LOG-IN
Displaying reports 1501-1520 of 3275.Go to page Start 72 73 74 75 76 77 78 79 80 End
KAGRA MIR (Polarization)
Print this report.
PengboLi - 22:16, Sunday 22 March 2020 (2083)Get code to link to this report
Birefringence measurement results on Shinkosha S5 substrate

Simon, Pengbo

we finished the birefringence characterization of the shinkosha s5, as can be seen from the attachment(the first four figures are the result this time , the last two figures  are the results last year.), it seems the offset is even larger than before.

Images attached to this report
2083_20200322140750_figure1.png 2083_20200322140759_figure3.png 2083_20200322140808_figure2.png 2083_20200322140818_figure4.png 2083_20200322140854_figures1.png 2083_20200322140906_figures2.png
KAGRA MIR (Polarization)
Print this report.
PengboLi - 21:12, Thursday 19 March 2020 (2082)Get code to link to this report
Birefringence measurement results on Shinkosha S1 substrate

Simon, Pengbo

Today we finished the charaterization of the birefringence properties on Shinkosha S1 substrate. For the s-pol, we got 87.77 +/- 0.25 deg, which is slightly better than last year's result, 87.1+/- 0.05 deg. For the p-pol, we got 1.93+/- 0.68 deg, because we didn't do this measurment last year, there are no comparision. From the maps, it seems this time the distribution is more homogenous.

Images attached to this report
2082_20200319122548_figure1.png 2082_20200319122556_figure2.png 2082_20200319122606_figure3.png 2082_20200319122614_figure4.png 2082_20200319122629_figure21.png 2082_20200319122636_figure22.png
KAGRA MIR (Absorption)
Print this report.
SimonZeidler - 00:10, Wednesday 18 March 2020 (2081)Get code to link to this report
Comment to Absorption measurements on S1 (Click here to view original report: 2078)

Here the absorption map of the second surface.

Images attached to this comment
2081_20200317161024_map20200317xysurf2.png
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 21:48, Tuesday 17 March 2020 (2079)Get code to link to this report
Comment to Viewport replacement (Click here to view original report: 2075)

The cryostat was pumped down againg and the pressure became less than 10-3 Pa.
This level is enough for the measurement at room temperature.
So I vented the chamber and opened in order to do the alignment of the cavity.

R&D (Cryogenic)
Print this report.
SatoshiTanioka - 19:40, Tuesday 17 March 2020 (2077)Get code to link to this report
Positioning of input optics

Background

Since I modified the mode matching lenses, I had to also modify the beam path.
Furthermore, in order to avoid putting the mode matching lenses between the steering mirrors, the layout should be reconsidered.

What I Did

I determied positions of input optics around the viewport.
They are a littele bit complicated as shown in the attached picture due to the limited space.
Then I aligned the mirrors in TEM00 path.

Next Step

  • Alignment of the TEM00 path.
  • Alignment of the input and output fused silica mirrors.
  • Need to purchase some mirror holders and pedestals for the HOM's paths (next FY).
Images attached to this report
2077_20200317105101_20200317.jpg
KAGRA MIR (Absorption)
Print this report.
SimonZeidler - 18:32, Tuesday 17 March 2020 (2078)Get code to link to this report
Absorption measurements on S1

Pengbo, Simon

Yesterday and today we repeated basically the measurements we did on S5 for S1. The results can be seen in the attachement.

Also for this one, it became obvious that there is a large absorption excess on the surface, This time, however, it seems that this excess is existent on both surfaces. Therefore, I had run the surface-absorption measurements on both (as the time I write this, the second one did not yet finish).
The center-map shows a distribution which is somewhat comparable to the measurements we did last year. However, the mean absorption coefficient has been increased by a factor of ~2.

The next step will be to do the birefringence measurements on both samples.

Images attached to this report
2078_20200317102534_map20200316xy.png 2078_20200317102550_dist20200316xy.png 2078_20200317102555_map20200317yz.png 2078_20200317103130_map20200317xysurf1.png
Comments related to this report
SimonZeidler - 00:10, Wednesday 18 March 2020 (2081)

Here the absorption map of the second surface.

