LOG-IN
Displaying reports 2021-2040 of 3201.Go to page Start 98 99 100 101 102 103 104 105 106 End
R&D (FilterCavity)
Print this report.
YuhangZhao - 10:41, Tuesday 11 June 2019 (1405)Get code to link to this report
Alignment of the amplified BAB into filter cavity

EleonoraP, EleonoraC and Yuhang

We achieved the matching of the BAB (IR probe beam co-aligned with squeezing) into the filter cavity.

In order to align BAB into the FC, we mainly followed the IR alignment procedure we did more than one year ago (entry #646). The main steps of this procedure are to align the cavity for the green, recover the reference on PR chamber, maximize IR reflection from the input mirror, not to look for the beam on the first target and check instead the second target and the end camera.

However, since BAB is only 250uW, it is too weak to see. So we used the green pump to amplify it. The maximum green we can give is obtained setting the offset on the MZ control servo at 4.3 and at that time we have roughly 30mW of IR going to FC. This measurement of power is done while we scanned the green pump phase. So actually we are sending a repeated segment of a sine wave.

By using this 30mW and following the procedure above, we found the flashes of IR. Check attached video.

As you can see in the video, this beam is too bright so, in order to avoid saturation, we decide to reduce the power to around 10mW. In this case, we need to use temperature of 7.202 for OPO, MZ offset of 4.1 and lock p-pol PLL on 150MHz.

After achiving the alignment with the amplified BAB we confirmed that flashes cannot be seen without amplification. (BAB power ~250 uW)

Next steps are to optimize the matching and to use AOM to make green and IR both resonant.

Non-image files attached to this report
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 20:31, Monday 10 June 2019 (1404)Get code to link to this report
FC GREEN transmission into DAQ

I used one of the few ADC channels still availabe to acquire the filter cavity green transmission. Channel name is K1: FDS-FC_GR_TRA

For the moment it is recorded as an epic channel, since I wanted to display it on the medm screen.

The attached picture show today's lock streches (Even it doesn't seem so, no realigment has been done during this period and the lock was stopped on purpose)

Since the trasmission PD is actually a PSD, I plan to acquire also the X and Y signals in order to  possibly correlate beam motion to suspesion resonances and understand what we should improve.

Images attached to this report
1404_20190610131148_greentra.png
General (General)
Print this report.
EleonoraCapocasa - 19:16, Monday 10 June 2019 (1403)Get code to link to this report
Water leakage from the ceiling of the north-south arm in the mid-arm station

As entitled. 

It is probably due to the heavy rain of these days. You can check the video. We put a bucket to collect the water.

Non-image files attached to this report
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 18:35, Monday 10 June 2019 (1402)Get code to link to this report
USB PCI-Express also doesn't work on new DGS computer

As reported in entry #1267 we could not setup the new DGS computer wtith gentoo linux HDD, as it was not able to detect the USB.  We bought a PCI express board to be used as USB driver.

Last week I installed it but I still couldn't make the PC to detect USB. I think the board might not be compatible with linux. We should look for another one or consider to change PC.

Images attached to this report
1402_20190610121722_5d0d5683b2594d20bfccab0a220250cd.jpeg
R&D (FilterCavity)
Print this report.
EleonoraPolini - 17:42, Monday 10 June 2019 (1400)Get code to link to this report
Installation of IR injection telescope

[Eleonora P, Yuhang]

We installed the IR injection telescope on the bench and we pre-installed the reflection telescope (putting all the optical mounts on the bench without fixing them).

We used the two 750 mm focal length lenses present in the lens box in the lab. Since we saw the beam focusing before the cavity, after the telescope, we thought one of the two focal lengths to be wrong. In fact the second one was actually wrong and it was replaced by another lens of 750 mm focal, located in the lab (Newport KPX121AR.18, PCX LENS, BK 7).

We put HR Laser Line mirrors, BK 7, ALTECHNA Co. Ltd.

In fig 1 the photo of the injection telescope on the bench. We used two of the new small mirror mounts.

Images attached to this report
1400_20190610104134_injectiontelescopeinstallation.jpg
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 17:15, Monday 10 June 2019 (1399)Get code to link to this report
Effect of rotary pump vibrarion reduced on BS

We notice that the pipe of the rotary pump close to BS was directly touching the vacuum pipe, transmitting a lot of vibrations.  We have moved the pump of  few centimeters so that the pipe was not touching anymore.

