LOG-IN
Displaying reports 601-620 of 3273.Go to page Start 27 28 29 30 31 32 33 34 35 End
R&D (FilterCavity)
Print this report.
YuhangZhao - 10:31, Monday 24 October 2022 (3093)Get code to link to this report
GRMC and OPO aligned back

Yuhang, Michael, Marc

After aligning back SHG, we worked on aligning back GRMC and OPO.

For GRMC, we noticed that the EOM which was used for GRMC and FC locking must be remained if we don't want to adjust mode matching. The influence of this EOM on mode matching is substantial. 

For OPO, we noticed that the beam is going through the two lenses at very low place but not the center. There is an iris on the beam path going to OPO, which is checked again to be centered well.

The green beam going to the filter cavity is still misaligned at the stage of AOM. However, this will require another RF source.

The new OPO could be replaced soon.

R&D (FilterCavity)
Print this report.
YuhangZhao - 11:44, Tuesday 18 October 2022 (3092)Get code to link to this report
Comment to SHG temporary power reference levels (Click here to view original report: 3091)

Yuhang and Marc

We found that the SHG is misaligned this morning. This misalignment cannot be corrected by only tilting the steering mirrors before SHG. This is a bit strange for us. The situation of the misalignment is shown in the attached figure.

It was found later that this is caused by: the small scan amplitude of the locking servo.

The scan amplitude is adjusted to be small so that the lock loop can be engaged around the TEM00 resonant peak. However, today, due to the drift of cavity length, the small scan range is located accidentally around the first order resonance peak. This created the confusion at the beginning.

Images attached to this comment
3092_20221018044410_2449420449222702925520221018114343.jpg
R&D (FilterCavity)
Print this report.
MichaelPage - 18:33, Monday 17 October 2022 (3091)Get code to link to this report
SHG temporary power reference levels

Yuhang and Michael

We locked the SHG in the new setup and took some temporary power level references. The locking is not perfect - the SHG temperature was optimised (reading 3.140 on TED200C) to maximise the amount of transmitted green power, and by leaving a cursor level we could see that the lock was offset a bit from the maximum level of the transmission peak. This requires some adjusting of the custom servo filter (i.e. ask Pierre Prat...). Anyway, that's why this is only temporary.

Figure 1 shows the new SHG setup. We measured the following power levels:

Incident IR (measured before dichroic): 684.3 mW
Reflected green (measured before FI): 237 mW
Reflected IR (measured before ND2 on photodetector): ~ 87 mW.

Compared to previous measurements, it is almost exactly the same. A 1:1 efficiency calculation p_green/p_IR gives 0.347.

Images attached to this report
3091_20221017112932_shgsetup.png
Comments related to this report
YuhangZhao - 11:44, Tuesday 18 October 2022 (3092)

Yuhang and Marc

We found that the SHG is misaligned this morning. This misalignment cannot be corrected by only tilting the steering mirrors before SHG. This is a bit strange for us. The situation of the misalignment is shown in the attached figure.

It was found later that this is caused by: the small scan amplitude of the locking servo.

The scan amplitude is adjusted to be small so that the lock loop can be engaged around the TEM00 resonant peak. However, today, due to the drift of cavity length, the small scan range is located accidentally around the first order resonance peak. This created the confusion at the beginning.

R&D (FilterCavity)
Print this report.
YuhangZhao - 11:12, Monday 17 October 2022 (3090)Get code to link to this report
SHG alignment and mode matching are basically recovered as the old situation

After replacing wedged crystal EOM with the old one, we had alignment and mode matching seriously messed up.

Now the alignment and mode matching have been recovered.

The higher order modes and TEM00 power are shown in the attached figures.

Now, the mode matching level is: 1.68/(1.68+0.036+0.052) = 95%

Images attached to this report
3090_20221017040843_wechatimg73.jpeg 3090_20221017040851_wechatimg74.jpeg 3090_20221017040857_wechatimg75.jpeg
R&D (FilterCavity)
Print this report.
YuhangZhao - 10:09, Thursday 13 October 2022 (3089)Get code to link to this report
Replacement of main laser old Newport EOM with a new QUBIG wedged crystal EOM (2)

After the beam size issue is solved, we proceeded to recover the alignment of SHG. Since we touched several mirrors for the path from new EOM to SHG, we had to make sure several mirrors and lens are in relatively good position.

