LOG-IN
Displaying reports 2741-2760 of 3200.Go to page Start 134 135 136 137 138 139 140 141 142 End
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 19:11, Friday 09 June 2017 (497)Get code to link to this report
Comment to HeNe probe alignment, profile characterization and waist position (Click here to view original report: 495)

I calculated with Jammt the best position and focal length in order to have a probe waist of 180um at the cross point. So I replaced the lens with  a f=300mm lens and measured again the profile.  Then I plot the pump and probe profiles together. The waists positions are aligned well, within 1cm to the crosspoint. Next step is to align the imaging unit and maximize the absorption signal of a reference sample adjusting the pump position.

Images attached to this comment
497_20170609114622_probeprofile300.jpg 497_20170609115813_pumpprobeprofile.jpg 497_20170609121120_p6090453.jpg
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 14:16, Thursday 08 June 2017 (495)Get code to link to this report
HeNe probe alignment, profile characterization and waist position

I used a 200um-diam pinhole to set the position of the crosspoint between the pump and the HeNe probe ().

First I checked the position of the pump waist, by looking for the position of the maximum transmitted power through the pinhole. I found 185mW/206mW  at 11.9cm +/- 0.5cm (14cm from the board) which is consistent with the previous measurement with the beam profiler and fit.

Then, using a crosshair, with the help of Matteo, I aligned the HeNe laser to be at 0.1rad  (5.7deg) with the pump beam and passing through the crosspoint at 14cm from the board.

I mounted the pinhole to the 2inches mount using the 1inch-to-2inch adaptor. I centered  the pinhole at the crosspoint and checked the absolute position in microsteps (1mm = 8063steps): 

@1-axis 2641973

@2-axis 279284

@3-axis 981325

Then I made a fine adjustment of the probe position to maximize the transmission through the pinhole.

I measured the beam profile of the HeNe laser without focusing lenses and with a f=150mm lens. See plot. The waist of the probe should be at the cross point (14cm from the board). There are two ways to move the probe waist: or move the lens, or change the lens focal length.  

Images attached to this report
495_20170608070237_p6060376.jpg 495_20170608070245_p6070377.jpg 495_20170608070251_p6070378.jpg 495_20170608070301_p6070379.jpg 495_20170608070815_probeprofile.jpg
Comments related to this report
ManuelMarchio - 19:11, Friday 09 June 2017 (497)

I calculated with Jammt the best position and focal length in order to have a probe waist of 180um at the cross point. So I replaced the lens with  a f=300mm lens and measured again the profile.  Then I plot the pump and probe profiles together. The waists positions are aligned well, within 1cm to the crosspoint. Next step is to align the imaging unit and maximize the absorption signal of a reference sample adjusting the pump position.

ManuelMarchio - 09:33, Thursday 15 June 2017 (501)

I had to adjust a bit the position of some parts of  the translation stage, sothe new pin-hole position at the cross point is 

@01-axis: 2646049 2646049

@02-axis: 279284 279284

@03 0 OK IDLE NC 978021
@03-axis: 978021
R&D (FilterCavity)
Print this report.
YuefanGuo - 03:18, Wednesday 07 June 2017 (496)Get code to link to this report
EOM installed
Today when we tried to install the EOM, we found out since the telescope changed, the beam size at EOM input is a little bit large. The EOM should be installed between the second lens and the MZ, and we still need some space for another mirror and beam splitter of MZ. So we checked the optical scheme and found out there is enough space to push the MZ further from the lens. We pushed the MZ three holes back and installed the EOM.

