LOG-IN
Displaying reports 881-900 of 3273.Go to page Start 41 42 43 44 45 46 47 48 49 End
R&D (FilterCavity)
Print this report.
YuhangZhao - 20:00, Friday 07 January 2022 (2783)Get code to link to this report
new OPO cavity closed with decent alignment

Michael and Yuhang

After the in-coupling mirror cleaning as reported in elog2774 and elog2776, we tried to close OPO again. Apart from mirror cleaning, this time, we tried to align OPO better than elog2769. To make sure a good crystal alignment, we paid attention to:

1. Making sure the flatness of laser beam in pitch/yaw direction using alignment tools such as rulers. Use power of few milliwat now.

2. Change to less than microwatt. Making sure camera is at good height and angle. Mark the beam position on camera by drawing a circle for the laser beam shown on TV screen. 

3. Change back to few milliwatt. Putting OPO crystal part on rotation stage with HR side facing laser. Checking injection/reflection overlapping and transmission reaching the circle on TV screen. If not, adjusting by hand the position of OPO crystal part. 

4. Adjusting camera lens to zoom in and inspect the image of crystal transmission. This image should be a circular spot without any clipping. If not, consider to adjust the position of OPO crystal part in transversal, longitudinal and tilt DOF.

5. Fixing OPO crystal part on rotation stage gently. In my case, the OPO didn't move after this. If it is moved a bit, bring it back by adjusting rotation stage. Be careful that the longer side of rotation stage should be aligned with laser. 

6. Putting in-coupling mirror part of OPO and half-fixing it. Half-fixing means that the screws are ~20 degrees before the feeling of tight screws. This is to make sure that we can adjust in-coupling mirror while keeping it not going to far after each adjustment.

7. Zooming out camera to find roughly the flash of OPO cavity. When zooming out, we can see the inner cavity reflection more clearly.

8. Moving incoupling mirror by hand in horizontal direction to make inner cavity reflection good for horizontal. Moving vertical adjusting screws to find flash.

9. Zooming in camera to see the Hermite-Gaussian mode of OPO cavity. The scan speed was set to be 3Hz.

10. Adjuting finely the incoupling mirror position until the mode shape looks to be good enough. In the end, maybe there is still some residual modes. But it will be fine.

11. Change power to few hundred of milliwatt. Putting a BS before camera and direct the light to a PD to see cavity scan spectrum on an oscilloscope.

12. Fixing in-coupling mirror by the sequence of diagonal so that the alignment can be less affected.

13. Adjusting the injection laser to finalize the alignment.

After doing this procedure, we got OPO cavity scan spectrum as Fig.1. The zoom-in of TEM00 is Fig.2. We haven't characterized how much power is on higher order modes. But it looks decent for us.

Images attached to this report
2783_20220107115932_wechatimg758.jpeg 2783_20220107115942_wechatimg759.jpeg
Comments related to this report
YuhangZhao - 21:37, Tuesday 11 January 2022 (2787)

The camera used in this set-up is found to be tilted today. Fig.1 shows that this tilt is introduced through the interface between camera and post.

During the alignment, we use camera to characterize cavity HG modes shapes. When the beam and camera height is aligned, we found beam appears not in the center of screen as Fig.2. We suspect the tilt between camera and post is the cause of the cavity modes mis-center as Fig.2.

Note that Fig.2 shows the non-ideal alignment of crystal, which has a black defect.

R&D (FilterCavity)
Print this report.
MichaelPage - 00:52, Thursday 06 January 2022 (2781)Get code to link to this report
OPO replacement incoupler cleaning

Yuhang and Michael

We checked the incoupling mirror again after Yuhang applied and peeled First Contact on each day of the holidays. It seems to be much cleaner now. Perhaps the light spots that can still be seen are internal scattering or something like that.

