fit the opto-mechanical transfer function and filter design, otherwise we cannot use zpk function in matlab. We did this by using proper poles and zeros in proper frequencies. 

The zeros I used (corresponding q factor): 7250,13800,500,14600,25500,42000,31000,18100,21500(25,14,0.01,4,15,40,35,12,5)

The poles I used(corresponding q factor): 7000,12000,5000,8000,14800,16500,18900,26500,30000,34000,35900,46000,44900(12,15,0.03,0.6,6,18,10,2,10,10,2,0.8,30)

The gain is 28 at 100Hz.

I tried to put a low pass filter at 200Hz(q=0) with a gain of 20. And an intergrator at 0Hz(q=0). Finally a notch filter between 6950Hz(q=6) and 7550Hz(q=10).

Then I got unity gain frequency of 376Hz with 34 degrees of phase margin. The plot is in attahced Fig. 2 and blue line is data, red line is OLTF, green line is filter.

Images attached to this report
867_20180709085953_fitoptomecha.png 867_20180709091705_oltf.png
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 20:52, Friday 06 July 2018 (873)Get code to link to this report
Fit of the actuator TF and of the open loop TF

Manuel, Eleonora

We plotted the transfer function of the measured actuator (plant) and fitted it with a Matlab script based on the zpk function.
We used two simple poles at 40kHz, a gain of 0.16 in DC, and a delay of 2.3e-6. see the first plot

We plot the measured open loop TF obtained using a sr560 set with a first order low pass filter at 100Hz, a gain of 200.
We fitted it as the product of the modeled plant TF and a filter TF.
The filter that best fits the data is a first order low pass at 100Hz  with a gain of 240. see the second plot
The UGF is at 3.8kHz with a phase margin of 76deg.

We verify that the loop becomes unstable for a gain of 2000 (as observed experimentally).
Indeed, the UGF becomes 23kHz and the phase margin 7deg. See the third plot.

We will use this model to design a better filter in order to have more gain at low frequencies, compatibly with the possible configurations of the sr560.

Images attached to this report
873_20180706133413_20180706actuatortffit.png 873_20180706133417_20180706oltffit.png 873_20180706133421_20180706oltffitgain2000.png
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 15:44, Friday 06 July 2018 (872)Get code to link to this report
open loop transfer function and intensity noise reduction

I set the loop to control the intensity of the 1310nm laser.
first I measured the actuator transfer function with a random noise from the spectrum analizer of amplitude 200mV

Then I set a low pass filter at 100Hz, a DC offset from the waveform generator to keep the correction signal around 0, and a gain of 200.
I closed the loop and measured the open loop transfer function.

I measured the noise in the photodiode with the loop closed and without loop.
then I divided the noise without loop by (1 + the open loop transfer function) and compared with the closed loop pd noise.

If I increase the gain above 1000 the laser stops for exceeding the current limit.

Images attached to this report
872_20180706084154_pm100dx2000withwoloop.png 872_20180706084159_20180706transferfunctions.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 11:48, Friday 06 July 2018 (871)Get code to link to this report
Finesse of green mode cleaner

Participant: Marc and Yuhang

Today I reveived the comment from Marc and I measured the Finesse again. This time I put the ramp signal to make sure I am looking at the correct part of signal. This is of great important to find the correct FSR(Actually here the fsr corresponding to time, so maybe not good to call it fsr)

Then I use this fsr to fit only one peak so that I can see more clearly the fit.

In the end, I got a more pleasuible result. Finesse is 381. But this result is much higher than the calculation result, which is 248.

Images attached to this report
871_20180706044759_fsr.png 871_20180706044803_ramp.png 871_20180706044809_finess.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 10:07, Thursday 05 July 2018 (869)Get code to link to this report
Realignment after tidy up in July 2018

Participant: Eleonora and Yuhang

Beccause we changed many cables and arranged the control devices space, we think the green should be misaligned. Yesterday, I checked that it is totally misaligned!

I did the procedure of standard alignment of green: check target on PR chamber, check the first iris, check the second iris and match incidence with reflection, finally end mirror. After these procedures, I got the new offset of each mirror's local control offset. Then I can lock green and infrared together. See attached Fig.1

  YAW PITCH
PR 0.3 -0.07
BS -0.11 -0.9
IM 0 1.1
EM 2.4 -2

I also found some problems and did some change during this process.

1. I connect the image of second iris to the third part of monitor. I also checked the light you can see in this second iris camera is corridor light. Because if I go to turn off the corridor light, it disappeared. See attached Fig.2

2. If you find problem in attached Fig.3, you need to reopen the program.

3. I tried to remote control of the second iris, but the network cable seems not working. 

Images attached to this report
869_20180705030510_wechatimg22.jpg 869_20180705030525_wechatimg21.jpg 869_20180705030536_wechatimg23.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 20:49, Wednesday 04 July 2018 (868)Get code to link to this report
Realignment mode cleaner and polarization check

Because we found the unexpected peak while we are scanning mode cleaner, I tried to realign it.

The mothod is to block the beam going to mirror without PZT in Mach-Zehnder. Then use the steering mirror to align and make the peak of mode-cleaner as high as possible. Then remove the block, and adjust the mirror without PZT in Mach-Zehnder.

After alignment, the unexpected peak becoms very small. So I tried to take the data and calculate Finesse. Firstly, I tried to fit with airy function. But I found the FSR here is very strange, see attached Fig.1. It's obvious that the software cannot tell us this is a good fit. So this fit is done by my hand. I don't know why this FSR can be this unstable. But maybe this can be interesting.

But I tried to fit only one peak, then I get the fit result of F=15100. See attached Fig.2. But note here, we use PD with amplification, it is 40dB means 100 times amplification. So the measurement of Finesse is only 152. If we consider R=0.992, the Finesse should be 248. So there is this discrepancy.

Beside, I checked the polarization again. This time I put a half-wave plate infront of mode cleaner, and change s to p polarization. I saw a increase of larger than 10 times of transmission on oscillscope. I talked with Matteo. From Fresnel law, p-pol has more transmissvity for the mirror now we use to dump beam. This is the main reason for this increase.

Images attached to this report
868_20180704133014_figure1.png 868_20180704133019_fitone.png 868_20180704134333_photodetector.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 09:39, Wednesday 04 July 2018 (866)Get code to link to this report
Comment to Green mode cleaner trasmission issue (Click here to view original report: 850)

I checked the input and output mirror. From the point view of marker on the side of mirror, I am sure the mirror is installed in a correct way.

I checked also the mirror from the same box, this arrow points to the HF side of this mirror.

Images attached to this comment
866_20180704023859_544263865.jpg 866_20180704023905_2125266136.jpg