LOG-IN
Displaying reports 1061-1080 of 3274.Go to page Start 50 51 52 53 54 55 56 57 58 End
KAGRA MIR (Absorption)
Print this report.
MarcEisenmann - 15:55, Wednesday 16 June 2021 (2580)Get code to link to this report
absorption map of the spare viewport

Despite 2 windows updates that turned off the computer I finally finished the measurement of the absorption map of the spare viewport (see fig 1).

In order to recover the lost time I changed the step size to 0.5 mm.

The mean absorption is at the order of 18 ppm while there is some dust/surface defects creating few spikes in the absorption.

Images attached to this report
2580_20210616085645_absorptionmap.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 15:24, Wednesday 16 June 2021 (2579)Get code to link to this report
A mcmc fit of FDS measurement in elog2546

Marc and Yuhang

We ran mcmc fit of measurement in elog2546 based on the emcee pakage. This run took about 3 hours by using a computer with 20 cores, which is ten times faster than the time of 30 hours by using a single core. This time consuming is reasonable since there will be some overhead while using multiprocessing.

In this run, we tried two situations: two free parameters (detuning, homodyne angle), four free parameters (squeezing level, optical losses, detuning, homodyne angle).

The mcmc run for two free parameters is shown in attached figure 1 and 2. This run is very successful since we see a clear contour distribution in figure 1 and parameters are always in the stable region as figure 2.

  least square mcmc
homodyne angle 5.3 +/- 0.2 4.9 +0.1026/-0.0987
detuning 72.1 +/- 0.6 78.25 + 0.5013/-0.4992

The mcmc run for four free parameters is shown in attached figure 3 and 4. This run needs some optimization.

Images attached to this report
2579_20210616081646_2pcorner.png 2579_20210616082435_2pvalues.png 2579_20210616082439_4pcorner.png 2579_20210616082443_4pvalues.png
R&D (FilterCavity)
Print this report.
NaokiAritomi - 22:20, Tuesday 15 June 2021 (2578)Get code to link to this report
IR injection mode matching with new beam spot

I measured the IR injection mode matching with new beam spot which has more stable IR transmission.

First I optimized IR injection alignment with new beam spot. I checked the IR injection alignment by changing AOM frequency. The mode matching is 95.6%. Note that AA and BS pointing were engaged during this measurement. 

After the IR injection alignment, IR transmission was 460 with 430uW injection. According to our experience, the IR transmission should be 490 with 430uW injection. The possible reason of the smaller IR transmission is that the IR transmission is off centered in transmission PD with new beam spot.

AOM frequency (MHz) mode IR transmission
109.53805 TEM00 460
109.93494 pitch 100
110.33183 LG 105
  offset 94
 
After the IR injection alignment, IR transmission was 460 with 430uW injection. 
According to our experience, the IR transmission should be 490 with 430uW injection. I'm not sure why it is smaller now.
R&D (FilterCavity)
Print this report.
MichaelPage - 16:46, Tuesday 15 June 2021 (2577)Get code to link to this report
Comment to OPO nonlinear gain revisited (Click here to view original report: 2569)

I recorded the thermistor (temperature) values that I used for the OPO nonlinear gain measurement.

Initially I searched in increments of +/- 0.01 kOhm on the thermistor, and then checked a bit within the optimal range. Perhaps there is further room for optimisation when zooming in on the oscilloscope though. At 100 mV/div ranges on the oscilloscope, the difference in voltage for thermistor change under 0.01 kOhm was hard to distinguish.

In the table I give the temperature, oscilloscope reading as well as the range value on the power meter - using a range of 1.6 mW on the power meter gives 10x more voltage on the oscilloscope than a range of 16 mW.

green power [mW] 0 10 15 20 25 30 35 40 50 60
MZ offset   3.9 4.0 4.1 4.2 4.3 4.4 4.5 4.7 4.9

Amplification reading [mV] (power meter range [mW])

456 (1.6) 1140 (1.6) 1710 (1.6) 234 (16) 308 (16) 408 (16) 578 (16) 816 (16)    
De-amplification reading [mV] (power meter range [mW]) 456 (1.6) 240 (1.6) 216 (1.6) 192 (1.6) 178 (1.6) 166 (1.6) 160 (1.6) 152 (1.6)    
Thermistor value [kOhm] 7.137 7.147 7.165 7.173 7.185 7.196 7.203 7.215 7.215 7.225
p pol PLL frequency [MHz] 190 180 195 190 200 205 200 205 185 185
Nonlinear gain 1 2.5 3.8 5.1 6.8 8.9 12.7 17.9    
KAGRA MIR (Absorption)
Print this report.
MarcEisenmann - 15:13, Monday 14 June 2021 (2575)Get code to link to this report
spare viewport measurement started

Today I increased the laser diode current to 3A.

