LOG-IN
Displaying reports 41-60 of 3272.Go to page 1 2 3 4 5 6 7 8 9 10 End
R&D (Speed meter)
Print this report.
YoheiNishino - 01:23, Thursday 31 October 2024 (3807)Get code to link to this report
Transfer function measurement with a full locking

Nishino,

2024.10.28 (see TF5)

I locked all degrees of freedom and measured the speed transfer function for the first time.

The total loss is estimated as 150 ppm.

Images attached to this report
3807_20241030172345_output1.png 3807_20241030172352_output2.png
R&D (FilterCavity)
Print this report.
MichaelPage - 00:04, Thursday 31 October 2024 (3806)Get code to link to this report
Infrared beams to OPO

I checked alignment of ppol to OPO. It was a bit bad (maybe 80% mode matching, fig 1). I realigned to 86.2% (fig 2). It seems in the past the alignment of ppol to OPO was never really that critical anyway.

Since the laser was just activated, the ppol lock was not so stable - this is normal for about 2 hours after laser activation. ppol PLL quite regularly unlocks and relocks when the laser is warming up but it seems to go back to normal after a while.

There is some odd behaviour where sometimes if I touch the steering mirrors a small amount there is an extra mode which flashes on and off. Normally the main ppol mode isn't that sensitive to the steering mirror tuning. Also in transmission of OPO I can see some flashing on the sensor card, displaced about 5mm in yaw from the TEM00 mode.

I looked a bit at BAB and it seems also ok, maybe just a small tweak. But I didn't touch it or take a picture yet.

R&D (FilterCavity)
Print this report.
MichaelPage - 00:03, Tuesday 29 October 2024 (3804)Get code to link to this report
IRMC auto lock

There has been a long running issue with IRMC lock ever since the main laser EOM was switched from 15 MHz to 88.3 MHz. Before, the error signal was acquired with a TAMA 15 MHz resonant PD but we since switched to a Thorlabs PDA05CF2 broadband RFPD with 75 MHz linewidth. From old logs, the error signal was about 1Vpp before the change and decreased to 100 mV after changing to new modulation scheme. However, the SHG underwent the same change and only saw a factor of 4 decrease in error signal. Initially I though there was some issue with weak IRMC modulation or demodulation signal due to weakening local oscillator synthesis from DDS, but the SHG uses the same mod/demod electronics and works mostly fine so that isn't the main problem.

We currently inject about 2.5 - 3 mW IR to IRMC with the goal of about 2 mW transmission on resonance. So off resonance about 3 mW makes it to IRMC REFL, which is a bit much for the PD. For whatever reason, at some point there was a weak ND filter Thorlabs ND03A, which resulted in a base level of > 5 V on the reflection spectrum (fig 1). With the PD close to saturation, the side spikes of the PDH signal become very small (fig 2). I changed the ND filter to Thorlabs ND1 which is what it was a while ago, which gives a base level of about 2.0 V and dip of 1.3 V (fig 3). This makes the error signal have the usual PDH shape but it also became quite small, ~ 70 mVpp.

I checked the PDH sidband amplitude by manually putting the IRMC near resonance (servo scan amplitude zero and PZT HVD offset adjusted to make the reflection signal drop by 1.3 V). I could see the 88 MHz sideband from IRMC REFL RF was about -27 dBm. I sent it to a spare RF amplifier port on the rack +18 dB and could see that the sideband was amplified to maximum -8 dBm. The error signal becomes 400 mVpp (fig 4). I adjusted the cable length a bit to improve the demodulation phase and now the IRMC can automatically lock quite stably. IRMC transmission is about 2.16 mW.

For now this is reasonable to operate. There is some issue with the servo electronics that is reducing the signal. One fix we discussed a while ago with Yuhang and Pierre Prat is to replace resistor R33 in the servo schematic (feedback resistor of an inverting op amp) with a ten times higher resistance to increase the op amp gain 10x. Also I think we could use a *slightly* weaker ND filter on IRMC REFL RF. Could also make an 88 MHz resonant PD but maybe that's too much effort. For now it's fine, we still have spare RF amplifier ports so we don't need to mess with it until we need those for other things.

