LOG-IN
Displaying reports 1061-1080 of 3201.Go to page Start 50 51 52 53 54 55 56 57 58 End
KAGRA MIR (General)
Print this report.
MarcEisenmann - 17:58, Friday 14 May 2021 (2496)Get code to link to this report
dirty viewport received

Today we received two viewports of EY TM OpLev.

I placed them inside ATC cleanroom as in figure 1 with :

left : EY TM OpLev viewport, Inner shield, -X +Y side, dirty surface up

right : EY TM OpLev viewport, Inner shield, +X +Y side, dirty surface up

I left them under their protective cover as in last figure.

Images attached to this report
2496_20210514105519_img20210514172603.jpg 2496_20210514105527_img20210514172641.jpg 2496_20210514105607_img20210514172636.jpg 2496_20210514105908_img20210514172922.jpg
KAGRA MIR (Absorption)
Print this report.
SimonZeidler - 16:22, Friday 14 May 2021 (2495)Get code to link to this report
Comment to 3d absorption map of 1cm thick SHINKOSHA sample (Click here to view original report: 2476)

I have some questions.

What is the meaning of the fit in the histograms? Do you have a reason to fit with specific distributions/densities or do you want to find a systematic pattern?

R&D (FilterCavity)
Print this report.
YuhangZhao - 15:14, Thursday 13 May 2021 (2494)Get code to link to this report
Comment to TAMA suspension angular drift investigation (Click here to view original report: 2493)

Marc and Yuhang

We have also checked one week data (from 2021/04/25/3pm to 2021/05/01/3pm)

We made two plots, the first plot includes all the data in this week. But the second plot excluded the last part of the data. This is because an earthquake might happen around the end of the one week data. This makes we see a peak in the first plot. 

To compare different mirrors angular drift, we calculated RMS value of data and summarized them as the attached table. [unit: urad]

 

PR pit

PR yaw

BS pit

BS yaw

INPUT pit

INPUT yaw

END pit

END yaw

angular drift RMS 

(including earthquake)

59

4.5

14.35

8.85

31.24

15.81

9.81

2.86

angular drift RMS     earthquake excluded

62.46

3.98

10.39

9.07

29.42

15.49

9.08

2.66

From this table, we see PR pitch moved the most. But INPUT pitch also moved a lot. However, END pitch didn't move as large as PR and INPUT pitch. Since PR, INPUT and END have the same configuration, their difference in pitch drift indicates that this is not a design problem.

We know that temperature change influences mirrors drift. But PR, BS, and INPUT are all in the TAMA central room, they should have the same temperature change. But they didn't show the same drift. Form the airconditioner location, PR chamber is almost facing an air conditioner, which may cause temperature related problem. To confirm this is not the fault of air conditioner, we plan to switch off it for one or two days during weekends to check.

Images attached to this comment
2494_20210513081439_oneweek.png 2494_20210513081442_6days.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 14:06, Thursday 13 May 2021 (2493)Get code to link to this report
TAMA suspension angular drift investigation

According to the OPLEV signal calibration value in elog1874, I plotted the four suspended mirrors drift during a time scale of one day(from 2021/05/12 1pm to 2021/05/13 1pm JST).

In figure 1, all four suspended mirrors are compared. It is very clear that PR pitch has a far larger angular drift compared with other mirrors.

The PR mirror changed by 100urad pk-pk during one day. Considering the distance from PR to END mirrors, the 1e-4*300 = 3e-2 m = 3cm.

Other mirrors other DOF changed by 10urad pk-pk during one day, which is ten times smaller than PR pitch.

Considering this measurement result, I think we should make beam pointing loop act on PR mirror.

Images attached to this report
2493_20210513070608_suspensiondrift.png
Comments related to this report
YuhangZhao - 15:14, Thursday 13 May 2021 (2494)

Marc and Yuhang

We have also checked one week data (from 2021/04/25/3pm to 2021/05/01/3pm)

We made two plots, the first plot includes all the data in this week. But the second plot excluded the last part of the data. This is because an earthquake might happen around the end of the one week data. This makes we see a peak in the first plot. 