I also took compared the output power as a function of HWP angle :

HWP angle [deg] power [W]
19 1.0561
29 0.2020
39 0.0258
49 0.613
59 1.691
69 2.764
79 3.310
89 3.093
99 2.207
109 1.067

I set the HWP to 49 degrees did a large z scan and could see some AC signal.

I checked the viewport center :

Z_center = 73.4 mm (from phase = 0)

Y_center = 120.95 mm (from ac/dc = 0)

X_center ~ 331.75 (one of the extrema was not possible to estimate because I placed a stop in Zaber to avoid burning the holder.

The X_center was estimated from the top X value of 383.9 mm and assuming a similar radius as the Y direction (52.15 mm)

I tuned the DC at this position and got as a preliminary absorption value :

corr = 1;
ac = 8e-4;
dc = 4.08;
Pin = 0.613;
R_surf = 17.033
absorption = ac/(dc*P_in*R_surf)*corr ~416 ppm 18 ppm (Edit : I took wrong power in previous estimation...)

The absorption map is now on-going with a 3 cm radius and 50 um step size.

Comments related to this report
MarcEisenmann - 11:10, Friday 18 June 2021 (2584)

I fitted the power as a function of the HWP angle.

Quite useful to know how to increase the power step by step.

R&D (FilterCavity)
Print this report.
NaokiAritomi - 13:01, Monday 14 June 2021 (2574)Get code to link to this report
100Hz bump came back

This measurement was done on June 10th. 

Recently the bump at 100Hz came back as shown in an attached figure.

Images attached to this report
2574_20210614060135_20210610fdsccfc.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 10:21, Monday 14 June 2021 (2573)Get code to link to this report
Find the center of mirror for beam in filter cavity

Marc and Yuhang

Marc and I were suspecting that the beam may not hit on the center of mirror. To check that, we decide to scan BS pitch/yaw and check IR transmission.

Today I did this check. I found two resonant beams while scanning BS, which can be seen from the attached video (two videos show the situation of pitch and yaw separately).

Yaw: https://drive.google.com/file/d/1ahDx6MTfYqxdoFXbyH0PbrLRPybCtIhp/view?usp=sharing

Pitch: https://drive.google.com/file/d/1N6N3BdUr3WIG_GeNLJeDQDmCbeHzdTXm/view?usp=sharing

By pointing the incident beam towards the center of mirror, these two resonant beams get closer. In the end, they still don't overlap and separate by roughly a HWHM. In the end, the pointing to the center has a precision of 2urad, considering the distance of 300m, this is 0.6mm precision.

After centering the beam, I measured the IR length noise by using BAB. Attached figure 1 shows their difference. We can see that the low frequency length change is greatly reduced. Note: the calibration of IR length is done by taking the slope of PDH signal as attached figure 2, 3, 4, 5 for on resonance and detuned case, which may introduce some error. Especially, the calibration of the green line (beam off-center and on resonance) was using a PDH signal with opposite slope compared with others.

  old center for pointing new center for pointing
pitch 24 4.5
yaw -8 16

The attached figure 6 shows the new location of beam now. This moves the beam by 6mm in both yaw and pitch on end mirror. We should go to end room and center beam on both camera and PSD.

In addition, the AOM frequency changed by 2415Hz, which indicates the GR and IR length difference is 2415/2e-12m = 1.2075nm. This tells us, with the residual pointing error of IR, we have 1nm GR/IR length difference after beam moves 6mm on mirror. This phenomon explains the IR detuning change issue for different alignment of GR.

Images attached to this report
2573_20210614052354_figure4.png 2573_20210614064947_wechatimage20210614122125.jpg 2573_20210614064954_wechatimage20210614122143.jpg 2573_20210614064959_wechatimage20210614122150.jpg 2573_20210614065004_wechatimage20210614122157.jpg 2573_20210614065119_wechatimage20210614135106.jpg
Comments related to this report
YuhangZhao - 10:48, Friday 25 June 2021 (2601)

I was using AOM scanning speed as 4000Hz/1.7s in the calibration. However, since the scanning speed for IR is 1/2 of the value for GR, the figure in the old elog was wrong.