Then the beam size changed a lot at the PR window and also it seems diverge very quickly after the BS suspension. We checked the manual of the EOM, did not find anything about this situation. For better understanding what happened to the beam after the EOM, we removed the third lens, take a few measurement before the EOM and after the EOM. Then did the fitting to check what changes. The origin set as the line of hole near the second lens. From the figure it seems only the beam waist shift from 27.2 to 20.2, the beam waist size changed a bit, but we should consider that the space is limited, so we need to take several points of measurement which is quiet close to each other. Also the z value is measured with meters which is not very precise. It seems this problem can be solved only with moving the third lens.
Images attached to this report
496_20170606201634_eom.jpg
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 11:01, Tuesday 06 June 2017 (494)Get code to link to this report
Comment to Pump beam profile (Click here to view original report: 484)

I replaced the 125mm lens with a 150mm lens. So the waist moved as expected to 0.11m. See the plot

Images attached to this comment
494_20170606040107_pumpprofile.jpg
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 23:40, Monday 05 June 2017 (493)Get code to link to this report
Beam motion at 290 m possibly affected by PR resonance

This video shows the movement of the green beam on the target at 290 m when the local control loops of the suspended injection mirrors (PR and BS) are closed. It seems to me that it is moving too much in the vertical direction. 

In the attached pdf there are the spectra of the PR and BS mirrors.

When the loops are closed, there is still a narrow peak at about 7.5 Hz in the pitch of the PR, that is cleary visible also looking at the loop error signal in time domain. According to what observed here, I would say that it is due to the intermediate mass touching the magnet holder plate. In the past we observed that it happened often after moving a lot the yaw picomotor. 

We should now try to understand to what extent the beam motion is affected by this and possibly solve the problem.

Non-image files attached to this report
R&D (FilterCavity)
Print this report.
MarcEisenmann - 16:26, Monday 05 June 2017 (492)Get code to link to this report
filter cavity input mirror wedge
The input mirror wedge sign was studied with an autocollimator callibrated with the reflection of the front surface (by superposing the reticle and the front face reflection).
To understand which reflection was the one from the front surface, we looked at the reflections due to a roof light. When this light hitted only the front surface, its reflection was red. When it was hitting both surfaces, we could see two shifted reflections, one green and one red. So the red reflection corresponds to the front surface reflection.

The callibration of the autocollimator means that the incident beam and the front surface reflection are on the same direction. Then, depending on the wedge sign, the back surface reflection will either be seen on the top or on the bottom of the front surface reflection.

By looking at the autocollimator, we were able to tell that the wedge sign corresponds to the second configuration on the scheme. This means that the "#4" sign on the mirror corresponds to the lower part of the wedge.
Images attached to this report
492_20170605091305_rooflightreflections.png 492_20170605091324_2wedgeconfigurations.png 492_20170605091333_reflectioncrossesfromtheautocollimator.jpg
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 16:15, Monday 05 June 2017 (484)Get code to link to this report
Pump beam profile

In order to place the waist of the pump beam on the right position, I characterized the pump beam using the beam profiler and moving it with the translation stage. 

Laser settings: I = 1A, P = 210mW , hwp_angle=0, not chopped.

I aligned the pump beam to be parallel to the z-axis translation stage within 0.2mm on the whole translation range 250mm (±0.8mrad).

The beam profiler position is set 0 at the position of the KAGRA-size sample first surface (see picture).

Without the focusing lens, the waist is 377um at -0.88m

Using the focusing lens f=125mm placed at -60mm, the waist becomes w0=53um  at 0.081m

According to this calculation the deepest measurable point is 0.08*1.81=0.149m for sapphire (and less for silica).  Since KAGRA-size thickness is 15cm, the cross point (which will be at the pump waist) cannot go beyond the second surface.  There is not enough space to place the lens closer to the sample, therefore I need to change the focal lens.

Making a quick simulation with JamMt, a focal lens of 150mm will move the waist to 0.111m (w0=62.8um) without sample, With a 15cm-thick sapphire substrate the waist will move to 0.178m, and with a 15cm-thick silica substrate, the waist moves to 0.16m, which is enough to measure the entire thickness of the sample.

