KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 23:13, Thursday 26 July 2018 (921)Get code to link to this report
Separate setup for noise investigation on 1310nm laser
I moved another optical table in the cleanroom because I didn't have enough space to properly test the laser noise reduction.
 
I removed the laser fiber from its collimator without moving the collimator, so that, when I will put it back, I will not loose the alignment.
 
I set the laser at 200mA which is half of the typical operation, this current gives 60mW in output. I put an OD2 filter after the collimator and a beam dumper to collect its reflection. Then a non-polarizing cube beam splitter 50:50 and a 50mm lens for each beam to focus the beam on 2 identical photodiodes (DET10N and DET10N/M). On one beam I measure the beam size at different positions. To do that, I put a blade moved by the old translation stage and controlled by the old software that makes scans of the DC. On the DC channel, I sent the analog output of the PM100D power meter .
 
I found a waist of 20micron at 54mm from the lens and I put the PD at about that position. I didn't repeat the waist measurement for the other lens but I put the other PD at the same distance from the lens. Each PD is mounted on a XY micrometric translation stage, so, by watching the signal at the oscilloscope I centered the beams in the PDs (moving the PDs instead of the beam).
 
I put a T with a load resistance of 7.5kOhm for both the PDs. This gives 2.5V DC signal for one PD and 2.8V for the other. The two signals, coupled in AC, overlap on the oscilloscope and show the "jumps" that are probably the noise reason.
 
I closed the loop with a lowpass at 3Hz and gain 1000. See the spectra.
 
The noise in the out-of-loop PD reduces by a factor of 10 (20dB)
 
I also connected the PD at the lockin and after demodulation at 380Hz (with the internal oscillator) the noise significantly reduces (by about 10 times by eye).
 
Next thing to do is to place the PDs in the absorption setup and check if the absorption signal noise reduces in the same way
Images attached to this report
921_20180726151806_z00mm.png 921_20180726151813_z05mm.png 921_20180726151817_z10mm.png 921_20180726151822_z15mm.png 921_20180726151827_z20mm.png 921_20180726151832_z25mm.png 921_20180726151837_z30mm.png 921_20180726151840_z35mm.png 921_20180726151844_z40mm.png 921_20180726151848_z45mm.png 921_20180726151852_z50mm.png 921_20180726151906_after50mmlens.png 921_20180726161329_011.jpg 921_20180726161334_01.jpg 921_20180726161340_001.jpg 921_20180726161345_00.jpg 921_20180726161442_20180726inloopdet10noutofloopdet10nm.png
Comments related to this report
RaffaeleFlaminio - 15:17, Sunday 29 July 2018 (923)
- There is a large peak at 35 kHz when the loop is closed. Is the loop gain too high?
- The dark noise measured above 1 kHZ is not as flat as one would expect. Is this real or is it a problem with the measurement?