LOG-IN
Displaying reports 2521-2540 of 3201.Go to page Start 123 124 125 126 127 128 129 130 131 End
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 00:06, Thursday 10 May 2018 (767)Get code to link to this report
Simulations of the bulk reference sample and the sapphire for different thicknesses

We are still investigating the reason for the calibration problem in the bulk absorption that gives a factor of 3 between the measurements done with our experiment and at LMA or Caltech on the same samples.

In order to understand how the interaction length (the crossing area of probe and pump) affects the measurement, I made some simulations of the scan changing the thickness of the samples and keeping constant the absorption/cm rate.

Images attached to this report
767_20180509170500_bulksample6338mm.png 767_20180509170508_sapphiresample6338mm3mm616mm.png
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 23:49, Wednesday 09 May 2018 (764)Get code to link to this report
Problem with the SURFACE reference sample

Before measuring the coating absorption of LMA samples and crystalline coating I wanted to test the surface reference sample in order to check if it had any damage. So I made a map of it.
The map shows a regular pattern of absorption that oscillates by a factor of 2.

I made some checks to understand the nature of this strange behavior. I report the steps in the attached pdf slides.

I'm going to look for a better alignment

Images attached to this report
764_20180509132306_surfrefproblem.png 764_20180509164639_43.png 764_20180509164644_30.png 764_20180509164648_44.png 764_20180509164655_26.png 764_20180509164701_35.png 764_20180509164706_45.png 764_20180509164714_00.png 764_20180509164720_57.png 764_20180509164725_53.png 764_20180509164730_18.png 764_20180509164738_36.png 764_20180509164809_41.png 764_20180509164813_43.png
Non-image files attached to this report
R&D (FilterCavity)
Print this report.
MarcEisenmann - 17:55, Wednesday 09 May 2018 (765)Get code to link to this report
Alignment of the collimator

Participants : Marc, Yuefan

 

To have a lower beam divergency (so easier working conditions) we installed a lens f=200mm 6 cm after the output of the auxiliary laser 1.

We also increased the laser power to 44.3 mW (0.9 A for the pump).

By carefully aligning the lens and the collimator vertical and transversal position (X and Y) we obtain a good aligment on this 2 directions.

As the beam is quite diverging after the collimator, this was done checking the beam positions around 20 cm after the collimator.

 

By unscrewing the 3 screws moving the Z position of the collimator (its distance to the laser),

we reach a power of 44.1 mW at the output of the collimator meaning 99.55% transmission of the collimator.

This seems good enough to go on the others steps (installation of the fiber, add another collimator at the output of this fiber and check the beam size at its output).

We also found another holder for the second collimator meaning we have all the needed components for the next steps.

R&D (FilterCavity)
Print this report.
YuhangZhao - 11:17, Wednesday 09 May 2018 (763)Get code to link to this report
Comment to Comparison of old and new servo (Click here to view original report: 760)

Here I attach the rms integration of the four error signal curves.

Images attached to this comment
763_20180509041738_rms.png
R&D (FilterCavity)
Print this report.
MarcEisenmann - 23:33, Tuesday 08 May 2018 (762)Get code to link to this report
Auxiliary Laser 1 installation and pre-alignment of the collimator

Participants : Marc, Yuefan

 

Auxiliary Laser 1

Today we installed the auxiliary laser 1 on the squeezer bench.

Compared to the lens on the IR path far from the FC, its output port is at 14 holes (vertically toward FC) and 4 holes to the left (toward the main) laser.

This position seems convenient to avoid to twist its power supply while letting some space for the future optics to be installed.

Using 2 mm spacers, the beam height at the output of this laser is 7.5 mm.

 

Pre alignment of the collimator

Because of the divergency of the laser and the divergency of the not-aligned collimator, the beam at the output of the collimator is quite diverging.

In order to follow the alignment proposed by Thorlabs, we need to measure the output power of the collimator and incremantally find its maximum.

The powermeter had to be placed few cm after the collimator making this task quite painfull...

We could reach an output power of 10.35 mW (compared to the 11.3mW of the laser). However, a sad mis fixation of the collimator post made it moved quite a bit.

Tomorrow we will add a converging lens between the laser and the collimator in order to have a lower divergency of the beam and better work condition.

Images attached to this report
762_20180508163247_20180508172616.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 23:03, Tuesday 08 May 2018 (761)Get code to link to this report
Mach-Zehnder installation simulation

After the test of Mach-Zehnder, we need to install it in reality. But we change the original design. Thanks to the optocad code of tomuta-san, I revise it and do the simulation.

