LOG-IN
Displaying reports 2761-2780 of 3344.Go to page Start 135 136 137 138 139 140 141 142 143 End
R&D (FilterCavity)
Print this report.
Yuhang, Eleonora, Tomura, Matteo - 19:42, Tuesday 23 January 2018 (638)Get code to link to this report
Measurement of IR beam parameter for simulation and more attempts to recover IR alignment

Since we found something wrong last week, we tried to carefully measure the beam parameters again after the first Faraday Isolator. As Yuefan did, we also set the origin as the BS on the bench.

For x direction, the beam waist is 118.2 um at the position z=-0.10 m. For y, the beam waist is 119.8 um at z=-0.093 m. See attached figure 1 for the fitting result.

Then we used this result to simulate the propagation of the beam using the software "Jammt". See attached figure 2 for this simulation.

We found that our beam parameter was very sensitive to the last lens position. We have fine tuned its position in order to have a reasonable beam dimension. By using a steering mirror on the beam path we could propagate it for several meters in the central area and check that the beam size should be reasonable at the input of the in vacuum faraday.

After that we have keep trying to recover the alignment of the infrared without success. We have collected as many cameras and screens as possible as use them to look insides the input vacuum chamber simultaneously. 

Despite many tries the situation is not different from that of the last time:

1) We can recover the references for the IR on the plastic film but for one of them the beam as a quite irregular shape.

2) We could see the beam on the mount of the 2 inch mirrors and we tried to center it.

3) We could see the beam hit the leg of the PR telescope mirror and tried to center it on the PR.

After that we cannot see the beam on the BS or on the input mirror but just some strange quite dim shape on the first target.

As already observed in entry 631, when moving the PR or BS mirror with the local controls we can move accordingly the shape on the target. This is different from what we expect if the beam was hitting the pipe, since normally in this condition we observe a change in the intensity and in the shape of the scattered light.

Conclusions: We are probably not understanding what is happening inside the chambers and we need some new ideas to go on. 

Images attached to this report
638_20180123114224_best.png 638_20180123114235_55.png 638_20180123114250_1734486301.jpg
R&D (FilterCavity)
Print this report.
AkihiroTomura - 03:00, Monday 22 January 2018 (635)Get code to link to this report
Test of grating (GR-25-1210)
We purchased a grating for separation of two beams, 532 nm and 1064 nm. This is a ruled reflective diffraction grating. I conducted a test use. I measured power of diffraction beams (zero & first order) as changing incident angles. The setup I used for the test is shown in figure 2. Basically, polarization of a beam was firstly determined and manipulated by a half wave plate. The incident beam power was always being monitored because of the fluctuation. In order to obtain proper results, the measured power was multiplied by (the average incident power)/(incident power at each moment).

For the definition of the incident angle, please refer to figure 3.
For the spec of the grating, please refer to the attached pdf file.

I am not really sure about what I see on the graph.
I need your opinions.
Images attached to this report
635_20180121185339_gratingtest.jpg 635_20180121185351_20180121172831173ios.jpg 635_20180121185358_20180121171619272ios.jpg
Non-image files attached to this report
R&D (FilterCavity)
Print this report.
YuhangZhao - 00:44, Saturday 20 January 2018 (634)Get code to link to this report
Removing the unwanted lens for infrared

Because yesterday's measurement doesn't match previous result. I decided to do the simulation for the infrared. However, I found there is a lens without label on it. So Eleonora suggested me to consult yuefan about this. And yuefan said it is used for the test of mode cleaner. So we decide to remove it. After removing it, we want to check the beam waist again.

 

See attached figure 1, we found the beam waist was very far from where we did the measurement.(The measurement position is shown on the attached figure 2) And the beam waist is about 1mm. The beam waist fit the result of OptoCad. However the beam waist position doesn't fit the result of OptoCad. As you can see the attached figure 3, the highlight part show that the beam waist is around 8 meters ahead of R1. This R1 is the dichroic mirror which is used to combine green and infrared. But the measurement of beam waist tells us the distance between them is around 6 meters.