We made several mistakes since the reflected beam from the 2-inch BS is not going along the holes on the bench. To make sure this reflected beam is roughly going along the good path, we used two lens just after the BAB pick-off BS as reference. But we found that these two lens are roughly in the imaginary plane with each other. So we need to find other reference for the rough alignment of the 2-inch BS. These references were selected to be the two lenses in front of OPO. After that we fixed 2-inch BS and the following lens.

Since the alignment was completely messed up, we also set up a camera after for SHG transmission. For monitoring higher order modes, a power meter was set up in the reflection of SHG. Now we started to align back SHG.

R&D (Cryogenic)
Print this report.
KoheiMitsuhashi - 11:11, Wednesday 12 October 2022 (3088)Get code to link to this report
Measured the beam size

Mitsuhashi

I measured the beam size and fitted beam propagation.
When I measured the beam size, I used ND(2.0) filter.
The result of it was as follows and attached.The width of error bar are 2σ.

x[mm] Radius of the beam [mm]
447 0.1518±0.0040
465 0.1480±0.0022
488 0.1629±0.0033
507 0.1819±0.0027
514 0.2059±0.0035
530 0.2306±0.0029
Images attached to this report
3088_20221012040922_resultoffitting.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 17:00, Saturday 08 October 2022 (3086)Get code to link to this report
Laser current dependent beam size variation

Attached figure 1 is a characterization of laser current dependent beam size measurement. The measurement is done at a point shown in the attached figure 2.

There was two abrupt beam parameter change points. They coincides with the change of half wave plate before Faraday isolator, which is used to reduce laser power going to the beam profiler. It seems that there is also depdence of the measured beam size on this half wave plate. Or maybe the measurement depends on the laser power going to the beam profiler. A slightly more investigation would be better.

The half wate plate angle was originally 38 degrees. 

Images attached to this report
3086_20221008095440_figure1.png 3086_20221008100652_308620221008100040eomreplacement1008.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 00:19, Saturday 08 October 2022 (3085)Get code to link to this report
Wired beam size behavior and new beam parameter measurements after EOM replacment

Yuhang and Marc

As reported in the last logbook 3084, where we found the beam size changed significantly. Fortunately, we have a reported beam parameter measurement four years ago (logbook 1079). Thus, to compare, we measured again the beam parameter (relative to the first hole after the 2-inch BS). The new measurement is shown in Fig.1. According to this new measurement, we made a comparison of beam waist positions on the scheme as Fig. 2. We note that although the beam waist position is measured to be very different, the beam waist size is the same as before.

Images attached to this report
3085_20221007171746_eomreplacementbeammeas.png 3085_20221008075739_308520221007171755eomreplacementbeammeas2.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 22:50, Friday 07 October 2022 (3084)Get code to link to this report
Replacement of main laser old Newport EOM with a new QUBIG wedged crystal EOM

Yuhang and Marc

It was verified in logbook 3081 that the main laser EOM (as shown in the attached figure) was introducing significant amount of phase noise for the squeezing measurement. However, a better way of reducing the modulation caused by EOM is to use mode cleaner to reduce it. For instance, the new EOM has modulation frequency of 88MHz, which will be filterted out with a factor of about 4000.

r1 = r2 = 0.992

fin = np.pi*np.sqrt(r1*r2)/(1-r1*r2)

l = 0.5844

FSR = c/l

spac = 88e6/FSR

r1 = 0.992
r2 = 0.992
fin = np.pi*np.sqrt(r1*r2)/(1-r1*r2)
l = 0.022+np.sqrt(0.281**2+0.011**2)*2
FSR = c/l
spac = 88e6/FSR
r1 = 0.992
r2 = 0.992
fin = np.pi*np.sqrt(r1*r2)/(1-r1*r2)
l = 0.022+np.sqrt(0.281**2+0.011**2)*2
FSR = c/l
spac = 88e6/FSR

1/(1+4*(fin/np.pi)**2*np.sin(np.pi*spac)**2) = 0.0002449

Thus the phase noise caused by EOM modulation should be removed to a negilible level.

