NAOJ GW Elog Logbook 3.2
[Aritomi, Yuhang]
First we tried DC balance of homodyne with s pol, but power unbalance of BS was large for s pol. Reflection of BS is 630 uW and transmission is 600 uW for s pol while it's almost balanced for p pol. So we decided to use p pol for the moment. We made LO p pol with HWP and confirmed its polarization with cubic PBS (Newport,10BC16PC.9). However, when we put plate PBS (thorlabs, PBSW-1064) instead of cubic PBS, most of the LO was reflected by plate PBS. Actually I didn't check the direction of plate PBS and maybe it's wrong, but even in that case we cannot explain this behavior. We'll check the direction of plate PBS and check the polarization with AMC tomorrow.
Anyway, today we managed to make LO DC balanced with "p pol".
Aritomi and Yuhang
Today, we checked together about the polarization again with PBS plate.
So we concluded that the beam we were using is p polarization. And we also need to pay attention next time about the surface of mirror and the angle of PBS plate.