LOG-IN
Displaying reports 661-680 of 3200.Go to page Start 30 31 32 33 34 35 36 37 38 End
R&D (FilterCavity)
Print this report.
NaokiAritomi - 16:49, Monday 16 May 2022 (2945)Get code to link to this report
One magnet of PR mirror was fixed

[Takahashi, Aritomi, Marc]

We opened PR chamber and fixed one falling magnet of PR mirror and released the suspension. During this work, we noticed the oplev laser was hitting off center of PR mirror. We aligned the oplev laser to make it center of PR mirror. Currently the oplev beam is not reaching the PD and we need to align it.

We will open BS chamber on Friday next week (5/27).

Comments related to this report
RyutaroTakahashi - 19:49, Monday 16 May 2022 (2946)

Two photos (under gluing and after gluing) are attached.

KAGRA MIR (Scattering)
Print this report.
DanChen - 07:33, Monday 16 May 2022 (2944)Get code to link to this report
Laser scattering measurement

Date: 2022/5/13

With Homare Abe and Takayuki Tomaru

At KAGRA

  • We took pictures of 2 sapphire samples with a green injection laser beam:
    • One of them showed a green scattering light
    • The other one showed a red scattering light
  • We also took pictures of ITMY mirror with the same green injection laser beam:
    • The results showed a green scattering light

Analysis will be performed.

Pictures: klog

R&D (FilterCavity)
Print this report.
MarcEisenmann - 18:15, Friday 13 May 2022 (2943)Get code to link to this report
Recovery of FC (2)

Aritomi, Marc, Michael

 

We installed the new picomotors drivers. They worked fine for PR picomotor but did not work for BS picomotor.

We need to open BS chamber to check the picomotors.

With PR picomotors, we recovered the old reference in BS chamber (eg in elog 2794). In that configuration (and maybe BS in a 'random' position, the green beam is hitting quite below the gate valve window between BS/input but at least we don't have anymore clipping).

However, the PR oplev laser is now hitting the edge of the steering mirror before the PSD. It means that the PR reference at BS chamber that we are using might not be valid anymore for the current alignment inside the PR chamber.

We realigned PR and END oplevs and ran the coil health check codes.

All magnets of PR, INPUT and END mirrors are fine except PR H2 and we can not assess BS situation as the oplev PSD is broken as reported in elog2775.

We found a spare PSD inside the cleanroom and tested its dark noise (see attached figure where red and blue curves are the dark noise).

It is fine so we plan to replace it after the recovery of BS picomotors.

Images attached to this report
2943_20220513130354_11608.jpg
R&D (FilterCavity)
Print this report.
NaokiAritomi - 17:56, Friday 13 May 2022 (2942)Get code to link to this report
Comment to Recovery of FC (Click here to view original report: 2940)

Since the beam spot at GV between BS/input depends on both PR and BS, we should align PR and BS at the same time by using the GV between BS/input and the first target as references.

R&D (FilterCavity)
Print this report.
NaokiAritomi - 19:36, Thursday 12 May 2022 (2940)Get code to link to this report
Recovery of FC

[Aritomi, Michael]

We found that one of the green references at PR chamber was not aligned as shown in Fig. 1, which means we need to align the green injection beam. We removed the cover of the optical bench and aligned the green injection beam as shown in Fig. 2.

Then we aligned PR to make the green beam at the center of GV between BS/input. However, in this situation, the green beam at PR reference in BS chamber is clipped by a mirror behind the BS as shown in Fig. 3,4.

Images attached to this report
2940_20220512123618_fig1.jpg 2940_20220512123627_fig2.jpg 2940_20220512123632_fig3.jpg 2940_20220512123639_fig4.jpg
Comments related to this report
NaokiAritomi - 17:56, Friday 13 May 2022 (2942)

Since the beam spot at GV between BS/input depends on both PR and BS, we should align PR and BS at the same time by using the GV between BS/input and the first target as references.

KAGRA MIR (Absorption)
Print this report.
MarcEisenmann - 17:19, Thursday 12 May 2022 (2939)Get code to link to this report
start of AZTEC #2 absorption measurement

Marc, Matteo

 

Following the recovery of the PCI, we installed the AZTEC #2 sample.

First we measured X_center = 400.735 mm, Y_center = 122.105 mm.

We did a long z scan and got the 2 surfaces at z = 26.28 mm and 119 mm meaning that Z_center = 72.64 mm.

We have incident power = 7.69W  and transmitted power = 6.636W.

We started a XY measurement at Z_center and hope to finish all 5 maps by monday.

Images attached to this report
2939_20220512101828_longzscan.png
KAGRA MIR (General)
Print this report.
MarcEisenmann - 17:16, Thursday 12 May 2022 (2938)Get code to link to this report
Check of the pci setup

Marc Matteo

This entry summarizes these past days activities.

