NAOJ GW Elog Logbook 3.2
Attached to this report are the XY maps and distribution histograms of the absorption coefficient from the Shinkosha #5((shown in the attached figure 1 and 2), and Shinkosha #6(shown in the attached figure 3 and 4), the color bar scale for the two maps is the same.
As can be seen, sample #6 has better absorption than sample #5, and we think this is due to the inhomogeneous of the absorption along the z-axis of the samples.
For these two samples, we choose the center of the z-axis. We can check our assumption by doing another XZ maps.
Aritomi, Yaochin, and Yuhang
For CC2 phase shifter, we could always hear the sound around it and we wanted to check what is the frequency of this sound. So we used a sound spectrum analyzer and it shows only one frequency component which is 6.9kHz. Note that we couldn't see this oscillation from the oscilloscope.
We checked the squeezing spectrum and found that this frequency appears in the spectrum of squeezing. Also in the squeezing spectrum we could find the peak at ~23kHz. These peaks come from the resonance of CC2 phase shifter.
However, we found there was also a clear peak at 16kHz. And we found this peak when we measure OLTF of CC2. So it seems this peak comes from servo? Maybe we can consult with Pierre.
[Aritomi, Yuhang, Yaochin]
Last Friday, we lost IR alignment, but the IR alignment could be easily recovered by moving one steering mirror on the bench. I moved both pitch and yaw, but especially pitch was misaligned. Now mode matching went back to ~90% as before.
Aritomi, Yaochin, and Yuhang
For CC2 phase shifter, we could always hear the sound around it and we wanted to check what is the frequency of this sound. So we used a sound spectrum analyzer and it shows only one frequency component which is 6.9kHz. Note that we couldn't see this oscillation from the oscilloscope.
We checked the squeezing spectrum and found that this frequency appears in the spectrum of squeezing. Also in the squeezing spectrum we could find the peak at ~23kHz. These peaks come from the resonance of CC2 phase shifter.
However, we found there was also a clear peak at 16kHz. And we found this peak when we measure OLTF of CC2. So it seems this peak comes from servo? Maybe we can consult with Pierre. (figured out later this 16kHz if from mechanics)
Aritomi, YaoChin and Yuhang
To have a feeling of IR and GR overlap. We checked three points.
1. First target(shown in the attached figure 1 and 2). In the first figure, IR is actually in the center. When I was checking it, I could see the difference if Aritomi-san blocks or unblock the IR beam. In the second figure, there is GR. But it is difficult to tell the overlap level.
2. Second target(shown in the attached figure 3 and 4): It seems IR is higher and a bit left.
3. End camera(shown in the attached figure 5 and 6): It seems IR is also higher and a bit left.
Eleonora, Yaochin, and Yuhang
Firstly, we tried to excite FC length by using H1 and H3 coil. But we found H1 is not working(because we sent excitation to H1 only but the mirror couldn't move accordingly). So we checked if we were sending the signal and whether the signal is reaching the coil driver. We found the signal is reaching the coil driver but with some noise. The frequency of this noise is measured as 700kHz.
So we decided to excited FC length by using H2 and H4 coil. The method we used to diagonalize is to measure the response of the mirror optical lever signal. We sent excitation to channel K1:FDS-INPUT_Z_CORR_fil_EXC with a frequency of 4Hz and amplitude of 4000. We set the matrix index for H2 as 1 and measured the spectrum of K1:FDS-INPUT_Y_fil_IN1. There was a peak of 126.9 at 4Hz. Then we set the matrix index for H4 as 1 and measured the spectrum of K1:FDS-INPUT_Y_fil_IN1. There was a peak of 173.4 at 4Hz. So we decide to put index 1 for H2 while index 0.73 for H4 to make the coupling to yaw to be zero.