Next:

  • measure IRMC tansfer function and optimise UGF
  • recover BAB alignment to OPO and homodyne
  • optimise green alignment to OPO via BAB nonlinear gain
Images attached to this report
3804_20241028155745_20241026151135.jpg 3804_20241028155749_20241026163013.jpg 3804_20241028155753_20241028181016.jpg 3804_20241028155757_20241028181038.jpg
BIGFOOT (cell birefringence)
Print this report.
ShalikaSingh - 22:44, Sunday 27 October 2024 (3803)Get code to link to this report
Measurement with translation stage

[Shalika, Katsuki]

Both hydrogel and force gauge are on translation stage. We moved the force gauge up using the actuator on translation stage.See Fig 1 and Fig 2.

hydrogel is 8-9cm in length and 1.5mm thick. Each measurement take 2:30 mins

The laser is at center

0.26 N

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Fri, Oct 25, 2024 4-02-01 PM.txt

0.28

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Fri, Oct 25, 2024 4-06-13 PM.txt

0.30

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Fri, Oct 25, 2024 4-09-53 PM.txt

0.32 N

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Fri, Oct 25, 2024 4-13-41 PM.txt

0.34 N

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Fri, Oct 25, 2024 4-18-30 PM.txt

0.36

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Fri, Oct 25, 2024 4-22-30 PM.txt

0.38

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Fri, Oct 25, 2024 4-26-43 PM.txt

0.40

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Fri, Oct 25, 2024 4-30-35 PM.txt

0.42 N

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Fri, Oct 25, 2024 4-34-02 PM.txt

0.44

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Fri, Oct 25, 2024 4-37-43 PM.txt

0.46

Fri, Oct 25, 2024 4-41-56 PM

0.48

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Fri, Oct 25, 2024 4-46-00 PM.txt

0.50

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Fri, Oct 25, 2024 4-50-26 PM.txt

 

we moved the hydrogel laterally--the beam is not in center

0.51

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Fri, Oct 25, 2024 4-54-17 PM.txt

0.48

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Fri, Oct 25, 2024 4-59-40 PM.txt

0.46

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Fri, Oct 25, 2024 5-03-11 PM.txt

0.44

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Fri, Oct 25, 2024 5-06-56 PM.txt

0.42

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Fri, Oct 25, 2024 5-11-10 PM.txt

0.40

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Fri, Oct 25, 2024 5-17-25 PM.txt

0.38

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Fri, Oct 25, 2024 5-21-26 PM.txt

0.36

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Fri, Oct 25, 2024 5-24-52 PM.txt

0.34

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Fri, Oct 25, 2024 5-29-02 PM.txt

0.32

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Fri, Oct 25, 2024 5-32-51 PM.txt

The hydrogel maybe is a bit dry. So, the 0.32 was the minimum the translation stage could move.

We moved the hydrogel laterally to another side.

0.32

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Fri, Oct 25, 2024 5-41-43 PM.txt

0.36

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Fri, Oct 25, 2024 5-45-22 PM.txt

we stop coz the force change by itself.

Images attached to this report
3803_20241027144929_20241025155014.jpg 3803_20241027144938_20241025165533.jpg
Comments related to this report
ShalikaSingh - 14:55, Monday 11 November 2024 (3827)

the results (birefringence and diattenuation) for both points are attached. 

Fig 1 , 2 - point where laser is at center

Fig 3, 4 - point where laser is on one side (not center)

BIGFOOT (cell birefringence)
Print this report.
ShalikaSingh - 15:33, Thursday 24 October 2024 (3800)Get code to link to this report
Comment to hydrogel meaurement with new force gauge (Click here to view original report: 3795)

The analysis done only for the 2nd hydrogel (the one which was measured after the first one broke). The point with force 0N is omitted, since the actual 0N should be the point of start of seeing the force change. The birefringence was measured at 2 points

Fig1:Birefringence vs Force of point 1

Fig 2: Diattenuation vs Force  of point 1

Fig 3:Birefringence vs Force of point 2

Fig 4: Diattenuation vs Force  of point 2

The number plotted is the median obtained from the data at each force value.

Update 1: I changed the plot becauase I had previously overconverted rad to rad, assuming it was degrees.