 

According to attached figure1, the beam size at Faraday Isolator should be around 2mm in radius. This measurement is totally different from the requirement of e-log 441. E-log 441 gave us the beam size around Faraday Isolator should be 1.368mm. No matter it is diameter or radius, it doesn't agree what we measured today.

 

Besides, I just found there is e-log 442 which tells us the setting of infrared telescope. We can see from there, the design is using leses f500, f300 and f350. However I found the actual lenses we are using is f500, f150 and f175. See attached last three pictures to check this.

Images attached to this report
634_20180119164310_newfitting.png 634_20180119164322_06.png 634_20180119164333_21.png 634_20180119164346_344794330.jpg 634_20180119164356_470690766.jpg 634_20180119164403_2116561758.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 14:23, Friday 19 January 2018 (632)Get code to link to this report
Comment to checking the infrared beam waist (Click here to view original report: 630)

Today I measured it again. Since the last measurement seems to be limited in a small region, I tried to expend the region this time. The result is shown here.

 

And now, we have four lenses on the infrared path. They are f500, f150, f175 and one without lable. Is this the same with previous situation?

Images attached to this comment
632_20180119062244_infraredfitting.png
R&D (FilterCavity)
Print this report.
EleonoraCapocasa, YuhangZhao - 23:21, Thursday 18 January 2018 (631)Get code to link to this report
Attempt to align the IR beam
Today we have tried to recover the IR alignment.
 
1) We have checked the references on the PR viewport. According to the attached scheme, they come from the transmission of the first and the second PBS of the Faraday.
 
We have observed that when the beam transmitted by the Faraday is well centerd on the cross on the plastic film out of the wieport, the other reference beam is slightly hitting the edge of the second PBS of the Faraday and this makes the beam to have a quite irregular shape on the cross. Since the irregurity is not so evident we wondered if it was like this also when the reference was done or if it a sign that the beam is not well aligned. 
 
2) We have put a camera to look inside the BS and we coud spot the beam on the 2 inch telescope mirror.  We did it by moving the steering mirrors on the bench in order to make it hit the edge of the mirror and we have verfied that it is reasonably circular.
 
3) Then twe continued to follow the beam path: we have looked at the PR mirror with a camera and we have done a scan in the horizontal direction. We could see a dim shape hitting the leg of the PR suspension and we could see it moving on the leg on the other side of the suspension while performing the scan.
 
4) We have out a camera to look at the input mirror but we coudn‘t see any light even when scanning the beam position in the two direction.
 
5) Finally we have used the camera to look at the first target. We coud observe a very irregular light shape, much larger than the expected beam. At the beginning we thought it was the light diffracted by the beam hitting the pipe but when we moved the PR we observed that the shape was mooving accordingly. (You can see a video here). It probably suggests that the beam is reaching the target but it is clipped along its path before the BS.
 
6) As reported in entry 630, we have also cheked the dimension of the beam on the bench, finding a difference of about a factor two with the value measured previously by Yuefan. Since the main laser has been mooved, it is likely that the the telescope has to be retuned. i don‘t know how much this discrepancy can contribute to the strange effect we observed.
 
We have also measured the power of the references, wich gives a lower limit on the losses of the Faraday. For an input power of 15.7 mW we found that the transmission of the fist BPS of the Farady is 270 uW and while that of the second one is 120 uW. We have verified that changing the IR polarization from s to p most of the light is transmitted by the the first PBS.
 
 
1) We have checked the references on the PR viewport. According to picture 1, they come from the transmission of the first and the second PBS of the Faraday. We have checked that when the beam
transmitted by the Faraday is wel centerd on the cross on the plastic ï¬Âlm out of the wieport, the other reference beam is slightly hitting the edge of the second PBS of the Faraday and this makes the
beam to have a quite irregular shape on the cross. Since the irregurity is not so evident we wondered if it was like this also when the reference was done or if it means that the beam is not well aligned.
2) We have put a camera to look inside the BS and we coud spot the beam on the 2 inch telescope mirror. By mooving the steering mirrors on the bench we moved it in order to make it hit the edge of
the mirror and we have veriï¬Âed that it is reasonably circular.
 