To compare different mirrors angular drift, we calculated RMS value of data and summarized them as the attached table. [unit: urad]

 

PR pit

PR yaw

BS pit

BS yaw

INPUT pit

INPUT yaw

END pit

END yaw

angular drift RMS 

(including earthquake)

59

4.5

14.35

8.85

31.24

15.81

9.81

2.86

angular drift RMS     earthquake excluded

62.46

3.98

10.39

9.07

29.42

15.49

9.08

2.66

From this table, we see PR pitch moved the most. But INPUT pitch also moved a lot. However, END pitch didn't move as large as PR and INPUT pitch. Since PR, INPUT and END have the same configuration, their difference in pitch drift indicates that this is not a design problem.

We know that temperature change influences mirrors drift. But PR, BS, and INPUT are all in the TAMA central room, they should have the same temperature change. But they didn't show the same drift. Form the airconditioner location, PR chamber is almost facing an air conditioner, which may cause temperature related problem. To confirm this is not the fault of air conditioner, we plan to switch off it for one or two days during weekends to check.

KAGRA MIR (Absorption)
Print this report.
MarcEisenmann - 13:57, Thursday 13 May 2021 (2492)Get code to link to this report
Comment to KASI sample absorption measurement (Click here to view original report: 2489)

Abe, Marc

Following the 3 maps measurements we performed again the bulk calibration where this time we moved the imagining unit by 0.32 mm in order to compensate the thickness difference between surface and bulk reference samples. We got the following result (also see last figure of this entry) :

AC_bulkref = 0.0731;
DC_bulkref = 4.164;
ACDC = 0.01755;
P_in = 29.5e-3;
P_t = 13.3e-3;
T_bulkref = P_t/P_in;
abs_bulkref = 1.04;
R_bulk = AC_bulkref/(DC_bulkref*sqrt(T_bulkref)*P_in*abs_bulkref) = 0.852 cm/W

Using this new calibration factor and using :

P_t = 6.25 W;
P_in = 7.322 W;

the absorption of this sample seems to be around 60 ppm/ cm for all 3 maps (see attached figures)

Today we will double check the bulk calibration as the change was quite larger than expected.

Images attached to this comment
2492_20210513065451_xy.jpg 2492_20210513065457_yz.jpg 2492_20210513065525_xz.jpg 2492_20210513065737_screenshotfrom20210513135724.png
KAGRA MIR (Absorption)
Print this report.
MarcEisenmann - 11:05, Thursday 13 May 2021 (2491)Get code to link to this report
Comment to 3d absorption map of 1cm thick SHINKOSHA sample (Click here to view original report: 2476)

Abe, Marc

 

We modified the analysis to better estimate the mean and standard-deviation of absorption measurements.

The corrected results are attached to this entry.

Today I will remove the first contact that we applied on this sample and cross-checked if it affected the absorption measurement.

Images attached to this comment
2491_20210513040509_20210513shinkoshaxy.jpg 2491_20210513040515_20210513shinkoshayz.jpg 2491_20210513040520_20210513shinkoshaxz.jpg
R&D (FilterCavity)
Print this report.
NaokiAritomi - 23:15, Monday 10 May 2021 (2490)Get code to link to this report
CCFC preparation

[Aritomi, Yuhang]

First we measured nonlinear gain again. We measured BAB transmission from OPO. When we used 40mW green, we decreased CC1 gain from 2 to 1. 

green power (mW) BAB maximum (V) OPO temperature (kOhm) p pol PLL (MHz)
0 0.0552 7.194 305
40.9 1.02 7.194 195

The measured nonlinear gain is 18.5 with 40.9 mW green while the theoretical value should be g = 1/(1-sqrt(40.9/80))^2 = 12.3. We don't know why they are different.

We found an oscillation in FC lock with green injection of 27mW and FC gain of 1.3. The green injection power was larger so we decreased FC gain from 1.3 to 1.

We measured CCFC with 40.9 mW pump green and 50% pick off. Fig 1 shows the CCFC error signal with some demodulation phase and Fig 2 shows CCFC calibration signal when CCSB are off resonance and CC1 is scanned. The amplitude of CCFC calibration signal is 524mVpp which is 10 times larger than before.

