ShalikaSingh - 11:58, Tuesday 24 September 2024 (3754)Get code to link to this report
Beam characterization for EOM

There is a scale on the table kept in backward direction, i.e increasing distance from laser is decreasing measurement on the scale. 

In reference to elog 3561 the beam was as attached in Fig 1. The beam characterization was originally done with f=100mm lens at 101.5cm. The waist of beam before the lens was at -47.1 mm, i.e at 101.5 cm +4.71 cm=106.21 cm

According to the elog the lens f = 100 mm was moved later by d=0.2m. But, now the setup has two f=50mm lens. This was never mentioned in elogs that the lens was changed. 

For the modified setup:

The two f = 50 mm, -50mm lens are placed at 96.5 cm and 90.5cm.Hence, the relative distance between lens and waist of beam is 9.71cm and 15.71cm respectively.

The new beam propagation is as in Fig 2. I propagated the original beam (one before f=100 mm lens) after the two f = 50mm lens. 

The beam power before filter is 64mW. 

I wish to replace the energy filter before the HWP on the EOM path, with a beam splitter or beam sampler and make a separate path for SLM. 

Images attached to this report
3754_20240924045757_completebeambeforeandafterf100mm.png 3754_20240924100951_completebeambeforeandafter2f50mm.png
Comments related to this report
ShalikaSingh - 13:26, Tuesday 24 September 2024 (3755)

The lens change was mentioned in 3563

ShalikaSingh - 17:09, Tuesday 24 September 2024 (3756)

The lens actually was f=50mm and f=-50mm. I have changed the beam characteristics accordingly.