On the other hand, the new wedged EOM will remove residual amplitude modulation. However, the old EOM doesn't have wedge so that the introduction of new EOM deflects the laser beam. This causes a substantial alignment change, although we expect small beam path length change. To handle the alignment change, we decide to put the wedge oriented in the horizontal direction. In this situation, the misalignment will happen without causing beam height change. In general, we would like to avoid beam height variation.

For the SHG/BAB path, we could correct alignment change by the 2-inch BS after EOM. For the LO/PLL path, we set up two one-inch mirror with small mirror mount to correct the beam tilt. Using this strategy, we started to align beam again to SHG and IRMC as the first step. In particular, we found that we resulted to have a very large beam in front of SHG by moving just the 2-inch BS to recover alignment. This is out of our expectation because we expect small beam path length change. However, now it seems to be not the situation. To solve this problem, we decided to measure beam parameter again after the 2 inch BS. This measurement will be reported in the next logbook since it's a slightly different topic.

Images attached to this report
3084_20221007162052_eomreplacement.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 16:59, Wednesday 05 October 2022 (3081)Get code to link to this report
Comment to Measurement of sqz/asqz when pump power is scanned (Click here to view original report: 3080)

The extracted optical losses and phase noise are 22.51+/-3.49 % and 8.88 +/- 55.25 mrad.

The error bar for phase noise is very large since points are scattered around the fitted curve. Either more precise measurement or more points would help to reduce the error bar.

Images attached to this comment
3081_20221005100007_s.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 16:29, Wednesday 05 October 2022 (3080)Get code to link to this report
Measurement of sqz/asqz when pump power is scanned

Measurement conditions:

MZ 4.6, CC1 44.4 temp 7.165, sqz 110, asqz 160 

MZ 4.7, CC1 51.2 temp 7.172, sqz 125, asqz 180

MZ 4.4,                 temp 7.163, sqz 110, asqz 150

Measurement results:

sqz: 5.41, 6.34, 6.03

asqz: 10.84, 14.63, 16.64

Images attached to this report
3080_20221005093204_s1.png 3080_20221005093211_s2.png 3080_20221005093220_s3.png
Comments related to this report
YuhangZhao - 16:59, Wednesday 05 October 2022 (3081)

The extracted optical losses and phase noise are 22.51+/-3.49 % and 8.88 +/- 55.25 mrad.

The error bar for phase noise is very large since points are scattered around the fitted curve. Either more precise measurement or more points would help to reduce the error bar.

R&D (FilterCavity)
Print this report.
MarcEisenmann - 14:46, Wednesday 05 October 2022 (3079)Get code to link to this report
Comment to investigation of spurious OPO seed from LO (Click here to view original report: 3076)

Marc, Matteo, Yuhang

We found out that we were only rotating the PBS circular dump and not the PBS itself..

We removed the circular dump and could see some beam. After a little realignment of the LO back-reflection to the FI, we measured :

Pinc = 3.54 mW, 12.6 uW and 8.4 uW rejected by the first PBS respectively towards top and bottom, 3.5 mW rejected by the second PBS toward the top and 0.4 uW transmitted by the FI.

This is closer to what we expect, especially because the 8.4 uW and 3.5 mW measured have some error due to space constraints to place the power-meter.

With this measurement, we can estimate the FI isolation to be about 40 dB which is in agreement with the specsheet.

We reinstalled the ND filters on the LO path, realigned the LO into the AMC, removed the steering mirror we used to increase the back-scatter light of LO (ie recovered the usual situation) and measured 3nW incident on the FI. It means that the LO seed should be about 0.3 pW.

Then, we measured the FI transmission using the BAB. We optimized the PLL frequency to be at maximum OPO transmission and got 342 uW incident on FI and 328 uW transmitted.

Finally, we were also able to remove one titled HWP on the squeezing path towards the Homodyne because we were able to use the HWP just after the FI to remove the p pol as seen in the AMC scan.

R&D (FilterCavity)
Print this report.
YuhangZhao - 11:06, Wednesday 05 October 2022 (3078)Get code to link to this report
Squeezing measurement after reducing the phase modulation on main laser

Yuhang, Marc, Matteo

The main laser phase modulation introduces phase noise for squeezing measurement. This phase modulation is reduced from DDS setting by a factor of 4. Before and after this modulation reduction, we measured squeezing level. The result of squeezing measurement is shown in the attached figures.

