NAOJ GW Elog Logbook 3.2
laser : 7 A and HWP = 0deg : ~30mW
surface calib
z_iu = 69.18mm
z = 33 mm, (x,y) = (327.093,119.965)
checked calibration factor from 33 to 36mm, the max is at z=35mm ~13.9/W
z scan : Fri, Mar 29, 2024 11-29-17 AM.txt with Pt = 20.8mW (Pt = 21.4mW corresponds to Pin = 31.4mW)
bulk calibration
Pin = 31.1mW, Pt = 17.1mW to compute transmission
during measurement Pt = 17.5mW
moved z_iu by -0.32mm, z = 33.8mm gives R = 0.6413cm/W
z scan : Fri, Mar 29, 2024 2-27-42 PM.txt
T50#1 (30mm diameter, 50mm length, #4/6)
ziu = 48.61mm
dc was heavily misaligned in vertical... due to holder?
check if center is (x,y) 327,119 we can do +/-10mm in all direction without problem
also dc does not change during this so holder seems fine
Pt = 26.6mW, Pin = 31.2mW
can change z between 20 to 90 mm
HWP = 28deg : Pt = 4.83W
long z scan at x,y center : Fri, Mar 29, 2024 5-33-48 PM.txt
z edges are at 26 and 60 mm :Z_center = 43mm
10 mm radius map at z_center with 5mV gain and Pt = 4.81W : Fri, Mar 29, 2024 6-53-59 PM.txt
I placed a pink tape on top close to IU side
Even though I already tried few days ago, this time unplugging/replugging the ethernet switch cables seems to have solve the ethernet connection issues of PCI and FC pcs.
The data is being saved in the format below. Also, a header indicating the name of column has been added using "set dynamic attribute", and by enabling the header function in saving measurement to file.
Data | Column Number |
LC1 voltage | 0 |
LC2 voltage | 1 |
LC1 temperature | 2 |
LC2 temperature | 3 |
Mean Transmitted Power | 4 |
STD Tansmitted power | 5 |
Mean Main laser power | 6 |
STD Main laser power | 7 |
Mean Azimuth | 8 |
STD Azimuth | 9 |
Mean Ellipticity | 10 |
STD Ellipticity | 11 |
In labview the averging filter was modified to include the standard deviation of data and also to save them simultaneously in file.
The data is being saved in the format below. Also, a header indicating the name of column has been added using "set dynamic attribute", and by enabling the header function in saving measurement to file.
Data | Column Number |
LC1 voltage | 0 |
LC2 voltage | 1 |
LC1 temperature | 2 |
LC2 temperature | 3 |
Mean Transmitted Power | 4 |
STD Tansmitted power | 5 |
Mean Main laser power | 6 |
STD Main laser power | 7 |
Mean Azimuth | 8 |
STD Azimuth | 9 |
Mean Ellipticity | 10 |
STD Ellipticity | 11 |
I checked the He pressure today and did not notice any change. I will add a mark for future reference.
I returened the adpaters and teflon tape borrowed for He recharge to Tama circuit room.
Took a measurement scanning from 0-25V with 1V step for both LC. The mirror/BB1EO3 was rotated and then the measurement was taken.
Mirror at 0 deg-
Input_foldername=r'C:\Users\atama\OneDrive\LC-Experiment\Measurement Data\Polarization states\20240325'
input_filename= 'Mon, Mar 25, 2024 7-49-56 PM.txt'
Output_foldername=r'C:\Users\atama\OneDrive\LC-Experiment\Measurement Data\Coating measurement\BB1-E03\20240325\0 deg'
output_filename= 'Mon, Mar 25, 2024 7-58-01 PM.txt'
Mirror at 28 deg -
Input_foldername=r'C:\Users\atama\OneDrive\LC-Experiment\Measurement Data\Polarization states\20240325'
input_filename= 'Mon, Mar 25, 2024 7-49-56 PM.txt'
Output_foldername=r'C:\Users\atama\OneDrive\LC-Experiment\Measurement Data\Coating measurement\BB1-E03\20240325\28 deg'
output_filename= 'Wed, Mar 27, 2024 8-33-59 PM.txt'
[Tomaru, Mitsuhashi, Futaba Kogyo Members, Rishabh]
We organized and cleaned up the ATC lab (outside the speed-meter cleanroom) to enhance floor space utilization. some photos.
If possible, we should avoid using the newly cleared area for storage purposes.
[Tomaru, Rishabh]
The regulator from KEK was delivered, so we recharged the compressor to 16.6 bar. I will monitor the pressure for a week and have a cool-down test around mid-April.
Date: 2024/03/19
[Tomaru, Rishabh]
While working to recharge the compressor, we noticed that the inside of the storage shed (and the compressor) was extremely dirty. This is because of the openings on three sides of the shed. I looked for aluminium sheets to cover these openings temporarily but couldn't find any. I will order some plastic or aluminium sheets to cover up the openings and clean up the inside of the shed.
[Tomaru, Rishabh]
Date: 2024/03/19
We tried to recharge the compressor with G1 Helium in the ATC lab. For this work, we used the flexible line brought from KEK. Since there was no adapter to evacuate the flexible line, we just purged it with helium before connecting it to the compressor's recharge port. After connecting the line, we tried to recharge the compressor but found the reading on the regulator fell to 0. We realized the regulator is only rated for 0-3kgf/cm^2, an order smaller than we need. So, we requested Node-san in KEK to send a 2-3MPa regulator to NAOJ. We will use it to recharge the compressor and purchase a new one in the next fiscal year.