Calibration for the measured spectrum should be: calibration = 2000/1.66666*11.5/11.2 #Hz/V (PDH: 11.2mV/11.5ms) (AOM: 4000Hz/1.66666s)

There was also problem for the calibration for off-center on-resonance, I modified the plot by using a more reasonable calibration. It comes from the center on-resonance. The new plot is shown in the attached figure.

We can see the new stable optical axis makes especially the low frequency length noise reduced. However, the high frequency noise is increased a bit.

NaokiAritomi - 04:32, Saturday 26 June 2021 (2603)

The high frequency noise is same for old and new beam spots, but is increased for 50Hz detuning compared with the one on resonance. This noise difference could be explained by the cavity pole effect. The cavity pole effect for 50Hz detuning (half detune) is smaller than the one on resonance by a factor of ~sqrt(2). Please check P.50 of LIGO-T1800447 for the cavity pole effect of detuned cavity.

KAGRA MIR (Absorption)
Print this report.
MarcEisenmann - 21:07, Sunday 13 June 2021 (2572)Get code to link to this report
Summary of preparation of the clean viewport characterization

This entry summarizes the (too many) measurements performed last Thursday and Friday...

I started by checking the bulk reference sample as it was the last calibration measurement performed ( see entry 2510).

First, I had to change the HWP that controls the ir laser power from 118 degrees to 119 degrees in order to get the usual Pin~28 mW.

Then, I found out that the calibration was lower than expected at 0.5 cm/W.

This, together with the (small) laser power change seems to indicate that the beam parameters changed during a 2 weeks window.

My guess is that, the temperature/humidity increase of last week affected the laser beam parameters (effects of temperature inside TAMA central building can be found in entry 2493 where PR pitch motion seems to follow trends similar to temperature change troughout the day).

I replaced the bulk reference sample by the surface reference sample and started several measurements...

To summarize :

I checked the IU optimal position : 68 mm is still good.

I checked the sample Z position : after several tests, I found out that a good position for the surface reference sample is at Z = 39.5 mm

This corresponds indeed to have the surface of this sample at the ir beam waist and crossing with the red beam (details in entry 2446)

With the good sample Z position, I still could not achieve 'optimal' calibration factor.

I checked again the IU position and confirmed than IU at 68 mm is optimal.

I found out that having this sample Y center (ie Y = 121.2 mm) at the beam crossing was giving :

y = 121.2

AC_surfref = 0.4075;

DC_surfref = 3.966;

acdc = 0.1026;

P_in = 27.6e-3;

R_surf = 16.8972

 

while moving the Y position to 120.2 mm gives

AC_surfref = 0.411;

DC_surfref = 3.96;

acdc = 0.1038;

P_in = 27.6e-3;

R_surf = 17.0949

I will need to check if this sample has some dust in the center...

Finally, I moved on the spare viewport measurement :

I placed it on the TAMA holder using 2 spacers to fix it.

The good Z position to have the surface at the 2 beams crossing point is z = 38 + 60/2 = 68 mm

The good IU Z position is 68 - 3.4 = 64.6 mm (here I assumed 1 cm thickness of the viewport).

Unfortunately, at that point I moved by mistake the last lens on the IR path... So I performed again the surface calibration as :

AC_surfref = 0.4125;

DC_surfref = 3.974;

acdc = 0.1038;

P_in = 27.7e-3;

abs_surfref = 0.22;

R_surf = 17.0331 (screenshot saved at 20h35)

I re-installed the spare viewport and tried to measure its absorption.

28 mW did not show any signal.

428 mW (ie HWP at 89 degrees)  did not show anything as well...

As this power is around the maximal that can be produced with the laser diode current of 1 A and because any current change requires ~1h to stabilize the laser power, I stopped the measurement at that point.

Anyway, this tells us that the absorption is quite low...

R&D (FilterCavity)
Print this report.
NaokiAritomi - 05:51, Saturday 12 June 2021 (2570)Get code to link to this report
Comment to OPO nonlinear gain revisited (Click here to view original report: 2569)

Good measurement! Now it is obvious that the OPO threshold is lower than before as I said. 

R&D (FilterCavity)
Print this report.
MichaelPage - 22:52, Friday 11 June 2021 (2569)Get code to link to this report
OPO nonlinear gain revisited

Retry of the OPO nonlinear gain measurement in 2528, motivated by the idea that the uncertainty in the detuning fit (i.e. 254425462550) may be caused by uncertainty in the fit parameters (see 25442549).

