LOG-IN
Displaying reports 861-880 of 3273.Go to page Start 40 41 42 43 44 45 46 47 48 End
KAGRA MIR (General)
Print this report.
MarcEisenmann - 20:03, Thursday 27 January 2022 (2806)Get code to link to this report
start of shinkosha7 measurement with normal incidence

Katsuki, Marc

We measured the distance between the blade at z = 25 mm and the last steering mirror.

From this distance and knowing the 2.5deg angle of the beam we found that the beam is hitting this last steering mirror at y = 134.4 mm (in the frame of the translation stage).

We placed the blade at this position and z = 110 mm and tweaked the steering mirror to get half of the power transmitted.

After little back and forth, we measured the horizontal angle of incidence of 0.024 deg and the vertical one to be 0.015 deg.

Because this is more than 10 times smaller than KAGRA requirement on the C-axis angle wrt to the surface normal we can consider that we are now in normal incidence.

We realigned the imaging unit , did the polarization calibration and started the measurement with shinkosha 7 (new center at x = 398 mm and y = 162 mm).

However we are not sure if this substrate has a wedge or not which means that we might see the fringes that Manuel observed...

Images attached to this report
2806_20220127120421_20220127horizontalangle.jpg 2806_20220127120433_20220127verticalangle.jpg
R&D (FilterCavity)
Print this report.
MichaelPage - 14:41, Thursday 27 January 2022 (2805)Get code to link to this report
OPO locking setup

Yuhang and Michael

We rearranged the OPO optical layout to use proper photodetectors as mentioned by Yuhang. The layout with breadboard spacing (not to scale) is shown in figure 1. A photo of the setup past L1 is shown in figure 2.

The following components are used. Some of them I'm not sure though:

Laser: Lightwave laser
WP1: ?
WP2: ?
KPX1: KPX094AR.33
BS1: ?
KPX2: KPX094AR.33
KPC: KPC034AR.33
L1: ?
---
Equipment placed for OPO setup
FI: Thorlabs IO-5-1064-VLP
HWP_FI: Half wave plate
EOM: New Focus 4003 Resonant EOM
AOM: AA Optoelectronic MT110 IR 27
L2: f = 40mm
SM1: Steering mirror 1
BS2: 50/50
SM2: Steering mirror 2
L3: f = 75mm
SM3: Steering mirror 3
BS3: ?
L4: f = 30mm
PD_R1: Thorlabs PDA05CF2
L5: f = 50mm
PD_R2: Thorlabs DET10N/M InGaAs biased
IN: OPO input mirror
OPO: OPO assembly block
BS4: ?
CCD: Camera
L6: ?
PD_T: Thorlabs PDA36A
 

Images attached to this report
2805_20220127064246_oposetup4.png 2805_20220127064252_opolayout20220126183355.jpg
KAGRA MIR (General)
Print this report.
MarcEisenmann - 11:38, Thursday 27 January 2022 (2802)Get code to link to this report
realignment of the pump beam to normal incidence

In order to avoid polarization coupling when measuring birefringence with a tilted IR beam (the pump beam) we plan to take future measurement with the pump beam at normal incidence.

Yesterday I removed the last 2 steering optics on the 1310 nm beam path to have enough space to add 2 steering mirrors on the pump beam path.

I also installed a new power meter head (S146C) that has enough range to be used for both absorption and birefringence measurements.

I started by aligning the vertical direction using both a marker on a IR viewing card and iris.

Note that the injection table height is 28 cm with beam height of 5.5 cm and imaging table height is 24.5 cm (so beam height there should be 9cm).

Then I started to align the horizontal direction but it is slightly more tricky as the injection and imaging tables are not aligned.

I installed the razor blade and measured 1.5 deg incidence angle in this plane.

Today we will check the relative position of the blade with respect to the last steering mirror on the pump beam path to fine tune this incidence angle.

R&D (FilterCavity)
Print this report.
YuhangZhao - 22:23, Monday 24 January 2022 (2801)Get code to link to this report
Photodetectors for OPO internal losses measurement

Michael and Yuhang

Last week, we replaced two QUBIG PDs for FC GR lock. An RF output only PD is currently used for FC GR lock, which was found to have flat spectrum across tens of MHz as Fig.2. The QUBIG PD with DC and RF outputs is taken out and used for new OPO reflection, which has spectrum as Fig.1. Comparing Fig.1 and Fig.2, we can also find that the RF only PD has a better SNR. However, after a test of DC RF QUBIG PD for new OPO, we found it cannot read phase modulation taken from the reflection of OPO. Although from its specification, PD bandwidth (1-100MHz) is enough to detect 40MHz signal, we decide to not use it since we couldn't find by checking either spectrum directly or demodulated PDH signal. On the other hand, we will use it for monitoring DC response. Considering a BW of 100MHz, which has a very small rise time of 3.5ns.