The vibration of BS in the region of 20 Hz seems improved. (See pic 1)

Images attached to this report
1399_20190610101532_screenshotfrom20190607171827.png 1399_20190610103319_4f21c91cfafd4a168faa8e31c7c4a57d.jpg 1399_20190610103327_6e52e4be3bab4eb09dec5db9682b2e08.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 21:23, Friday 07 June 2019 (1398)Get code to link to this report
mirror check for squeezing path

EleonoraP and Yuhang

Today we checked the PBS and the two following mirrors for OPO transmission. PBS has a transmission of 0.2% while we measured reflection more than incidence.  And the two mirrors are fine as well. We didn't use wrong coating components or they have some issues.

R&D (FilterCavity)
Print this report.
YuhangZhao - 10:58, Friday 07 June 2019 (1396)Get code to link to this report
The measurement of squeezing before and after cleaning optics

Chien-Ming and Yuhang

After the last very good measurement of squeezing, we start to think about improving coherent control 1 PD. Then we measured squeezing again. The result is attached in figure 1 and 2. As you can see the low frequency is covered by classical noise while high frequency seems fine. However, the squeezing level was only 4.9dB at that moment.

Then Chien-ming suggested cleaning optics. And then we checked optics together. We found optics are tremendously dirty. Especially one of the lenses is very dirty on one side of the surface. And for others, there are lots of dust on the surface of the mirror. However, after the cleaning of optics, the balance of BS is destroyed. After the alignment of this BS, we measured squeezing again. However, we didn't improve the squeezing. We observed squeezing level almost the same with last time. See attached figure 3 and 4.

There is some consideration after this measurement:

1. The fluctuation of measurement is about 4dB pk-pk. Is this coming from the measurement device?  

2. The low-frequency noise of the second measurement of squeezing is much higher than the first one. We also know that for the second measurement, the balance of the SQZ part is very bad. The unbalance signal on PD is about 50mV. Although we know this balance is not as important as the balance of local oscillator, we still need to make it stay at a decent level.

3. To keep a clean clean room is really important for squeezing to avoid some potential losses.

4. We also tried to change coherent control power, but we noticed that after we decrease coherent control power we may also increase phase noise. This measurement is better done when we have a better coherent loop. Also with some characterization of the coherent control loop.

Images attached to this report
1396_20190607130347_wholef.png 1396_20190607130352_highf.png 1396_20190607130405_wholef.png 1396_20190607130418_highf.png
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 23:22, Thursday 06 June 2019 (1394)Get code to link to this report
Suspensions spectra with vacuum pump off in the central area

[Yuhang, Eleonora]

Today we temporarily swiched off both the rotary and turbo pump in the central area (close to BS).

The attached plot show the comparison of the spectra of the suspensions in the central area, with pumps on and off.

We see that BS is strongly affected by the pumps vibration while PR and INPUT seem not.

[Note that at the time when I developed local control for the first time, the pump where swiched off].

The pumps have been switched off for about 3 hours ( from 5PM to 8PM).

Images attached to this report
1394_20190606125646_vacuumpumpoff.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 22:26, Thursday 06 June 2019 (1395)Get code to link to this report
Some number for filter cavity locking in June 2019

Yuhang, Aritomi, EleonoraP, and EleonoraC

On 5th of June 2019, we locked the filter cavity again after almost half a year we haven't done this. We checked several references including irises on the bench, target on the film of PR chamber, target on the film of BS chamber, the first and second target inside the filter cavity vacuum tube. To recover these checking points, we used pico-motors with almost no failures. But it seems there are still some problems with end mirror pico-motor. Also, we observed some large mirror motion after we moved pico-motor of BS. But it went back to static after several minutes.

The alignment is also done with the standard procedure. It is first to misalign input mirror and find filter cavity transmission. Then center it on the screen. Then bring back input mirror and make injection and reflection overlap. Finally, align end mirror to have TEM00 dominant. After the lock, change the control point of each mirror to see if the transmission goes up to maximize.

Power:

Before AOM 36.5mW
Before EOM 212mW
Before MZ 166mW
To filter cavity 34mW
Before FC GR ref PD 136uW

Signal:

FC tra PD 4V
FC GR ref PD DC 80mV
FC locking error signal 260mV

The demodulation phase for filter cavity locking is 65deg.

The setting of FC locking servo:

Input attenuation 3.6
Gain 1 4.0
Images attached to this report
1395_20190606152639_wechatimg428.jpeg 1395_20190606171052_fcpdtragreen.jpeg
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 19:36, Thursday 06 June 2019 (1393)Get code to link to this report
DAC recovery after death

Today suddenly DAC died. Since I was working around the DGS I supposed I accidentaly disconnect some cables, but I check carefully and everything seemed fine.