The previous measurement of OPO nonlinear gain was too imprecise in a way that user error had a lot of impact. In the previous process, we left the OPO unlocked while sending a sine wave to GRPS to modulate between amplification and deamplification. We then used the oscilloscope persist function and triggered the resonant peak to appear on the screen. The way this was intended to work is shown in figure 1, however, it seemed like the peak was fluctuating in the horizontal axis too much due to the cavity unlock, and obtaining the minimum value was quite imprecise. On top of that, the OPO temperature was not optimised for different green injection power to compensate for absorption. This of course causes an error in the co-resonance of p-pol/BAB and green.

This time I measured again with the OPO locked, and optimised the temperature and PLL frequency for each value of green injection power. With the cavity locked, the output transmission is just a sine wave due to GRPS modulation, going between amplification and deamplification. While there is less uncertainty in just reading off the values now, there is of course some user uncertainty in optimising the temperature and PLL frequency, especially because being too careful about it is very time consuming. Juding by the optimisation process, I would say that the temperature is easily optimised to about +/- 0.01 kOhm on the thermistor, with differences in this range amounting to about +/- 2-8 mV on the amplification reading on the oscilloscope, for values between 100-1000 mV.

The values were measured using a power meter in transmission of the OPO. There are a lot of green junk beams coming out of the OPO as well, so they were removed with a laser line filter to leave only the BAB transmission. I measured a calibration factor of 0.00451 mW/mV for the correspondance between power meter readout and oscilloscope voltage, but in retrospect this is unnecessary, since only the relative power matters, and the oscilloscope voltage is linear with power meter incident optical power. Aritomi also has a more accurate calibration in 2566.

Figure 2 shows the OPO nonlinear gain with simple curve fitting. I will further consider the error of the input parameters in a future update.

Images attached to this report
2569_20210611155138_persistfunction.png 2569_20210611155152_nonlineargain1.jpg
Comments related to this report
NaokiAritomi - 05:51, Saturday 12 June 2021 (2570)

Good measurement! Now it is obvious that the OPO threshold is lower than before as I said. 

MichaelPage - 16:46, Tuesday 15 June 2021 (2577)

I recorded the thermistor (temperature) values that I used for the OPO nonlinear gain measurement.

Initially I searched in increments of +/- 0.01 kOhm on the thermistor, and then checked a bit within the optimal range. Perhaps there is further room for optimisation when zooming in on the oscilloscope though. At 100 mV/div ranges on the oscilloscope, the difference in voltage for thermistor change under 0.01 kOhm was hard to distinguish.

In the table I give the temperature, oscilloscope reading as well as the range value on the power meter - using a range of 1.6 mW on the power meter gives 10x more voltage on the oscilloscope than a range of 16 mW.

green power [mW] 0 10 15 20 25 30 35 40 50 60
MZ offset   3.9 4.0 4.1 4.2 4.3 4.4 4.5 4.7 4.9

Amplification reading [mV] (power meter range [mW])

456 (1.6) 1140 (1.6) 1710 (1.6) 234 (16) 308 (16) 408 (16) 578 (16) 816 (16)    
De-amplification reading [mV] (power meter range [mW]) 456 (1.6) 240 (1.6) 216 (1.6) 192 (1.6) 178 (1.6) 166 (1.6) 160 (1.6) 152 (1.6)    
Thermistor value [kOhm] 7.137 7.147 7.165 7.173 7.185 7.196 7.203 7.215 7.215 7.225
p pol PLL frequency [MHz] 190 180 195 190 200 205 200 205 185 185
Nonlinear gain 1 2.5 3.8 5.1 6.8 8.9 12.7 17.9    
R&D (FilterCavity)
Print this report.
NaokiAritomi - 15:27, Friday 11 June 2021 (2568)Get code to link to this report
Comment to FDS with CCFC around optimal detuning (Click here to view original report: 2556)

I analyzed again with correct shot noise level (-133.7dBVrms/rtHz) and 12% loss from visibility. The readout loss includes 12% loss from visibility, 2% loss from homodyne (dark noise, quantum efficiency, AR), and 20% loss from pick off BS.

sqz_dB = 11.8;                  % produced SQZ (dB)