 

We have considerations:

1. The stable lock of filter cavity. We decided to use MZ only for the Mode cleaner and OPO. 

2. The focal length of lense is limited. So I asked yuefan to give me the list of lense we have. I select focal length of 50mm and 75mm

3. We need to adjust the position of these two lenses to have a good mode matching.

 

The change can be seen in the attached picture 1. We will start to install it tomorrow.

Images attached to this report
761_20180508160323_benchchange.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 19:19, Tuesday 08 May 2018 (760)Get code to link to this report
Comparison of old and new servo

The result of previous comparison is wired. So we decide to change back the old servo and measured error signal again. At the same time, we also did calibration again.

Note here, the calibration method for green likes entry 750. The only difference is that we consider frequency much higher than unity gain frequency. Then we got result like attached picture 1 and 2.

The calibration for infrared got from attached picture 3 and 4. For the problem of saturation, we changed the demodulation phase.

Finally we got result as attached picture 5. We can make them overlap by multiplying a factor of 11 like picture 6.

As a result, we find the new servo make noise level lower than before. So now I put back the new servo, we can get the green transmission as 1.5V and very stable like picture 6.

Images attached to this report
760_20180508121118_greenca.png 760_20180508121133_greencafinal.png 760_20180508121341_calibrationinfrared.png 760_20180508121357_infraredca.png 760_20180508121525_comparison.png 760_20180508121538_overlap.png 760_20180508121949_721319689.jpg
Comments related to this report
YuhangZhao - 11:17, Wednesday 09 May 2018 (763)

Here I attach the rms integration of the four error signal curves.

EleonoraCapocasa - 08:04, Monday 14 May 2018 (773)

The error signals for green and infrared account for the closed loop laser frequency noise filtered by the pole of the cavity  (which is different for green and IR).

In the fist attached plot, the error spectra has been divided for the corrispondig pole in order to go back to the close loop laser frequency noise.

freq nois =  err sig * ( sqrt( 1+(f/f0)^2)     with f0 = 55 Hz for IR and 1.45 kHz for green

The two curves obtained shoud be coincindents. The discrepancy (about a factor 2.5) suggests that there is maybe an issue with the calibration.

YuhangZhao - 21:45, Wednesday 16 May 2018 (780)

Corresponding to the comment of Eleonora, the bandwidth of filter cavity for infrared is 114Hz but not 55Hz. Then I think we can explain the result (almost).

bandwidth=FSR/Finesse=500000/4355=114

Matteo Barsuglia - 05:58, Friday 18 May 2018 (782)

I think there is a factor 2 missing in the formula: the pole of the cavity is FSR/(2*F) = 500000/(2*4355) = 57 Hz

R&D (FilterCavity)
Print this report.
EmilSchreiber, YuhangZhao - 10:06, Tuesday 08 May 2018 (759)Get code to link to this report
Mach-Zehnder control loop design and testing
We finished and tested the electronics for the Mach-Zehnder controller. The loop performs well with a control bandwidth of 3.2kHz. We will need to further investigate whether the resulting amplitude stability of the transmitted green beam is sufficient.
 
The servo electronics are designed to give a simple 1/f open-loop transfer function. We use the fact that the PZT driver already includes a low-pass filter with a corner frequency of 77Hz (see entry 585). Additionally, the servo includes a low-frequency integrator to give infinite gain at DC, which can be switched off for lock acquisition.
 
We found that a mechanical resonance at 12.4kHz was limiting the achievable unity-gain frequency. Including a low-Q notch filter for this frequency allows for a slightly higher UGF, but it is then limited by another mechanical feature at 4.9kHz. There is yet another instability at around 600Hz, but this only occurs if the gain is set too low and can thus be completely avoided.
 
I measured noise spectra of the error signal for the in-loop case, free running noise (set to the same operating point by hand) and dark noise (green beam path blocked). The residual in-loop noise is dominated by features at 600-1000Hz. At these frequencies there is currently not much loop gain to suppress the noise. The dark noise is surprisingly high, in particular at low frequencies. We will have to look into this further.
 
Things to check:
  • Can we identify the 4.9kHz resonance and maybe reduce it mechanically (e.g. by tightening screws)?
  • Are the 600Hz noise features actual amplitude fluctuations coming from the SHG or are they added by the Mach-Zehnder? (This can be tested by putting the PD before Mach-Zehnder.)
  • Is the low-frequency dark noise caused by ambient light or electronics? Can it be reduced?
- Is the low-frequency dark noise caused by ambient light or electronics? Can it be reduced?We finished and tested the electronics for the Mach-Zehnder controller. The loop performs well with a control bandwidth of 3.2kHz. We will need to further investigate whether the resulting amplitude stability of the transmitted green beam is sufficient.
 