Images attached to this report
2781_20220105164951_cr67618.jpg 2781_20220105165005_cr67615.jpg 2781_20220105165039_cr67612.jpg 2781_20220105165053_cr67609.jpg
KAGRA MIR (General)
Print this report.
MarcEisenmann - 17:07, Wednesday 05 January 2022 (2780)Get code to link to this report
Comment to Start of SHINKOSHA 7 birefringence measurement (Click here to view original report: 2777)

Here is the result of the vertical and horizontal angles of incidence that will be used during the measurement :

vertical ; 4.75 mrad or 0.272 deg

horizontal ; 41.94 mrad or 2.403 deg

KAGRA MIR (Absorption)
Print this report.
MarcEisenmann - 15:08, Wednesday 05 January 2022 (2778)Get code to link to this report
spare ETMY absorption

Here is the result of spare ETMY bulk absorption with the 2 ears flats.

The measured value agrees with the one measured at Caltech and reported in entry 1583

Images attached to this report
2778_20220105070807_xy.png 2778_20220105070812_yz.png 2778_20220105070816_xz.png
KAGRA MIR (General)
Print this report.
MarcEisenmann - 20:11, Tuesday 04 January 2022 (2777)Get code to link to this report
Start of SHINKOSHA 7 birefringence measurement

Abe, Marc and help from Yuhang to move safely the mirror

We removed the spare ETMY but did not placed it inside its box as we might put first contact before the shipping back to ICRR.

We checked the surface and bulk calibration without tuning the pump nor the probe beams alignment and got R_surface = 17.87 /W and R_bulk = 0.6503 cm/W

We measured the pump beam incidence angle and installed the SHINKOSHA 7 on the translation stage (X_center = 398 mm, Y_center = 122 mm and Z_center = 71 mm).

After the usual calibration we started the measurement with s polarization at the input.

Comments related to this report
MarcEisenmann - 17:07, Wednesday 05 January 2022 (2780)

Here is the result of the vertical and horizontal angles of incidence that will be used during the measurement :

vertical ; 4.75 mrad or 0.272 deg

horizontal ; 41.94 mrad or 2.403 deg

R&D (FilterCavity)
Print this report.
MichaelPage - 16:47, Tuesday 28 December 2021 (2776)Get code to link to this report
OPO incoupling mirror cleaning

Yuhang and Michael

We removed First Contact from the mirror and inspected the cleaning. It seems like some residue of the rubber stain remains and there is a lot of scattered light (fig 1 and 2).

After this we applied alcohol solution from a bottle in the ATC cleanroom (isopropanol based, I think it was called "Clarity" or something like that) in an attempt to remove water soluble stains (fig 3). Then after removing we reapplied First Contact.

Images attached to this report
2776_20211228084404_cr67559.jpg 2776_20211228084411_cr67561.jpg 2776_20211228084435_cr67590.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 19:48, Monday 27 December 2021 (2775)Get code to link to this report
Comment to BS seems to touch suspension frame again (Click here to view original report: 2767)

The BS issue has been checked again today by switching off the oplev laser since I was suspecting that BS PSD maybe broken. Indeed, after switching off BS oplev laser, the BS oplev signal is as Fig.1 which is the same as what is taken last Thursday. So the strange signal we are seeing is from the broken BS PSD.

To confirm the health of BS suspension, we need to take an another PSD and use it to characterize BS suspension.

Images attached to this comment
2775_20211227115020_bsoplevlaseroff.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 19:35, Monday 27 December 2021 (2773)Get code to link to this report
Comment to Guardian deployed on TAMA DGS (Click here to view original report: 2122)

Today, we have some progress about guardian deployment in TAMA. Thanks to the help from Yamat. The detailed information can be found in the #filter_cavity channel of gokagra in Slack. Here, I put a summary of my activity today.

1. Yamat helped us check simulink file and found no problem.

2. Yamat suggested to run caget K1:FDS-FC_GR_TRA to check potential issue on client workstation of desktop1 and k1grd0. Running it on desktop1, I got -27.7212 for K1:FDS-FC_GR_TRA. But running it on k1grd0, I got channel connect timed out for K1:FDS-FC_GR_TRA.

