LOG-IN
Displaying reports 2641-2660 of 3244.Go to page Start 129 130 131 132 133 134 135 136 137 End
R&D (FilterCavity)
Print this report.
YuhangZhao - 01:00, Tuesday 20 February 2018 (669)Get code to link to this report
Characterization of transmission stability
Today I follow the procedure of simultaneous lock of IR and GREEN. After locking, I took the data from the oscilloscope.

I use the data to calculate the standard deviation of IR transmission and GREEN transmission. For green, it is 0.0286. For IR, it is 0.1235.

The result is also shown as attached figure. Next step is to figure out the reason of instability.
Images attached to this report
669_20180219165936_iranderror.png 669_20180219165945_granderror.png
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 10:34, Monday 19 February 2018 (668)Get code to link to this report
Preliminary measurement of filter cavity round trip losses and decay time

Participants:  Eleonora, Matteo L., Raffaele, Tomura

In the past days we kept working on the characterization of the IR lock, with the main purpose of measuring the cavity losses and the decay time.

ROUND TRIP LOSSES

We observed that in good alignement condition (IR trasmission above 1.5 V), the fluctuations of the transmitted and reflected power were much less than what observed before.

In this condition we were able to measure a change in the cavity reflectivity when the cavity is resonant and when it is not and give a preliminar estimation of the round trip losses (RTL)

In Fig.1 the trasmission and the reflection of the IR are shown when a set of lock/unlocks of the cavity was done. The reflected light has been focused on a photodiode using a lens with f = 50 mm. Et the beginning of the measurement the IR light has been blocked to measure possible offsets of the photodiodes.

The technique used to switch from resoant to not resoant state was to suddenly change the driving frequency of the AOM of 5 kHz. By using the values of the reflected power in the two states (resonant and not resonant) as explained in detail in the attached pdf we estimated the RTL to be about 80 +/-12 ppm, corresponding to 0.26 ppm/m. The error is mainly do to the residual fluctuations of the refected power when the cavity is locked. The associated squeezing degradation is reported in FIg 2.

The presence of light not coupled in the cavity (mismatching/misalignement) normaly reduces the measured losses and has to be compensated in order to have a real estimation of them. In the previous calculation I assumed a mismatching of 15%. 

[An idea of the impact of the mimastch compensation: assuming no mismatching the computed losses are 70 ppm while with 20% of mismatching they becomes 85 ppm. (Details about this can be found in my thesis at pag.101)]

According to the simulation we expected about 55 ppm of RTL  (40 ppm from flatness, 10 ppm from rougness/point defects and 5 ppm from trasmission and absorpition). Note that losses from small angles scattering (between mrad and few degrees) have not been considered in this loss budget.

DECAY TIME and FINESSE

A preliminar estimation of the decay time has also been done. To do that we used different tecniques: bringing the cavity suddenly out of resonance (by stopping the lock with the servo or changing the AOM driving frequency) or cutting the light in input. The transmission and reflection in this 3 cases are reported in the second pdf attached) 

A fit of the transmitted power for the first measurement shows a decay time of 0.0027 s, corresponding to a finesse of 4250  (Finesse = pi*FSR*decaytime ). See third figure attached.

Assuming the nominal reflectivity of the mirrors, this value is compatible with RTL of about 100 ppm.

A better analysis of the decay time, with an estimation of the error bars will be done soon.

Images attached to this report
668_20180219023529_rlt1602.png 668_20180219023610_80ppm.png 668_20180219023648_decayfit.png
Non-image files attached to this report
R&D (FilterCavity)
Print this report.
AkihiroTomura - 07:46, Tuesday 13 February 2018 (666)Get code to link to this report
Characterization of Aux lasers and Faraday isolators

Participants: Eleonora, Tomura

We conducted characterization of  two auxiliary lasers and faraday isolators.

For two aux. lasers, beam dimension evolution along propagation were measured. Laser power during measurements were approx. 260 mW, which corresponds to 1.2A in laser current.  Attached figures shows bean waist sizes and position. The origins of x-axes were set at aperture of laser housing. The waist positions were somewhat different from what expected.

The beam polarizaiton purity was confirmed with half/quarter waveplate. It was 96.7% and 99.9% for aux1 and aux2, respectively.