 

We found a mistake in the fitting code (the wavelength was hard coded to 633 nm instead of 1064 nm making the previous estimation wrong by a factor sqrt(1064/633))

To avoid this issue, we coded a more flexible function 'fit_blade.m' that is saved in the PCI scripts folder on the desktop

With this function, we could finally tuned the pump beam telescope and recover Manuel's parameters (fig 1)

We also check the vertical and horizontal angle of incidence to be -0.2 deg and 2.5 deg respectively.

We also checked the probe beam parameters by placing a power meter in front of the absorption PD.

We also got same parameters as Manuel (fig 2)

The angle of incidence is 3.47 deg meaning that the relative angle between the probe and pump beam is also correct.

 

We installed the surface reference sample and tried to maximize the R coefficient by changing both the translation stage and imaging unit z positions.

While doing these motions, we could find the expected maximum at x = 35 mm and z_iu = 68 mm.

However, R = 14 /W meaning that we still had not optimal conditions.

After investigating other possible issues (chopper frequency, laser power, pd, lock-in are all fine) and repeating several times this calibration, we could not improved this value by much...

We installed the bulk reference sample and got R = 0.5536 cm/W.

Both surface and bulk calibration are 10 % lower than expected but the reason is still not clear to us...

 

We decided to use the 1.5 inches sapphire that Manuel used to check the setup calibration (eg check entry 1132).

We did the exact same measurement and got similar values (fig 3)  : mean absorption ~ 40 ppm/cm.

However, the sample was really dirty so it we cleaned this sample by wiping alcohol on it and repeated the measurement that gave identical result (fig 4)

 

It means that despite some issue with the setup, our calibration gives us reasonable results and can be considered to be working again.

Images attached to this report
2938_20220512101306_20220510pumpprofile.jpg 2938_20220512101318_20220511probeprofile.jpg 2938_20220512101458_15inchdirty.png 2938_20220512101503_15inchcleaned.png
KAGRA MIR (General)
Print this report.
MarcEisenmann - 19:29, Monday 09 May 2022 (2937)Get code to link to this report
Check of pump beam

Today I installed the razor blade that cuts the beam vertically and measured the pump beam parameter.

The blade is at 90 mm from the edge of the breadboard and I used about 100 mW of input power.

I measured a waist size of 22 um at 58 mm (in Manuel's unit) while we expect 36 um at 61 mm.

I started the tuning of the telescope but it is not concluded yet.

Images attached to this report
2937_20220509122933_20220509pumpprofilebefore.jpg
R&D (FilterCavity)
Print this report.
NaokiAritomi - 15:27, Monday 09 May 2022 (2936)Get code to link to this report
Subtraction of dark noise from locking accuracy measurement

I subtracted the dark noise from the locking accuracy measurement done in elog2864. I calculated sqrt(locking accuracy^2-dark noise^2). In the point where the dark noise is larger than the locking accuracy, the subtraction is set to 0. The rms after the subtraction is almost the same as one before subtraction.

Images attached to this report
2936_20220509082738_ccfclockingaccuracy20220303subtraction.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 17:07, Wednesday 27 April 2022 (2935)Get code to link to this report
Some checking reference for beam position on filter cavity iris inside the arm

We have two iris inside filter cavity arm to check the direction of green laser beam. To make sure the beam can arrive at the filter cavity end mirror, we need to make the green beam hit on a point checked in the past when filter cavity is aligned.

Remember that, for the first step, we should make the green beam hit on several reference points on PR and BS chambers.

The first attached picture shows the beam position on the first iris. In this case, the iris needs to be rotated so that the stick is not visible (hidden on the back side of iris). At this moment, the beam position should be just above the hole.

For the first iris, when the stick is visible, the situation is shown in Fig. 2. At this moment, the beam is above the hole and located a bit the left side.

If the beam hits on the first iris like what is shown here, it should be not diffcult to find the green beam on the second iris. If not visible on the second iris, we should move the green beam around with little adjustment. Then what we need to do is to just make the green beam go through the hole of it.

Images attached to this report
2935_20220427095727_wechatimg46.jpeg 2935_20220427100236_wechatimg47.jpeg
KAGRA MIR (Scattering)
Print this report.
DanChen - 11:41, Wednesday 20 April 2022 (2934)Get code to link to this report
Test experiment for scattered light observation of sapphire

Information

Date: 14th Apr 2022

Members: Dan Chen, Satoru Ikeda

Place: Kamioka Hokubu-kaikan 1F meeting room

Background

The mirror substrates made for KAGRA O5 may have high internal scattering, and we are developing a simple measurement method for this purpose.
The idea is using a camera to take a picture.
At first we used a laser pointer (Green, 1mW) as a laser beam source and a sapphire sample from Mitaka to check the measurement principle.

