LOG-IN
Displaying reports 1781-1800 of 3275.Go to page Start 86 87 88 89 90 91 92 93 94 End
R&D (FilterCavity)
Print this report.
NaokiAritomi - 23:04, Thursday 24 October 2019 (1769)Get code to link to this report
IR locking accuracy

[Aritomi, Yuhang, Yaochin, Eleonora]

We measured IR error signal of filter cavity using BAB with TAMA PD which is same as CC1 PD in the reflection path from filter cavity. Pic. 1 shows IR error signal and IR transmission when filter cavity is scanned with AOM (setting is Pic. 2). Calculation method of calibration factor is same as entry 750.

Calibration:
peak-peak voltage of error signal: 26.2 mV
peak-peak time difference of error signal: 98 ms
slope = peak-peak voltage of error signal/peak-peak time difference of error signal = 26.2 mV/98 ms = 0.267 V/s
AOM scan frequency: 500 mHz
AOM deviation: 1 kHz
AOM scan speed for green: AOM deviation/AOM scan frequency = 2 kHz/s
AOM scan speed for IR: 2kHz/s /2 = 1 kHz/s
 
Calibration factor:
AOM scan speed for IR (Hz/s)/(2*slope(V/s)) = 1000(Hz/s)/(2*0.267(V/s)) = 1873 Hz/V

Note that factor of 2 in calibration factor is because slope of error signal on resonance is as twice as slope of peak-peak of error signal.

Then we measured spectrum of IR error signal when IR is locked on resonance and calculated rms (Pic.3). IR locking accuracy is 4.2 Hz which corresponds to 4.5 pm of locking accuracy. Pic. 4 shows squeezing degradation with this locking accuracy. In addition to mode mismatch, locking accuracy is also dominating. We need IR locking.

Images attached to this report
1769_20191024160409_error.jpg 1769_20191024160414_aom.jpg 1769_20210724223331_lockingaccuracy20191024.png 1769_20210724223338_degradation20191024.png
Comments related to this report
EleonoraCapocasa - 13:21, Sunday 27 October 2019 (1771)

Most of the RMS is accumualted below 10 Hz. We observed a correlation between the PDH error signal for IR (demodualtion of 15 Mhz sidebands) and the lock correction signal from RAMPEAUTO.

At the time of the measument we know the IR alignment was not good. This can be the cause of such behaviour.  We should repeat this measurement in good alignment condition.

R&D (Cryogenic)
Print this report.
SatoshiTanioka - 18:46, Thursday 24 October 2019 (1768)Get code to link to this report
Work on Cryostat

[Namai-san, Ueda-san from KEK, Sato-san, and Tomaru-san]

We worked on the cryostat in ATC to install adapters for windows on 80K shield.
What we did is as follows:

  1. Attached the adapters on 80K shield
  2. Adjusted the insulator on 80K shield not to touch the 300K shield
  3. Removed 300K shield for temporary and install 80K shield
  4. Re-installed 300K shield
  5. Re-located the chamber to install the cold head and bellows
  6. Installed an optical breadboard and connected to cold head with grease

Since we don't have indium sheet, we used grease to contact with cold head.
Though this grease can be used in vacuum (~10^-7 Pa), if the contamination on mirrors become a problem, we will remove the grease and use indium sheet.

The remaining tasks are:

  • Install modified 4K shield, and windows on 80K shield
  • Cabling
  • Vacuum check
Images attached to this report
1768_20191024114542_20191024adapter.jpg 1768_20191024114547_20191024bar.jpg
Comments related to this report
SatoshiTanioka - 18:12, Monday 28 October 2019 (1775)

Today I temporary installed the adapter flange and 4K shiled inside the cryostat chamber to confirm the screw hole position is correct.
The adapter flange and shield seemed to be O.K., but I need 10 more ultrasonic cleaned M6 screws.

In addition to that, we need super insulator on 4K shield and tape to fix cables.

