LOG-IN
Displaying reports 2861-2880 of 3273.Go to page Start 140 141 142 143 144 145 146 147 148 End
R&D (FilterCavity)
Print this report.
YuefanGuo - 17:13, Friday 07 April 2017 (444)Get code to link to this report
Try to send the video signal from the end room to the central room
I tried to find a way to send the video signal from the camera at 290m to the central room.

Firstly, I tried the optical fiber used to send the PD signal,connected it to the camera at the end room, and connected it to the screen in the central room. There is only some flash when I turned on and off the receiver, but seems not capable to get the real signal from the camera.

Then we found out there is another system like a recorder that can be used. I plugged the recorder with the camera and the output to the screen at the end room. It works pretty well. And also there is a sticker with an IP address on it. We found the right internet cable for it, and tried to connected it with the computer in the center room. We successfully connected it, but cannot see the video signal.

Then we checked the old TAMA board which can convert the video into optical signal, send through the fiber and there is another this kind of board in the central room to convert it back, then connect to the screen, then we should see the video on the screen. At first, we tried with the cable 1-13 and 1-14, but we are not able to see anything from the screen. Then we tried to change the cable into 1-15 and 1-16, then we got the video in the central room.

The first picture is taken on the screen near the camera at 290m, the second picture is taken in the screen in the central room, both taken by my phone. It seems there are some loss during the signal transmission. Then we adjust the camera's position and also the focal lengths, and inject the red light into the arm, the third and the forth pictures are taken at 290m and central room.

The fifth picture is the board we are using now.
Images attached to this report
444_20170407100734_20170407170602.jpg 444_20170407100926_20170407170902.jpg 444_20170407101102_20170407170608.jpg 444_20170407101302_p4070352.jpg 444_20170407101318_20170407170555.jpg
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 21:44, Thursday 06 April 2017 (443)Get code to link to this report
Picomotor control issue

I upload some slides I have prepared to summarize the picomotor control issue we are facing.

Non-image files attached to this report
R&D (FilterCavity)
Print this report.
YuefanGuo - 15:59, Wednesday 05 April 2017 (442)Get code to link to this report
Finish the design of telescope on infrared path
With Matteo's correction, I found out I made some mistakes in the dimension of the final beam size we need on the 2 inch mirror and the FI.

The right dimension should be about 1mm when it arrives at the 2 inch mirror and a little bit larger like around 1.4mm at the FI, so the beam waist should almost at the mirror.

With these new conditions, I found some combination, chose the best one in the size and position, and check the position of the lens not overlap the mirror position we already mounted on the bench. The result shows as below.

The origin is set at the front surface of the BS mirror as I mentioned in my last entry.

L1: z=48.5cm f=500mm(already mounted)
L2: z=173.5cm f=300mm
L3: z=236cm f=350mm

With these lens, the final beam size at the FI is 1.368mm and the size of the beam waist is 0.97mm at the position of 648.57cm with the mirror's position at 648.15cm.

Also check again with JamMT, it shows the same result with the calculation.
Images attached to this report
442_20170405085811_telescopefin.jpg
R&D (General)
Print this report.
YuefanGuo - 20:29, Tuesday 04 April 2017 (441)Get code to link to this report
The telescope of infrared design

We measured the infrared beam size last Friday and calculated the beam waist size and position:(set the front surface of the BS mirror as origin) the position of the beam waist is z=-12.0623 cm, size is 98.0277 um.

Then we measured the distance between the origin and FI in the PR chamber, and between the origin and 2inch mirror in the BS chamber.

The calculation process is as below:

The distance between the O and FI(z1)=277.5(The path length on the bench)+63.65(From the last hole of the bench to the window)+13.5(the depth of the viewport)+26.5(the distance between the chamber wall and the first mirror)+5.5(the length between and the FI)=386.65cm

The distance between the O and mirror(z2)=z1+16(the FI length plus the distance between FI output and the second mirror)+16(the second mirror to the dichoric mirror)+37(dichoric mirror to the chamber wall)+162(the distance between two chambers)+15(the BS chamber wall to the chamber bench)+57.5(the edge of the bench to the mirror)=648.15cm

The first lens is already put with focal length of 500mm at 48.5cm. Then there are other conditions need to be satisfied.