In the end, we measured the transfer function from the excitation of the INPUT mirror length to FC correction. There should be a response. We also measured the TF from the excitation of INPUT mirror length to INPUT mirror yaw optical lever. There should be not a response. Also the coherence of the above two. The result is shown in the attached figure 1. We have a response from the FC correction signal 10 times larger than the response from the yaw optical lever. Also, the coherence is better for FC correction. This can be a good start for the implementation of the feedback for CC2 phase noise.
Eleonora, Pengbo, Yaochin, and Yuhang
We found green transmission DC dropped to ~600 counts at the beginning of today's work.
First, we checked the GR higher-order modes we have for the filter cavity.(As shown in the last figure) It was fine.
Then we went to the end room and found the GR_tra PD was tilted almost by 45deg. We think the PD cables have been accidentally pulled during the work to repair air conditioner which took place this morning in the end room. After we correct the angle of this PD, we found the signal went back to ~2700 counts.
We also checked green power at two points(shown in the attached picture 1) in today's situation (the green injected power is 12mW). The first point is just before GR_tra PD and it is 45uW(shown in the attached picture 2 and measured as shown in the attached picture 3). The second point is just after the green BS and it is 1mW(shown in the attached picture 4).
Simon
I recalculated the absorbance values to take into account the small transmittances we got from the spectra taken at the ATC for all colored Sapphire samples.
The new spectra, now also with reference values at 633nm, can be found in the attachement.
Pengbo, Simon
We removed the OSTM and packed it again. It is now located in the small shelf inside the clean room.
After that, we recalibrated the system.
The calibration values are:
R_surf = 17.86 1/W
(AC = 0.54V, DC = 4.58V, P_in = 0.030W, abs_surfref = 0.22)
R_bulk = 0.8 cm/W
(DC = 0.105V, AC = 5.42V, P_in = 0.030W, abs_bulkref = 1.04/cm)
Then, we exchanged the calibration sample with the first of the 5 small Shinkosha Sapphire samples: S5 (2" x 20mm)
We located the center of the sample as [X,Y,Z] = [327.25, 122, 33.5].
The transmittance of the sample is T ~ 0.85
Since we cannot set 10W as an input power and leave it as it is while moving the sample into the beam and out again (the sample holder will cut the beam otherwise), we set the power to its minimum trough the polarizer while it is feeded with 7A current and used the beam-shutter for moving the sample holder. In the center position, we set the power to maximum. The transmitted beam power has been measured to be 8.4W.
Now the map is being taken with 15mm radius.
Aritomi and Yuhang
We found the problem of the small RF signal of the AA's quadrant (elog1670). Matteo T suggested checking the bias. We turned on the button on the quadrant box and measured the voltage at the end of the power cable. We checked with multi-meter and it shows 150V reached quadrant.
So bias seems not to be a problem.
I installed 3 steering mirrors for HOMs beam paths to roughly decide where to put PDs for intensity stabilization.
2 HOM-beams are roughly alinged into STMs which are for the alignment for cryogenic cavity.
Attached is the picture of current situation of optical table.
- fluctuation of IR transmission from filter cavity (entry 1701)
- large CC2 phase noise from filter cavity (entry 1695)
- large bump in shot noise spectrum at low frequency (entry 1529)
[Aritomi, Yuhang]
We measured mode matching when filter cavity is aligned with pitch dithering. The result is as follows. Mode matching was around 90%. We found that when pitch dithering was engaged, pitch misalignment became less, but yaw misalignment became more.
Mode | AOM frequency (MHz) | IR transmission |
TEM00 | 109.03607 | 3000 |
HG10 | 109.43128 | 350 |
HG01 | 109.43219 | 200 |
IG02 | 109.8288 | 115 |
offset | 94 |
IR TEM00 transmission was fluctuating even when dithering was engaged (attached movie). Time scale of the movie is 2s and DC offset is 94.
We measured BAB reflection when BAB is on/off resonance. Off resonance reflectivity is 82% and the reflection might be still cutted.
I modified the simulink model to include the possibility to feedback the CC2 correction signal to the length d.o.f. of the input mass.