We will try with 20% pick off and thorlabs BSS11 seems good for BS (20% reflection for p pol).

Images attached to this report
2490_20210510161453_img8905.jpg 2490_20210510161501_img8906.jpg
KAGRA MIR (Absorption)
Print this report.
MarcEisenmann - 20:14, Monday 10 May 2021 (2489)Get code to link to this report
KASI sample absorption measurement

The measurement performed during the golden week were using too low power (I_laser =3A ie P_laser~3.5W) that made it hard to really acquire signals.

The measurements have been restarted with higher power (I_laser=6A ie P_laser~8W?).

We'll have to check carefully the incident and transmitted power after the measurements.

Comments related to this report
MarcEisenmann - 13:57, Thursday 13 May 2021 (2492)

Abe, Marc

Following the 3 maps measurements we performed again the bulk calibration where this time we moved the imagining unit by 0.32 mm in order to compensate the thickness difference between surface and bulk reference samples. We got the following result (also see last figure of this entry) :

AC_bulkref = 0.0731;
DC_bulkref = 4.164;
ACDC = 0.01755;
P_in = 29.5e-3;
P_t = 13.3e-3;
T_bulkref = P_t/P_in;
abs_bulkref = 1.04;
R_bulk = AC_bulkref/(DC_bulkref*sqrt(T_bulkref)*P_in*abs_bulkref) = 0.852 cm/W

Using this new calibration factor and using :

P_t = 6.25 W;
P_in = 7.322 W;

the absorption of this sample seems to be around 60 ppm/ cm for all 3 maps (see attached figures)

Today we will double check the bulk calibration as the change was quite larger than expected.

MarcEisenmann - 16:12, Tuesday 18 May 2021 (2510)

It seems that the bulk calibration was overestimated. This is especially apparent when computing its transmission that was 45% instead of the expected 55%.

I performed again the bulk calibration and got :

AC_bulkref = 0.062;
DC_bulkref = 4.14;
P_in = 26.4e-3;
P_t = 13.1e-3;
T_bulkref = P_t/P_in
abs_bulkref = 1.04;
R_bulk = AC_bulkref/(DC_bulkref*sqrt(T_bulkref)*P_in*abs_bulkref) = 0.7743 W/cm

I used this new calibration to compute again the absorption map of the sample (see the 3 attached figures.

In the figure, the absorption is extracted from a fit using 2 normal distributions.

Here I also add the overall mean and standard deviation of each map (ie without any fittting) :

  XY YZ XZ
mean [ppm] 70 50 48
std [ppm] 8 35 37

 

KAGRA MIR (Absorption)
Print this report.
MarcEisenmann - 20:12, Monday 10 May 2021 (2488)Get code to link to this report
SHINKOSHA evaluation plate cleaning

Today I putted first contact on one side of the SHINKOSHA evaluation plate.

We'll remove and clean the other side tomorrow.

KAGRA MIR (Absorption)
Print this report.
MarcEisenmann - 14:52, Monday 10 May 2021 (2487)Get code to link to this report
Comment to 3d absorption map of 1cm thick SHINKOSHA sample (Click here to view original report: 2476)

Thank you.

I have updated the absorption value using the bulk calibration measured just after removing this sample (see elog 2480) and the corrected absorption is 302 +/- 108 ppm.

Images attached to this comment
2487_20210510075055_shinkopshaevaln1.jpg 2487_20210510075100_shinkopshaevaln1hist.jpg
R&D (FilterCavity)
Print this report.
MichaelPage - 19:27, Friday 07 May 2021 (2486)Get code to link to this report
OPO replacement - placement of modulators and beam profile

Yuhang and Michael

 

We finished setting up the modulators.

Set up and aligned FI: 5.15 mW transmission, 5.54 mW incident (92.96% tranmission vs Thorlabs 92% specification)

Set up HWP and PBS to check vertical polarisation required for EOM: 5.17 mW transmit, 0.03 mW reflect, 5.35 mW incident (PBS), 5.97 mW incident (HWP)

Set up EOM: 5.63 mW transmission, 6.04 incident. 