1.The beam waist should be around the 2inch mirror in BS chamber, and the size should be less than 103 um

2. The size of the beam at FI should also less than 103um.

3. All the lens should on the bench which means should located in the position z<=270 cm, because the last mirror is at 270cm.

I am trying to find a good combination with other two lens, but until now I did not get any result, I will try with more combination or I will add one more lens. Also today I measured the green beam path, but actually I did not know if I should trust the measure result or not, since the green beam is almost the same size or even larger than the size of the beam profiler, so I will try to put another lens first and try to measure it again.

R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 01:07, Tuesday 04 April 2017 (440)Get code to link to this report
Beam injection in the south arm

Participants: Eleonora, Manuel, Yuefan, Raffaele

Today we have worked on the injection of the beam, exiting the telescope, in the south arm cavity.

1) We have moved the 2" mirror telscope in the BS chamber 5 cm closer to the PR in order to have a distance of 272.5 cm between the two telescope mirrors. 

   distance  = f1 + f2  =  272.5 cm  where  f1 = 302.5 cm (see entry 419)  and f2 = -30 cm  (see entry 431)  

2) We have used a spare TAMA PR mirror (ROC of about 9 km, almost flat)  placed on a tripod, as steering mirror to send the beam in the corridor beside the pipe in order to check its features. We manage to send it into the end room and obseved its shape after 300 m. We remark that:

     a)  The beam is well collimated as it keeps roughly the same dimension after 300 m.

     b)  The beam moves a lot because of the air 

3)  We injected the green beam through the first viewport before NM2 chamber and tried to observe it on a target placed at 10 m,  without success. We changed the beam direction moving the yaw and the pitch of the last telescope mirror.

4) We injected a red beam in the viewport and we were easily able to make it impinge on the target at 10 m. We observed that, in order to be centered both on the target and on the viewport the beam has to start from a position about 4 mm higher than the green beam position (21.3 cm from the stack top plate)

Conclusion: It is not clear the reason why we cannot see the green beam on the target at 10 m. The only two possible reasons we could imagine are that we were not able to properly align the beam or that it is not enough transmitted by the viewports or by the temporary mirror installed in the NM2 chamber (a former LISM mirror ). Effects of a possibile antireflective coatings for the infrared light on the green light should be investigated.

Attached pictures:

1- Green beam at 150 m

2- Green beam at 300 m

3- Red beam impinging to the target at 10 m

Images attached to this report
440_20170403175024_img20170403184921.jpg 440_20170403175102_end.jpg 440_20170403175222_img20170403175947.jpg
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 00:05, Friday 31 March 2017 (439)Get code to link to this report
PR spectra: possible evidence of intermediate mass touching the magnet holder plate

I have measured transfer functions and spectra for the PR mirror.

- Figure 1: TF when injecting noise in YAW

- Figure 2: TF when injecting noise in PITCH

- Figure 3: spectrum YAW (comparison open and closed loop)

- Figure 4: spectrum PITCH (comparison open and closed loop)

NB. It seems that there is some acoustic noise: I did the measurement with the light switched off but the chamber was open and I didn't put any boxex around the laser and the PSD

While looking at the error signals in the past days I have noticed that often some strong oscillations at 1 Hz and 3 Hz were present (see figure 5). We observed it also in the past and we were suspecting the stack vibratiion introduced by the steering mirrors used for the optical lever ( see entry 276:  the attached spectra taken last july are very similar to those of fig 5)

Since this oscillation is not present when the intermadiate mass is well centered on the magnet older plate (as it was just after the mirror installation or today after that I performed the check on the yaw motion) and it appeared after we tried to move as much as possible the yaw picomotor yesterday, I think it is caused by the intermediate mass touching the magnet holder plate.