Update 2: Added Error bar to fitted data

Images attached to this comment
3800_20241025060710_birefringencevsforcepoint1.png 3800_20241025060714_diattenuationvsforcepoint1.png 3800_20241025060718_birefringencevsforcepoint2.png 3800_20241025060723_diattenuationvsforcepoint2.png
BIGFOOT (Characterization)
Print this report.
ShalikaSingh - 15:49, Tuesday 22 October 2024 (3797)Get code to link to this report
BB1E03 at 45 deg

with BS removed, I aligned the BB1EO3 at 45 degree. This is to check if Polarization ray tracing analysis is correct for coatings. 

LC goes from 0-3.5V with 0.1V step and 100 averaging

C:\Users\atama\Dropbox\LC-Experiment\Measurement Data\Birefringence Measurements\Coating measurement\BB1-E03\20241022\Tue, Oct 22, 2024 3-37-46 PM.txt

BIGFOOT (cell birefringence)
Print this report.
ShalikaSingh - 23:37, Monday 21 October 2024 (3795)Get code to link to this report
hydrogel meaurement with new force gauge

[Katsuki, Shalika, Marc]

1.5 mm hydrogel, recipe same as previous. The error on force can be ±0.01 N. The measurement is taken in transmission of LC without BS. 

20 avg, 0.5V-3V with 0.25V

0 N

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Mon, Oct 21, 2024 5-15-53 PM.txt

0.28 N

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Mon, Oct 21, 2024 5-22-16 PM.txt

0.31 N

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Mon, Oct 21, 2024 5-27-33 PM.txt

0.36 N

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Mon, Oct 21, 2024 5-31-40 PM.txt

0.41 N

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Mon, Oct 21, 2024 5-36-20 PM.txt

hydrogel broke -- Also, there could be an issue of hydrogel breaking due to it getting dried. So, reduced to 10 avg for faster measurement.

0 N

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Mon, Oct 21, 2024 5-44-04 PM.txt

0.28 N

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Mon, Oct 21, 2024 5-47-38 PM.txt

0.30 N

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Mon, Oct 21, 2024 5-50-56 PM.txt

0.37 N

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Mon, Oct 21, 2024 5-54-23 PM.txt

0.42 N

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Mon, Oct 21, 2024 5-58-02 PM.txt

0.45

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Mon, Oct 21, 2024 6-02-41 PM.txt

0.51

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Mon, Oct 21, 2024 6-06-23 PM.txt

we moved the translation stage by (23.2-20.2) cm

0.51 N

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Mon, Oct 21, 2024 6-12-19 PM.txt

0.45 N

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Mon, Oct 21, 2024 6-15-46 PM.txt

0.42 N

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Mon, Oct 21, 2024 6-20-16 PM.txt

0.37 N

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Mon, Oct 21, 2024 6-23-50 PM.txt

0.31 N

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Mon, Oct 21, 2024 6-27-47 PM.txt

0.28 N

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Mon, Oct 21, 2024 6-31-09 PM.txt

0 N

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Mon, Oct 21, 2024 6-34-17 PM.txt

Input without any sample

C:\Users\atama\Dropbox\Cell Birefringence\Measurement data\hydrogel\Force gauge v2\Mon, Oct 21, 2024 6-37-59 PM.txt

Comments related to this report
ShalikaSingh - 15:33, Thursday 24 October 2024 (3800)

The analysis done only for the 2nd hydrogel (the one which was measured after the first one broke). The point with force 0N is omitted, since the actual 0N should be the point of start of seeing the force change. The birefringence was measured at 2 points

Fig1:Birefringence vs Force of point 1

Fig 2: Diattenuation vs Force  of point 1

Fig 3:Birefringence vs Force of point 2

Fig 4: Diattenuation vs Force  of point 2

The number plotted is the median obtained from the data at each force value.

Update 1: I changed the plot becauase I had previously overconverted rad to rad, assuming it was degrees.

Update 2: Added Error bar to fitted data

R&D (Speed meter)
Print this report.
YoheiNishino - 20:17, Friday 18 October 2024 (3792)Get code to link to this report
Transfer function measurement with a near-optimal PCC and signal phase

Nishino

In my calculation, the best PCC round-trip phase is 1) when the transmittance light from ETM gets maximum AND 2) monitoring the transmittance light from the PCM, the relative angle between the signal and main beam becomes pi/4. 