Beam transmission through FI was also measured using aux laser as a light source. Pictures of two FIs were attached.

For first FI (thorlabs), transmitted power was 213 mW out of 265 mW, which means 80.4%.

For second FI (Gsaenger), it was 79%.

Images attached to this report
666_20180212140055_20180212auxlaser1.jpg 666_20180212140102_20180212auxlaser2.jpg 666_20180212140240_fi1.jpg 666_20180212140251_fi2.jpg
R&D (FilterCavity)
Print this report.
MatteoBarsuglia - 21:13, Sunday 11 February 2018 (664)Get code to link to this report
Comment to IR and green beams both resonant in the cavity (Click here to view original report: 661)

Great!

R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 20:43, Sunday 11 February 2018 (661)Get code to link to this report
IR and green beams both resonant in the cavity

Participants: Eleonora, Raffaele, Matteo L.

We have installed in the end bench a thorlab passband filter (FL1064-10) on the path of the IR beam after the harmonic beam splitter. This allowed us to get rid of the residual green light and finally to monitor the beaviour of the IR light when the cavity is locked.

By changing the driving frequency of the AOM installed on the green path we were able to induce a frequency shift between the the green and the IR frequency in order to have both resonant at the same time.

NOTE THAT: since the AOM is put on the green path, the change in the frequency which it induces is compensated by the servo with a change on the IR which is half of the frequency change in the AOM. This means that a shift of 1 MHz in the driving frequency of the AOM corresponds to a shift of 500 kHz in the frequency of the IR light. 

In the following are reported some preliminary results that we were already able to obtain:

TEM00 resonance and FSR:  The resonance frequency of the TEM00 has been found at about 109.366 MHz. (The standard driving frequency of the AOM is 110 MHz). We have verified that it occurs every time we shift the AOM frequency of 1 MHz, meaning that the FSR is 500 kHz, as expected.

Cavity finesse for IR: We did a rough estimatin of the IR linewidth by slightly changing the AOM frequency in order to scan the IR resonance. We drove the AOM in order to find the maximum in the transmission (about 1.5 V) and then we checked the frequency shift necessary to get half of the maximum. Repeating this procedure few times, we found a FWHM of about 110 Hz +/- 10. This corresponds to a finesse of about 4500 +/- 450, which is consistent with what we expect. The big error is due to a large fluctuation in the trasmitted power.

HOM resonance frequency: The first order modes have been found at a AOM frequency of 108.970 MHz (and FSR multiples). It means that the the difference in the resonance frequency betwen the fundamental and the first HOM is about 198 +/-1 kHz. This value is in very good agreement with what expected, considering the cavity g-factors computed by using the RoC values measured at LMA.

According to LMA measurement, we have R_in = 436.7 m  and  R_end = 445.1 m which correspond to a frequency shift given by

δν(FSR/pi)* acos (sqrt (1-L/R_in)*(1-L/R_end) = 198.1 kHz

We also observed that different modes of the first order resonate at slightly different frequencies, which is likely to be an effect due to the astigmatism. More investigations have to be done. 

Residual misalignment: by comparing the trasmitted power of the funtamental mode with that of the first HOM we estimated the misaligment to be between 20% and 25%

Lock accuracy: Since we don't have an error signal for the IR lock, it is not easy to evaluate the lock accuracy. Anyway the fact that resoanance can be scanned by shifting the AOM frequency gives us a lower limit that is surely better than few tens of Hz, that is much better than the accuracy on the green lock, estimated to be about 100 Hz. This is compatible with the hypothesis proposed by Matteo B, according to which the IR pole (at much lower frequency than the green one) contributes to filter the high frequency laser noise, improving the lock accuracy.

Stability: Once put on resonace the IR seems to be stable as long as the lock is mantained, anyway we observed that when the cavity unlocks and relocks it is not always possible to get the IR resonant again. This is probably due to the fact that only half of the green locking points are also good locking points for the IR.

The attached picture shows the transmitted beam at resonance for the IR (top screen) and green (bottom screen)

Images attached to this report
661_20180211123431_irgreenlocked.jpg
Comments related to this report
MatteoBarsuglia - 21:13, Sunday 11 February 2018 (664)

Great!

R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 20:31, Sunday 11 February 2018 (662)Get code to link to this report
Realignment of the green beam and spikes issue (solved) on PR yaw

Participants: Eleonora, Matteo L.