Images attached to this report
484_20170605090937_screenshotfrom20170605155946.png
Comments related to this report
ManuelMarchio - 11:01, Tuesday 06 June 2017 (494)

I replaced the 125mm lens with a 150mm lens. So the waist moved as expected to 0.11m. See the plot

R&D (FilterCavity)
Print this report.
YuefanGuo - 23:30, Friday 02 June 2017 (491)Get code to link to this report
Comment to Green telescope on bench changed and simulation of the beam (Click here to view original report: 490)
Add the picture
Images attached to this comment
491_20170602163022_beampropa.jpg
R&D (FilterCavity)
Print this report.
YuefanGuo - 23:29, Friday 02 June 2017 (490)Get code to link to this report
Green telescope on bench changed and simulation of the beam
Yesterday after tried moving many times all three lenses in the bench, it seems the first 100mm lens has very large effect on the beam shape. So we decided to change this lens into a large focal length one and re-design the telescope.

This is the new telescope parameters:(origin still at the first line of hole out of the cavity)
L1 f1=200mm z1=8cm
L2 f2=200mm z2=56.5cm
L3 f3=175cm z3=99cm

With this new telescope we had a better beam at the window of PR chamber and also before the beam goes inside the arm, the previous one produced a beam a little bit longer in horizontal direction. So we decided to go on with this configuration.

Then today after we realign everything on the bench, took some measurement to check the difference between the theoretical value and practical one is inside the tolerance or not. All the points we measured are equal or smaller than the result of the calculation. We also added the two telescope mirror inside the chamber to check if the beam in the tube has the size we want or not. At the two inch mirror we had a beam radius about 1mm which is good, and the largest size inside the arm is about 1.3cm.

The other thing we did is we measured several points on the bench after the third lens. With these data, we calculated the beam waist and simulate the propagation of beam in 300mm, but without the two telescope mirror. From the picture we can see that the beam shape on the bench can not produce astigmatism at 300m.
Comments related to this report
YuefanGuo - 23:30, Friday 02 June 2017 (491)
Add the picture
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 21:29, Friday 02 June 2017 (488)Get code to link to this report
investigation on the green beam astigmatism: effects of the windows

Today we keep investigating on the beam astigmatism. As a first thing we realign all the optics on the bench and measured the beam in different position. There results will be posted in a dedicated entry by Yuefan which has also calculated the astigamism expected at 300 m on the basis of the measuremts done on the bench. (Very small).

Then we sent the beam in the pipe and open the first windows between the BS and NM2. The change in the shape when the windows is open is shown in this video.

After that we let the beam propagate for about 30 m in the corridor and move the small telescope mirror in order to have the beam as collimated as possible. We succeded in finding a position allowing to have a beam which remains pretty small (2-3 cm) up to the end of the tunnel.

After the small telescope tuning, the beam (both with closed and open window) looks different on the 290 target. Here there is another viedo who recorded the beam during the closure of the BS-NM2 windows. The beam with open widows looks still very astigmatic while with the close window is seems aftected by diffused light and very difficult to judge.

[I have the impression than after opening the windows for the first time it becames somehow dirtier and once we close it again, it affects the beam more than it was doing before.]

Some preliminary conclusions:

1) Tuning the 2"  telescope mirror position have a great impact on the beam shape at 290 m.

2) The windows seem to have a great impact on the beam shape at 290 m.

3) The beam seems well collimated and not significantly astigmatic when it propagates in the corridor.

R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 10:53, Friday 02 June 2017 (486)Get code to link to this report
Picomotor remote control interface

In order to speed up the remote control of picomotors that are largely used for the beam aligment, I wrote a custum vi which avoid to inserting each time command lines to select drivers, motors, velocity and steps (pic 1).  The vi (picocontrol.vi) is in a dedicated labview project (C:Digital/picomotor_control.lvproj)

We remark that BS yaw picomotor still doesn't work.