Images attached to this report
439_20170330165540_noiseyaw.png 439_20170330165646_noisepitch.png 439_20170330165750_yaw.png 439_20170330165843_pitch.png 439_20170330165920_03.png
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 18:06, Thursday 30 March 2017 (438)Get code to link to this report
Investigation on the PR yaw motion

In order to better quantify the yaw motion of the PR suspended mirror I made the optical lever reflected beam imping on a ruler put at a distance of about 88 cm from the mirror.(See attached picture 1)

I started from one side of the picomotor range (left column of the second attached picture) and I kept turing the screw as far as I was able to see a beam movements on the ruler. 

I had to stop well before that the end of the range of the picomotor (right column of the second attached picture) since the beam was not moving anymore (and the mirror was oscillating a lot)

Even if it difficult to clearly see it, I think that the intermiate mass was touching the magnet holder plate preventing a further motion.

The total motion of the beam on the ruler is about 1.5 cm which corresponds to 8.5e-3 rad.

Images attached to this report
438_20170330110126_setup.png 438_20170330110426_finecorsa.png
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 00:54, Thursday 30 March 2017 (437)Get code to link to this report
Picomotor driver issue

After the problem reported yesterday here.  I have replaced the picomotors GPIB driver with a similar one that I have found in the south end room.

At the beginning it seems to work fine and I was able to move the 3 degree of freedom of the 2'' telescope mirror in both direction both with the pad that using the PC.

Then, I have noticed that while connecting more than one cable to the driver output ports,  it starts showing some problems. 

Its behaviour becomes very strange and it is no loger possibile to move properly the selected channel neither from remote nor with the pad.

For exemple: if I send a remote command to move a specific channel the driver display shows correctly the name of the selected channel but it cannot move it and will instead make other picomotor ( connected to a different channel) to produce some noise.

I could not find any repetable pattern in such kind of strange behaviour while trying to test different combination of slots and cables. 

I only noticed that everything is working fine when only one cable is connected.

Since Raffaele suggested that it can be a cabling problem, I spent some times to check the connections but I haven't find anything wrong so far.

In the attachement there is a picture of the connection we made between the 2" telescope picomotor and the invacuum cable.

Images attached to this report
437_20170329175317_img0863.jpg
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 00:28, Thursday 30 March 2017 (436)Get code to link to this report
telescope installation and PR yaw motion issue

Participants: Matteo Tacca, Eleonora

Today we installed the 2" telescope mirror in the BS chamber. The infrared and green beams have been superposed. They impinge the first telescope mirror and are reflected on the second one (PR suspended mirror). Beams are both hitting the PR mirror almost in the center. We observed that the green beam is slightly elliptical (also before hitting the first telescope mirror)

Then we tried to send the beams back on the BS mirror but we didn't manage to move the yaw of the PR mirror enough to do that.

We spent some time investigating this issue.

Some remarks:

1) The PR mirror is very sensitive to pich motion: few turns of the pitch picomotors screw are enough to remarkably change the vertical angle of the beam.

2) The yaw picomotor screw works fine as we can observe the relative slide to move (in both directions) when turning the screw.

3) We checked that the mirror is free to move and it not touching the earthquake stops

4) It was possible to observe the pitch motion using the error signal of the optical lever. The displacement is big enough to make it sweep the all the signal range  (from -5V  to +5V). Assuming a calibration of 2.2e-4 rad/V this means that the yaw motion achievable is al least larger than 2.2e-3 rad. 

5) Another critical point is the intermidiate mass. During the mirror istallation we noticed that one side of it was almost touching the magnet holder plate. We cannot rule out the possibility that the yaw motion is limited by this.

Images attached to this report
436_20170329172501_img0851.jpg 436_20170329172530_img0854.jpg 436_20170329172611_img0856.jpg
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 00:02, Wednesday 29 March 2017 (435)Get code to link to this report
BS suspension moved and picomotor control issue

Participants: Matteo Tacca, Eleonora

Today we moved the BS suspension in order to have the reflecting surface of the mirror 10 cm south with respect to the center of the chamber, according to the telescope design.

We have also installed the fixed mirror support and connected the in-vacuum and out-of-vacuum cable to move the picomotors. We had some issues in moving the pitch and the translation picomotors in one direction (the other direction works fine).