Last Thurday, while measuring the hight of some optical component on the bench, the green beam got a bit misaligned. When we realigned it, we observed that the trasmission when the cavity is locked is higher than usualy (1.5 V). The beam seemed still centered on the cavity mirrors and the IR is still well aligned. 

We observed that the references of the PR and in particualr BS have slightly changed in yaw and we marked a new one.

At some point we had a problem of spikes with PR yaw error signal. We observed that this was the only one very far from zero. The spikes disappear as soon as we put it backto zero. 

R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 10:40, Thursday 08 February 2018 (659)Get code to link to this report
Optimization of SM2 position for IR alignment

Participants: Tomura, Eleonora

As reported in entry 654, we had to move the yaw of SM2 dicroic mirror in PR chamber to avoid the IR reflection to touch the viewport side. Since after the last 1000 steps done, we were not able to get a good IR alignent anymore, yesterday we moved back SM2, hoping to find a good postion where we could have both a good alignement and the maximum of the power in reflection. 

The input power is about 16 mW.

The situation is such that in the original position (before last thurday) the aligment can be very good but the beam hits the viewport side and the reflected power is very low (less than 1 mW). In order to have the maximum reflected power possible (about 12.5 mW) SM2 has to be moved so much and it is not possibile to recover a good alignent. (We remarked that in this case the cavity seems very mismatched, but we where not able to improve the situation by moving the last IR lens on the bench).

Today we tried to put back SM2 towards the original position and to fine-tune its position in order to have the best compromise between good alignement and high reflected power.

We did the followoing series of steps (velocity 500)

-300, -300, -300, -1000, -1000, -1000, +1000, +1000, +500, +300, -300, -300, +200, -100

In the end, we stopped in a position where we can have quite a good alignement (but not the best possible) and about 11 mW of reflected power. By looking with the camera it is not evident that the reflection is touching the viewport side.

We also moved the last IR lens back, from 20 mm to 17 mm.

In the attached files we recoreded the new reference for the green beam: the white cross corresponds to the orginal position and the black point corresponds to the maximum displacement done by SM2.

Considering the maximum reflection we could get (12.5 mW over 16 mW), we estimated the injection losses in vacuum to be about 22% (11% single pass) which is not much higher than we expected.

Images attached to this report
659_20180208024214_s97468421.jpg
Non-image files attached to this report
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 00:19, Thursday 08 February 2018 (658)Get code to link to this report
3 beams alignment, profiles, and references scan

We measured the beam profiles of all the 3 lasers we have. See the plot

We aligned them at the cross point and maximized the signal using the surface reference sample.

Then we made a scan of the surface and the bulk reference samples.

Images attached to this report
658_20180207155226_3profile.png 658_20180207161824_1310nmprobesurfref.png 658_20180207161829_henesurfref.png 658_20180207161835_henebulkref.png
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 22:56, Tuesday 06 February 2018 (656)Get code to link to this report
Preliminary test of AOM operation with the green beam lock

Yesterday I did a very preliminary check of the AOM operation.  After locking the filter cavity I have changed the driving frequency of the AOM (which is normaly set at 110 MHZ) and I have looked at the correction sent from the servo to the laser piezo.

In the attached picture you can see in yellow the trasmission of the filter cavity and in blue the laser correction (from PZT mon).

I changed the AOM driving frequency in steps of 100 kHz each time. Most of the time this change causes a very short unlock of the cavity, as it can be seen from the picture. Nothing can be noticed in the correction signal.

Since the piezo calibration is 2 MHz/V and the channel PZTmon has an attenuation of 100, a change of 100 kHz should result in a change of 2 mV in the correction, which is likely not visible.  I tried to do bigger steps but the cavity loses easly the lock.

I observed that the correction is about a factor 3 bigger than that measured this summer. (See last picture of entry 475)

Images attached to this report
656_20180206145601_aompreliminarytest.jpg
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 22:54, Tuesday 06 February 2018 (654)Get code to link to this report
Yaw picomotor of SM2 moved to try to recover IR reflected beam

Participants: Raffaele, Eleonora 

As reported in entry 649, while improving the IR aligment we realized that the power of the reflected beam was very small.  After some investigations we were able to see with the IR camera that the reflected beam was hitting a side of the inner part of the viewport. in order to avoid this we decide to slightly change the positon of the diachroic mirror SM2 (where green and IR recombine) in order to change the way the beam passes through the faraday.