3) Then twe continued to follow the beam path: we have looked at the PR mirror with a camera and we have moved the beam. We coud see a dim shape hitting the leg of the PR suspension and we
could see it moving on the leg onother side of the suspension.
 
4) We have out a camera to look at the input mirror but we coudn‘t see any light even when mcoving the beam.
 
5) Finally we have used the camera to look at the first target. We coud observe a strange, irregular light shape, much larger than the expected beam. At the beginning we thought it was the light
diffracted by the beam hitting the pipe but when we moved the PR we observed that the shape was mooving accordingly. (You can see a video here). It probably suggest that the beam is reaching the
target but it is getting clipped along its path.
 
6) As reported in entry. we have also chech the dimension of the beam on the bench finding a difference about a factor two with the value measured previously by Yuefan. Since the laser has been
mooved. it is likely that the the telescope has to be retuned. i don‘t know how much this discrepancy can contribute to the strange effect we observed.
 
We have also measured the power of the referencies. wich give a lower limits on the losses of the Faraday. For an input power of we find that the translission of the fist BPS of the Farady is and the
second one is. We have veryfled that chaging the IR polarization from s to p almost all the light is transmitted by the the first PBS
 
CONCLUSION: we suspect that the beam is clipped somewhere in its in vacuum path e have been working on the IR aligment.
 
1) We have checked the references on the PR viewport. According to picture 1, they come from the transmission of the first and the second PBS of the Faraday. We have checked that when the beam
transmitted by the Faraday is wel centerd on the cross on the plastic ï¬Âlm out of the wieport, the other reference beam is slightly hitting the edge of the second PBS of the Faraday and this makes the
beam to have a quite irregular shape on the cross. Since the irregurity is not so evident we wondered if it was like this also when the reference was done or if it means that the beam is not well aligned.
2) We have put a camera to look inside the BS and we coud spot the beam on the 2 inch telescope mirror. By mooving the steering mirrors on the bench we moved it in order to make it hit the edge of
the mirror and we have veriï¬Âed that it is reasonably circular.
 
3) Then twe continued to follow the beam path: we have looked at the PR mirror with a camera and we have moved the beam. We coud see a dim shape hitting the leg of the PR suspension and we
could see it moving on the leg onother side of the suspension.
 
4) We have out a camera to look at the input mirror but we coudn‘t see any light even when mcoving the beam.
 
5) Finally we have used the camera to look at the first target. We coud observe a strange, irregular light shape, much larger than the expected beam. At the beginning we thought it was the light
diffracted by the beam hitting the pipe but when we moved the PR we observed that the shape was mooving accordingly. (You can see a video here). It probably suggest that the beam is reaching the
target but it is getting clipped along its path.
 
6) As reported in entry. we have also chech the dimension of the beam on the bench finding a difference about a factor two with the value measured previously by Yuefan. Since the laser has been
mooved. it is likely that the the telescope has to be retuned. i don‘t know how much this discrepancy can contribute to the strange effect we observed.
 
We have also measured the power of the referencies. wich give a lower limits on the losses of the Faraday. For an input power of we find that the translission of the fist BPS of the Farady is and the
second one is. We have veryfled that chaging the IR polarization from s to p almost all the light is transmitted by the the first PBS
 
CONCLUSION: we suspect that the beam is clipped somewhere in its in vacuum pathhave been working on the IR aligment.
 
1) We have checked the references on the PR viewport. According to picture 1, they come from the transmission of the first and the second PBS of the Faraday. We have checked that when the beam
transmitted by the Faraday is wel centerd on the cross on the plastic ï¬Âlm out of the wieport, the other reference beam is slightly hitting the edge of the second PBS of the Faraday and this makes the
beam to have a quite irregular shape on the cross. Since the irregurity is not so evident we wondered if it was like this also when the reference was done or if it means that the beam is not well aligned.
2) We have put a camera to look inside the BS and we coud spot the beam on the 2 inch telescope mirror. By mooving the steering mirrors on the bench we moved it in order to make it hit the edge of
the mirror and we have veriï¬Âed that it is reasonably circular.
 
