NAOJ GW Elog Logbook 3.2
[Aritomi, Yuhang, Eleonora, Matteo]
By reducing green pump power from 26.7mW to 18mW, we managed to reproduce the last Friday's FDS measurement. The detuning is around 100-130Hz. It seems that squeezing level at low frequency is still limited by locking accuracy.
The parameters are as follows.
sqz_dB = 7; % produced SQZ
L_rt = 150e-6; % FC losses
L_inj = 0.20; % Injection losses
L_ro = 0.11; % Readout losses
A0 = 0.05; % Squeezed field/filter cavity mode mismatch losses
C0 = 0.05; % Squeezed field/local oscillator mode mismatch losses
ERR_L = 13.3e-12; % Lock accuracy [m]
ERR_csi = 80e-3; % Phase noise [rad]