We moved the yaw picomotor of SM2 several times (velocity 500, direction +), trying to recover the IR aligment with the two steering mirrors on the bench after each picomotor movement.

We did the following series of steps:

10, 10, 100, 300, 300, 600, 600 (Thursday) and 600,1000 (Friday)

Before the last movement (1000 steps) we were able to have a quite good alignment and a reflection between 11 and 12 mW  (the input was about 16 mW), nevertheless we were still able to see some IR scattered light inside the viewport.  After the last movement (1000 steps) the scatterd light was no more there but we were not able to recover a good aligment.

Sometimes, while recovering the aligment we had the impression that the matching was worsened (more LG modes), so we have changed the postion of the last IR lens different times in a range from 7 mm (original postion) to 22 mm.

Note that since we moved SM2, one of the two green references (the reflection from SM2) has changed accordingly. See attached picture.

Images attached to this report
654_20180206145011_changerefgreen.png
General (General)
Print this report.
EleonoraCapocasa - 22:27, Tuesday 06 February 2018 (655)Get code to link to this report
Removal of the optics from the optical table down the stairs in TAMA central building

Participants: Manuel, Eleonora, Tomura

Today, after asking Takahashi-san, we removed most of the optics intalled on the optical table placed at the bottom of the stairs in TAMA central building. There were three lasers ( two 532 and one 1064) and many other optics which have been temporarily stored in a plastic box below the table. Currenty the table is being used to test the two auxiliary lasers and faraday isolators for the filter cavity experiment.

I attach some pictures taken to the table before we removed the optics.

Images attached to this report
655_20180206142436_tableb1.jpg 655_20180206142448_tableb2.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 16:50, Monday 05 February 2018 (652)Get code to link to this report
Comment to SHG characterization (Click here to view original report: 645)
Yuefan told me that it is not SEMITEC 103JT-050, but SEMITEC 103JT-025. Their difference is only the shape.
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 10:09, Thursday 01 February 2018 (649)Get code to link to this report
More work on the IR alignment

Participants: Yuhang, Tomura, Eleonora

Today we kept working on the alignemet of the infrared beam:

1) We have centered the green beam on the end mirror. To do this we have made a scan in pitch and yaw with the BS in order to reach the edge of the mirror and see the green beam on the coils. We recorded the corresponent offsets and pick the intereidiate value as the good one. 

2) We have investigated the origin of the bright halo visible on the camera for the IR trasmission (see pic1). We have looked insiede the end bench putting a camera in the viewport behind the end mirror (the one through which the cavity transmission passes) (See picture 2).  The bright halo seems to come from the contour of the end mirror. Raffaele suggested that it may be the light diffused by the input mirror at small angle, which is trasmitted by the most external part of the end mirror. In fact the HR coating has not bean applied there and it is almost trasparent. The fact that we see it very bright could be due to the comparison with the very small transmissivity (few ppm) of the HR coating anywhere else on the mirror surface.

3) We tried to maximize the alignement of the IR beam by moving the steering mirrors on the bench. As usualy it was done after having aligned the cavity for the green. We could obtain quite a good alignement but we observed that in this condition the reflection from the faraday was not maximized anymore and it became actually very small (less the 1 mW, with 16 mW in input). This suggest that the axis of the faraday is not well aligned with that of the cavity and so an additional work to align them will be required.  To my knowledge we have never payed attention to the faraday reflection while aligning the IR beam in the past, so it is likey that it was the case also before.

 
Images attached to this report
649_20180201020638_endcamera.png 649_20180201020850_endchamber.png
R&D (FilterCavity)
Print this report.
MatteoLeonardi - 16:14, Wednesday 31 January 2018 (648)Get code to link to this report
Comment to SHG characterization (Click here to view original report: 645)

The NTC used seems to be SEMITEC 103JT-050.
According to the datasheet, the R25 = 10kΩ ± 1% and B = 3435K ± 1% (note that this value is correct only between 25deg to 85deg Celsius).

R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 13:27, Wednesday 31 January 2018 (646)Get code to link to this report
IR realigned!