3) Then twe continued to follow the beam path: we have looked at the PR mirror with a camera and we have moved the beam. We coud see a dim shape hitting the leg of the PR suspension and we
could see it moving on the leg onother side of the suspension.
 
4) We have out a camera to look at the input mirror but we coudn‘t see any light even when mcoving the beam.
 
5) Finally we have used the camera to look at the first target. We coud observe a strange, irregular light shape, much larger than the expected beam. At the beginning we thought it was the light
diffracted by the beam hitting the pipe but when we moved the PR we observed that the shape was mooving accordingly. (You can see a video here). It probably suggest that the beam is reaching the
target but it is getting clipped along its path.
 
6) As reported in entry. we have also chech the dimension of the beam on the bench finding a difference about a factor two with the value measured previously by Yuefan. Since the laser has been
mooved. it is likely that the the telescope has to be retuned. i don‘t know how much this discrepancy can contribute to the strange effect we observed.
 
We have also measured the power of the referencies. wich give a lower limits on the losses of the Faraday. For an input power of we find that the translission of the fist BPS of the Farady is and the
second one is. We have veryfled that chaging the IR polarization from s to p almost all the light is transmitted by the the first PBS
 
CONCLUSION: we suspect that the beam is clipped somewhere in its in vacuum path
Images attached to this report
631_20180118151525_prscheme.png
KAGRA MIR (Absorption)
Print this report.
iManuelMarchio - 18:52, Wednesday 17 January 2018 (628)Get code to link to this report
Status of 633nm probe alignment

we found the pd DET10A for the 633nm probe didn't give any output. Replacing the power supply with a 12V battery solved the problem. We checked the power supply with a multimeter and it looks fine (9V of output). We figured out that it's the adaptor bad contact. We continue working using the battery.

We made a scan of the surf ref sample and maximized the AC signal adjusting the pump alignment.

Current experimental parameters:

Laser current 1.3A Power meter value: 31mW (without sample)

DC signal: 3.26 (with sample) (3.4V at the max of the scan)

AC signal max: 0.2V

Images attached to this report
628_20180117070208_20180117surfrefscan.png
R&D (FilterCavity)
Print this report.
AkihiroTomura, YuhangZhao - 13:41, Tuesday 16 January 2018 (627)Get code to link to this report
SHG characterization
We measured the output power of SHG (green) with respect to the LiNb temperature and the input power (IR). A figure attached shows a result. The x-axis of the right graph remains to be converted to actual temperatures (we don't know the name of the thermistor we are using at this moment).
Images attached to this report
627_20180116053714_20171226lncharacter.png
Comments related to this report
AkihiroTomura - 11:07, Wednesday 31 January 2018 (647)
I did the SHG output power vs input power measurement again. I used HWP and FI pair after the main laser to change the input power in order not to get mode-hops. During the measurement, the current value of the main laser was 2.004 A. The SHG crystal temerature was 3.170 kOhm.
Images attached to this comment
647_20180131030428_20180123lncharacter4.png
R&D (FilterCavity)
Print this report.
YuhangZhao, EleonoraCapocasa - 23:02, Monday 15 January 2018 (626)Get code to link to this report
Recovery of the cavity lock after Christmas break
Today I and Eleonora have aligned and locked the cavity again. We found that the beam was pretty misaligned (not reaching any of the two in-vacuum targets) and the range of the optical levers was too small to recover the alignement. So we had to move PR picomotor to send the beam at 300 m and INPUT mirror picomotor to superpose the reflected beam on the incoming one.
After doing that we have put to zero the local control error signals and we have closed all the local controls loops. Then, we could fine tune the alignment with the local controls and lock the cavity stably.
We took pictures of each local control panel while the cavity was locked (See figure from 1 to 4: END, INPUT, BS, PR). The lock can last for more than 30 minutes (we had to unlock on purpose before going away). See figure 5 for the transmitted beam during the lock. We took photos of the oscilloscope showing the transmitted power just after achieving the lock (pic 6) and after 30 min of locking (pic 7). In the best alignment condition it is about 1 V.
Images attached to this report
626_20180115150135_em.jpg 626_20180115150145_im.jpg 626_20180115150159_bs.jpg 626_20180115150208_pr.jpg 626_20180115150217_locking.jpg 626_20180115150229_justafterlocking.jpg 626_20180115150240_after30min.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 23:16, Friday 12 January 2018 (625)Get code to link to this report
Second time testing of grating with infrared
After talking with Matteo, I found the last measurement was needed to be improved. Firstly, the rotation direction should make sure that the incident direction is against the arrow direction marked on the edge of grating. Secondly, to find a good place to do measurement with enough times. Thirdly, to check the polarization of infrared beam. Fourthly, we also need to know the diffraction angle.