A PDA05CF2 will be used for acquiring PDH signal for locking. This PD will be placed in the reflection of OPO. It has been proved to be able to provide large enough error signal.

We will use PDA36A-EC in the OPO transmission. If 0dB is enough to measure transmission signal, we can have 10MHz bandwidth, which gives 35ns rise time. However, NIR may have larger rise time since this PD is made of Si. According to Isogai paper, transmission doesn't need fast response. So maybe PDA36A-EC is enough. We will see this later. On the other hand, I took some dark noise measurement, which is summarized in Fig.3. The dB value in the legend shows different chosen gain of this PD. This measurement shows much less dark noise indicated in the specification of this PD. 0dB should be 300uV, 20dB should be 250uV, 40dB should be 340uV, 60dB should be 800uV. One issue is that the RMS I got is only from 10Hz to 3.2kHz. Higher frequency noise must be integrated but I didn't measure. But anyway, we can see that with a factor of 10 gain increase, the noise is not increased by the same factor.

Images attached to this report
2801_20220124142250_wechatimg776.jpeg 2801_20220124142256_wechatimg775.jpeg 2801_20220124142517_figure1.png
R&D (FilterCavity)
Print this report.
NaokiAritomi - 15:55, Monday 24 January 2022 (2800)Get code to link to this report
A magnet of input mirror fell down

[Takahashi, Aritomi, Yuhang]

Today we opened the input chamber and checked the suspension of input mirror. Because of misunderstanding, we vented not only the input chamber, but also the PR/BS chambers.

Takahashi-san found that one magnet of the input mirror fell down. According to Takahashi-san, the magnet could touch the coil holder and the input mirror could be tilted because of it. In this situation, we tried to recover the tilt by picomotor and then the magnet could fall down. Takahashi-san will fix the magnet and close the input chamber next Monday.

After Takahashi-san adjusted the suspension, we could move the picomotor for input mirror and the green reflection overlaps with the injection now.

Finally, we started to evacuate the PR/BS chambers.

Comments related to this report
NaokiAritomi - 14:36, Monday 31 January 2022 (2811)

Takahashi-san glued the magnet to the input mirror and closed the input chamber. Since it takes one day for the glue to be fixed, we will evacuate the input chamber after tomorrow.

RyutaroTakahashi - 22:44, Monday 31 January 2022 (2814)

Pictures of gluing.

NaokiAritomi - 16:18, Wednesday 02 February 2022 (2822)

I started the evacuation of input chamber. First I used rotary pump until the pressure in the input chamber reaches below 0.1mbar. After I removed the rotary pump, I closed the small gate valve close to arm and opened the small gate valve close to the input chamber. I will wait until the pressure in the input chamber reaches the similar one in the BS chamber.

NaokiAritomi - 15:23, Thursday 03 February 2022 (2827)

Before I opened the gate valves between input/BS and input/arm, the pressures in input, BS, arm was 1.3e-6 mbar, 9.9e-10 mbar, 3.1e-8 mbar, respectively.

First I opened the gate valve between input/BS. After 1 hour, the pressures in input and BS chambers became 5.5e-7 mbar and 1.9e-8 mbar, respectively.

Then I opened the small/large gate valves between input/arm. The pressures in input, arm became 5.4e-7 mbar,4.5e-7 mbar, respectively.

R&D (FilterCavity)
Print this report.
NaokiAritomi - 17:46, Friday 21 January 2022 (2799)Get code to link to this report
Strange oplev spectrum of input mirror

Since the oplev beam reflected by the input mirror was very high as reported in elog2798, I tweaked a steering mirror just after the oplev laser source for input mirror (Fig. 1). The oplev sum for input gets around 5000.

Fig. 2 shows the oplev spectrum of input mirror (top right). There are large peaks at 2 Hz and 3.4 Hz. We need to open the input chamber and check the suspension.

Images attached to this report
2799_20220121094645_img9115.jpg 2799_20220121094651_20220121input.png
R&D (FilterCavity)
Print this report.
NaokiAritomi - 19:52, Thursday 20 January 2022 (2798)Get code to link to this report
Input mirror issue

[Aritomi, Yuhang, Michael]

We found that the green reflection from input mirror is very high and hits on the upper edge of gate valve between BS/input chambers. The oplev light for input mirror is also very high as shown in the attached picture. We tried to move input mirror with picomotor, but the input mirror did not move. We will align oplev and measure the input oplev spectrum tomorrow.