The final configuration for the picomotors control is 

CONTROLLER 133.40.121.13

A1

  1. PITCH BS
  2. YAW BS
  3. empty

A2

  1. PITCH INPUT
  2. YAW INPUT
  3. empty

A3

  1. PITCH 2" TELESCOPE
  2. YAW    2" TELESCOPE
  3. LENGTH 2" TELESCOPE

CONTROLLER 133.40.121.14

A1

  1. PITCH PR
  2. YAW PR
  3. PITCH SM1

A2

  1. PITCH SM2
  2. YAW SM2
  3. YAW SM1

CONTROLLER 133.40.121.15  

A1

  1. PITCH END
  2. YAW END
  3. empty

 

Where 1, 2, 3 correspondes to motor 0, motor 1 and motor 2

Images attached to this report
486_20170602035148_picocontrol.png
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 22:53, Thursday 01 June 2017 (485)Get code to link to this report
Green beam astigmatism investigation

In the past days we have been working in order to improve the green beam shape at the end of the arm. 

1) We have realigned the optics on the bench many times in order to be well centered on the optics. After adding each optics we measured the beam and propagate it far (about 5-6 m) to check if we could notice any astigmatism.

2) We have tried to be as centered as possibile in the sferical mirror of the telescope and on the two gates between BS-NM2 and NM2-PIPE

3) We have change the telescope configuration for the grean beem in order to have longer focal length whit respect to the dimension of the beam on the lenses and avoid sferical aberations as much as possible.

In this condition we are able to have a beam which looks failry circular at the input of the pipe and also on the first target but which becomes very bad when observed on the  290 target. (The main effect is elongation in the orizontal direction). This video shows what we see on the far target while scanning the orizontal direction with the BS local controls.  While moving from one side to the other, at first we can observe some scattered light, than we don't see anything for a while and finally we see the beam on the target. Continuing the scan, the beam disapears and after another moment of darkness we can see again the scatterd light. From this I would say that what we observe is the direct beam and not its reflection on the pipe wall.

In order to better understand the evolution of the propagating beam we have put two steering mirrors at the level of the input mirror in order to send the beam in the corridor. (As steering mirrors we used to 5" mirrors used as PR ans END dummy mirrors, which should be both former TAMA PR mirrors with long RoC )

The beam at 300 m is shown in picure. It is moving a lot because of the air but it doesn't look elongated in the orizontal direction. There are no evident reasons why the beams propagating in the corridor and in the vacuum look so different. Here some hypothesis:

1) The air is affecting the propagation, masking the astigmatism observed in vacuum.

2) The steering mirrors are not flat enough and they modifiy the beam masking the astigmatism.

3) The astigmatism is introduced by the window at the input of the pipe and becomes evident only after a long propagation (on the target at 10 m the beam stil looks good) 

4) Effect due to possible multiple reflections from the target and the input window (it could be ruled out by looking at the beam reaching the end chamber) 

We remark that the beam on the target observed by the camera looks better than in reality. The same is true for the pictures taken to the green beam where small deviation from the circular shape are difficult to be apprieciated.

Images attached to this report
485_20170601154436_jpg.jpg
R&D (FilterCavity)
Print this report.
YuefanGuo - 14:41, Wednesday 31 May 2017 (483)Get code to link to this report
Green beam characterization
The first plot shows the measurement of the green beam out of the SHG cavity and the fit result of it. The red spot is the position(-7.3cm) and the waist size of the green beam(26um). Origin is set at the first line of hole in front of the cavity.

In the second attachment, we put the position and the focal length of the three lenses we using now. We used the result of the first plot as initial value of Matlab program to get these result. We did some measurements of the beam size near the lenses this morning and compared them to the theoretical values, seems they fit well. Also if we considering the threshold of causing the aberration from the report, with 2cm beam waist, convex-plano lens should has a focal length large than 2m. The lenses we are using now is 100mm, 200mm, 1000mm, the beam size we have now is in the safe range.
Images attached to this report
483_20170531071632_grfit.jpg
Non-image files attached to this report
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 01:19, Wednesday 31 May 2017 (481)Get code to link to this report
BS local controls restored

In the past days I have restored the local control on the BS mirror (used as steering mirror after the telescope). The optical scheme (similar to that described here) is shown in picture 1. We have noticed that there are two reflected beams (from the first and the second surface) which are very close. In this configuration the two reflections cannot be separated and they arrive almost superposed on the PSD. Moreover the beam is impinging at a distance of about 4.5 cm from the mirror center. According to my computation (reported here), it should induce an error on the yaw measurament of about 3.5%.