The procudure for remaining tasks are:

  1. Prepare windows for 80K shield
  2. Install windows on 80K shield
  3. Cabling on 4K shield
  4. Attach SI on 4K shield
  5. Install 4K shield
SatoshiTanioka - 14:08, Wednesday 30 October 2019 (1779)

This entry is log on 28th Oct.
I cut some teflon sheet for cushioning between glass window and adapter.

R&D (FilterCavity)
Print this report.
NaokiAritomi - 16:36, Thursday 24 October 2019 (1767)Get code to link to this report
14MHz signal at homodyne

I checked homodyne RF signal if I can get 14MHz CCSB beat note, but I couldn't find 14MHz signal in homodyne RF. 14MHz signal may be filtered out by low pass filter after demodulation inside homodyne. We have to modify homodyne if we want to detect 14MHz at homodyne...

R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 16:29, Thursday 24 October 2019 (1766)Get code to link to this report
Squeezing degradation budget

[Aritomi,Yuhang,Yao-chin, Eleonora]

The attached plot shows a very tentative degradation budget for the squeezing injection into FC detuned at 50 kHz (entry #1751)

PRODUCED SQUEEZING: 16 dB  (from anti squeezing measurement) (entry #1587)

TOTAL INJECTION LOSS = 33 %

- from inside OPO to 1st PBS = 18% (loss of frequency independent squeezing: 21% - visibility: 2% - quantum efficiency: 1%)

- after 1st PBS before PR chamber = 10%

- from PR viewport up to FC  = 5 %

TOTAL READOUT LOSS = 12 %

- from FC to PR viewport: 10%

- after viewport to Homodyne: 1%

- Homodyne Q. E. = 1%

MISMATCH from SQZ to FC = 10 %

MISMATCH from SQZ to HOM = 5 %

FILTER CAVITY LOSSES : 100 ppm

PHASE NOISE = 100 mrad (very tentative, when CC2 doesn't unlock)

LOCK ACCURACY = 5 pm (very tentative, to be measured)

Conclusion: it seems that the degradation is dominated by mismatching 

Images attached to this report
1766_20191024092915_sqzdeg2410.png 1766_20191024094610_175320191018072640lossbudget.png
R&D (FilterCavity)
Print this report.
RyutaroTakahashi - 17:44, Wednesday 23 October 2019 (1764)Get code to link to this report
Comment to BS TMP trouble (Click here to view original report: 1761)
I baked out the TMP for only 20 min today. It was hard to keep the temperature of heater manually. Thermostat is necessary to control the heater automatically.
 
In the other hand, the TMP was working through two days. The outgassing rate should be decreased even though the baking was not implemented. The penning gauge shows 1.9~2.0 x10-6 Torr in each case;  open/close the gate valve. It's strange. The gauge would not be working.
R&D (FilterCavity)
Print this report.
NaokiAritomi - 16:34, Wednesday 23 October 2019 (1763)Get code to link to this report
CCSB error signal for filter cavity locking (CCFC)

[Aritomi, Yaochin]

We put TAMA PD in the reflection path from filter cavity and detected CCSB 14MHz beat note and demodulated it with TAMA demodulator. Filter cavity is locked with green and CCSB are off resonance. For demodulation LO, we used CC1 demodulation LO. Then we scanned green phase going to OPO which scans relative phase of CCSB. This relative phase scan of CCSB emulates CCFC error signal.

Note that we only have 90:10 RF BS and when we used this RF BS to divide CC1 LO and used 10% of CC1 LO for this filter cavity error signal, there was no error signal. We'll buy 50:50 RF BS.

Pic.1: CCFC 14MHz signal 

Pic. 2: CCFC 14MHz signal after RF amplification (34dB)

Pic. 3: CCFC error signal for filter cavity when CCSB are off resonant and green phase is scanned

Images attached to this report
1763_20191023093239_ccrf.jpg 1763_20191023093247_ccrfamp.jpg 1763_20191023093259_img7941.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 16:11, Wednesday 23 October 2019 (1762)Get code to link to this report
Nikhef quadrant characterization

[Yuhang, Yaochin, Aritomi, Eleonora]

We wanted to check the gain of the quadrant PD by shining some light on it and look at the response of the RF channel with a spectrum analyzer. We assumed that amplitude fluctuations at that high frequencies are shot noise dominated and thus have a flat spectrum. It this way it would be possible to measure the gain TF of the PD and indentify its resonances.