The servo electronics are designed to give a simple 1/f open-loop transfer function. We use the fact that the PZT driver already includes a low-pass filter with a corner frequency of 77Hz. Additionally, the servo includes a low-frequency integrator to give infinite gain at DC, which can be switched off for lock acquisition.
 
We found that a mechanical resonance at 12.4kHz was limiting the achievable unity-gain frequency. Including a low-Q notch filter for this frequency allows for a slightly higher UGF, but it is then limited by another mechanical feature at 4.9kHz. There is yet another instability at around 600Hz, but this only occurs if the gain is set too low and can thus be completely avoided.
 
I measured noise spectra of the error signal for the in-loop case, free running noise (set to the same operating point by hand) and dark noise (green beam path blocked). The residual in-loop noise is dominated by features at 600-1000Hz. At these frequencies there is currently not much loop gain to suppress the noise. The dark noise is surprisingly high, in particular at low frequencies. We will have to look into this further.
 
 
Things to check:
- Can we identify the 4.9kHz resonance and maybe reduce it mechanically (e.g. by tightening screws)?
- Are the 600Hz noise features actual amplitude fluctuations coming from the SHG or are they added by the Mach-Zehnder? (This can be tested by putting the PD before Mach-Zehnder.)
- Is the low-frequency dark noise caused by ambient light or electronics? Can it be reduced?
Images attached to this report
759_20180508030432_transferfunction.png 759_20180508030444_noisespectrum.png 759_20180508030451_machzehnderservocircuit.png
Comments related to this report
EleonoraCapocasa - 10:46, Friday 20 July 2018 (897)

We verified that the noise feature at 600 Hz which appears in many error signal spectrum: SHG (entry #620), MZ (entry #759), IR FC error signal (entry #750) and some power spectrum (entry #772) is coming from the turbo pump of the BS in the central bulding. 

R&D (FilterCavity)
Print this report.
YuhangZhao - 19:38, Saturday 05 May 2018 (757)Get code to link to this report
Comment to More activity of cavity control loop (Click here to view original report: 756)

It seems like the peak of infrared error saturates on oscilloscope from picture 3. Maybe we can put an attenuator for it?

R&D (FilterCavity)
Print this report.
YuhangZhao - 21:23, Thursday 03 May 2018 (756)Get code to link to this report
More activity of cavity control loop

Participaint: Emil and Yuhang

After change the new servo and new infrared demodulation board, we did a rough phase adjustment. We decided to make it as best as possible, so we changed the green and infrared demodulation phase. 

For infrared phase, we change the demodulation phase and find a really small error signal. Then we add 90 degree to get a good phase. However, we cannot get a clear green error signal and we cannot see the difference when we change the phase. For the green locking, we did like this:

1. We lock green.

2. Change the phase until we can see the oscillation of error signal.

3. Decrease the gain till oscillation disappears

Because the gain of SR560 is 1 now, so we connect the demodulation signal directly to Rampotu servo. Now the gain is 4.5 on the Rampeauto board.

After change the demodulation phase, we measured open loop transfer function and error signal again. I put the result here.

(Note: this time I also attach the error signal before calibration)

Images attached to this report
756_20180503142052_error.png 756_20180503142124_beforecalibration.png 756_20180503142706_infraredfinal.png 756_20180503142733_green.png
Comments related to this report
YuhangZhao - 19:38, Saturday 05 May 2018 (757)

It seems like the peak of infrared error saturates on oscilloscope from picture 3. Maybe we can put an attenuator for it?

R&D (FilterCavity)
Print this report.
MarcEisenmann - 13:58, Wednesday 02 May 2018 (755)Get code to link to this report
Comment to Simplified Optical Scheme (Click here to view original report: 754)

Here is the 2nd version :

 

Changes :

OPO, homdyne and squeezed vacuum beam path have been added

Images attached to this comment
755_20180502065828_tamasimplifiedv2.png
R&D (FilterCavity)
Print this report.
MarcEisenmann - 15:57, Tuesday 01 May 2018 (754)Get code to link to this report
Simplified Optical Scheme

For the article preparation as well as for future meetings, it could be useful to have a simplified optical scheme of the experiment.

 

Here we tried to do one following Oelker example ( Audio-Band Frequency-Dependent Squeezing for Gravitational-Wave Detectors  ).

In the scheme attached to this entry is a preliminary scheme.