3. Yamat suggested to check environment values in guardian computer. To do that, I used 'env > tama_filter_cavity_k1grd0_env_out.txt' to save the environment variable of k1grd0 workstation. Then I used 'scp tama_filter_cavity_k1grd0_env_out.txt controls@192.168.11.110:/home/controls/Desktop' to copy environment variables from k1grd0 workstation to desktop1 workstation. To share with Yamat on Slack, I firstly uploaded the txt file to dropbox and then share link with him. (Note that I didn't just take screenshot because the environment variable information is large.)

4. After Yamat checking env output, he found that a variable called 'EPICS_CA_ADDR_LIST' seems to have a wrong IP address. He suggested to change this from '/home/controls/.bashrc or /kagra/apps/etc/client-user-env.sh.' and reboot k1grd0.

5. I found the channel 'EPICS_CA_ADDR_LIST' is in /kagra/apps/etc/epics-user-env.sh. So I did the modification and reboot k1grd0. But the problem is not solved.

6. I found after modifying from /kagra/apps/etc/epics-user-env.sh, the env output in k1grd0 is still as before and didn't change. Maybe this is why the problem is not solved. Now I am waiting for the answer from Yamat.

R&D (FilterCavity)
Print this report.
MichaelPage - 19:21, Monday 27 December 2021 (2774)Get code to link to this report
OPO incoupling mirror cleaning

Yuhang and Michael

This is a followup to the issues discussed in 2769

The OPO cavity was aligned and we attempted to optimise the mode structure of the transmission spectrum. However, there were two second order HOMs we could not get rid of by altering the input steering mirrors, and inspection of the transmission camera showed a lot of scattered light around the main beam.

Today, when we looked at the setup, we saw a lot of dirt on the incoupling mirror, which was visible when shining a phone flashlight below it. So we took it out to apply First Contact on both sides. At this point, we also saw a mark from the rubber o-ring that was mounted on the mirror. There may be some inspection needed for the OPO as well. However, I do not personally recall seeing the scattered light problem when the OPO was being aligned without the incoupling mirror, the most I remember seeing was something like a diffraction pattern (attached figure).

We were also in doubt of the position of the OPO inside the holder apparatus. There is some freedom in the beam propagation direction with respect to the OPO placement inside the holder, and there was no specific instruction given in the assembly directions. Given that the beam size required is extremely small , it seems like there would be strict mode matching required here, and yet we also did not see any LG modes when scanning the cavity.

For now, we applied clear First Contact to both sides of the incoupling mirror. We also want to redo the OPO alignment. Most of previous works have ensured the beam propagation axis is very closely aligned to the OPO axis. But, the incoupling mirror position was not that finely adjusted last time.For example, if the beam is hitting the incoupling mirror off-center while not passing through the center of curvature, the wavefronts will not be perpendicular to the curved surface of the mirror, so there will be mode mismatch introduced. The micro adjusters for the incoupling mirror only adjust transverse x and y position, not tilt.  So we should optimise the beam position on the incoupling mirror, and then for locking make fine adjustments to the input beam steering mirrors.  

Images attached to this report
2774_20211227111306_opoaligndiffract.jpg
KAGRA MIR (Absorption)
Print this report.
MarcEisenmann - 15:56, Monday 27 December 2021 (2772)Get code to link to this report
spare ETMY absorption measurement started

Today I did 2 long Z scans (between 20 and 100 mm then 100 mm to 110 mm) with 0.05 mm step size and median/average filters order 10 and could see the phase transition at Z = 37.6 mm and Z = 104.95 mm.

It means Z_center = 71.275mm.

The XY map started with 70 mm radius, median/average filters order 10 for a ~14h long measurement Pt = 6.541 W and Pin = 7.582 W.

It means that laser will be on up to December 30th for the 3D measurements.