We sent to one sector of the quadrant about 1.5 mW of green light in reflection from FC and we amplified the RF signal of 32dB. We couldnt' see any difference in the spectrum with respect to the case when we don't send any light, except for the light at the modulation frequency that we could not switch off (15 MHz, needed for SHG).

We decided to take the same measuremnt with a red laser diode, in the same configuration for differen power value. We found that the RF spectrum level increases with the power showing a peculiar shape.

We observed that the RF signal stop increasing already at about ~250uW, while DC seems linear (see first plot) 

Not that the responsivity of the sensor is similar for green and red (0.2 [A/W] wrt 0.25 [A/W])

In order to confirm that the laser diode has not any strange amplitude feature at high frequency, we tried it on a QUBIG PD and found a reasonable RF spectrum with a cut off at 100 MHz ( as reported by the data sheet)

Conclusion: something is wrong. The quadrant seems to repond too differently to green and red light. But the gain doesn't show any resonace. We will investigate more.

Images attached to this report
1762_20191023091057_dccal.png 1762_20191023092005_nikhefpd.png 1762_20191023092513_qubigpd.png
Non-image files attached to this report
R&D (FilterCavity)
Print this report.
RyutaroTakahashi - 09:41, Wednesday 23 October 2019 (1761)Get code to link to this report
BS TMP trouble

(Log on 21st)

As reported in the Elog, the BS TMP was failed on 18th. The pump was restarted, but it has problems.

  1. Started the RP in the back side of TMP.
  2. Opened the angle valve of RP.
  3. Reset the TMP and started it.
  4. The pressure in BS chamber was 1.8x10-6 Torr.
  5. After the rotation speed of TMP reached 600 Hz, opened the gate valve for BS chamber.
  6. The pressure was increased to 2.0x10-6 Torr. This means that the outgassing rate of TMP itself is too large.
  7. Closed the gate valve and the angle valve.
  8. Stoped the TMP and the RP.
  9. Checked the RP. The inline filter between the TMP and the RP was filled by oile. The vapor of oile might contaminate the TMP.
  10. Replaced the RP system (RP itself, inline filter, bellows, angle valve) to the other from the EW arm.
  11. To declease the outgassing rate of TMP, a baking is necessary. Binded the TMP with the ribbon heater and attached the thermometer.
  12. Restarted the TMP and the RP closing the gate valve.
  13. The temparature must be lower than 120 oC. The heater was turned on/off manually around 110oC for 30 min.
  14. According to the manual of TMP, 4 hours are necessary for the baking. Will continue the baking on 23rd.
Comments related to this report
RyutaroTakahashi - 17:44, Wednesday 23 October 2019 (1764)
I baked out the TMP for only 20 min today. It was hard to keep the temperature of heater manually. Thermostat is necessary to control the heater automatically.
 
In the other hand, the TMP was working through two days. The outgassing rate should be decreased even though the baking was not implemented. The penning gauge shows 1.9~2.0 x10-6 Torr in each case;  open/close the gate valve. It's strange. The gauge would not be working.
R&D (FilterCavity)
Print this report.
NaokiAritomi - 17:47, Monday 21 October 2019 (1759)Get code to link to this report
IR reflection measured on 20191015 after closing chamber

Injection: 234 uW

Reflection: 200 uW (off resonance)

Reflection: 135-139 uW (on resonance)

Off resonance reflectivity is 85.5%. 