At this point the power meter was seen to be not particularly reliable for alignment, since it seems to register too much power at the edge of the detector compared to the center. The EOM was centered by using the IR detector card and seeing that the beam passes through the center. Then, we added AOM.

 

Afterwards, we measured the beam profile coming out of the modulators. In 2421, the beam profile was measured coming from the lens that was previously fixed on the bench, as indicated in that entry. From that measurement I used a beam waist of 150 µm with position z = 0 defined by the position of the last preset lens. However, that beam waist was an average of the horizontal and vertical beam waist measurements, which were quite different. But perhaps that could have also been due to the alignment of the beam on the beam profiler, which in 2425 was seen to be off centered in the measurements close to z = 0. Using Jammt with Thorlabs lens database, an f = 100 mm lens placed at z = 0.125 m gives a waist of 218 µm at z = 0.272 (figure 1). However, as mentioned in 2482, the lenses shown were slightly adjusted in order to properly level the beam at 76 mm height. So the prediction may not be completely accurate anymore.

The measured beam profile is shown in figure 2. The predicted beam seems to be a bit too far forward and too large.

Readjusting the mode matching telescope with a waist of 193 µm at 13.3 cm with respect to the f = 100 mm lens gives the result shown in figure 3. The spacing between the two modematching lenses (f = 40 mm and f = 75 mm), as well as the space from the f = 75 mm to the OPO waist should still be sufficient.

Images attached to this report
2486_20210507122538_prediction100.png 2486_20210507122544_atclaserfit0507.jpg 2486_20210507122551_atcmmt5.png
R&D (FilterCavity)
Print this report.
MichaelPage - 12:17, Friday 07 May 2021 (2482)Get code to link to this report
OPO replacement - ATC setup

Yuhang and Michael

We just did some adjustments to the beam using the lenses shown, to make sure the beam is travelling at about 76mm height along the screw hole line. It rises by about 1mm over ~50cm of travel. The mounts have also been adjusted to match the beam height.

This work will be on hold while the KAGRA ETMY is being cleaned.

Images attached to this report
2482_20210507051606_b4ddc66fd7b146419764fc5187f2eb25.jpeg
KAGRA MIR (Absorption)
Print this report.
SimonZeidler - 19:12, Sunday 02 May 2021 (2481)Get code to link to this report
Comment to 3d absorption map of 1cm thick SHINKOSHA sample (Click here to view original report: 2476)

Nice results!

It seems that Shinkosha still has its main feature: the prominent spider-web absorption.

But the absorption is really high this time.

KAGRA MIR (Absorption)
Print this report.
MarcEisenmann - 13:03, Saturday 01 May 2021 (2480)Get code to link to this report
bulk calibration and tama size KASI sample measurement started

Abe, Aritomi, Marc

Yesterday we removed the shinkosha evaluation plate and its holder from the translation stage.

We installed back the bulk reference sample and got R_bulk ~0.773 W/cm with

AC_bulkref = 0.0555 V;
DC_bulkref = 4.103 V;
P_in = 24.2e-3 W;
P_t = 11.7e-3 W;
abs_bulkref = 1.04;

Then, we brought the KASI sample inside the clean room and cleaned it with alcool while checking with the green light.

With this sample, the IU is now at z=43.1mm and the mirror centers are :

X_center = 327.5 mm

Y_center = 122.95 mm

Z_center = 62.5 mm (we had to increase the laser power to more than 1 W (~3W) to clearly see the phase change at the sample surfaces)

We added new X limit (lower vertical) on Zaber and started a circular map at Z_center with 2 cm radius.

Our initial guess of the absorption is abs~78 ppm/cm given by

AC = 2.3e-4 V;
DC = 4.242 V;
P_in = 3.24 W;
P_t = 2.77 W;
T_sample = P_t/P_in;
R_bulk = 0.7730 W/cm;
mat_correction=3.34;

abs = AC/(DC*P_in*sqrt(T_sample)*R_bulk)*1e6*mat_correction

R&D (FilterCavity)
Print this report.
MichaelPage - 23:29, Friday 30 April 2021 (2479)Get code to link to this report
OPO replacement - optical setup revisited

Yuhang and Michael

We did some more inspection of the OPO replacement setup at ATC. Perhaps we will need some more low f lenses.