PARTICIPANTS: Yuhang, Matteo, Raffaele, Tomura, Eleonora

Finally today we managed to realign the IR beam.

First we aligned and locked the cavity with the green light and then we cut the green.

We checked the IR reflection from the invacuum faraday on the bench and we maximized it by moving the two last steering mirrors for the IR on the bench. 

The laser power is set to the standard value (current: 1.2 A),  the incoming IR beam just before the windows is about 16 mW and the maximum reflection we could find is 12.8 mW. Increasing the laser power (2 A) we have an incoming beam of about 33 mW and the maximum reflection is about 25.7 mW

We checked that the reflection was moving accordingly to the IM motion, so we were pretty sure that it was coming back from it.

After maximizing the reflected power the beam was good on the PR reference but still we couldn't see anything on the first target, except for a very dim scattered light.

Then we put down the first target and rised the second target where we found a very strange beam, showing some fringes. We observed that fringes were reduced and the shape changed when we misaligned the end mirror.

After that, we reinstall the IR camera in the end bench and we center it by using the residual transmission of the green (when it was locked). By changing the yaw of the BS of a small amount we were able to see IR flashes on the camera. Then we put the BS back in the initial position (the one that makes the green to flash) and we adjust the yaw of the two last IR steering mirrors on the bench in order to make the IR flashes appear again. In this condition we have both green and IR aligned in the cavity.

The IR aligment was not optimized: we were still able to see HOM flashing but it is not easy to fine tune the alignment by moving the steering mirror on the bench

We remarked that there is a very bright circle of IR light on the end camera around the flashes. It resembles the contour of the end mirror and it is not clear why we can see it. Since we suspect a mismatching we tried to move a bit the last IR lens on the bench but it didn't improve the situation. (We put it back to the previouse position.)

Lessons learned:

1) The reflected beam is a good reference for the alignment of the beam and it should be maximized as a good starting point.

2) The behaviour of the IR beam on the target when it is aligned is much diffent from that of the green

    - we can't see anything on the first target, so we should probably not consider it in the future

    - when the second target is up proobably same unpredictable reflections between the mirrors and the target produce a very strange beam. 

More work will be done to understand the origin of the bright circle and to improve the alignment.

R&D (FilterCavity)
Print this report.
AkihiroTomura - 11:07, Wednesday 31 January 2018 (647)Get code to link to this report
Comment to SHG characterization (Click here to view original report: 627)
I did the SHG output power vs input power measurement again. I used HWP and FI pair after the main laser to change the input power in order not to get mode-hops. During the measurement, the current value of the main laser was 2.004 A. The SHG crystal temerature was 3.170 kOhm.
Images attached to this comment
647_20180131030428_20180123lncharacter4.png
R&D (FilterCavity)
Print this report.
YuhangZhao,AkihiroTomura,YuefanGuo - 02:17, Tuesday 30 January 2018 (645)Get code to link to this report
SHG characterization

Thanks to thermistor data sheet from yuefan and the data tomura-san and I took(see e-log 627). I finished the SHG characterization work.

Since the data sheet doesn't give us a formula but only a few points, I use them to make a cubic interpolate. Then I use the interpolation data to make the plot so that you can see the relationship between crystal temperature and Laser power.

Images attached to this report
645_20180129181741_shgcharacterization.png
Comments related to this report
MatteoLeonardi - 16:14, Wednesday 31 January 2018 (648)

The NTC used seems to be SEMITEC 103JT-050.
According to the datasheet, the R25 = 10kΩ ± 1% and B = 3435K ± 1% (note that this value is correct only between 25deg to 85deg Celsius).

YuhangZhao - 16:50, Monday 05 February 2018 (652)
Yuefan told me that it is not SEMITEC 103JT-050, but SEMITEC 103JT-025. Their difference is only the shape.
R&D (FilterCavity)
Print this report.
YuhangZhao - 00:19, Sunday 28 January 2018 (643)Get code to link to this report
OptoCad simulation

Today I finished the OptoCad simulation with practical scenario. I used the real measurement distance for optical components on the bench, and the in-vacuum distance from yuefan's e-log entry 441. I tried also to fine tune(means to move it with like two milimeters) the position of the last infrared lens on the bench. However, the results turn out to be similar with each other when you just look the simulation output picture. And also you can find the output data is similar. For example, the beam size around FI for them are 1.06mm and 1.36mm individualy. See attached file1~4.