L_rt = 120e-6;                  % FC losses

L_inj = 0.32;                   % Injection losses

L_ro = 0.31;                    % Readout losses

A0 = 0.05;                      % Squeezer/filter cavity mode mismatch

C0 = 0.06;                      % Squeezer/local oscillator mode mismatch

ERR_L =   1e-12;                % Lock accuracy (m)

ERR_csi = 30e-3;                % Phase noise (rad)

Images attached to this comment
2568_20210611115004_20210603ccfcfdsnew.png
R&D (FilterCavity)
Print this report.
NaokiAritomi - 08:28, Friday 11 June 2021 (2566)Get code to link to this report
visibility measurement

[Aritomi, Michael]

We measured the visibility between LO and BAB reflected from FC for 10s. The BAB was half detuned.

To know the BAB power during the measurement, we also measured the BAB pick off power in CCFC port. Fig 1 shows the BAB power before AMC and the BAB pick off power. From this measurement, we can know the calibration factor from the BAB pick off power to the BAB power before AMC (bottom plot in Fig 1). The calibration factor is 0.246. The LO power was 1.16V. The offset of visibility was 9mV and the offset of pick off was 5.8mV. Note that the offset subtraction is important for visibility calculation.

The measured visibility and BAB pick off are shown in Fig 2. The 10Hz modulation was applied to IR phase shifter. As you can see, both of visibility and BAB pick off are fluctuating. To calculate the visibility, we divided this data into 1000 segments with 0.01s step (Fig 3) and calculated the visibility in each segment (Fig 4). From the histogram, the visibility is 0.94(1) which corresponds to optical loss of 12(2)%.

In FDS measurement in PRL paper, we assumed that the visibility is 0.98 which corresponds to optical loss of 4%. It seems that the visibility got worse and we have 8% more loss than before...

 
Fig 1 shows the BAB power before AMC and the BAB pick off power. From this measurement, we can know the calibration factor from the BAB pick off power to the BAB power before AMC (bottom plot in Fig 1). The calibration factor is 0.257. The LO power was 1.16V.
 
The measured visibility and BAB pick off are shown in Fig 2. As you can see, both of them are fluctuating. To calculate the visibility, I divided this data to 1000 segments with 0.01s step (Fig 3) and calculated the visibility in each segment (Fig 4). From the histogram, the visibility is 0.92(1) which corresponds to optical loss of 15(2)%. 
 
In FDS measurement in PRL paper, we assumed that the visibility is 0.98 which corresponds to optical loss of 4%. It seems the visibility got much worse than before and we have 10% more loss than before...
Images attached to this report
2566_20210611112754_20210610visibilitycalibration.png 2566_20210611112804_20210610visibility.png 2566_20210611112811_01s.png 2566_20210611112821_20210610visibilityhistogram.png
R&D (FilterCavity)
Print this report.
NaokiAritomi - 19:54, Thursday 10 June 2021 (2565)Get code to link to this report
BAB power fluctuation after OPO

I measured BAB power fluctuation after OPO for one minute (attached picture). The BAB power fluctuation is 0.7% (Ratio of Max/Min = 1.007).

Images attached to this report
2565_20210610125451_img8955.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 17:22, Thursday 10 June 2021 (2564)Get code to link to this report
homodyne LO intensity may have long term change

Aritomi and Yuhang

In the past few weeks, we measured homodyne shot noise few times, we found the shot noise level changes from -133.2dB to -133.7dB.

We monitor the BAB intensity for almost one hour on Monday this week. We found BAB intensity changed by about 5% during this time (attached figure 1). By doing 20*log10(sqrt(1.05))-20*log10(sqrt(1)), we found this can introduce 0.2dB shot noise change.

Although BAB and LO both come directly from main laser, this is not a test done for a real LO. We decide to check the power of LO before and after IRMC soon.

Images attached to this report
2564_20210610102255_wechatimg141.jpeg
R&D (FilterCavity)
Print this report.
MarcEisenmann - 14:48, Tuesday 08 June 2021 (2563)Get code to link to this report
AA open loop transfer functions

Marc, Yuhang

In entry 2545 the filter was not exported properly and was actually just a gain...

This time we used the proper filter and got the correct 4 open loop transfer functions of automatic-alignment as shown in the 4 attached figures.

It seems that for input & end pitch, the 11 Hz resonance is really too close to unity gain.

And the frequencies around 1 Hz have similar behavior.

We'll tune the filter in order to reduce this effect.