The picomotors seems to receve signals from the drivers since they make a noise but most of the time they don't move and sometimes they move very slowly.

NB: Ony few days ago I I have checked that the 3 picomotors can work in both directions (entry 420)

We made many checks in order to find out the cause problem:

1) We checked carefully all the connections in and out of the vacuum chamber. They were all fine.

2) We changed the picomotor with a spare one but it didn't solved the problem. So it is not a problem of the picomotor itself.

3) We have tried to control them by PC using labview: nothing changed.

4) We have changed the pad of the driver: nothing changed.

5) We have changed the slot at which the cable was connected in the driver: nothing changed.

6) We have inverted the pins of pitch and yaw in the in vacuum connector and this time the pitch was working fine in both directions while the yaw was now showing the issue. 

This last test made as to suspect the driver which, by the way, showed some other problems ( i.e the knob to change the controlled slot gets often stuck and it is necessary to switch it on and off).

I found a similar driver in the instrument shelf at the first flor but it didn't switch on.

Tomorrow we will:

1) Check the driver on other picomors, as those of the PR suspension which were also succesfully tested before.

2) Look for other GPIB picomotor drivers in TAMA. (NB according to their label these drivers have been bought in 1999)

R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 00:02, Wednesday 29 March 2017 (434)Get code to link to this report
Optical lever restored on PR telescope mirror

Today I restored the optical lever of the telescope suspended mirror in PR chamber. Error signals look fine even if, since the chamber is open, they are a bit noisy than usual due to the light.

In order to close the loops I had to increase a bit the gain of the yaw loop and change both the loop signs. I have also tweaked the driving matrix to reduce the exicitation of yaw when injecting noise on pitch. In the previous configuration I have added a rotation of the sensing of 0.04 rad (as reported here), now I have removed it since it seems to be not needed anymore.

I will upload soon spectra and transfer functions.

R&D (General)
Print this report.
EleonoraCapocasa - 00:00, Wednesday 29 March 2017 (433)Get code to link to this report
Leaking pipe issue in the central building

Today while tiding up the area close to the BS chamber I noticed that there was same air leaking from a pipe between BS and NM2 chamber (see attached picture).

As soon as I touched the pipe tying to tight it better, it came off, loosing a lot of air. With the help of Matteo, Fujii and Ishizaki-san we managed to reconnect and tight it. Now the pipe seems well attached but there is still some leakege even if it is smaller than it was before. 

It has to be checked by an expert as soon as possible.

Images attached to this report
433_20170328154300_45.png
R&D (General)
Print this report.
YuefanGuo - 11:57, Tuesday 28 March 2017 (432)Get code to link to this report
Alignment of the beam inside the PR chamber
Last Friday and yesterday we did the alignment in the PR chamber and also the beam path from PR chamber to BS chamber. The mirror is still in almost the same position as the picture I showed last time. We adjust the green and infrared beam surpassed until the BS chamber.

Then install the reflect mirror just after the dichoric mirror temporarily to adjust the reflect beam of infrared to goes the position it should be at the optical bench as the new optical scheme.

We also check the power loss of the beam goes through the Faraday and comes back to the optical bench, it is about 6% and 10%.

The we put back the middle part of the PR chamber, since then the beam will first go through the thick window mirror, if it is not exactly perpendicular to the mirror, there is at least some shift of the beam. And also it seems the mirror also is a wedge mirror, now the beam in the chamber moves a lot. One of the idea is to move the last mirror before the beam goes into the chamber on the bench, but then there is the risk that the beam will be move out from the mirror. Matteo will work on this in these days.

And on Friday, we plan to measure the green beam, and do some calculation to decided the lens we are to use, and we should buy them, since now in the lab the shortest focus length is 100mm.