Adjusting the PCC angle to meet 1) and the signal beam angle to meet 2) condition by hand, I obtained the result shown in Fig. 2). Red is the speed state and blue is the position state. Somehow it behaves weird at high-frequencies but taking their ratio (green) it could be fitted well with a model. I guess the photo-detetor calibration went wrong for some reason.

Images attached to this report
3792_20241018131207_pcsmexperiment68.jpg 3792_20241018131216_output.png
BIGFOOT (LC)
Print this report.
ShalikaSingh - 08:30, Thursday 17 October 2024 (3778)Get code to link to this report
BS Front face and BB1EO3

I rotated BS to align the arrows with the beam propagation direction. 

BS Font face

C:\Users\atama\Dropbox\LC-Experiment\Measurement Data\Birefringence Measurements\Beam Splitter Calibration\20241006\front_refl\Tue, Oct 8, 2024 5-18-51 PM.txt

C:\Users\atama\Dropbox\LC-Experiment\Measurement Data\Birefringence Measurements\Beam Splitter Calibration\20241006\front_trans\Tue, Oct 8, 2024 6-59-48 PM.txt

The large birefringence(23 deg) in reflection was confusing. So, alignment was improved more. Redid measurement in reflection of BS front face. 

C:\Users\atama\Dropbox\LC-Experiment\Measurement Data\Birefringence Measurements\Beam Splitter Calibration\20241009\front_refl\Wed, Oct 9, 2024 10-26-57 PM.txt

I mounted mirror(BB1EO3 sample 1) in the transmission of the BS front face. The mirror is aligned to obtain reflection at optimal position. Then I flipped BS. This was to ensure that in case the BS moved, I can understand that its from BS flipping and not mirror misalignment. After flipping BS, the beam maintained its position after reflection from coating. We don't care about the reflection from the back face of the BS anymore. 

Polarization states generated with BS back face transmission:

C:\Users\atama\Dropbox\LC-Experiment\Measurement Data\Polarization states\20241010\Thu, Oct 10, 2024 4-20-30 PM.txt

In the reflection of the mirror, after BS:

C:\Users\atama\Dropbox\LC-Experiment\Measurement Data\Birefringence Measurements\Coating measurement\BB1-E03\20241010\Thu, Oct 10, 2024 5-40-42 PM.txt

Also, the measurement takes only 1 hour. Maybe it improved sometime from labview modifications. I never noted the time of measurements after the 1st labview was made, which was taking rather 2 hrs. 

R&D (Speed meter)
Print this report.
YoheiNishino - 22:11, Tuesday 15 October 2024 (3791)Get code to link to this report
Transfer function measurement of the speed meter

Nishino,

I measured transfer functions of the speed and position meter. The PCC is not locked but just scan the PCM to minimize and maximize the transmittance light which is a consequence of interference between the first and second circulation light. 

The result is shown in the figure. Blue is the position state and red is speed state. Their ratio is fitted with a function with a cavity loss of 500 ppm. 

Images attached to this report
3791_20241015151132_tfs.png
KAGRA MIR (Polarization)
Print this report.
MarcEisenmann - 18:54, Tuesday 15 October 2024 (3790)Get code to link to this report
birefringence measurements

with 1 mm step size, 14mm radius

Input = ['Fri, Oct 11, 2024 11-36-41 AM.txt', 'Fri, Oct 11, 2024 1-34-57 PM.txt' , 'Fri, Oct 11, 2024 2-13-44 PM.txt' , 'Fri, Oct 11, 2024 2-51-14 PM.txt' , 'Fri, Oct 11, 2024 3-41-43 PM.txt' , 'Fri, Oct 11, 2024 4-52-41 PM.txt']

iLM/LMA ; white dot placed on top toward laser source

Output = ['Fri, Oct 11, 2024 11-40-34 AM.txt' , 'Fri, Oct 11, 2024 1-36-56 PM.txt', 'Fri, Oct 11, 2024 2-17-22 PM.txt' , 'Fri, Oct 11, 2024 2-53-16 PM.txt' , 'Fri, Oct 11, 2024 3-46-31 PM.txt' , 'Fri, Oct 11, 2024 4-55-02 PM.txt']

AZTEC 30mm number 1 (number is attributed randomly by me).