The "Not Yet installed" part should contain OPO PLL (?) and homodyne readout.

We haven't yet put the OPO, the PLL and the homodyne readout. The question being how to add them without making the scheme to difficult to read.

All the other main components are indicated.

Maybe it could be also useful to add the control loop?

Images attached to this report
754_20180501085537_tamasimplified.png
Comments related to this report
MarcEisenmann - 13:58, Wednesday 02 May 2018 (755)

Here is the 2nd version :

 

Changes :

OPO, homdyne and squeezed vacuum beam path have been added

R&D (FilterCavity)
Print this report.
MarcEisenmann - 20:38, Friday 27 April 2018 (753)Get code to link to this report
Losses Measurement

Participants : Yuefan, Yuhang

 

When we tried to recover the FC lock, we had to act quite a lot on the BS control.

The pitch was saturating below -0.7 so we had to play with BS and IM yaw in order to reduce the saturation on the BS control while keeping a good beam position.

[ When the IM is misaligned it is really difficult to see the green transmitted beam because of another beam splitter has been installed on the green path in the squeezed bench]

We could finally performed a losses measurement still using the lock/unlock technique which gives us : 60.4 ppm +/- 7.3

With 2.54% misalignment and 0.25% mode-mismatching.

 

We also plotted the SHG stability over 1 000 s ("shgstability.png")

It seems to be quite stable around 1.5V even though some low frequency variations can be seen.

The last part from around 750s corresponds to the time we started to try to lock the FC.

It seems that some of the light came back towards the SHG.

Images attached to this report
753_20180427131231_lockunlock.png 753_20180427131241_april27th2018.png 753_20180427133252_shgstability.png
R&D (FilterCavity)
Print this report.
MarcEisenmann - 19:43, Friday 27 April 2018 (752)Get code to link to this report
Comment to Losses Measurements (Click here to view original report: 751)

When the "noisy measurement" lock was performed I forgot to check if the error signal was at 0 ...

R&D (FilterCavity)
Print this report.
MarcEisenmann - 15:36, Friday 27 April 2018 (751)Get code to link to this report
Losses Measurements

Participants : Eleonora, Emil, Yuhang

 

We performed 3 new losses measurements using the "lock-unlock" techique.

The IR reflected power can be seen in fig "23to26.pdf".

To extract the losses value for the measurement made on April 23d 2018  I only extracted the firsts values to avoid the noisy part of the signal.

The April 26th measurement was really noisy. However, 3.69% of the power were coupled to the FC 1st order mode and 2.49% with the 2nd order.

In order to better investigate the source of this noise, we will add a beam splitter and a photodiode on the IR injection path in order to see if there is any coupling between the laser power fluctuations and  the IR reflected power or if we should investigate other sources.

 

It also seems that the IR reflected power fluctations are higher when the frequency of the AOM is such that the IR 0 order mode is resonant in the FC.

However, if we change slightly this frequency, the fluctuations seems to disappear.

We will install a camera on the IM viewport where a 2" mirror was installed to see if we can extract some informations about the scattered light.

 

The computation of the 3 measurements can be seen on "23to26meas.png".

The noisy measurement leads to the huge error bar.

Images attached to this report
751_20180427082925_23to26meas.png
Non-image files attached to this report
Comments related to this report
MarcEisenmann - 19:43, Friday 27 April 2018 (752)

When the "noisy measurement" lock was performed I forgot to check if the error signal was at 0 ...

R&D (FilterCavity)
Print this report.
YuhangZhao - 12:04, Friday 27 April 2018 (750)Get code to link to this report
Comparison between old and new green/infrared error signals

After change the new servo, we tried a lot to make it work well. Now we can get a stable operation by using this new loop. So it's time to characterize our new locking.

The green calibration factor was got in this way:

1.Measure the ratio between the point before servo(Y) and the point after servo(X). Actually the second point is PZT monitor, so we need to multiply PZT monitor by 100 to get X.

2.According the loop flow chart, we can present Y/X by using transfer function of plant and filter. And we know the laser will actuate with 10^6 V/Hz. The SHG gives a factor of 2. We also measured the open-loop transfer function and we call it G. You can refer to attached picture 1, the blue line is Y/PZTmonitor. We can see from the phase, only high frequency has good shape. So we trust only high frequency and we use it for the calculation afterwards.