R&D (FilterCavity)
Print this report.
NaokiAritomi - 17:40, Monday 21 October 2019 (1758)Get code to link to this report
SHG, OPO parameters after typhoon

SHG temperature: 3.115 kOhm

green power (mW) OPO temperature (kOhm) p pol PLL (MHz) BAB maximum (V)
0   305 0.088
20 7.17   0.416
25 7.18   0.576
30 7.18   0.792
35 7.18   1.06
40 7.19 160 1.5
45 7.19   2.06
50 7.2   2.82
55 7.209   3.52
60 7.209 160 4.36
R&D (FilterCavity)
Print this report.
YuhangZhao - 09:54, Monday 21 October 2019 (1757)Get code to link to this report
Correction of PD's position for FC_tra IR / IR transmission issue

Yao-Chin and Yuhang

We found filter cavity transmission IR PD is placed not on the waist of the detection beam. To avoid the clipping noise and also increase the detection ability, we corrected it. Look at the attached picture 1 and 2. Now the PD is closer to BS and the beam is well within PD.

Also, we measured the filter cavity transmitted power.  Compare this transmitted 0.5uW with incident 400uW. The measured transmission is about 0.6%. While the calculated value from the LMA specification is 1%. Since the measured IR after Dichroic and laser line filter goes thorough a lot of optics, this should be consistent.

Images attached to this report
1757_20191021025403_beforemove.jpeg 1757_20191021025417_now.jpeg 1757_20191021025433_fctrair.jpeg
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 19:58, Sunday 20 October 2019 (1756)Get code to link to this report
Sudden failure of BS turbo pump
[Yuhang, Yao-chin, Eleonora ]
 
Last friday 18/10 at about 16.30, we had a sudden issue with turbo pump of BS.  While we were working in the clean room, it started to make a strong noise and to decrease its rotation frequency down to zero. On the screen it appeared the message “ERROR 089”.  We checked on the manual (link below)  and found that it corresponds to “ Rotor vibration” error.
 
 
We couldn't contact Takahashi-san. We closed the gate valve of turbo pump and rotary pump and switch rotary pump off.  We didn't close any main gate valve.  Today the vacuum level was still the same of Friday (2.8e-6 Torr)
KAGRA MIR (Polarization)
Print this report.
PengboLi - 00:10, Sunday 20 October 2019 (1755)Get code to link to this report
P-polarization Maps of OSTM (SK)
[Simon, Pengbo]

We have started to analyze the coated OSTM.

First, we took out the OSTM and inspected the mirror visually. We found pencil marks on the barrel and among them an arrow that indicates the thicker side of the wedged substrate and shows toward the HR side of the mirror.

The orientation which we choose for the measurement is to put the thicker side upside first and then rotate it by 90 degrees(with the thinner side facing the PC). Thus we only need to consider the horizontal deflection of the beam. After the adjustment of the sample, we found no output beam from the other side of the sample which mainly due to the coating absorption. So we removed one of the two ND filters to increase the laser power.

Second, we started the adjustment of the IU.

Due to the deflection effect, we first changed the lens position to make sure the beam goes through its center. Secondly, we rotated the second PBS slightly to have a perpendicular incident of the beam. Then, we adjusted the position of the two photodiodes so that the s-polarized beam and the p-polarized beam can hit the center area, respectively. Because we were doing the P-polarization measurement, the s polarized beam is nearly invisible, so we used the oscilloscope to finish the adjustment.

Then we ran the polarization map scan in the center with a 15mm radius. Attached to this report is the P-polarization map.
Images attached to this report
1755_20191019170251_figurep1.png 1755_20191019170259_figurep2.png
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 16:01, Friday 18 October 2019 (1754)Get code to link to this report
Not responding coil circuit seems open

[Yuhang, Yaochin, Eleonora]

We continued investiagation about not responding coil of the input mirror. (See entry #1742)

We measured the resistance across each coil loop at the flange. For all the working coils it was about 2.4 Ohm for the not working one we found the loop is open.

This suggests the magnet coud be fine and the problem could come from the coil circuit. 

R&D (FilterCavity)
Print this report.
YuhangZhao - 14:29, Friday 18 October 2019 (1753)Get code to link to this report
Measurement of current loss budget

Aritomi, Eleonora, Yao-Chin, and Yuhang

We summarized the loss budget in the first attached figure.