I had some trouble in diagonalizing the driving. After many tries, I have found that the best diagonalization is achieved without using the two lower coils (see coils disposition in the last attached picture). This fact is very strange and should be better investigated. With the following driving matrix 

  YAW PITCH
COIL 1 1 0.95
COIL 2 -1 1.1
COIL 3 0 0
COIL 4 0 0

I found the mechanical TFs shown in picture 2 and 3 (when exiciting yaw and pitch respectively). They look similar to those measured last november.

The open loop transfer functions are shown in picture 4. Due to the second narrow resoance in pitch at about 10 Hz, the UGF is crossed two times. I'm not sure about the phase of the pitch open loop TF for the second crossing point. Anyway I was able to close the two loops and they look stable. We also added offsets and we were able to observe the change in the beam position on the target at 290 (both in vertical and orizontal direction according to the loop to which we add the offset) 

The comparison between the open and closed loop spectra are shown in picture 5. Maybe the UGF can be increased a bit.

The calibration is  0.37 mrad /V. It has been computed as done here. where V_SUM of the PSD is 13.5 V and the lever arm is about 0.7 m

Some remarks

1) In order to investiate the driving issue, I have injected a line at 5 Hz with the same amplitude in each coil (one by one). The spectra in the four cases are shown in the attached pdf and seem pretty much the same.

2) I have observed that while measuring mechanical TFs, at the begining I was not able to find a good coherence at low frequency (below 1 Hz) and the resonances both in pitch and yaw where always very excited. I found out that this was due to the air flow inside the cleanbooth. I have temporarily disconected it to make the measurements. 

3) I have observed an oscillation at 50 Hz in the four signals sent to the coil driver. The amplitude is quite high (about 600 mV pp) and it is not present if i look at the signals just at the output of DAC. I'm not sure about how to get rid of this. 

4) While glueing the magnets we checked that the north-south polarization was in agreement with the convention reported in the last attached picture. Now, observing the driving matrix it seems that the sign should be inverted. I remember that this happened also for the dummy BS. 

Images attached to this report
481_20170530180058_img20170530193526316.jpg 481_20170530180126_noiseyaw.png 481_20170530180147_noisepitch.png 481_20170530180221_oltf.png 481_20170530180242_yawoc.png 481_20170530180648_pitchoc.png 481_20170531023727_26020160627054235mirrormagnet.jpg
Non-image files attached to this report
R&D (General)
Print this report.
YuefanGuo - 10:43, Tuesday 30 May 2017 (480)Get code to link to this report
Check the beam shape
Since last Friday we found out when we sent the beam to the end, the shape of the beam is very bad, much worse than the beam shape we saw at the beginning of last week, both the green and the infrared, the beam both are large and long in the horizontal direction. Last Friday we tried to align the beam to the center of both the suspension mirrors, but the situation did not become better, so we tried to move the 2inch mirror a little bit by the picomotor, and found out to one direction the beam gets better but not very obvious, it seems we did not have enough range of that picomotor, so we decided to move 2inch mirror by hand.

Yesterday before we start to move the 2inch, we checked the beam shape at different place of the beam path with beam profiler,the beam seems circular, although at some place the beam is fluctuated very quickly, but during the fluctuation there are some moments that the beam is round. Then we tried to adjust the beam more center on the lens of green beam, and finally we had to adjust from the very beginning, we aligned all the component to let the beam go straight and checked each time we put back a component,almost circular everywhere,(except there seems have some dust on one of the component of green, so when the beam reached the window, there is one line shape shadow on the top of the beam, we tried to clean all the component with gas, but cannot remove that) after everything was done, the beam shape at the 290m target did not get any better. So then we can focus on the two vacuum chamber.