However, we can see from the above attached simulation pdf file that the entering beam of PR chamber has some anomalous behaviours. You can see from attached file5, it shows only the incident beam and it is really good. That means bad thing is reflection.  I also checked the output txt file and confirmed the anomalous comes from the reflection of cavity. However, since now I haven't found that why it is like this. I would like to talk with Matteo next week.

Non-image files attached to this report
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 16:35, Saturday 27 January 2018 (636)Get code to link to this report
1310nm probe alignment; lens changed

The usual procedure to align the beam for a best absorption signal is:

- to make the 2 beams pass through the pinhole
- maximize the DC centering the probe on the detector
- maximize the AC aligning the pump.

Adding the second probe beam makes things a bit more difficult because if we move the pump, we lose the alignment with the other probe. So, the maximization of the 1310nm probe AC signal have to be done without moving the pump. 

Since the 1310nm probe turns out to be difficult, we decided to follow the usual procedure for the 1310nm laser, moving the pump to maximize it. Then try to recover the alignment on the red probe without moving the pump.

38.85mm is the position of the maximum of the scan with the red probe. Then we move the pump to find the maximum with the 1310nm probe.

filename:
Mon, Jan 22, 2018 3-01-33 PM.txt ; pd distance 83mm. Stage pos: 9mm
Mon, Jan 22, 2018 3-53-55 PM.txt ; pd distance 95mm (not better)
Mon, Jan 22, 2018 3-59-17 PM.txt ; pd distance 65mm

----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

After trying the alignment many times, I noticed that the 1310nm probe is not focused properly, so we decided to measure the beam profile.
We discovered that the waist position is about 6 cm after the crossing point.
So we decided to change the focusing lens from f200mm to f300mm and to move the lens backward.

Images attached to this report
636_20180127083424_probeprofiles1310.png
R&D (FilterCavity)
Print this report.
EleonoraCapocasa, ZhaoYuhang - 17:14, Thursday 25 January 2018 (642)Get code to link to this report
filter cavity lock characterisation

Yesterday we have worked on the characterization of the filter cavity lock.

First we have experienced some trouble with the labview ADC of the local control signals for the input mirros. The error signals for the pitch and the yaw looked the same. After some attempts the problem was solved simply by switching off and on the CPU unit. We have already observed this problem in the past but I could not understand what triggers this kind of behaviour. 

After recovering the local control we could align and lock the cavity. 

At first we wanted to measure the calibration factor for the error signal. In order to do that we summed a line to the piezo correction (input channel " RAMP")  above the UGF.

We monitored the amplitude of the line on the monitor channel "PZT mon" finding a value of V_RMS = 17.84e-6 V.

The correspondent amplitude in Hz is 

S_Hz = V_RMS (V) * 100 * sqrt(2) * 2e6 Hz/V = 5045 Hz         (1)

The factor 100 is the attenuation of the "PZT mon" channel

The factor sqrt(2) is to pass from V_rms to the line amplitude

The factor 2e6 Hz/V is the gain of the piezo after the SHG

The correspondent amplitude observed in the error signal is

Err_V = K(V/Hz) * S_Hz/sqrt(1+ (f/f_0)^2) =   sqrt(2)* 0.465 V       (2)

% (2020/09/04) Note that this formula should be Err_V = K(V/Hz) * S_Hz/sqrt(1+ (f/f_0)^2)

Where we have taken into account the effect of the cavity pole assumed to be at f_0 = 1.45 kHz.

Comparing the amplitude of the two lines (see picture 1), inverting Eq. 2. We could find the calibration factor K (V/Hz) = 2.5 e-3

We have used it to calibrate the error signal into Hz and we obtain a lock accuracy of about 120 Hz, (see picure 2) which is similar to what we obtained in July.

We have also performed a noise injection to measure the open loop TF finding a UGF at about 10 kHz  with a phase margin of 50 deg. (See picture 3)

We have tried to change the gain but we observed that it was already optimized to have the error signal as small as possible.

In picure 4 the transmission of the filter cavity is plotted.

Images attached to this report
642_20180125085624_line.png 642_20180125085654_calibratederrorsignal.png 642_20180125091302_tf2401.jpg 642_20180125091324_trans.jpg