3.Now we can get the the correction of PZT(in other word, the slope of error signal). The unit of it is V/Hz. The equation to get it is K=(Y/(X*100))*(1+abs(rho(G)*e^(-iphi(G))))*sqrt(1+f^2/f_0^2)/10^6/2

The result is calibration factor(green)=1/K=552 Hz/V.(Here f is frequency, f_0 is the pole of filter cavity)

 

The infrared calibration factor was got in this way:

1.We give a tri-angular modulation for AOM, it is pk-pk 4000Hz, T 5s. This means 1600Hz/s.

2.We save the error signal of infrared, we calculate the pk-pk of this error signal. Then we divide it by their corresponding time. Let's say the result is slope(V/s). The calibration factor is 2*slope(V/s)/1600(Hz/s)

The result is calibration factor(infrared)=13.33 Hz/V

 

Finally, we get the error signal.

Images attached to this report
750_20180427050409_comparision.png 750_20180427165334_calibrationinfrared.png 750_20180427165409_calibration.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 09:36, Tuesday 24 April 2018 (749)Get code to link to this report
Mach-Zehnder interferometer testing

Participant: Emil, Eleonora, Marc, Yuefan,Matteo

Yesterday, We used the green light to test the Mach-Zehnder. After giving a ramp to its PZT, we can get the picture shown as below. After adjusting, we can maximize it to this case shown in the picture. The contrast is 4.04/4.72=85.6%

Images attached to this report
749_20180424023612_541970613.jpg
R&D (FilterCavity)
Print this report.
PierrePrat - 19:22, Wednesday 18 April 2018 (748)Get code to link to this report
Things to do
Matteo, PIerre, Yuefan and Yuhang

We discussed about the possible interaction between the SHG control loop and the Filter Cavity control loop to explain the 6.56 kHz oscillation at the error signal of the FC control loop.
If noise appears on the SHG control loop, the power level of the green beam will vary.
The gain of the FC control loop is power beam dependent and can vary.
As the FC control loop is controlled to 0 crossing of the error signal, a variation of gain has a 2nd order influence.

But if the mixer produces an offset, this influence increases with the level of this offset.

Things to do:
0) Measure the FC loop offset (at EPS1 divided by 26.5)
1) Inject a DC into PERTURB to see the influence on the oscillation
2) Measure noise in SHG without and with FC control loop
3) Match the impedance (50 Ohm) at output of the LP filter (set at the output of the FC demodulation miser)
4) Measure LO and RF levels on FC demodulation mixer
5) Remove cables between LP filter and the input of the Stanford Research SR560 amplifier (connect directly Mixer, filter and SR560 without cable)

R&D (FilterCavity)
Print this report.
PierrePrat - 18:52, Wednesday 18 April 2018 (747)Get code to link to this report
Synoptic of the "Rampeauto" servo
The attached image file shows a synoptic of the "Rampeauto" servo.

The gain of the input gain G1 has changed and is set to G1 = 5.1

So the output EPS1 (AC2) shows the input signal with a gain G4 = G1 * G2 = 5.1 * 5.2 = 26.5

If you measure the offset at EPS1 output, it should be divided by 26.5 to obtain it at the input (Detect Sig).


Remark:
The attenuator "Gain Piezo" has an effect on the dynamic of "Piezo Sig" output.
For instance, if "Gain Piezo" is set to position 0.7 as currently (on 10 max), the dynamic will be:

D = +/- 0.7/10 * 14 * 10 = +/- 10V

(Gain Ampli HT = 10)
Images attached to this report
747_20180418112757_20170303140317.png
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 18:06, Wednesday 18 April 2018 (746)Get code to link to this report
Preparation for crystalline coating measurement

The sample we are going to measure is 0.5mm thick and 2inches diameter large. Our mount has a grub screw that is not able to secure such a thin sample. So I had to add a thick ring behind the sample. I did a mounting test with a "not for use" (not polished) sapphire piece with the same dimensions as the sample. I add a ring (which is the 1.5-to-2 inches adaptor), leaned the ring on the sample, and fixed the ring with the grub screw. The ring doesn't push the sample, it only touches it. This reduces the risk of breaking the sample.

In order to test the IR probe on the LMA samples I previously cleaned them and applied the first contact.

I checked the crystalline coating sample (the one on silica substrate) under a powerful green light. I carefully blew some spray air on it to remove the dust and the best I could do is showed in the picture.

To check which side is the coated one, I put the sample on a clean tissue on the table, and I looked at the shadow below the border of the coating. We the shadow is larger, the coating faces up.

Images attached to this report
746_20180418110601_img20180418155537.jpg 746_20180418110613_img20180418161238.jpg 746_20180418110616_img20180418161246.jpg 746_20180418110622_img20180418155544.jpg 746_20180418110629_img20180418165116.jpg