The ADC channel where to inject the CC2 signal is the n 13 in the top BNCtoDsub converter in the clean room (named ADC0 Ch16-32).
I modified the medm screen accordingly. Next step will be to optimize the length driving of the input mirror.
On Wed 02/10 we had another timed-out error on diaggui. It was solved by restarting the standalone. It happened ~13 days after the previous one. (entry #1650).
Pengbo, Simon
We have started to analyze the coated OSTM from Shinkosha regading the absorption of the HR side.
First, we took out the OSTM and inspected the mirror visually. We found pencil marks on the barrel and among them an arrow that indicates the thicker side of the wedged substrate and shows toward the HR side of the mirror (see attached photos).
The actual orientation of the sample inside the sample holder was a little bit tricky mainly because of the wedge and the size of the sample-holder which is basically too large for such a mirror.
- At first, we tried to put the thicker side upside to have a somewhat parallel orientation between sample-holder and mirror-plane (as can be seen from the pictures, we are using a - with optical tissue - covered ruler as a spacer). However, we recognized that this will distract the pump-beam so that it cannot be measured anymore regarding its power.
- Therefore, secondly, we rotated the mirror by 90 degrees so that the distraction of both pump and probe beam is only parallel to the optical table which can be counter measured by a respective relocalisation of the IU and the photometer. However, with this position, the HR coating is facing the pump but also the spacer, which we originally wanted to avoid
- Especially regarding the probe, we had to change the IU position by ~3mm less than it would have been the case for a wedge-free substrate
After the alignment of the sample, we looked for the exact position of the HR-coating by applying Z-scans. We carefully increased the laser power to take care that the coating is not damaged (initially, we did this on the outer edges of the mirror of course).
Then, we ran a map-scan in the center with 15mm radius in the position were we identified the coating (Z = 48.8). The results of that scan can be seen also in the pictures attached. Our main result is a quite homogeneous mean absorption of 16 ppm (+/- 3ppm) with some point-like excesses indicating the positions of either dust or defects within the coating, most likely.
Now we could lock CC2 loop with unity gain frequency of 2kHz. To see the difference of CC2 phase noise with different gain. We measured phase noise with different gain.
As expected, higher gain make noise lower at low frequency. But also the higher gain excites resonance at higher frequency.
From the measurement, it seems the gain of 0.2 is the best case. (Although they are all quite similar)
[Aritomi, Yuhang]
We measured CC2 free running phase noise when filter cavity is locked and aligned with dithering (attached picture). We cannot lock CC2 stably since piezo actuation range is not enough. We'll try to feedback CC2 error signal at low frequency to input mirror of filter cavity.
R_surf = AC_surfref/(DC_surfref*P_in*abs_surfref) = 18.1 [1/W]
where AC_surfref = 0.425V, DC_surfref = 3.55V, P_in = 0.030W and abs_surfref = 0.22
R_bulk = AC_bulkref/(DC_bulkref*sqrt(T_bulkref)*P_in*abs_bulkref) = 0.741 [cm/W]
where AC_bulkref = 0.072V, DC_bulkref = 4.2V, T_bulkref = 0.55, P_in = 0.030W and abs_bulkref = 1.04/cm
Aritomi and Yuhang
We checked power at several points this Monday and make IRMC transmission set at 1.7mW.
Actually, this means we increased also the error signal(or increase gain). Today we checked the error signal, and actually, it was quite close to oscillation. So we adjusted the gain while looking at the error signal and measured transfer function again. In the end, we put the gain value of the control board from 1.3 to 0.8.
The open-loop transfer function now is as the attached figure 1.
We also measured the IRMC error signal spectrum while IRMC is unlocked and locked. As shown in the attached figure 2. From this locking performance, we could see that the loop suppresses the even harmonics of a fundamental 9Hz oscillation while there are still some odd harmonics left. Also, the 50Hz and its harmonics are introduced after closing the loop. So the PD doesn't introduce any 50Hz noise.