R&D (Cryogenic)
Print this report.
SatoshiTanioka - 22:59, Monday 16 March 2020 (2075)Get code to link to this report
Viewport replacement

Preface

As the viewport attached the cryostat was not AR coated at the wavelength of 1550 nm but 1064 nm, there was a 10% loss for beam power at 1550 nm.
So I decided to replace the viewport for 1550 nm AR coated.
The cost, however, was a little bit expensive.
Therefore, I bought a AR coated window and a viewport without a window separately and they were supposed to be assembled.

What I did

I assembled the viewport as attached pic. 1.
Then I connected to the conversion flange though I was not confident to use M8 bolts and washers stored in the ATC.
After that I installed to the chamber and vented as a test.
Before venting, I measured the beam power before the viewport and after.

Result

First, the transmitted beam power was 2.73 mW against 2.74 mW input.
So the total loss was about 0.5% which is dramatically reduced compared to the previous one.
Moreover, the viewport does not have any serious leakages and the pressure reached less than 10-2 Pa.

Images attached to this report
2075_20200316123903_viewport1.jpg 2075_20200316123907_viewport2.jpeg
Comments related to this report
SatoshiTanioka - 21:48, Tuesday 17 March 2020 (2079)

The cryostat was pumped down againg and the pressure became less than 10-3 Pa.
This level is enough for the measurement at room temperature.
So I vented the chamber and opened in order to do the alignment of the cavity.

R&D (Cryogenic)
Print this report.
SatoshiTanioka - 22:58, Monday 16 March 2020 (2076)Get code to link to this report
Work on input optics

What I Did

  • Tweaking the lens and mirror position in a double-pass AOM path
  • Mode matching for TEM00 mode

Details

As I replaced a lens in double-pass AOM path, I tweaked the position of the lens and the mirror.
Current lens is f=70mm one.

Since the transmitted flash could not see due to the poor mode matching, I tried to improve the mode matching.
The lenses were replaced and f=-75mm and f=300mm ones are put now.
Their positions are about 0.226m and 0.659m from the beam waist, respectively.
After that I could get beam size about 65um at its waist where it is reasonable position for the apex mirror of folded cavity.

Next Step

  • Install some mirrors to inject the TEM00 beam into the chamber.
  • Alignment of the input and output mirrors.
  • Tweaking.
KAGRA MIR (Absorption)
Print this report.
SimonZeidler - 15:42, Monday 16 March 2020 (2074)Get code to link to this report
Absorption results on annealed S5

Pengbo, Simon

On the weekend, Pengbo finished the measurements on the annealed S5 sample.
Already last week, we recognized by doing the Z-scans that a large absorption excess must exist on one side of the sample (actually the side which has NO damages). So we decided to take actually 3 maps: in the center, on the suspicious surface and along the Z-axis. The results can be seen in the attachement.

As can be clearly seen, there is a very prominent structure on the suspicious surface which leads to an excess in the absoption data. This structure is partly visible also in the center-map, probably due to interference effects coming from the absorption excess (at least judging from the phase map). Please note that the sample for the center-map has been flipped so that the suspicious surface is on the out-going side because we had large problems in obtaining a meanigful result when the beam got influenced by the excess before reaching the targeted position.

As can be also seen from the map along the Z-axis, this excess is apparently not limited to the single surface area but has a depth which is hard to quantify (given the strong disturbance in phase and AC) but we estimate the affected depth to be several millimeter.

As a reference, the histogram for the absorption coefficient taken at the center is also given. The actual mean-value is not so far away from the measurements last year (however biased by the mentioned structure), which is at least one good news.

Images attached to this report
2074_20200316073857_20200314.png 2074_20200316073902_20200315.png 2074_20200316073906_20200316.png 2074_20200316073912_map20200314pengbo.png 2074_20200316073917_map20200315pengbo.png 2074_20200316073921_map20200316pengbo.png 2074_20200316073934_dist20200314pengbo.png
KAGRA MIR (Absorption)
Print this report.
SimonZeidler - 09:56, Thursday 12 March 2020 (2073)Get code to link to this report
Measuring absorption on annealed S1 and S5

Pengbo, Simon

We received the samples S1 and S5 back from France after their annealing.
Before we started measuring the absorption coefficient, we inspected the samples and we discovered some damages on the edges of both samples. Interestingly, the damages are only on one side (see attached photos).

The measuremetns started with S5 and are ongoing. However, before setting the sample into the sample holder, we cleaned it with FC on both sides.