KAGRA MIR (General)
Print this report.
MarcEisenmann - 11:10, Monday 27 December 2021 (2771)Get code to link to this report
angles of incidence for spare ETMY birefringence measurement

Abe, Marc

Here are the vertical and horizontal angle of incidences used during the briefringence measurement of the spare ETMY.

vertical : 4.78 10^-3 rad or 0.27 deg

horizontal : 4.33 . 10^-2 rad or 2.48 deg

Images attached to this report
2771_20211227031009_horizontalangleincidence.jpg 2771_20211227031012_verticalangleincidence.jpg
R&D (FilterCavity)
Print this report.
NaokiAritomi - 20:38, Sunday 26 December 2021 (2770)Get code to link to this report
IR locking accuracy with 1/f^4 and 1/f filters in FC green lock

To compare the IR locking accuracy (CCFC error signal) for different FC green gain, I compared the 1/f^4 and 1/f filters of FC green lock. The 1/f^4 filter has much larger gain than the 1/f filter at low frequency. The nominal filter is 1/f^4 and the nominal gain is 1.5. For the 1/f filter, I used the FC gain of 5 to have the UGF of 14kHz. Fig 1 shows FC OLTF for 1/f^4 and 1/f filters. Both filters have UGF of 14kHz.

Fig 2 shows the FC green error signal (EPS1, 1Vpp fixed range) with 1/f^4 and 1/f filters. The FC green error signal with 1/f^4 filter at low frequency is below the dark noise and limited by the spectrum analyzer noise. This means that increasing the green gain just reinjects the dark noise.

Note that BS pointing and AA were engaged during the measurement, but Z correction was not engaged due to the problem. I found that 1/f filter with gain 5 is stable even without Z correction, but 1/f^4 filter is unstable without Z correction.

Then I aligned BAB to FC. The optimal p pol PLL frequency without green was 275MHz and BAB power before FC was 466uW. The maximum IR transmission of FC was 490.

Then I checked the CCFC error signal. I optimized the p pol PLL frequency to maximize the CCFC error signal. The optimal p pol PLL frequency with 20mW green was 230MHz and the CCFC error signal was 122mVpp.

Fig 3 shows the CCFC error signal with 1/f^4 and 1/f filters. We can see that increasing the green gain improves the IR locking accuracy above 10Hz, but not below 10Hz. The IR locking accuracy is dominated by the FC length noise below 10Hz and the laser noise above 10 Hz. By increasing the green gain, the laser noise can be reduced because the laser noise is common for both IR and green, but the FC length noise cannot be reduced because the FC length noise is different for IR and green.  

Images attached to this report
2770_20211226123807_fcoltf.png 2770_20211226123823_fcgreenerrorsignal20211226.png 2770_20211227112553_lockingaccuracy20211227.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 23:42, Friday 24 December 2021 (2769)Get code to link to this report
New OPO closed and cavity scan

Michael and Yuhang

Today, we have successfully closed OPO cavity. The OPO no.28 component issue is solved by replacing it with a longer screw.

After cavity closed, we optimized OPO cavity alignment relative to its incident beam. However, we found the scan spectrum is as Fig.1.

1. We can see that there are apparently two higher order modes in the scan spectrum. By scanning OPO cavity with 10mHz ramp and looking at a designated camera , we found that the higher order modes are second-order Hermit Gaussian modes. (The cavity scan video is here https://drive.google.com/file/d/1Od2jI40D_3oZuZksuI474tunwtt7znIs/view?usp=sharing , note that it seems cavity mode is not centered in a circle. But this circle is just a mirror. The cavity mode looks to be well centered when mirror is removed.)

2. We can also see that there are modes overlapping with TEM00 mode. It indicates that we are having a cavity length which offers a non-ideal Gouy phase.

3. When we misalign pitch or yaw of the incident beam, we found the first order mode appears but the existing second order mode decreases. This is wired for me.