Result

ASI camera (ASI224MC) is not suitable for this measurement because of the noise.
The reason can be the insufficient noise reduction in the camera or poor lens we used or both.

A digital camera (Canon EOS Kiss M2) successfully captured the scattered light from the sapphire sample with 61s exposure time and ISO6400.

Detail

Laser source

Model: UC-S1
Wave length: 532nm
Max power: 1mW

ASI camera

Camera: ASI 224MC
Lens: CCTV LENS 2.1mm 3MP
Application: Planetary Imager

Result: we tried several conditions for exposure time and gain, but the scattered light in the sample was not clear because of noise.

Digital camera

Camera: Canon EOS Kiss M2
Lens: EF-M15-45 F3.5-6.3 IS STM
Exposure time: 61s
ISO: 6400
Color temperature: 6000K
F: 16
Focus length: 45mm
Recorde image quality: best (about 15MB/picture)
Output image format: jpeg

Result:
With the above condition, the scattered light was observed. In order to make it clear, we took a picture with the Green laser ON and OFF, then we made a diff image using "imageJ". The diff image shows the internal scattered light clear.

What we learned:
Because the scattered light we want to measure is very faint, we need to avoid environmental light and stray/scattered light generated outside of the sapphire sample interfering with the measurement.

Images attached to this report
2934_20220420043952_img0865comp.jpg 2934_20220420044011_img0866comp.jpg 2934_20220420044016_substructrgbchannelscompcomp.png 2934_20220420044026_img6701comp.jpg
KAGRA MIR (Absorption)
Print this report.
MatteoLeonardi - 16:55, Tuesday 19 April 2022 (2933)Get code to link to this report
Comment to AZTEC #1 (KAGRA size) (Click here to view original report: 2919)

Additional maps.

Here the position and average.
Z=105.3, mean absorption: (61.6 +/- 13.95)
XZ map, mean absorption: (50.1 +/- 18.1)
YZ map, mean absorption: (56.8 +/- 19.1)

Images attached to this comment
2933_20220419095153_3.png 2933_20220419095203_xzmap.png 2933_20220419095221_yzmap.png
R&D (FilterCavity)
Print this report.
NaokiAritomi - 17:55, Monday 18 April 2022 (2932)Get code to link to this report
FC recovery trial

[Aritomi, Michael]

After opening the gate valves between input/end and arm, we tried to recover FC. We centered the PR reference and first target, but we could see only the scattered light at second target... We checked that the green beam is roughly center of the gate valve between BS/input with the PR reference.

Note: After the gate valve between BS/input is used as PR reference, don't forget to open the gate valve.

R&D (FilterCavity)
Print this report.
NaokiAritomi - 14:45, Monday 18 April 2022 (2931)Get code to link to this report
RF amplifier for AOM is working

In elog2865, it is reported that the RF amplifier (ZHL-2) for AOM stopped outputting. I checked the output of the RF amplifier and confirmed that it is working. I injected the input of 5.5dBm and the output of the amplifier is 23.4dBm, which is reasonable.

R&D (FilterCavity)
Print this report.
NaokiAritomi - 14:34, Monday 18 April 2022 (2930)Get code to link to this report
Comment to input/end chamber evacuation (Click here to view original report: 2929)

I opened the gate valves between input/end and arm. Before I opened the gate valve between input/arm, the pressure of input and arm were 1.2e-6 mbar and 3.7e-8 mbar, respectively. After I opened the gate valve, both of them became 1e-6 mbar. Before I opened the gate valve between end/arm, the pressure of end and arm were 2e-7 mbar and 1.5e-7 mbar, respectively. After I opened the gate valve between end/arm, the pressure of end and arm became 4e-7 mbar and 5.4e-7 mbar, respectively. 

R&D (FilterCavity)
Print this report.
NaokiAritomi - 19:03, Thursday 14 April 2022 (2929)Get code to link to this report
input/end chamber evacuation

[Aritomi, Michael]

We started the evacuation of input/end chambers. After evacuation with a rotary pump, we opened small gate valves close to input/end chambers. The current pressure of end chamber and arm are 3.7e-4 mbar and 3.4e-8 mbar, respectively. The current pressure of input chamber and arm are 4.1e-4 mbar and 3.1e-8 mbar, respectively. We will open small gate valves close to arm, and large gate valves between input/end and arm next week.