Since there are peaks appearing in the squeezing measurement, to compare measuremet in a fair manner, I did an average of data from 9.6kHz to 16kHz for each measurement separately. We see that the squeezing level differs by 0.2dB, it seems that this gain of 0.2dB could come from the reduction of phase noise. To confirm this phase noise, we would make a squeezing measurement at a higher pump power situation (the current measurement used 40mW pump with 4.5 MZ offset). CC1 pk-pk was 41.2mW

Images attached to this report
3078_20221005040823_capture.png
R&D (FilterCavity)
Print this report.
MichaelPage - 14:34, Monday 03 October 2022 (3077)Get code to link to this report
KAGRA filter cavity - TAMA PLL noise

Yuhang and Michael

This measurement is more for the purpose of the KAGRA filter cavity project. We would like to optimise the design of the length control for the filter cavity. This involves setting the parameters of the control loop (bandwidth, filters, RMS noise etc) as well as choosing a control scheme of one of the three available for frequency dependent squeezing (A+ Resonant Locking Field, AdV+ Subcarrier, TAMA CCFC).

The performance of the filter cavity length control is in principle limited by two noise types:

  • backscattering - the length noise of the filter cavity causes fluctutations of incident coherent fields, most notably the residual interferometer dark port power that leaks through the squeezing path Faraday isolators. This light has length noise imprinted on it by the filter cavity which degrades squeezing at frequencies where mirror motion is dominant (~ < 10 Hz). Backscatter can be decreased by increasing the gain or bandwidth of the control loop. However, increasing the gain at low frequency causes issues with stability, and increasing the bandwidth causes issues with...
  • sensing noise - primarily caused by phase (frequency) noise of the voltage controlled oscillator that generates the sensing field for filter cavity locking. The phase noise of the sensing light causes locking fluctuations of the filter cavity length, which in the case of sensing noise can't be suppressed by increasing the gain of the control loop. In the A+ case, the phase noise of the AOMs that generate the resonant locking field is quite severe and the reinjection of sensing noise becomes prominent at and above about 10 Hz.

Changing the control bandwidth will raise one of these noises and lower the other, and the "bucket" is of the order 10-30 Hz for ~ 100m filter cavities with km scale GW detectors. Thus, we would like to measure the frequency or phase noise of the TAMA PLL at low frequency, as a placeholder for the KAGRA filter cavity sensing control noise. The PLL generates a sine wave that is the frequency difference of the main laser and CC/p-pol. Obviously we would like to keep this difference as stable as possible. We can beat this sine wave with a different one in order to check the stability. When the frequency of the signal and LO is the same, provided that the LO is stable enough, the voltage coming out of the mixing between the local oscillator and PLL is proportional to the phase noise. The relevant equations for this principle of measurement are given in Yuhang's thesis, 4.25 to 4.31. By default, we are using 190 MHz for the CC PLL and 250 MHz for p-pol, but the digital system in TAMA can only generate sufficiently clean signals up to 100 MHz. So we lowered them to 50 MHz. 

The spectrum of the phase noise is related to the spectrum of the mixed voltage by:

PSD (rad/rtHz) = PSD (V/rtHz) / Apk^2

Where Apk^2 is the amplitude of the oscillation that occurs when we change the LO by a small amount.

For the actual measurement, I used the P-POL PLL MON channel as the signal and the CC PLL LO (DDS3 board, Channel 0) as the local oscillator. The level of the DDS3 Ch0 output at 50 MHz was found to be 7.61 dBm after removing the attenuator that is normally attached to it. The signal and LO were send to a small mixer. By setting DDS3 Ch0 to 50.0001 MHz, the Apk is found to be 0.083 V. Then, resetting it back to 50 MHz, the signal and LO are combined at the mixer to give the phase noise spectrum. Of course, you should make sure that the PLL has remained locked throughout the measurement :).

Figure 1 shows the spectrum of PLL phase noise. Figure 2 shows a previous measurement from Yuhang's thesis. In general seems mostly consistent but I think we have more noise in the tens of Hz band right now.

Images attached to this report
3077_20221003072635_tamapll0930.jpg 3077_20221003073403_yuhangpll.png
R&D (FilterCavity)
Print this report.
MarcEisenmann - 15:46, Saturday 01 October 2022 (3076)Get code to link to this report
investigation of spurious OPO seed from LO

Marc, Yuhang

As described in McKenzie PhD, a possible coupling mechanism of pump RIN coupling to squeezing measurement is the presence of a spurious IR seed.