4. We have identified a potential beam clipping in the optical path before OPO. After fixing the clipping issue, we didn't get better OPO scanning spectrum.

5. We suspect that we may fix either crystal or in-coupling mirror too tightly, so that mirror has distortion and introduces higher order modes. After loosing a bit the screws, we didn't see improvement.

6. We suspect that there maybe some contamination. Especially, the crystal was in air for several months. But we are not sure if contamination is really true because the cleanroom in ATC is designed to be a class 10 cleanroom.

Personally, I think we need to open OPO again and adjust maybe crystal position.

Images attached to this report
2769_20211224153705_wechatimg26.jpeg
KAGRA MIR (General)
Print this report.
MarcEisenmann - 21:40, Friday 24 December 2021 (2768)Get code to link to this report
preparation of spare ETMY absorption measurement

Abe, Marc (with help of Michael and Yuhang to safely move the spare ETMY)

First we removed the spare ETMY.

Then, we used the razor blade and high power power meter connected to the lock-in amplifier DC to check the pump beam angle of incidence in vertical and horizontal planes. (analysis to follow)

We installed the surface reference sample and got R ~15.7 /W without tuning the alignment.

We tuned the alignment and got R_surface = 18.537 /W with Z_translationStage = 42 mm and Z_IU = 66 mm.

We also measured R_bulk = 0.6703 cm/W.

We reinstalled the spare ETMY with the 2 ears flat (ie same configuration as in entry 2755.

We tuned Z_IU to 5.4 mm taking into account the thickness after repolishing (14.3 cm instead of 15 cm).

We started to look for the 2 surfaces signal increasing the input pump power from 2W to about 7W (in transmission of the spare ETMY).

At that point we may have seen the expected signal with median and average filters order 10 but because the step size was 0.2mm we could not clearly see them.

For reference, the last time that the spare ETMY absorption has been measured the input pump power was about 10 W (see for example entry 1601)

As it was getting late on Christmas eve we stopped there and will start absorption measurement on monday.

R&D (FilterCavity)
Print this report.
YuhangZhao - 14:33, Friday 24 December 2021 (2767)Get code to link to this report
BS seems to touch suspension frame again

Aritomi, Yuhang

There was an earthquake on 12th Dec. which caused mis-alignment of suspended mirrors. We worked on the re-alignment on 22th Dec. During the re-alignment, we found that the BS picomotor can be only moved in pitch direction but not in yaw direction. Fortunately, we were able to bring back the filter cavity alignment even in this situation.

Since BS picomotor couldn't be moved in yaw direction, we looked into the BS oplev spectrum. The result is shown in Fig.1 and we can see that the oplev signal is totally off from the nominal value. Note that input oplev was different because we took pre-amplifier from input temporarily.

For the time being, we can still work with this condition. But maybe we should correct BS mirror from touching in the near future.

Images attached to this report
2767_20211224063701_20211223vischeck.png
Comments related to this report
YuhangZhao - 19:48, Monday 27 December 2021 (2775)

The BS issue has been checked again today by switching off the oplev laser since I was suspecting that BS PSD maybe broken. Indeed, after switching off BS oplev laser, the BS oplev signal is as Fig.1 which is the same as what is taken last Thursday. So the strange signal we are seeing is from the broken BS PSD.

To confirm the health of BS suspension, we need to take an another PSD and use it to characterize BS suspension.

R&D (FilterCavity)
Print this report.
NaokiAritomi - 21:16, Thursday 23 December 2021 (2766)Get code to link to this report
FC lock problem (2)

[Aritomi, Yuhang]

We continued the investigation of the 20kHz peaks in FC green error signal. We checked the IRMC error signal and there are no 20kHz peaks in the IRMC error signal. So the 20kHz peaks should not come from laser.

Then we measured SHG OLTF and found that there is a peak around 20kHz as shown in Fig 1. In this measurement, the SHG gain was 1.3. We decreased the SHG gain from 1.3 to 1 to avoid the 20kHz peak, but the FC green error signal is still noisy.