Images attached to this report
2563_20210608074725_aaoltfinp.jpg 2563_20210608074729_aaoltfiny.jpg 2563_20210608074732_aaoltfendp.jpg 2563_20210608074735_aaoltfendy.jpg
R&D (FilterCavity)
Print this report.
NaokiAritomi - 12:24, Tuesday 08 June 2021 (2562)Get code to link to this report
Latest setting for 20mW green
green power (mW) OPO temperature (kOhm) p pol PLL (MHz) BAB maximum (mV) nonlinear gain
0 7.164 240 56.4 1
20 7.164 180 254 4.5
R&D (FilterCavity)
Print this report.
YuhangZhao - 12:47, Saturday 05 June 2021 (2561)Get code to link to this report
Comment to Filter cavity pointing loop control precision (Click here to view original report: 2555)

Marc, Michael, and Yuhang

We have done a calibration for pointing loop with a new method. We move the locking point of pointing loop and check how much BS needs to compensate.

Pointing pitch:

we move pointing pitch, as attached figure 1 [1,0] subplot, from  -24 to -34 [counts]. Then we found BS pitch, as attached figure 1 [0,0] subplot, needs to move from 275.8 to 326.0 [counts] to compensate. Taking into account the calbration of oplev, we got  calibration factor for pointing loop on the base of BS.

The pointing precision was estimated by a RMS integral of pointing loop error signal spectra. We also found that high frequency spectrum can be better if we center beam on PSD, which is shown in attached figure 1 [0,1] and [1,1] subplots. However, the RMS of these two seneriaos are the same, which is reasonable since the dominating noise is at low frequency. In conclusion, we got pitch pointing precision as 0.43urad.

pitch pointing loop calibration factor 1.3 urad/count.

Pointing yaw:

We have done the pointing yaw calibration in the same manner with pointing pitch. The details can be found in attached figure 2. In conclusion, we got yaw pointing precision of 0.65urad.

yaw pointing loop calibration factor 1.5 urad/count.

Note: to take a quicker measurement, we have increased pointing loop gain by factors of 3 and 5 for pitch and yaw. So you can see the BS compensation finishes within 1 minute. We have also tested the spectrum of poiting pitch/yaw don't show difference when gains are increased.

Images attached to this comment
2561_20210605054700_pit.png 2561_20210605054704_yaw.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 17:47, Friday 04 June 2021 (2560)Get code to link to this report
Delete data from \full directory in TAMA DGS

I consulted with Takahiro Yamamoto-san about if it is OK to delete data from \full directory and he said there is not important setting but just saved data inside here.

So I deleted the data from 2020/07/30 to 2020/11/21.

Now we have 63% in use, which means 37% space is released.

R&D (FilterCavity)
Print this report.
YuhangZhao - 09:45, Friday 04 June 2021 (2559)Get code to link to this report
Comment to Detuning stability with fixed homodyne angle (Click here to view original report: 2558)

A similar measurement was done in elog2350, I and Michael got quite similar result even with a time scale of 1 hour. However, if I remember and understand well, Aso-san had a comment that different homodyne angle may affect our measurement, which is what we have seen. It seems important to understand why detuning becomes not so stable if we measure with different homodyne angle.

However, the same measurement may be interesting to be made for without CCFC but with AA/pointing.

R&D (FilterCavity)
Print this report.
NaokiAritomi - 22:40, Thursday 03 June 2021 (2558)Get code to link to this report
Detuning stability with fixed homodyne angle

To investigate the detuning stability, we fixed the homodyne angle to squeezing quadrature and measured CCFC error signal and FDS three times with half an hour interval (attached figures). As you can see, the CCFC error signal and FDS are stable and the detuning fluctuation is ~2Hz.

The strange thing is that although the CC detuning is almost optimal, the CCFC error signal with demodulation phase of 23deg seems good detuning for FDS.

Images attached to this report
2558_20210808054043_20210603ccfc.png 2558_20210808054056_20210603fdsstability.png
Comments related to this report
YuhangZhao - 09:45, Friday 04 June 2021 (2559)

A similar measurement was done in elog2350, I and Michael got quite similar result even with a time scale of 1 hour. However, if I remember and understand well, Aso-san had a comment that different homodyne angle may affect our measurement, which is what we have seen. It seems important to understand why detuning becomes not so stable if we measure with different homodyne angle.

However, the same measurement may be interesting to be made for without CCFC but with AA/pointing.