Images attached to this report
2073_20200312015508_43.jpg 2073_20200312015515_24.jpg 2073_20200312015524_35.jpg 2073_20200312015528_15.jpg 2073_20200312015535_16.jpg
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 23:26, Wednesday 11 March 2020 (2072)Get code to link to this report
Mode Matching

I tried to see the transmitted flash but I could not on Monday.
Indeed, the beam size was at its waist was much larger than the design value i.e., 100 um though 50 um is required.

I decided to change the mode matching lenses in order to achieve the designed spatial mode.
To be honest, the obtained minimun size was about 80 um radius.

I gonna try improving the mode matching until end of this week.

Note

The lenses are biconvex or biconcave in order to reduce the aberration of reflected beam.

R&D (FilterCavity)
Print this report.
YuhangZhao - 18:29, Wednesday 11 March 2020 (2071)Get code to link to this report
Measurement of QPD shot noise from filter cavity reflection with filter cavity locked

We performed the measurement which is close to the real case when we use QPD. The difference is only light is adjusted into only one segment of QPD.

We could see the modulation frequency we are interested in can be seen.

Images attached to this report
2071_20200611123305_figure3.png
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 22:27, Monday 09 March 2020 (2070)Get code to link to this report
Alignment work and PDH servo modification

I tried to see the transmitted flash with scanning the laser frequency, but I could see nothing.
I decided to remove the input and output mirrors and do the alignment work again.

I also modified the PDH servo.
The schematic of the servo will be uploaded on wiki.

R&D (FilterCavity)
Print this report.
YuhangZhao - 16:01, Monday 09 March 2020 (2069)Get code to link to this report
Measurement of FDS after new FI

Pengbo and Yuhang

We performed the measurement although the mirror is not very stable. As you can see from the attached figure 2. This makes the low frequency measurement of shot noise(with the corruption of backscattering) much worse.

The measurement of squeezing level is only 2.5dB.

Images attached to this report
2069_20200309080127_figure11.png 2069_20200309080319_wechatimg641.jpeg
R&D (FilterCavity)
Print this report.
YuhangZhao - 15:10, Monday 09 March 2020 (2068)Get code to link to this report
Measurement of QPD shot noise with different beam size and different power(second measurement)

We measured again with 5mW green power, but this time we make the beam smaller to see if it will decrase or not.

The result is that shot noise will decrease if the power density is higher than 47mW/mm2.

There is a discrepancy between this measurement and the one measured last time(entry 2067). I think the reason is that, the beam size is highly related to position. And this time we removed QPD and we may put it back to a slightly different position. Then it leads to this discrepancy.

We need to use a more robust telescope if we want to have a more precise measurement.

Images attached to this report
2068_20200309071659_figure1.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 21:14, Thursday 05 March 2020 (2067)Get code to link to this report
Measurement of QPD shot noise with different beam size and different power

Pengbo and Yuhang

We measured shot noise spectrum with 5, 15, 26mW seperately. For each power, we measured also with different beam size ranging from to.

The set-up is using the second segement of QPD2. We checked DC voltage with oscillscope. We also checked RF channel after an amplification of 32dB with spectrum analyzer operates in 1MHz RBW(this time we average for 10 times so that the noise vurve is smooth). The noise spectrum of RF signal is plot for each case.

1. 5mW case(attached as the first picture): the noise floor is the same for all the beam size.

2. 15mW case(attached as the second picture): the noise floor reaches maximum when the power density is below ~25 mW/mm2

3. 26mW case(attached as the third picture): the noise floor reaches maximum when the power density is below ~22mW/mm2

We didn't measure the changing point for 5mW, however, from the beam density we measred, all the measurement we did for 5mW has low power density relative to the threshold (roughly between 22 and 25mW/mm2). We should see the noise floor decrease when the beam size is smaller than 500um.

Conclusion: The QPD response will saturate and decrease if the power density exceeds around 23mW/mm2.

(Notice: the beam size in this entry is diameter)

Images attached to this report
2067_20200305131521_5mw.png 2067_20200305131526_15mw.png 2067_20200305131531_26mw.png
KAGRA MIR (Absorption)
Print this report.
PengboLi - 14:02, Thursday 05 March 2020 (2066)Get code to link to this report
Absorption results on KAGRA-size Shinkosha#7 after setup upgrade

Simon, Pengbo
 

Attach to this report I show the result on shinkosha #7 sample with different polarization input beam(S-pol and P-pol).