Regarding the Z correction, we found that the dc_damp2 filter was completely different from the one written in elog2608. We changed the dc_damp2 filter to the one written in elog2608. We set the dc_damp2 filter gain as 10 in foton and 4 in medm. However, FC is still not very stable.

Images attached to this report
2766_20211223131556_shgoltf20211223.png
KAGRA MIR (Polarization)
Print this report.
MarcEisenmann - 16:38, Thursday 23 December 2021 (2765)Get code to link to this report
Comment to Spare ETMY rotated by 30deg (Click here to view original report: 2757)

Marc, Michael

To facilitate the removal of the mirror, we used the knob controller of X direction to move the mirror.

After that, we found out that the Zaber reading got good so we started the measurement with 45 deg input polarization angle.

R&D (FilterCavity)
Print this report.
NaokiAritomi - 23:23, Wednesday 22 December 2021 (2764)Get code to link to this report
FC lock problem

We found that FC green error signal is very noisy. Fig 1,2 show the FC green error signal before/after lock. Fig 3 shows the spectrum of the FC green error signal after lock. There are two peaks around 20 kHz in the FC green error signal. It seems that the two peaks come from SHG. The SHG error signal is also shown in Fig 3. The SHG gain was 1.4. We decreased the SHG gain from 1.4 to 1, but the peaks around 20 kHz did not change.

During this investigation, we checked green power and OLTF of FC green lock. The measured green power is as follows. The green power seems reasonable. The UGF was 14kHz, which is also reasonable. Note that FC gain was 1.6.

green power before AOM 48.6 mW
green power before FC 24.3 mW

We also found that the Z correction is not very stable. We quickly checked the filters in filter bank of Z correction as follows.

not bad filters damp, damp2
bad filters DCdamp, dc_damp2, dc_damp3 

The bad filters immediately caused FC unlock when they were engaged. The not bad filters didn't cause the immediate unlock, but FC was not very stable for long term. We need to investigate the Z correction.

Images attached to this report
2764_20211222152336_img9082.jpg 2764_20211222152343_img9081.jpg 2764_20211222152353_fcshgerrorsignal20211222.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 22:50, Wednesday 22 December 2021 (2763)Get code to link to this report
Fixing cable connection issue for new OPO assembly

Michael and Yuhang

Some cables need to be connected to the new OPO, such as PZT cables, thermometer cables, and Peltier cables. The old connection was not firm enough and we decide to use a new way to connect cables. This new way is suggested by Tomaru-san and it uses a copper tube as interface. After putting the two sides of cables inside the copper tube (Fig.1,2), we use a plier to clamp two cables so that the cables are fixed (Fig.3,4). Then we use Kapton tape to isolate the copper tube from environment (Fig.5).

After fixing, we found the connection is firm enough. We also tried to drive PZT and found the driving is successful.

Images attached to this report
2763_20211222144919_1640180757637.jpg 2763_20211222144926_1640180757696.jpg 2763_20211222144952_1640180757722.jpg 2763_20211222144957_1640180757742.jpg 2763_20211222145001_1640180757765.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 22:14, Wednesday 22 December 2021 (2762)Get code to link to this report
An additional issue about one of the new OPO mechanical components (component number 28)

Michael and Yuhang

Today we found a new issue about new OPO component No.28. The issue is described in the attached PDF file. The issue is that a counter bore hole is made from a wrong side of no.28 component.

We also took photos about no.28 component with wrong and correct holes, as shown in the attached Fig. 1 and 2. (The one with correct hole actually has wrong thread as reported in elog2724)

For solution, we found that a longer screw seems to be able to solve the problem. We will look for some longer screws and clean them. Then try to use them for the new OPO.

Images attached to this report
2762_20211222141404_1640176191686.jpg 2762_20211222141409_1640176191608.jpg
Non-image files attached to this report