Output = ['Tue, Oct 15, 2024 10-21-16 AM.txt' , 'Tue, Oct 15, 2024 9-45-08 AM.txt' , 'Fri, Oct 11, 2024 9-05-53 PM.txt' , 'Fri, Oct 11, 2024 8-26-53 PM.txt' , 'Fri, Oct 11, 2024 7-53-34 PM.txt', 'Fri, Oct 11, 2024 7-17-27 PM.txt']

AZTEC 30mm number 2 (number is attributed randomly by me).

Output = ['Tue, Oct 15, 2024 12-03-02 PM.txt' , 'Tue, Oct 15, 2024 1-20-44 PM.txt' , 'Tue, Oct 15, 2024 2-01-50 PM.txt' , 'Tue, Oct 15, 2024 2-40-34 PM.txt' , 'Tue, Oct 15, 2024 3-27-17 PM.txt' , 'Tue, Oct 15, 2024 4-20-30 PM.txt']

R&D (Speed meter)
Print this report.
YoheiNishino - 12:19, Tuesday 15 October 2024 (3784)Get code to link to this report
Open loop gain measurement of the IR cavity lock

Nishino,

This is a result on 2024/10/11.

I measured the open loop transfer function (OLTF, Fig.1) of the GR lock loop. The filter funtions are measured independently (Fig. 2 and 3).

Setting of SR560 (Fig. 5) and Moku (Fig. 6):

SR560: 0.3 Hz, first order low pass, gain=10000

Moku PID:  I=2.87 kHz, DS=25.0 dB

The inferred optical gain (not optical gain itself but A*H) is plotted in Fig. 4. From this result, I reduced the overall gain just by -2 dB, then the modified OLTF looks like Fig. 7, with UGF=2 kHz and phase margin=20 degrees.

Images attached to this report
3784_20241015045830_oltf.png 3784_20241015045838_lpfsasr560.png 3784_20241015045846_moku.png 3784_20241015045852_ah.png 3784_20241015050000_img1036.jpg 3784_20241015050800_img1105.jpg 3784_20241015051741_oltfmodified.png
R&D (Speed meter)
Print this report.
YoheiNishino - 11:55, Tuesday 15 October 2024 (3783)Get code to link to this report
Open loop gain measurement of the GR cavity loop

Nishino,

This is a result on 2024/10/10.

I measured the open loop transfer function (OLTF, Fig.1) of the GR lock loop. The filter funtion and actuator efficiency is measured independently (Fig. 2 and 3, Fig. 4 is their sum).

Setting of SR560 (Fig. 7) and Moku (Fig. 8):

SR560: 3 Hz, first order low pass, gain=100

Moku PID: P=10 dB, D=4.118 kHz, DS=13.0 dB

The inferred optical gain is plotted in Fig. 5. From this result, I modified the shape of filter by Moku PID controller as:

P = 10 dB, D = 20 dB, DS = 8 kHz.

Then the modified OLTF is plotted in Fig. 6 with UGF=4 kHz and phase margin=50 degrees.

Images attached to this report
3783_20241015044504_oltf.png 3783_20241015044514_lpfsasrmoku.png 3783_20241015044523_pztamp.png 3783_20241015044528_lpfsasrmokupztamp.png 3783_20241015044533_h.png 3783_20241015044609_oltfmodified.png 3783_20241015045223_img0237.jpeg 3783_20241015045229_img0238.png
General (General)
Print this report.
ShalikaSingh - 10:24, Wednesday 09 October 2024 (3777)Get code to link to this report
Old elogs are looking strange

It seems there is some issue with all elogs made before the server change. They are looking weird. For instance https://gw-elog.mtk.nao.ac.jp/osl/index.php?r=3342

Due to this issue, sometimes the search filter fails, because it is unable to find the word(since it looks strange). The issue happens only if I search and go to the elog. Perhaps not if I go directly without searching through keyword option. 