Since there was no chain, the He tank was tied to a white rack using a crane belt for safety. Also, random items stored around the cryogenic optical table and DAQ rack were moved around for this work and left that way. They shouldn't be in anyones way. I want to clean up/move unused so they are not in the way.
[Tomaru, Rishabh]
The regulator from KEK was delivered, so we recharged the compressor to 16.6 bar. I will monitor the pressure for a week and have a cool-down test around mid-April.
I checked the He pressure today and did not notice any change. I will add a mark for future reference.
C:\Users\atama\OneDrive\LC-Experiment\Measurement Data\Polarization states\20240314\
'Thu, Mar 14, 2024 1-34-25 PM.txt' for voltage from 0 to5V for both LC
'Thu, Mar 14, 2024 2-45-11 PM.txt' for voltage from 0 to 10V for LC1 and 0 to 5V for LC2
sweep done with 1V resolution for both LC to have unique azimuth and ellipticity and also avoiding the multiple of angles.
I moved three boxes of Swagelok adapters from TAMA circuit room to ATC lab. (image)
The boxes are now next to the DAQ rack near the cryostat. (image)
I will use them while recharing helium gas and should return them in a week or two.
I returened the adpaters and teflon tape borrowed for He recharge to Tama circuit room.
BB1EO3 was installed after the BS. The reflected beam divergence is np.arctan(1/200)=0.29 deg. The LC are scanning roughly from 0-5V with 1V resolution. This is intended to directly estimate the birefringence of the mirror.
measurement:
C:\Users\atama\OneDrive\LC-Experiment\Measurement Data\Coating measurement\BB1-E03\20240311
Mon, Mar 11, 2024 5-25-22 PM
[Marc (remote), Mitsuhashi]
There seems to be some issue with TAMA ethernet switch as no PC are accessible from remote but WIFI is available.
Mitsuhashi-san went to check ethernet connection on LC PC, on new DGS switch and ethernet switch and everything seems plugged and properly on.
We also turned on/off the ethernet switch without success.
In the end the LC PC is connected to NAOJ open and we could recover the remote connection.
We need further check as FC and PCI pcs are still not accessible..
This is a report on the measurement we did on 03/07.
We measured the open-loop transfer function of the cavity by injecting a swept-sine disturbance signal between plant H and actuator A and taking the ratio of measurement data from directly before and directly after signal injection. Goal of this measurement was to assess lock stability by means of the phase margin. The lock of the main laser is achieved by a fast actuation with a piezo with servo transfer function F_pzt*A_pzt combined with slow thermal actuation with servo F_th*A_th, both actuations performed on the laser itself. While F_pzt, A_pzt and A_th are basically fixed , we can influence the stability of the lock by adjusting the servo F_th with the MokuLab. Based on trial and error, we designed the thermal servo to have the following parameters:
Integrator gain (saturation limit) and UGF: g_i = 43.6 dB, f_i = 160 Hz
Differentiator gain (saturation limit) and UGF: g_d = -45.5 dB, f_d = 315.6 kHz
With the above information and the knowledge of F_pzt, A_pzt and A_th, the optical gain can be inferred and found to be constant. UGF lies around 100 Hz and the phase margin amounts to around 85 deg. The lock seems reasonably stable, at least, enough for our purposes, so we will continue. A plot of the described analysis can be found here.
7 mixers and their low pass filters, 2 DC blocks and 1 splitter has been fixed to the aluminium frame.
I also found a board with LEMO 'i-style' feedthrough so I'll take them from this board for the front panels.
Next step is to prepare front panel and several short length sma cables.
For interpolation verification, did a small finer scan.
V1- 1-1.01V, 1mV
V2- 5-5.005V, 1mV
data is here:
C:\Users\atama\OneDrive\LC-Experiment\Measurement Data\Polarization states\20240308\
Fri, Mar 8, 2024 6-36-13 PM.txt
Since the failure of DDS boards last year we have too messy cabling around NIM rack.
I started to prepare a box to enclose all mixers and PLL monitor boxes.
I picked a 2U box and fitted a small aluminium plate to host these components.
The BNC mixers are now properly fixed.
For the SMA mixers I would like to use custom length SMA cables.
I would like to use LEMO feedthrough on the front panel but up to now I can't find the proper washers...
7 mixers and their low pass filters, 2 DC blocks and 1 splitter has been fixed to the aluminium frame.
I also found a board with LEMO 'i-style' feedthrough so I'll take them from this board for the front panels.
Next step is to prepare front panel and several short length sma cables.
[Marc, Katsuki]
Thanks to Kanzawa-san we got the new pusher for the glass cell container.
We measured again birefringence versus pusher position. This time we saw small increase in ellipticity while it was strongly decreaseing with the plastic cell container.
One possibility is that the pusher pressure was inducing stress birefringence on the plastic walls.
We then measured mass versus pusher discplecements for both plastic and cell container with the 3d printed holder.
We still had some issue of hydrogel escaping from the space between the pusher and cell container walls so we then put the pusher into one finger of a plastic glove that mostly solved this issue.
We repeated measurement of hydrogel birefringence versus pusher position but this time the restoring force of hydrogel was stronger and pushing up the translation stage despite zip tie.
We should prepare a better enclosure for the translation stage.
I tried to apply large displacement while holding the translation stage but the glass cell container broke..
In any case it seems that hydrogel is sensitive enough to cell induced birefringence.
All measurements were done with 10% hydrogel mix.
All data and analysis are in 'hydrogel_birefringence.py' file
pt _ 4.65W
map at z_53 mm , Pt _ 4.65W; Sat, Mar 30, 2024 11-35-05 AM.txt
map at 33mm ' Sat, Mar 30, 2024 8-01-36 PM.txt