We can see a 1 percent difference from these two maps, which is even smaller than the difference for TAMA-size sample.

Images attached to this report
2066_20200305060051_kagras1.png 2066_20200305060054_kagrap1.png 2066_20200305060059_kagras2.png 2066_20200305060102_kagrap2.png
KAGRA MIR (Absorption)
Print this report.
PengboLi - 18:42, Wednesday 04 March 2020 (2065)Get code to link to this report
Absorption measurement on TAMA#1 and Shinkosha #7

Simon, Pengbo

After the birefringence measurement, we change to the absorption system with a controllable polarization of the laser. First, we did two XY-plane absorption measurements on TAMA#1 with different polarization, which is P-pol and S-pol. Then we choose a small area of the mirror and did another measurement under the S-pol incident beam. We can see a very clear structure in the map. Then we did another YZ-plane absorption measurement with an S-pol incident beam. The distribution is quite homogenous.

We change the sample to Shinkosha #7, then follow what we did before, checking whether the polarization might have some influence on the absorption. We already have the result of the absorption map with the S-pol incident beam. The result is almost the same compare with the former result.

Images attached to this report
2065_20200304093548_tamap1.png 2065_20200304093602_tamas1.png 2065_20200304093606_tamal1.png 2065_20200304093612_tamap2.png 2065_20200304093617_tamas2.png 2065_20200304093621_tamal2.png 2065_20200304103301_tamayz.png 2065_20200304103308_figure1.png 2065_20200304103312_figure2.png
R&D (FilterCavity)
Print this report.
YuefanGuo - 05:56, Wednesday 04 March 2020 (2064)Get code to link to this report
Measurement of the QPD did in Virgo

We did the same measurement as in logbook entry 1875, the QPD response to different size of the green beam

The measurement was done by using the green beam reflected by the green mode cleaner of the Virgo squeezer(1500W). 

First, we measured the green beam size without any lenses and saved the data file from the beam profiler.  The beam profiler has 1928*1448 pixels corresponding to active are of 7.1mm*5.3mm, we got power value at each data point, and we did a 2-d Gaussian fit to these data points. One of the fit shows in figure 1. From the beam plotted in this figure, it is clear that the beam has astigmatism, so the final beam waist size and position are quite different in two axes. Below is the number (position zero is just a random point we chose easy for the measurement) 

 

Beam waist size

Beam waist position

X

242um

0.0924m

Y

367.6um

0.6262m

By checking the 2d plot of the raw data, we found out the beam profiler is saturated (fig 2 shows the top cut shape). But in the 2d fit, we were not able to remove these points by substituting them into 'NaN' while using 'lsqcurvefit' function in Matlab, because this function needs 'double' format input.  Then to check the quality of the fit, we plot the difference between the fit the original data, result shows in figure 3. It seems the fit is fine. Since the data is very noisy, I was guessing maybe we actually got the peak of the Gaussian, and the saturation part is just the noise. 

Anyway, after measuring the beam size, we put a 50mm lens and measured the beam with the QPD in different positions. We did two groups of measurements with different power by changing the density. the results show below.

Beam size

Group 1

Group 2

X direction (um)

Y direction(um)

DC (V)

RBW/VBW/MHz (dBm)

DC (V)

RBW/VBW/MHz (dBm)

35

23

1.35

-150.5

2.344

-150.6

148

256

1.35

-148.4

2.46

-148.7

277

496

1.309

-146.9

2.401

—145.6

407

738

1.307

-146.5

2.390

-144.8

538

979

1.307

-146.7

2.383

-144.7

668

1220

1.3

-146.6

2.360

-144.6

Images attached to this report
2064_20200303215458_2dfit30cm.jpg 2064_20200303215513_data2dplot.jpg 2064_20200303215541_resfitdata.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 16:54, Tuesday 03 March 2020 (2063)Get code to link to this report
Measurement of filter cavity reflection beam height

Eleonora and Yuhang

We measured the filter cavity reflection beam height. It is 74.5mm, which is 1mm lower than injection beam measured three months ago.

Images attached to this report
2063_20200303085325_wechatimg639.jpeg