I recalculated the error signal and mode matching shown in 2469.

We can obtain a much larger error signal in reflection when the meniscus is used as the input mirror. Even with <10 mW incident power it should not be a problem. Here I use 4 mW incidence, 40 MHz modulation and 0.3 modulation depth to achieve ~160 mVpk error signal in the linear range. Figure 1 shows the error signals and power outside of the cavity. Figure 2 shows the scan of demodulation phase versus reflection and transmission photodetector signals. Figure 3 shows the circulating power of 160 mW, for 4 mW input.

I also show a beam profile with a more complete mode matching telescope into the OPO. I did a mode matching telescope calculation using JamMT and the database of Thorlabs lenses. We wish to obtain a beam waist of 20.66 µm within the OPO cavity using two lenses. Judging by the OPO setup in TAMA, it would be good to leave approximately 15cm OPO to lens and lens to lens.

In 2469 I used an f = 100mm lens placed at 175mm from the last preset optic on the ATC bench. The beam waist of 150µm is located within 1cm of the last preset optic on the ATC bench This creates another 150µm beam waist 175mm from the lens. The rationale for doing this was:

i) to have a couple of steering mirrors before this lens, so that the beam would be level at 76mm height when going through the FI/EOM/AOM. However, Yuhang suggested to simply move the 100mm lens to achieve beam levelling.

ii) to reduce the size of the beam waist, and make the beam less than 1/5th of the EOM aperture size for a reasonable distance. 

This time I used an f = 100mm lens placed 125mm from the last preset optic (as per photograph and measurement in 2439). The mode matching telescope, calculated by JamMT, is shown in figure 4. We should have enough space for the modulators between the f = 100mm and f = 40mm (90mm FI, HWP, 56mm EOM, 22mm AOM). Steering mirrors can fit in the space between the f = 40mm and f = 75mm, and then there should be a beam splitter between the f = 75mm and the OPO cavity.

Images attached to this report
2479_20210430162935_errorsignal40403.jpg 2479_20210430162940_demodulationphase40403.jpg 2479_20210430162944_intracavity.jpg 2479_20210430162955_atcmmt4withmodlens.png
KAGRA MIR (Absorption)
Print this report.
MarcEisenmann - 17:34, Wednesday 28 April 2021 (2478)Get code to link to this report
Comment to 3d absorption map of 1cm thick SHINKOSHA sample (Click here to view original report: 2476)

Marc, Matteo, Yuhang

We used the green light to check possible dust on the mirror and took a picture after already using the air dust once.

The mirror is really dirty...

We then used again the air dust and could remove most of the large dust particles but not the smallest one (not visible on picture so no 'after picture').

We can still see the ethanol/acetone trace at the mirror edge so the mirror got dirty either during shipping or during the few weeks it stayed inside the clean room without cover.

We will clean it with first contact after the golden week.

Images attached to this comment
2478_20210428103424_img20210428163900.jpg
KAGRA MIR (Absorption)
Print this report.
MarcEisenmann - 17:30, Wednesday 28 April 2021 (2477)Get code to link to this report
Comment to 1cm thick SHINKOSHA absorption measurement started (Click here to view original report: 2474)

Before the measurement we also moved the IU to z=68.6mm.

I'll add to the wiki the IU position for surface and bulk reference as well as TAMA size, 1cm shinkosha and KAGRA samples.

KAGRA MIR (Absorption)
Print this report.
MarcEisenmann - 10:28, Wednesday 28 April 2021 (2476)Get code to link to this report
3d absorption map of 1cm thick SHINKOSHA sample

Abe, Marc, Matteo

Following the measurement of the absorption map at z=51.4 mm, we performed 2 others map in the XZ and YZ planes.

They correspond respectively to figure 1 and 2.

An important point to notice is that there seems to be a point absorber on the surface that distorts the absorption scale (up to 4000 ).

We decided to turn off the ir laser just in case.

Anyway looking at the histogram, it seems that the nominal absorption is around 400 ppm/cm (more precise analysis to come).