We tried to push or pull the 2 inch mirror by hand and align the green beam to see it on the target, although the beam shape is still strange but at some point the beam is much smaller than the other place, and except the main beam there is another small beam on the side of it which is quiet good shape, maybe we can find out where it comes. Then when we cut the green and tried to see the infrared, the infrared is almost what we want to see, a small beam with almost circular shape, so this means with moving the 2 inch mirror, we can get the situation better. Today we are going to find a good position of that mirror to get a better green beam.

When we tried to move the BS yaw picomotor yesterday, it does not work with the computer command again, it is better to change the picomotor before we close the chamber.
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 00:46, Tuesday 30 May 2017 (479)Get code to link to this report
Comment to wedge measurement and magnets/standoff glueing on the input mirror (Click here to view original report: 471)

Participants: Eleonora, Marc

Few days ago we observed that on one of the stand-off we glued on the input mirror the glue seemed to have overfowded on the groove where the wire is supposed to stay (picture1). Today we made some tests with a wire of the same diameter of those used in the suspension and verified that it was the case. After trying to remove the excess of glue without good results, we decided that was safer to remove the stand-off and to glue a new one. (picture 2)

Images attached to this comment
479_20170529174014_standoff2.png 479_20170529174040_img20170529115628244.jpg
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 14:15, Saturday 27 May 2017 (478)Get code to link to this report
Filter cavity end mirror successfully installed

In the past days we worked in order to suspend the filter cavity end mirror in the end room vacuum chamber. 

As a preliminary activity we performed a major cleaning of the end room.

On Thursday 25th, we opened the vacuum chamber and removed the dummy mirror (an old TAMA PR mirror).

On Friday 26th, we installed the final filter cavity mirror (end #1 substrate)

The procedure followed to change the mirror is sketched in the attached pdf.

After the mirror substitution, we sent the beam from the central bulding to the end mirror and tried to move it with picomotrs in order to send the beam back. Unfortunately the dispalcement in yaw achievable with picomotors was not enough to allow it and we were forced to slightly turn the whole suspension. (The problem of the small yaw range was already observed for the PR telescope mirror. Also in this case we solved it by slighty moving the whole suspension). After this operation the reflected beam seems reasonably centered on the small gate window (that cannot be open utill the end chamber is evacuated).

We remark that the green beam is quite big, still too astigmatic and also moving a lot. The work in the next days will be devoted to improve it.

OTHER REMARKS

1) Picomotors for controlling pitch and yaw are working fine. They can be remotely controlled with labview (IP adress 133.140.121.15) and also locally using the joystick (pictures 2-3)

2) Thanks to Yoshizumi-san, we now have working wi-fi (naoj-open) and working phone (number 3472) in the end room!  (picture 4)

3) Dummy mirror (old TAMA PR) is currently stored in the filter cavity mirror case of substate #1 (picture 5)

Images attached to this report
478_20170527070243_img20170527133400561.jpg 478_20170527070254_img20170527134253451.jpg 478_20170527070300_img20170527133754405.jpg 478_20170527070305_img20170527133248469.jpg 478_20170527070321_img20170527132733128.jpg
Non-image files attached to this report
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 17:14, Friday 26 May 2017 (477)Get code to link to this report
Comment to Sesnsitivity curve with crystalline coatings and squeezing (Click here to view original report: 476)

I calculate the horizon of BBH and BNS  for the sensitivity curves:

- Amorfous coating ; no squeezing      BBH = 3.28 GPc, BNS = 360 MPc;

- Amorfous coating ; squeezing           BBH =  4.42 GPc, BNS = 509 MPc; 

- Crystalline coating ; no squeezing     BBH =  3.46 GPc, BNS =  378 MPc; 

- Crystalline coating ; squeezing          BBH = 4.90 GPc, BNS = 566 MPc;  

d_H = (G^5/6 * M^1/3 * mu^1/2) / (c^3/2 * pi^2/3 * rho) * sqrt( 5/6 * int_f1^f2 f^(-7/3) / S(f) df)

M is the sum of the 2 masses, mu is the reduced mass, rho is the SNR, f1 and f2 are the frequency range for the event signal, S(f) is the noise spectrum (square of the equivalent strain)