From the power measurement as following, we could know this part contributes about 10% of loss

measured after first PBS after OPO 265uW
measured after last mirror before PR chamber 198uW

From the power measurement as following, we could know this part contributes about 1% of loss

measurement of IR reflection after PR chamber 221uW
measured before homodyne 218uW

We also measured the IR reflection(IR_ref) beam shape as in the attached figure 2. It has a bit ellipse in the horizontal direction. Because of this elongation, we had a second order peak when we match this IR_ref to alignment mode cleaner(as shown in the attached figure 3,4). By comparing this higher-order peak(12mV) and TEM00 peak(276mV), we could deduce the homodyne readout efficiency should be (1-12/276)^2 = 91.5%. But this IR_ref has beam jittering, so the real readout efficiency should be less than 91.5%.

We also measured the filter cavity round trip loss again. By using the same method we used last year, we found now this value is around 100ppm. We will do more investigation to make sure of this. The measurement is attached in the figure 5. 

In the last figure, the achievable squeezing without considering phase noise is attached.

Images attached to this report
1753_20191018072640_lossbudget.png 1753_20191018072906_irrefshape.jpeg 1753_20191018072915_tem00.jpeg 1753_20191018072922_higherorder.jpeg 1753_20191018072935_rtl.jpeg 1753_20191018151254_sqzdeg20191018.png
R&D (FilterCavity)
Print this report.
NaokiAritomi - 08:27, Friday 18 October 2019 (1752)Get code to link to this report
Saturation of CC2 correction signal

[Aritomi, Yuhang, Yaochin, Eleonora]

We have saturation of CC2 correction signal (8Vpp) and it seems coming from servo since we have this saturation without connecting piezo. HV amplifier we are using can take up to 10V and the gain is 30 and piezo for phase shifter can take up to 1000V.  So we are using 240V out of 1000V piezo range.

R&D (FilterCavity)
Print this report.
NaokiAritomi - 22:38, Thursday 17 October 2019 (1751)Get code to link to this report
Frequency dependent squeezing at 50kHz

[Aritomi, Yuhang, Yaochin, Eleonora]

Today we measured frequency dependent squeezing at 50kHz.

Setting is same as yesterday (AOM frequency is 109.13698MHz). Since CC2 lock is unstable, we made average number 5 while it was 50 yesterday.

Images attached to this report
1751_20191017153827_fds20191017.png
R&D (FilterCavity)
Print this report.
NaokiAritomi - 22:45, Wednesday 16 October 2019 (1747)Get code to link to this report
Frequency dependent anti squeezing at 50kHz

[Aritomi, Yuhang, Yaochin, Eleonora]

Today we measured frequency dependent anti squeezing at 50kHz.

We set AOM frequency 109.13684 MHz which is larger than carrier frequency by 100kHz and corresponds to 50kHz detuning. CC2 demodulation phase is 105deg for squeezing, 155deg for anti squeezing and 135deg for intermediate.

Images attached to this report
1747_20191017105535_fds20191016.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 19:12, Wednesday 16 October 2019 (1746)Get code to link to this report
Power measurement when PR chamber is open(measured on 20191011)

Aritomi, Matteo, Yao-Chin, and Yuhang

All the measurement is inside PR chamber. We sent BAB to do this measurement.

injection power 394uW
before dichroic 379uW
after dichroic 375uW
transmission from the first PBS 3.4uW
transmission from the second PBS 3uW
before going back to bench 334uW

Injection loss: 4.8%

Reflection loss: 10.9%

R&D (FilterCavity)
Print this report.
YuhangZhao - 19:00, Wednesday 16 October 2019 (1744)Get code to link to this report
IR reflected from filter cavity cut issue solved

Aritomi, Matteo, Yao-Chin, and Yuhang

We opened the PR chamber on last Friday. But moving the last IR back reflection steering mirror, we solved the cut issue of this beam.

Attached figure 1: We just opened PR chamber, we checked with IR viewer. We could see a beam on the right side of chamber window.

Attached figure 2: After moving the last IR back reflection steering mirror, we made IR back reflection hitting on the window. And this IR back reflection is between IR(injection) and GR.