So I did experiment today. Fortunately, I found a PBS without costing a lot of time. Although yuefan told me that the most consuming part of job in TAMA is finding something.

Then I improvised the testing set-up. I found most of the light is reflected by the PBS. However, I didn't know the ratio between s and p polarization should be around 1000, which I learnt from Matteo latter on. But this can explain why the energy doesn't conserve if I put half-wave plate in-between. You can refer to attached picture 1. I will take the picture of weak light and measure it next time. As Matteo suggested, the weak signal will have less fluctuation which is important for measurement.

The power evolution is shown as attached picture 2. The attached picture 3 is about the diffraction angle.

I will do the measurement of green soon and compare it with this one. Before that we will refine the alignment of some wave plates.
Images attached to this report
625_20180112151344_polarization.jpg 625_20180112151357_power.png 625_20180112151405_angle.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 00:47, Friday 12 January 2018 (624)Get code to link to this report
Test of grating for infrared
These days I tested the grating we will use for separating green and infrared.

I use the holes on the bench to estimate the incident angle. The test are divided by two sorts. For the first one, the incident beam is aligned with the arrow on the grating. For the second one, it is opposite.

I also test it by using a wave plate. It is used to see if polarization can affect the diffraction and if so, how it can affect.
Images attached to this report
624_20180111164644_rotation1.png 624_20180111164656_rotation2.png
R&D (FilterCavity)
Print this report.
AkihiroTomura - 01:07, Wednesday 10 January 2018 (623)Get code to link to this report
Comment to Mode cleaner telescope 1 (Click here to view original report: 613)
This is a correction of the entry 613.
In the entry 613, I made a miscalculation. This entry is a correct one.


I re-calculated optimal lens pairs for a telescope for the mode cleaner (green) with accurate dimensions (see here for the drawing). Figures attached show lens pairs that seem acceptable and the parameters I used.
In the case of the first figure, the mode matching factors are 99.718% and 99.601% for vertical and horizontal axes, respectively.
In the case of the second figure, the mode matching factors are 100% and 99.727% for vertical and horizontal axes, respectively.


The lens with 75.6 mm focal length that we already have is one without any surface coating thus low transmissivity. So I think the first case is preferable.
Images attached to this comment
623_20180109165435_20180109sol1.jpg 623_20180109165443_20180109sol2.jpg 623_20180109165450_20180109sol3.jpg
R&D (FilterCavity)
Print this report.
AkihiroTomura - 16:43, Wednesday 03 January 2018 (621)Get code to link to this report
Comment to Characteristics of SHG control loop with a new servo (Click here to view original report: 620)

Figures were missed in the entry. They are shown here.