Images attached to this report
2798_20220120115254_img9114.jpg
R&D (FilterCavity)
Print this report.
NaokiAritomi - 19:12, Thursday 20 January 2022 (2797)Get code to link to this report
Comment to Broken camera driver and pico-motor driver channel found during TAMA blackout recovery (Click here to view original report: 2742)

I connected the END YAW to the motor A of driver although I don't know the motor A is working or not.

Images attached to this comment
2797_20220120114801_img9113.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 21:58, Wednesday 19 January 2022 (2796)Get code to link to this report
Error signal for new OPO

I took a function generator and an oscilloscope from filter cavity clean room to modulate EOM and acquire data in ATC.

The RF PD is moved to OPO reflection to acquire PDH signal. A minicircuit zx05-1l-s+ is used for RF signal demodulation. 10dBm 40MHz signal is sent to New focus 4003 EOM to get a phase modulation around 0.07~0.21rad. 3dBm signal is used as LO. We got error signal as attached figure. Some glitches are found and we will try to understand it better tomorrow.

One possible reason is that I should put a DC block between RF PD and mixer. Note that demodulation phase is not optimized yet.

Images attached to this report
2796_20220119135820_wechatimg774.jpeg
R&D (FilterCavity)
Print this report.
YuhangZhao - 22:55, Tuesday 18 January 2022 (2795)Get code to link to this report
OPO aligned when laser is injected from the in-coupling mirror side

To measure effectively the internal optical losses of OPO, we need to inject laser from the in-coupling mirror side (elog2784).

Therefore, I flipped OPO and moved OPO closer to 75mm lens as described in elog2790. However, I found that moving 54mm is far too much, whose reason is not understood yet. After moving OPO farther from 75mm lens, the mode matching situation is better. However, I found that the beam is going into the in-coupling mirror but close to the edge of its hole. So I loosed in-coupling mirror and moved it vertically down. After this, I found it diffcult to recover alignment at the first glance.

However, I realized that it is actually very easy to re-align OPO using transmission camera and injection/reflection beam overlaping (without using periscope/translation stage). The method is bascially the same with the procedure provided in elog2783 apart from different beam parameters and almost not seperating the alignment of crystal and in-coupling mirror but aligning them as almost a whole. And I found it is possible to choose a personal preferred position of in-coupling mirror and just fix it with the crystal side of OPO. Then align them as a whole.

After this easier alignment procudure, checking by eye, I found the laser is far enough from the edge of OPO two sides holes. Then I aligned OPO until mode matching level is around 95%. Now OPO is ready for internal optical losses measurement.

R&D (FilterCavity)
Print this report.
NaokiAritomi - 20:19, Tuesday 18 January 2022 (2794)Get code to link to this report
FC alignment recovery

[Aritomi, Yuhang]

After the evacuation of PR/BS chambers, we aligned PR/BS picomotors to center the PR reference and the first target. Then we checked the second target. Although we scanned BS alignment a lot, we could see only scattered light on the second target.

Instead of the nominal PR reference, we used the green beam position in the gate valve between BS/input chambers as PR reference (Fig.1). We centered the green beam in the gate valve between BS/input chambers by PR picomotor and centered the green beam at first target by BS picomotor. Then we found the green beam in the second target. Since the green beam position in the nominal PR reference changed, we made a new PR reference as Fig.2.

Images attached to this report
2794_20220118134919_wechatimg773.jpeg 2794_20220118134927_wechatimg772.jpeg
R&D (FilterCavity)
Print this report.
YuhangZhao - 18:55, Monday 17 January 2022 (2792)Get code to link to this report
Comment to BS picomotor check (Click here to view original report: 2791)

We firstly used Rotary Vane Vacuum Pump (Alcatel 2100A) to evacuate. The speed is controlled to be less than 1mbar/s to avoid large pressure applied on in-vacuum components. Its ultimate achievable pressure is 0.03mbar.

Since the turbo pump (STP1003) requires outlet pressure smaller than 0.1torr (0.13mbar), we waited for rotary pump to evacuate PR/BS chambers until pressure reached 0.1mbar.

The small rotary pump and turbo pump were kept on with gate valve closed between turbo pump and PR/BS chambers. So we just opened the gate valve after PR/BS chambers pressure is lower than 0.1mbar.