In our case, this could come from spurious LO back-reflected from Homodyne to the OPO or from IR coming from the SHG.

We started investigation on the first case by checking if the FI after the OPO has good attenuation factor.

We removed the Homodyne BS and installed a reflective mirror before the AMC.

We injected BAB and overlapped these 2 beams.

Then, we removed BAB and check the LO power around the FI.

We checked the OPO Trans PD and found really low power. We tuned the steering mirror just before and got TEM00 peak height at about 2V.

We got about 420 uW incident, 400 uW reflected (to the top) but only 10 nW transmitted... Despite alignment tuning we could not find where the missing 20uW are..

We removed the OD filters on the LO path but we had similar power ratio.

We tried to tune the polarizer of the FI (furthest from OPO) by about 20 deg but the transmitted power did not change at all...

Comments related to this report
MarcEisenmann - 14:46, Wednesday 05 October 2022 (3079)

Marc, Matteo, Yuhang

We found out that we were only rotating the PBS circular dump and not the PBS itself..

We removed the circular dump and could see some beam. After a little realignment of the LO back-reflection to the FI, we measured :

Pinc = 3.54 mW, 12.6 uW and 8.4 uW rejected by the first PBS respectively towards top and bottom, 3.5 mW rejected by the second PBS toward the top and 0.4 uW transmitted by the FI.

This is closer to what we expect, especially because the 8.4 uW and 3.5 mW measured have some error due to space constraints to place the power-meter.

With this measurement, we can estimate the FI isolation to be about 40 dB which is in agreement with the specsheet.

We reinstalled the ND filters on the LO path, realigned the LO into the AMC, removed the steering mirror we used to increase the back-scatter light of LO (ie recovered the usual situation) and measured 3nW incident on the FI. It means that the LO seed should be about 0.3 pW.

Then, we measured the FI transmission using the BAB. We optimized the PLL frequency to be at maximum OPO transmission and got 342 uW incident on FI and 328 uW transmitted.

Finally, we were also able to remove one titled HWP on the squeezing path towards the Homodyne because we were able to use the HWP just after the FI to remove the p pol as seen in the AMC scan.

R&D (FilterCavity)
Print this report.
YuhangZhao - 19:04, Thursday 29 September 2022 (3074)Get code to link to this report
Checking the PDH signal demodulated at the refection of GRMC at 15.2MHz and 78MHz

In the past, we always demodulate the reflection of GRMC at 78MHz to get PDH and lock GRMC. However, the 15.2MHz modulation on the main laser which is well within the linewidth of SHG, which makes this 15.2MHz modulation present also on green. To check what is the quality of this modulation, we did a demodulation of GRMC reflection at 15.2 MHz today.

The PDH demodulated at 15.2MHz and 78MHz are shown in attached figures. We can see that the PDH at 15.2MHz is even larger than the signal at 78MHz. This indicates that the SHG process seems not to degrade the sideband of green. This could make the lock of GRMC easier in the future.

Images attached to this report
3074_20220929120045_grmc15.png 3074_20220929120052_grmce.png
R&D (FilterCavity)
Print this report.
MarcEisenmann - 17:14, Wednesday 28 September 2022 (3072)Get code to link to this report
LO RIN and phase noise coupling to squeezing measurement

Marc, Yuhang

We investigated the possible RIN and phase noise coupling from the LO to the squeezing measurement.

Pump RIN :

To make this measurement, we injected 900mVPk of white noise to IRMC.

Results are attached in the first 5 figures. We can see a huge degradation of the squeezing level below 8kHz. However, the coherence is quite low (about 0.4).

In the 5th figure we also showed the contribution of this noise to the squeezing level but the shape is too different to precisely estimate the 'real squeezing' we are generating.

Pump phase noise :

To make this measurement, we injected 20mVPk of white noise to CC2.

Results are attached in figure 6 to 11.

This time, there seems to be no contribution to the squeezing measurement.

