Warning: Undefined variable $s in /var/www/public_html/osl/classes/DAO.php on line 959
NAOJ GW Elog Logbook
LOG-IN
Displaying report 1-1 of 1.
General (General)
Print this report.
RyutaroTakahashi - 16:24, Thursday 20 February 2025 (3924)Get code to link to this report
Replacement of Rotary Pump

I replaced the RP (2015) for theTMP with new DRY pump (NeoDry30G) in the west end. The TMP is running with the DRY pump now.

Images attached to this report
3924_20250220082319_neodry30gwest.jpg
Comments related to this report
RyutaroTakahashi - 18:12, Friday 21 February 2025 (3926)

The TMP in the west end has stoped in error E050 (Excess tenperature). I found the FAN for the TMP was not working (broken). I stoped the DRY pump as well.