We will keep turbo pump on until the PR/BS chambers pressure reaches 1e-6 mbar. Then we will open the gate valve between PR/BS chambers and input mirror chamber.

R&D (FilterCavity)
Print this report.
NaokiAritomi - 17:07, Monday 17 January 2022 (2791)Get code to link to this report
BS picomotor check

[Takahashi, Aritomi, Yuhang]

Since the BS picomotor did not move, we opened the BS chamber and checked the BS picomotor. The BS picomotors for pitch and yaw were connected to the motors A and B in driver 1, respectively. However, we found that the motors A and B in driver 1 were broken. We decided to use the motor C in driver 1 for both BS pitch and yaw temporarily.

We noticed that the BS picomotor did not move in yaw direction although we heard the sound of the picomotor. Takahashi-san found that it was because the BS was touching the stopper of the mirror. After Takahashi-san fixed it, the BS picomotor could move in yaw.

After we adjusted the BS picomotor, the green beam is hitting on center of the first target. Finally, we started to evacuate the PR/BS chambers.

Comments related to this report
YuhangZhao - 18:55, Monday 17 January 2022 (2792)

We firstly used Rotary Vane Vacuum Pump (Alcatel 2100A) to evacuate. The speed is controlled to be less than 1mbar/s to avoid large pressure applied on in-vacuum components. Its ultimate achievable pressure is 0.03mbar.

Since the turbo pump (STP1003) requires outlet pressure smaller than 0.1torr (0.13mbar), we waited for rotary pump to evacuate PR/BS chambers until pressure reached 0.1mbar.

The small rotary pump and turbo pump were kept on with gate valve closed between turbo pump and PR/BS chambers. So we just opened the gate valve after PR/BS chambers pressure is lower than 0.1mbar.

We will keep turbo pump on until the PR/BS chambers pressure reaches 1e-6 mbar. Then we will open the gate valve between PR/BS chambers and input mirror chamber.

R&D (FilterCavity)
Print this report.
YuhangZhao - 23:10, Sunday 16 January 2022 (2790)Get code to link to this report
New OPO aligned when laser is injected from crystal side

Michael and Yuhang

This Friday (20220114), we changed optical set-up from what described in elog2788. This is because we found one of the steering mirrors in front of OPO cannot effectively align laser beam inside OPO. We checked beam parameters in elog2486, which shows that the non-effective steering mirror is located just around the beam waist after 40mm lens. This explains why this mirror shows anomaly while OPO alignment procedure.

To solve this problem, we modified the optical set-up. We add one steering mirror just before OPO while leaving enough space for a BS between it and OPO (Fig.1). To confirm this set-up is capable of aligning beam and the OPO still has good internal alignment, we aligned OPO while laser beam is injected from the crystal side of OPO. We found the newly add steering mirror helps to optimize alignment.

In the end, we achieved a decent alignment by moving the position of lenses. The TEM02 mode is optimized by moving position of lenses. The position of lens are shown in Fig.2 and Fig.3.

We also characterize the mode matching level by measuring the TEM00, TEM01 and TEM02 modes. Their seperate power is shown in Fig.4, 5, and 6. This indicates a mode-mismatch of (5.6+18.8)/(5.6+18.8+282) = 8%

Using parameters of OPO, we simulate the beam parameters as Fig.7. From this simulation, we see the beam waist should be located in front of crystal when injected from the crystal side. But if we inject from in-coupling mirror side, the beam waist should be located inside OPO. Therefore, to measure optical losses and flip OPO housing, we need to flip OPO and move it OPO by 54mm closer to 75mm lens.

Images attached to this report
2790_20220116150934_wechatimg766.jpeg 2790_20220116150955_wechatimg767.jpeg 2790_20220116151010_wechatimg768.jpeg 2790_20220116151018_wechatimg769.jpeg 2790_20220116151027_wechatimg770.jpeg 2790_20220116151035_wechatimg771.jpeg 2790_20220116151044_22.png
R&D (FilterCavity)
Print this report.
NaokiAritomi - 17:23, Friday 14 January 2022 (2789)Get code to link to this report
BS picomotor does not move

[Aritomi, Yuhang]

Last year, we found that the BS picomotor in yaw direction does not move as reported in elog2767.

Today, when we tried to move the BS picomotor in pitch direction with step of 50, the picomotor moved a lot and got stuck. We will open BS chamber to check the picomotor next Monday.