Images attached to this report
3072_20220928101204_losqz.png 3072_20220928101209_lorin.png 3072_20220928101326_locoh.png 3072_20220928101330_lotf.png 3072_20220928101403_loclas.png 3072_20220928101409_cc2sqz.png 3072_20220928101413_cc2pn.png 3072_20220928101417_cc2coh.png 3072_20220928101425_cc2tf.png 3072_20220928101431_cc2class.png
R&D (FilterCavity)
Print this report.
MarcEisenmann - 17:00, Wednesday 28 September 2022 (3071)Get code to link to this report
pump RIN and phase noise coupling to squeezing

Marc, Yuhang

We investigated the possible RIN and phase noise coupling from the green pump to the squeezing measurement.

Pump RIN :

We repeated the measurement of elog 3065 but injected this time larger noise (400mVPk instead of 100mVPk).

Results are attached in the first 5 figures. The results are compatible with the previous measurement : to match the projection of pump RIN to our estimation of classical noises, we should have about 9.4dB of squeezing instead of the 6dB that we are seeing at high-frequencies.

Pump phase noise :

To make this measurement, we injected 20mVPk of white noise to CC1.

Results are attached in figure 6 to 11.

This time, we can see a broadband degradation of the squeezing level (by at least 2dB). However, the coherence is really low between our excitation and the squeezing measurement..

Also, it seems that the contribution of pump phase noise create some similar pattern to the contribution of pump RIN to the squeezing measurement..

Images attached to this report
3071_20220928095034_pumprinsqz.png 3071_20220928095057_pumprinnoise.png 3071_20220928095104_pumprincoh.png 3071_20220928095111_pumprintf.png 3071_20220928095118_pumprinclass.png 3071_20220928095603_cc1sqz.png 3071_20220928095607_cc1noise.png 3071_20220928095612_cc1tf.png 3071_20220928095616_cc1tf.png 3071_20220928095624_cc1coherence.png 3071_20220928095629_cc1class.png
R&D (FilterCavity)
Print this report.
MarcEisenmann - 16:43, Wednesday 28 September 2022 (3070)Get code to link to this report
unaccessible PC + loud noise from DDS

Marc, Yuhang

These past days it was not possible to access the clean room pc from remote.

We found out that the was turned off, the DDS was emitting a really loud noise and lasers were all off.

It might have been caused by the typhoon this past week-end which might have created some electricity shut-down.

We restarted the pc and DDS rack before reloading the appropriate configuration files.

Everything seems fine for now.

R&D (Cryogenic)
Print this report.
KoheiMitsuhashi - 22:15, Saturday 24 September 2022 (3069)Get code to link to this report
RF driver test

I checked the RF driver(1080AF-AEN0-2.5) working in atc.

Tuning Voltage was fixed 10V, and Analog input was set from 0.1V to 1.0V.
I recorded voltage and frequency of AO Modulator, and also recorded voltage and ampere of Power Supply.
Before recording it , I ran electricity only from Power Supply for 5 minute to warm up the RF driver.
And I used a 20dB/50OHM attenuator(Mini-Circuits model CAT-20) to prevent the oscilloscope from saturating.

The result about Analog input and AO Modulator was as follows.

Analog input
[mV]
AO Modulator
voltage[mV]
AO Modulator
frequency[MHz]
100 93±1 81.46±0.24
200 190±3 81.20±0.66
300 285±2 81.13±0.22
400 385±5 81.00±0.27
500 478±6 81.15±0.23
600 567±9 81.09±0.13
700 665±3 80.89±0.22
800 759±6 80.97±0.14
900 852±3 81.10±0.19
1000 938±11 81.08±0.32

I graphed and fitted linearly with respect to voltage of the AO Modulator and voltage of the Analog input.
The figure was attached. The width of error bars is 1σ.


The result about Analog input and Power Supply was as follows.

Analog input
[mV]
Power Supply
voltage[mV]
Power Supply
ampere[A]
100 2804±5 0.659±0.002
200 2804±5 0.654±0.001
300 2808±4 0.649±0.001
400 2808±5 0.645±0.005
500 2804±5 0.638±0.001
600 2812±7 0.634±0.001
700 2808±7 0.629±0.001
800 2808±7 0.627±0.001
900 2808±7 0.624±0.001
1000 2808±7 0.625±0.001
Images attached to this report
3069_20220924100126_rfdriver.jpg 3069_20220925023327_resultoffitting.png