Images attached to this comment
621_20180103083821_20171221shgwithnewservo.png 621_20180103083831_20171221shgpowspeden.png 621_20180103083847_20171220061727925ios.jpg
R&D (FilterCavity)
Print this report.
AkihiroTomura, MatteoLeonardi - 16:35, Wednesday 03 January 2018 (620)Get code to link to this report
Characteristics of SHG control loop with a new servo

Open-loop transfer function

On 12/21, we measured open-loop transfer function of SHG control, because a new servo was designed and installed by Matteo Leonardi. For the previous situation, please refer to an entry 585.Fig.1 shows open loop transfer function of the SHG control. We have three switches on the new servo, each of them is corresponding to a different integrator. Measurements were conducted for three combinations of these integrators. Noise level inserted were different depending on the conditions in order to keep the cavity being locked. PZT resonance point is located aroun 25 kHz. Unity gain frequency is shown on each graph by a black line (770 Hz, 1380 Hz, and 1610 Hz). You can see a high phase margin (approx. 90 degree) when the mid integrator ON.

We also measured power spectra of an error signal on the loop and DC signal from the photo detector which is monitoring transimitted light from the SHG (Fig.2). You can see a spike around 600 Hz. I am not sure what it means.

All spectra were taken by Agilent 36540A.

Output power stability

Fig.3 shows a long term (1000 second) stability of the SHG output. Stability of the SHG is much better than before. In this moment, we do not find a large fluctuation in the SHG output (532 nm) as we did so far.

Comments related to this report
AkihiroTomura - 16:43, Wednesday 03 January 2018 (621)

Figures were missed in the entry. They are shown here.

Images attached to this comment
621_20180103083821_20171221shgwithnewservo.png 621_20180103083831_20171221shgpowspeden.png 621_20180103083847_20171220061727925ios.jpg
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 01:22, Friday 22 December 2017 (618)Get code to link to this report
Comment to abnormal behavior of local control, beam position, determination of PDH phase (Click here to view original report: 617)

A comment on the local controls large output: the voltage range at the output of the DAC is +/- 10 Volts. So If the correction signal is out of this range (as in the case of picture 1) the output is saturating and the control loop is likely not to be working properly (as can be seen from the large RMS in picture1).

More in general, looking at the error singnal, it seems to me that we are still affected by the "spikes issue". It could be useful to compare the open loop rms with the reference values in the absence of spikes.

R&D (FilterCavity)
Print this report.
YuhangZhao - 22:47, Thursday 21 December 2017 (617)Get code to link to this report
abnormal behavior of local control, beam position, determination of PDH phase
This week we installed steps for our optical bench, then we did some arrangement around the bench. It includes the rearrangement of equipments and cables.

We know for PDH control, the local oscillator phase and signal phase must match. But the cable length can affect this phase. So after changing cables, we need to set the phase of signal generator again.

Firstly, we need to close the loop for local control to make the light beam interferes well inside the cavity. However, I found the local control goes oscillation even with the previous value. The previous value means the value we used for locking yesterday. I should mention this abnormal case happened the day before yesterday.

After a very long time of adjusting, I found the the local can be fairly stable while the output of local is large(Fig 1). I adjusted it a little and make the output much smaller(but it is still around 40). Then I could find the beam through the end room camera.

Secondly, I took the photo of beam on the first iris(Fig 2,3) and the second iris(Fig 4).

Thirdly, I went to set the PDH phase. I set different value to see how the error signal looks like. I took three pictures(Fig 5,6,7) for this processes. We can deduce from them the worst case should between 80 and 90 degrees. So I set the phase as 175 degrees.

Finally, I found the transmitted signal decreases from 1V to 0.6V(Fig 8). And I found this is not caused by the SHG. Now the SHG locking is pretty pretty good! This is caused by the bad local control. The position of mirror drifts away fast. However, this maybe also caused by the large output of local control. Besides, I and Matteo measured the TF of BS pitch. Matteo said it is not a good control system according to this TF. So the local control definitely should be improved.
Images attached to this report
617_20171221144521_171221localcontrol.jpg 617_20171221144533_1irisgood.jpg 617_20171221144545_1iriswell.jpg 617_20171221144634_2irisgood.jpg 617_20171221144649_70degree.jpg 617_20171221144658_80degree.jpg 617_20171221144707_90degree.jpg 617_20171221144716_degradation.jpg
Comments related to this report
EleonoraCapocasa - 01:22, Friday 22 December 2017 (618)

A comment on the local controls large output: the voltage range at the output of the DAC is +/- 10 Volts. So If the correction signal is out of this range (as in the case of picture 1) the output is saturating and the control loop is likely not to be working properly (as can be seen from the large RMS in picture1).