Comments related to this report
NaokiAritomi - 14:34, Monday 18 April 2022 (2930)

I opened the gate valves between input/end and arm. Before I opened the gate valve between input/arm, the pressure of input and arm were 1.2e-6 mbar and 3.7e-8 mbar, respectively. After I opened the gate valve, both of them became 1e-6 mbar. Before I opened the gate valve between end/arm, the pressure of end and arm were 2e-7 mbar and 1.5e-7 mbar, respectively. After I opened the gate valve between end/arm, the pressure of end and arm became 4e-7 mbar and 5.4e-7 mbar, respectively. 

R&D (FilterCavity)
Print this report.
MichaelPage - 21:03, Wednesday 13 April 2022 (2928)Get code to link to this report
TAMA DGS upgrade - frontend PC installation

Yuhang and Michael

We went through the instructions outlined in T2012111, for installing the Debian 10 operating system on the frontend computer for the central station. As described in 2905, this overrides the previous OS installations on the frontend computer. We have a list of errors/ambiguities in the installation procedure which will be sent to Takahiro Yamamoto.

1.1 OS installation
OS was installed in English. Actually I have tried installing Japanese input packages on my Debian 10 workstation in the past and found it extremely poorly documented, so I didn't even bother with that and will just ask Yamamoto-san.

1.2 Additional packages
No issues. We used 1.0.6 version of cdssoft on the ALIGO Caltech repository. Python and Python3 are called separately, just to let it be known (Python version is 2.7.12, Python3 version is 3.7.3)

1.3 OS settings
As mentioned previously, there was an issue with mounting network locations in the /etc/fstb file. We did not follow through with the rest of the instructions since they all seem to be tied to things appearing in the /users/ directory that was created. I feel like there is also some abiguity in the instructions on where these mount directories should be created. I created them inside /etc/, but maybe I should try again just starting at /.

1.4 Bash settings
No issues. Should mention for clarity that .bashrc is a hidden file located in /home/controls.

1.5 miniconda3 installation
We did not follow through with the last step since it seems the conda command doesn't work. In the first step here, a symbolic link is created between /home/controls and a folder that should be created in /users/DGS..., but nothing is present inside the /users/ folder right now.

1.6 ssh settings
No issues

1.7 NTP settings
No issues

1.8 Vinagre setting
No issues

1.9 Guardian setting
We were not sure where the bugfix is supposed to be applied. But upon looking again there is the blue line at the top that indicates which file is changed, so I will check again before sending my questions to Yamamoto-san.

1.10 Camera settings
The procedure says "camera code itself is old version" but I'm not sure what it is referring to or even if anything is installed.

1.11 SDF settings
I guess the blue line at the top indicates which files should be changed. We haven't tried yet.

1.12 z command
No issues

1.13 Zoom setting
No issues

1.14 Skippy-XD installation
We weren't sure what counts as the "1st PC". 

 

R&D (FilterCavity)
Print this report.
NaokiAritomi - 17:10, Wednesday 13 April 2022 (2927)Get code to link to this report
Comment to Squeezing level and CC2 signal pk-pk as a function of CC2 demodulation phase (Click here to view original report: 2926)

Similar measurement was done in elog2646. In this measurement, the minimum(maximum) CC2 amplitude did not correspond to squeezing(anti-squeezing) quadrature.

R&D (FilterCavity)
Print this report.
YuhangZhao - 13:56, Wednesday 13 April 2022 (2926)Get code to link to this report
Squeezing level and CC2 signal pk-pk as a function of CC2 demodulation phase

Michael and Yuhang

As suggested by Matteo, we took squeezing level and CC2 signal magnitude as a function of CC2 demodulation phase.

They are both normalized, either by shot noise level or CC2 signal without pump.

Fig 1 is squeezing measurements at different CC2 demodulation phase. The phase is changed as linspace(0, 180, 19). It is not understood why sometimes some peaks appear around 5-20kHz.

Fig 2 is squeezing and CC2 signal magnitude as a function of CC2 demodulation phase.

Images attached to this report
2926_20220413065703_figure3.png 2926_20220413065708_figure4.png
Comments related to this report
NaokiAritomi - 17:10, Wednesday 13 April 2022 (2927)

Similar measurement was done in elog2646. In this measurement, the minimum(maximum) CC2 amplitude did not correspond to squeezing(anti-squeezing) quadrature.

R&D (FilterCavity)
Print this report.
YuhangZhao - 13:41, Wednesday 13 April 2022 (2925)Get code to link to this report
Squeezing measurement while the bench is covered

Michael and Yuhang

The optical bench was closed and some squeezing measurements are done with different pump power.

The result is attached. Compared with previous characterization, this measurement has more phase noise of 30mrad. It is not understood why phase noise is increased.

Images attached to this report
2925_20220413064207_figure1.png 2925_20220413064213_figure2.png