In figure 3 you can see a combination of the 3 maps where the colorbar scale has been limited to the maximum of the circular map (ie 1500 ppm/cm).

Images attached to this report
2476_20210428032407_202104271cmshinkoshaxz.jpg 2476_20210428032417_202104271cmshinkoshayz.jpg 2476_20210428032425_202104271cmshinkoshaoverview.jpg
Comments related to this report
MarcEisenmann - 17:34, Wednesday 28 April 2021 (2478)

Marc, Matteo, Yuhang

We used the green light to check possible dust on the mirror and took a picture after already using the air dust once.

The mirror is really dirty...

We then used again the air dust and could remove most of the large dust particles but not the smallest one (not visible on picture so no 'after picture').

We can still see the ethanol/acetone trace at the mirror edge so the mirror got dirty either during shipping or during the few weeks it stayed inside the clean room without cover.

We will clean it with first contact after the golden week.

SimonZeidler - 19:12, Sunday 02 May 2021 (2481)

Nice results!

It seems that Shinkosha still has its main feature: the prominent spider-web absorption.

But the absorption is really high this time.

MarcEisenmann - 14:52, Monday 10 May 2021 (2487)

Thank you.

I have updated the absorption value using the bulk calibration measured just after removing this sample (see elog 2480) and the corrected absorption is 302 +/- 108 ppm.

MarcEisenmann - 11:05, Thursday 13 May 2021 (2491)

Abe, Marc

 

We modified the analysis to better estimate the mean and standard-deviation of absorption measurements.

The corrected results are attached to this entry.

Today I will remove the first contact that we applied on this sample and cross-checked if it affected the absorption measurement.

SimonZeidler - 16:22, Friday 14 May 2021 (2495)

I have some questions.

What is the meaning of the fit in the histograms? Do you have a reason to fit with specific distributions/densities or do you want to find a systematic pattern?

MarcEisenmann - 10:27, Monday 17 May 2021 (2504)

The absorption distribution is fitted with 2 normal distributions.

I thought it could be useful for the case of XZ and YZ maps (where there are measurement points outside the sample) because it allows to remove the effects of absorption outside the sample and point defects/dust on the surface.

But I agree that it might not be the most suitable distribution, especially for the shinkosha samples...

KAGRA MIR (Absorption)
Print this report.
MarcEisenmann - 12:48, Tuesday 27 April 2021 (2475)Get code to link to this report
1 cm shinkosha absorption map

The absorption map and absorption distribution of the 1 cm thick shinkosha sample is attached to this entry.

It has a mean absorption of 355 +/- 127 ppm/cm

Images attached to this report
2475_20210427054737_202104271cmshinkosha.jpg
KAGRA MIR (Absorption)
Print this report.
MarcEisenmann - 17:01, Monday 26 April 2021 (2474)Get code to link to this report
1cm thick SHINKOSHA absorption measurement started

Abe, Marc, Matteo

This morning, we installed the 1cm thick SHINKOSHA sample.

We recentered the beam dump on the laser unit and covered back this unit.

Then, we increased the power to ~300 mW to check the sample center and inclinaison.

We checked its X,Y and Z centers :

X center = 398.195 mm

Y center = 121.5 mm

Z center = 51.4 mm

Then, we checked the z center at the left, right, bottom and top position of the sample with extremal positions separated by 13 cm.

We found a 0.8 mm vertical tilt of the mirror and a 0.4 mm horizontal tilt.

Finally, we increased the laser power to 3.46W and started the measurement of absorption map of this sample.

We also did a preliminary evaluation of the absorption as

AC = 3.6e-3 V;
DC = 4.71 V;
P_in = 3.46 W;
P_t = 2.94 W;
T_sample = P_t/P_in;
R_bulk =0.7414 W/cm;
mat_correction=3.34;

abs = AC/(DC*P_in*sqrt(T_sample)*R_bulk)*1e6*mat_correction ~1080 ppm/cm
 

Comments related to this report
MarcEisenmann - 17:30, Wednesday 28 April 2021 (2477)

Before the measurement we also moved the IU to z=68.6mm.

I'll add to the wiki the IU position for surface and bulk reference as well as TAMA size, 1cm shinkosha and KAGRA samples.