More in general, looking at the error singnal, it seems to me that we are still affected by the "spikes issue". It could be useful to compare the open loop rms with the reference values in the absence of spikes.

R&D (FilterCavity)
Print this report.
MatteoLeonardi - 16:41, Thursday 21 December 2017 (616)Get code to link to this report
Comment to Mode cleaner telescope 1 (Click here to view original report: 613)
I upload the drawings of the MC (made by Uraguchi-san) as a reference for dimensions and specs.
Non-image files attached to this comment
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 04:10, Thursday 21 December 2017 (615)Get code to link to this report
Pump beam waist at different powers

see the attached document

Non-image files attached to this report
R&D (FilterCavity)
Print this report.
YuhangZhao - 22:14, Wednesday 20 December 2017 (614)Get code to link to this report
Locking of filter cavity again
After the installation of shield case and step around, we reconnect the cable. This cause the beam goes away. Now we can lock it again like figure 1.

While I check the first iris, I find the pattern on it is really similar with the pattern we get from infrared. So I think the problem should come from the bench. Refer to figure 2.

But the locking is not stable enough, I will check the phase of two signals sending to EOM. Refer to figure 3.
Images attached to this report
614_20171220141420_lock1220.jpg 614_20171220141437_iris1.jpg 614_20171220141448_signal.jpg
R&D (FilterCavity)
Print this report.
AkihiroTomura - 18:47, Wednesday 20 December 2017 (613)Get code to link to this report
Mode cleaner telescope 1
I measured dimensions of the green beam after Mach-Zehnder to design a telescope for a mode cleaner. One of attached pictures shows a result. The x-axis of this graph starts from the output of EOM. 
 
An eigenmode of the mode cleaner was calculated using values and a formula shown in a picture. A cavity length was roughly estimated from the number of holes on the optical table because I could not find it in e-logbook.
 
Based on these information, I searched for a optimal lens pair for telescope. One example is shown in a picture. In this case, mode matching factor for vertical and horizontal axis of the beam are 99.37 % and 99.56 %, respectively. 
Images attached to this report
613_20171220102445_20171220gausbeamfit.jpg 613_20171220102834_20171220092640911ios.jpg 613_20171220104537_newsol1.jpg
Comments related to this report
MatteoLeonardi - 16:41, Thursday 21 December 2017 (616)
I upload the drawings of the MC (made by Uraguchi-san) as a reference for dimensions and specs.
Non-image files attached to this comment
AkihiroTomura - 01:07, Wednesday 10 January 2018 (623)
This is a correction of the entry 613.
In the entry 613, I made a miscalculation. This entry is a correct one.


I re-calculated optimal lens pairs for a telescope for the mode cleaner (green) with accurate dimensions (see here for the drawing). Figures attached show lens pairs that seem acceptable and the parameters I used.
In the case of the first figure, the mode matching factors are 99.718% and 99.601% for vertical and horizontal axes, respectively.
In the case of the second figure, the mode matching factors are 100% and 99.727% for vertical and horizontal axes, respectively.


The lens with 75.6 mm focal length that we already have is one without any surface coating thus low transmissivity. So I think the first case is preferable.
Images attached to this comment
623_20180109165435_20180109sol1.jpg 623_20180109165443_20180109sol2.jpg 623_20180109165450_20180109sol3.jpg
R&D (FilterCavity)
Print this report.
AkihiroTomura - 13:32, Wednesday 20 December 2017 (612)Get code to link to this report
Caution about AOM alignment

After the installation of the coverage on optical bench, we are now checking the situation.

Note that our AOM is very easily misaligned by touching the cable.

Temporarily, I put a post right next to the cable in order not to make it move a lot (see a figure attached).

Images attached to this report
612_20171220053042_20171219064718302ios1.jpg