R&D (FilterCavity)
Print this report.
RaffaeleFlaminio - 22:08, Monday 27 March 2017 (431)Get code to link to this report
Measurement of the telescope convex mirror ROC
In order to measure the ROC of the convex mirror I proceeded as follows:
I placed a lamp at a distance D from the convex surface and I looked
for the image reflected by the second face of the mirror (which is flat).
The image is real and placed at a distance D' from the convex surface.
I changed D until D'=D. In this situation the focal length f is
f = D = D' = 1.305 +/- 0.01 m
I can deduce the ROC from the expression ROC = f * (n-1) where n is the
fused silica index of refraction in the visible (green) i.e n=1.4607 at
lambda = 532 nm. I found ROC=0.601 +/- 0.005 m.
In conclusion, within the error of the measurement the mirror ROC
is equal to the required value (which is 0.6 m).
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 12:31, Saturday 25 March 2017 (430)Get code to link to this report
Magnets and standoffs glued on BS telescope mirror

This morning I have removed the supports used to glue the standoffs on the BS telescope mirror that I glued last wednesday evening. For the moment, they look well attached to the mirror.

The magnets were glued on tuesday and their supports were removed on wednesday in order to rotate the mirror and allow for the glueing on the standoffs.

Since BS mirror has a different disposition of magnets with respect to test masses, the glueing has to be done in 2 times.

http://www2.nao.ac.jp/~gw-elog/osl/uploads/260_20160627054235_mirrormagnet.jpg

Both magnets and standoff have been glued using a new set of MasterBond EP30-2, stored in the fridge of TAMA.

Images attached to this report
430_20170325043251_p3250291.jpg 430_20170325043313_p3250289.jpg 430_20170325043325_unnamed.jpg
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 11:48, Friday 24 March 2017 (427)Get code to link to this report
What if the optical lever beam is not impinging on the center of the mirror

In the optical lever scheme, a beam is sent to the center of the the mirror with an angle α.

A rotation of the mirror around its vertical axis (a yaw) of an angle β, corresponds to an angular displacement of the reflected beam of 2β.

This produces a shift of the beam's position on a PSD placed at a distance L from the mirror of

X = L tan (2β)  2Lβ

Now, if the beam impinges (with an angle α)  at a distance d from the center, the shift on the PSD becomes 

Xd =  L tan (2β) + d (sinβ/cos(α+β))  2Lβ + d β (1/(cosα βsinα))

In the case of the beamsplitter, the optical lever beam is currently sent with a very small incidence angle (α ≈ 0)  and impinges at a distance d (in the orizontal direction) of about 3.5 cm from the mirror center.  The distance between the mirror and the PSD (arm of the optical lever) is about 60 cm.

The relative errore due to the mis-centering is   (Xd-X) / X = d/(2L) = 3%

which is small. So it seems not so crucial to hit the center of the mirror. 

For what concers the pitch, I would say that as long as the beam is on the orizontal axis (as in our case), its mesurement is not affected by the mis-centering.

 

KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 17:45, Thursday 23 March 2017 (426)Get code to link to this report
HEPA filter check for replacement

I disassembled one of the two HEPA filter of the cleanbooth for the absorption measurement system ( Model MAC-103 ) in order to check the conditions and to figure out whether and how to change it. The circular prefilter is very easy to replace and very dirty, I think we should change it. The real HEPA filter inside is quite dirty but is not easy to replace because the filter is kind of glued at the frame, so maybe we can contact the company. I hope there is the possibility to change only the filter, and not the whole fan.

Images attached to this report
426_20170323094129_circularprefilter.png 426_20170323094137_hepadarkside.png 426_20170323094151_hepacleanside.png
Comments related to this report
ManuelMarchio - 16:11, Monday 19 June 2017 (511)

The filter's fan part was fine, so it was enough to buy only the filter part. New HEPA filters were delivered to Tama last friday.  I washed the prefilters with water, I cleaned the fans, replaced the filters and placed them back to the top of the absorption bench clean booth.

R&D (General)
Print this report.
YuefanGuo - 20:12, Wednesday 22 March 2017 (425)Get code to link to this report
Low efficiency of the green
We tried to do the better alignment today. While we were adjusting the EOM, we found out the EOM has better transmission rate when the beam inject near the edge. Another problem is that the beam path seems tilt somewhere, since we aligned all the mirror in the green path last Friday, but now it does not work anymore.