Images attached to this report
3777_20241009032547_screenshot20241009102640.png
BIGFOOT (Characterization)
Print this report.
ShalikaSingh - 16:21, Tuesday 08 October 2024 (3773)Get code to link to this report
Comment to Re-measuring BS birefringence (Click here to view original report: 3766)

I made a mistake in measuring the BS. I rotated only 90deg. It was supposed to be rotated by 180 deg.

BIGFOOT (cell birefringence)
Print this report.
MarcEisenmann - 10:16, Monday 07 October 2024 (3772)Get code to link to this report
adapter for force gauge

I designed some adapters between force gauge and hydrogel holders.

They're 3D printed with help of Kanzawa-san and models are saved in the dropbox/mount and holder folder.

BIGFOOT (Characterization)
Print this report.
ShalikaSingh - 08:14, Monday 07 October 2024 (3771)Get code to link to this report
Comment to S7 birefringence measurement (Click here to view original report: 3753)

The birefringence obtained is Fig 1. Since there were too few points (112/6050) after filtering. I will re take measurements. 

Images attached to this comment
3771_20241007011455_s7ret.png
BIGFOOT (cell birefringence)
Print this report.
ShalikaSingh - 08:13, Monday 07 October 2024 (3770)Get code to link to this report
Comment to Hydrogel measurement with displacement (Click here to view original report: 3760)

The diattenuation vs. Displacement is as in Fig 1. Doesn't change with increasing displacement. 

Images attached to this comment
3770_20241007011318_diattenuationvsdisplacement.png
BIGFOOT (Characterization)
Print this report.
ShalikaSingh - 08:05, Monday 07 October 2024 (3767)Get code to link to this report
Comment to Re-measuring BS birefringence (Click here to view original report: 3766)

The birefringence of BS changed from 17deg to 10 deg after alignment tuning. See the picture for its Jones matrix (Fig1) and birefringence (Fig2). As you can see the JM doesn't have -ve sign any more in the real parts, and is closer to identiy matrix.

The condition on physical realizablity had error bar of 1E-4 (to lower two peaks on BS birefringence).

Polarization ray tracing to correct coordinate systems seems correct. Perhaps can tune BS alignment a bit more. 

Images attached to this comment
3767_20241007010421_bsjm.png 3767_20241007010426_bsret.png
BIGFOOT (Characterization)
Print this report.
ShalikaSingh - 18:36, Friday 04 October 2024 (3766)Get code to link to this report
Re-measuring BS birefringence

Owing to the spurious 17deg of circular retardation observed in BS birefringence characterisitics after correction using Polarization ray tracing, I redid calibration.

1. First it seems the alignment was off by 0.2 cm before the BS. I corrected this using the last steering mirror before the LC path. The beam height is now 7.5cm all along. The alignment was also checked with BS and was slightly(0.1cm) tuned to obtain good alignment in its reflection. 

2. The BS is removed. LC voltage changes from 0-3.5V with 0.1V step and 100 avg.

Polarization states without BS:

C:\Users\atama\Dropbox\LC-Experiment\Measurement Data\Polarization states\20241004\Fri, Oct 4, 2024 6-35-14 PM.txt

3. The BS is rotated by 180 degrees to measure its back face.

Polarization states after relfection from back side of BS:

C:\Users\atama\Dropbox\LC-Experiment\Measurement Data\Birefringence Measurements\Beam Splitter Calibration\20241006\back_refl\Sun, Oct 6, 2024 5-32-19 PM.txt

Comments related to this report
ShalikaSingh - 08:05, Monday 07 October 2024 (3767)

The birefringence of BS changed from 17deg to 10 deg after alignment tuning. See the picture for its Jones matrix (Fig1) and birefringence (Fig2). As you can see the JM doesn't have -ve sign any more in the real parts, and is closer to identiy matrix.

The condition on physical realizablity had error bar of 1E-4 (to lower two peaks on BS birefringence).

Polarization ray tracing to correct coordinate systems seems correct. Perhaps can tune BS alignment a bit more. 

ShalikaSingh - 16:21, Tuesday 08 October 2024 (3773)

I made a mistake in measuring the BS. I rotated only 90deg. It was supposed to be rotated by 180 deg.