R&D (FilterCavity)
Print this report.
YuhangZhao - 21:52, Tuesday 11 January 2022 (2788)Get code to link to this report
Preparation for OPO internal losses measurement

Michael and Yuhang

As discussed in elog2784, we need to inject laser from OPO in-coupling mirror side to distinguish different OPO internal losses values. This is the reason of work reported in this elog. To achieve the measurement, we have done the followings:

1. Remove periscope and rotation stage for OPO

2. Prepare a 50:50 BS placing in front of OPO to extract OPO reflection signal information

3. Lower the height of OPO transmission BS, PD and camera

(these works are shown in Fig.1)

4. Confirm and adjust the flatness of laser beam after removing periscope

(this work is shown in Fig.2,3,4)

5. Put OPO in place and optimize its position by hand. Try to get a reasonable OPO transmission mode. We have found a TEM01 mode now.

Next step: Find OPO transmission signal on PD. Optimize OPO injection beam alignment.

Images attached to this report
2788_20220111135159_wechatimg762.jpeg 2788_20220111135210_wechatimg763.jpeg 2788_20220111135218_wechatimg760.jpeg 2788_20220111135227_wechatimg761.jpeg
R&D (FilterCavity)
Print this report.
YuhangZhao - 21:37, Tuesday 11 January 2022 (2787)Get code to link to this report
Comment to new OPO cavity closed with decent alignment (Click here to view original report: 2783)

The camera used in this set-up is found to be tilted today. Fig.1 shows that this tilt is introduced through the interface between camera and post.

During the alignment, we use camera to characterize cavity HG modes shapes. When the beam and camera height is aligned, we found beam appears not in the center of screen as Fig.2. We suspect the tilt between camera and post is the cause of the cavity modes mis-center as Fig.2.

Note that Fig.2 shows the non-ideal alignment of crystal, which has a black defect.

Images attached to this comment
2787_20220111132916_wechatimg764.jpeg 2787_20220111133444_wechatimg765.jpeg
KAGRA MIR (Polarization)
Print this report.
MarcEisenmann - 11:01, Tuesday 11 January 2022 (2786)Get code to link to this report
SHINKOSHA 7 birefringence measurement

This entry reports the measurement of polarization angle and birefringence of the shinkosha 7 sample measured in the same orientation as Manuel absorption measurement (ie arrow at the top pointing towards the imaging unit).

It is in quite good agreement with the previous measurement.

Note that the birefringence results reported here neglect the angle of incidence of the pump beam

Images attached to this report
2786_20220111030032_0deg.png 2786_20220111030036_15deg.png 2786_20220111030040_30deg.png 2786_20220111030044_45deg.png 2786_20220111030047_60deg.png 2786_20220111030051_75deg.png 2786_20220111030055_s0.png 2786_20220111030059_s1.png 2786_20220111030103_dntheta.png 2786_20220111030108_dn.png 2786_20220111030111_theta.png
R&D (FilterCavity)
Print this report.
NaokiAritomi - 01:11, Saturday 08 January 2022 (2785)Get code to link to this report
Estimated locking accuracy with CCFC

Considering the CCFC+green OLTF shown in elog2758 and the IR locking accuracy (CCFC error signal) without CCFC, we can estimate the expected IR locking accuracy with CCFC. I compared the estimated IR locking accuracy with CCFC and the measured one as shown in the attached figure. As you can see, the two spectra agree well. The discrepancy between 10-100 Hz could be because the measured spectrum would be limited by the spectrum analyzer noise between 10-100 Hz.

Images attached to this report
2785_20220130155604_20210622lockingaccuracyoltf.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 20:18, Friday 07 January 2022 (2784)Get code to link to this report
Comment to Expected ring-down measurement of OPO (Click here to view original report: 2418)

Since OPO is finally closed, the next step is to characterize the intra-cavity losses. This is important for us because we are suspecting some of the optical losses are from OPO (current estimated loss budget for FDS). So this is an important step to understand the loss budget in the frequency dependent squeezing experiment.

I modified a bit the code to see the difference of measurement for different OPO intra-cavity losses.

Now, the laser is injected from the crystal side of OPO. I did a simulation of this case as Fig. 1. In this case, we will miss the information of OPO reflection. The blue and orange curves overlap for reflection.

If laser is injected from in-coupling mirror, as shown in Fig.2, we find that although decay time is not enough to indicate optical losses. We can extract losses from reflection signal.

So we will rotate OPO next week and inject laser from the in-coupling mirror side.

Images attached to this comment
2784_20220107122022_opoinverse.png 2784_20220107122029_oposim.png