And we could not kill one higher order mode to get better mode matching, and also there is one mode jump out from time to time when we try to align the cavity. The final efficiency we got is about 20%(260mW goes into the cavity and 55mW green received at the transmission of the dichoric, 12mW received by the reflection of the dichoric)
R&D (General)
Print this report.
YuefanGuo - 23:49, Tuesday 21 March 2017 (424)Get code to link to this report
Found the possible reason for high frequency oscillation
The first check we did today is to install a PD on the after the first BS mirror to look at the infrared beam came out from the laser,so the first picture is the error signal when the output power of the laser about 500mW, and the second is the signal from the new-set PD which shows that the infrared signal from the laser somehow also has the high frequency oscillation. And the situation are same with or without the green.

Then we increased the power a little bit, do the alignment again, then the error signal and the infrared signal shows in the third picture. There is a flashing spot in the infrared signal.

But the changing part is when we lock the cavity, the 500mW situation has the oscillation also(picture 4), but the second seams not.(pic 5)

According to this circumstance, we at least got one conclusion is that the high frequency oscillation has nothing to do with the green.

Then we rethought the polarization again, found out no matter s or p polarization reflect back, it is should stopped by the Faraday Isolator, but now it seems not.

We decided to check the polarization again from the beginning. Firstly one thing for sure is that, the input infrared beam to the cavity is in p polarization, and the reflect infrared beam is also p polarization.

Then we try to put another FI and half wave plate after the BS mirror and before the telescope to cancel the reflect infrared beam from the cavity. We use the same FI as we used before, but found out the two polarization cube point to different direction of the original one. Then the new put FI did reflect out the reflection beam, and get rid of the oscillation.

Then we guess that the original FI did not do its job maybe because the two wave plate before it did not clean the polarization as they should do.So we put a PBS after the first quarter wave plate and second wave plate, found out exactly what we suspected, so we adjusted them firstly then the Faraday and the other two half wave plate, got pure p polarization inject into the cavity.

Now the oscillation seems not there but the efficiency of green beam is only about 10%, so we tried to increased the alignment, but it seems we should do it from the start. After we make a progress in the alignment, we can check again if the oscillation really disappear. Then we can be sure that it is caused by the polarization problem.
Images attached to this report
424_20170321152735_500withgreenerr.jpg 424_20170321152753_500withoutgreenin.jpg 424_20170321152826_gt500withgreeninerdetail.jpg 424_20170321152840_500withgreenlock.jpg 424_20170321152855_gt500gtalignlock.jpg
R&D (General)
Print this report.
YuefanGuo - 01:40, Tuesday 21 March 2017 (423)Get code to link to this report
Beam goes into the chamber
The table leg arrived on Friday, so we install them under the original leg, increase the bench to the height we want. Show in the first and second picture.

After doing this, we remove all the tall mounts for infrared and do the alignment again both in the infrared and green path. During the alignment, we found out maybe the cavity itself is a little bit tilt up, with one mirror we recover the beam to the right height.

Then we inject both the green and infrared beam into the chamber.

Firstly, adjust the position of G1 mirror in the picture try to get the green beam in the middle of it in horizontal direction. Then use adjust it slightly to get the reflection beam goes to the GR mirror.

Then we adjust the infrared beam, use the last two mirror on the bench in infrared path to adjust the infrared go through the Faraday Isolator, reflect by the R2 mirror and also goes to the GR mirror.

Then after the GR mirror we try to do our best to overlap the infrared and green beam and get a good result, next step is let these two beams goes into the BS chamber and try to overlap them there.

Then we installed the RP1 mirror, this is a mirror with high reflection with 0 zero degree. Then also install the RP2 mirror, this is the mirror has high reflection in P polarization. With this two mirror we got the p polarization reflection of the infrared beam and it arrives about the right position in the bench.
Images attached to this report
423_20170320173911_img2514.jpg 423_20170320173929_img2515.jpg 423_20170320173945_img2519.jpg