I used f1=10Hz; f2_BBH=73Hz; f2_BNS=1571Hz

M_BH = 30M_sun (M=60 M_sun)

M_NS = 1.4M_sun (M=2.4 M_sun)

rho = 8

KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 10:56, Thursday 25 May 2017 (476)Get code to link to this report
Sesnsitivity curve with crystalline coatings and squeezing

I calculated how the coating brownian thermal noise will change in the case KAGRA mirrors will employ crystalline coatings. The mechanical loss I used is 4.5e-6 at cryogenic temperature (from G.Cole, et.al, "Tenfold reduction of brownian noise in high-reflectivity optical coatings", Nature photonics, 2013)

I took the LCGT design sensitivity curve contributions from KAGRA website. I replaced the brownian coating thermal noise with the one for crystalline coatings, and I replaced the quantum noise with the one calculated by Eleonora with the frequency dependent squeezing.

Images attached to this report
476_20170525035601_sensitivityplot2.jpg
Comments related to this report
ManuelMarchio - 17:14, Friday 26 May 2017 (477)

I calculate the horizon of BBH and BNS  for the sensitivity curves:

- Amorfous coating ; no squeezing      BBH = 3.28 GPc, BNS = 360 MPc;

- Amorfous coating ; squeezing           BBH =  4.42 GPc, BNS = 509 MPc; 

- Crystalline coating ; no squeezing     BBH =  3.46 GPc, BNS =  378 MPc; 

- Crystalline coating ; squeezing          BBH = 4.90 GPc, BNS = 566 MPc;  

d_H = (G^5/6 * M^1/3 * mu^1/2) / (c^3/2 * pi^2/3 * rho) * sqrt( 5/6 * int_f1^f2 f^(-7/3) / S(f) df)

M is the sum of the 2 masses, mu is the reduced mass, rho is the SNR, f1 and f2 are the frequency range for the event signal, S(f) is the noise spectrum (square of the equivalent strain)

I used f1=10Hz; f2_BBH=73Hz; f2_BNS=1571Hz

M_BH = 30M_sun (M=60 M_sun)

M_NS = 1.4M_sun (M=2.4 M_sun)

rho = 8

R&D (General)
Print this report.
YuefanGuo - 23:15, Monday 22 May 2017 (474)Get code to link to this report
Telescope installed for green and infrared
Today we received the optics we need so we started to install the telescope for both infrared and green path. Since green is easier to see and two beam are more or less superposed at the 2inch mirror, so we use the green for the reference.

Then we started to installed the 150mm and 175mm lens at the calculation place with rails and moved a little bit along the rail to make the beam have a good size at the 2inch mirror.

For the green path, we tried to remount everything. But according to my calculation, the first lens should be at 7.5cm from the cavity output, but we discovered today, the limited we can do is 8cm, cannot get closer because of the screw of the dichroic mirror. So we put it at 8cm, and did the simulation again, the result shows below,

L1: z=8cm f1=100mm
L2: z=49cm f2=200mm
L3: z=117cm f3=1000mm

In the order of L1, two dichroic, Faraday Isolator, half-wave plate, L2, L3, Faraday Isolator, half-wave plate, we put all of them and align the beam again. During the installation, one problem is that the third lens is very close to one mirror, and also there is an aperture very close to this mirror, so we put another aperture firstly just after the previous one but provide enough space for the rail of the lens.

Then we tried to measure both the green and infrared beam with the beam profiler, did some adjustment to have the size we want(1mm in radius). But the infrared beam looks have some astignatism. Checked the beam at 290m target, the green is larger than the infrared, and also there are two green beams, one is larger, one is smaller but more circular(pic 1), we tried to move the picomotor of BS to make this two green beam superpose(pic 2). The infrared is much smaller(pic3), when it moves to some position of the target, it looks good.
Images attached to this report
474_20170522161421_20170522225456.jpg 474_20170522161514_20170522225518.jpg 474_20170522161533_20170522225511.jpg