We rebooted the standalone computer and the problem was solved. Note that in order to switch it on we had to disconnect and reconnect the power cables and to disconnect the timing singnal (pic 1). Timing cable was reconnected after the restart.

Images attached to this report
1393_20190606123552_timing.jpeg
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 19:27, Thursday 06 June 2019 (1392)Get code to link to this report
HUGE 50 Hz on END signal after connecting picomotors driver to the power line

We observed that connecting end room picomotor driver to the power line in the same mupliple socket of OPLEV laser and PSD brings a HUGE 50 Hz in the signals.

Since picomotors are only used temporarily we have now disconneted them. We may try to connect them to another socket next time that we need them.

Comments related to this report
Federico Paoletti - 20:09, Saturday 15 June 2019 (1412)

You could use some remote "smart" PDUs as done in virgo.


See as example

https://tds.virgo-gw.eu/?content=3&r=15139

(the name of the PDU is ENERGENIE EG-PMS2-LAN, see attached PDF file)

R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 19:11, Thursday 06 June 2019 (1388)Get code to link to this report
Local control for END mirror with new DGS

Local control on END mirror were implemented.

Pic 1 Pitch TF

Pic 2 YawTF

Pic 3 Comparison between open and closed loop spectra.

Pic 4, 5 photon model and damp for pitch  

Pic 6, 7 photon model and damp for yaw

I found a minimum of the sensing coupling when input signals (pitch and yaw) are rotated of -0.06 rad. So I updated accordingly the rotation matrix.

I used the following driving matrix  (pitch, yaw -> coils)   [0, 1, 0, 1; 1 0 -1, 1].  Note that the coil disposition is not the usual one!

For a comparison with the old control check entry #238.

Images attached to this report
1388_20190606121047_endpitchtf310519.png 1388_20190606121052_endyawtf310519.png 1388_20190606121117_endspectra300519.png 1388_20190606121558_endpitchdamp.jpeg 1388_20190606121607_endpitchmodel.jpeg 1388_20190606121615_endyawdamp.jpeg 1388_20190606121622_endyawmodel.jpeg
R&D (FilterCavity)
Print this report.
NaokiAritomi - 18:55, Thursday 06 June 2019 (1391)Get code to link to this report
Power of BAB transmission from filter cavity
 
I roughly calculated power of BAB transmission from filter cavity. When the injection of BAB is 200uW, transmission power at PD is 1.2uW. If we use parametric amplification, we can increase it by a factor of 20. Here is the assumption.
 
BAB injection: 200uW
Injection loss: 10%
Transmission of filter cavity: 1%
Transmission of IR line filter in End bench: 70%
BAB transmission at PD: 1.2uW  
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 18:04, Thursday 06 June 2019 (1390)Get code to link to this report
Scattered light and vibration of chamber

Today I did alignment work to see the flash of transmitted beam using PD.
I though that PD detected transmitted beam, but it turned out that it was scattered beam...

Then I re-aligned so that the reflected beam reach to FI, and scanned laser frequency with monitoring reflected beam.
At that time, the reflcted beam power was fluctuated.
Actually, it was caused by vibration of cryostat chamber.
So far I have used long pedestal to hold silicon mirrors, and this may affect the vibration of cavity.
Using fixed spacer may reduce the vibration effect, but it might be better to consider about vibration isolation system.

Images attached to this report
1390_20190606105114_img9334.jpg
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 20:06, Wednesday 05 June 2019 (1389)Get code to link to this report
Install PD and alignment work

I installed a PD for monitoring transimtted beam of optical cavity with one STM. (forgot to take a photo...)
Though the transmitted power is very weak, the PD detects the transmitted beam.

Then I did alignment and scanned the laser frequency.
However, I have not seen any fringe so far.

I will continue to align the beam with scanning laser frequency and try to find good alignment.

R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 13:00, Wednesday 05 June 2019 (1387)Get code to link to this report
Suspensions behavior during EQ

Taking advatage of the the DAQ I playbacked the data and check the oplev signal during a 4.9 Earthquak occured in Chiba prefecture, last saturday.

https://earthquaketrack.com/quakes/2019-05-31-22-58-08-utc-4-9-20

The EQ is cleary seen by all the sensors. Some dof ( in paricular  PR and END pitch) didn't go back to the original position.

The data on DAQ are referred to UTC but the time of standalone PC is not well sincronized (about 3 min ahead the correct one).

Images attached to this report
1387_20190605060035_eq310519.png
R&D (FilterCavity)
Print this report.
ChienMingWu - 01:13, Wednesday 05 June 2019 (1386)Get code to link to this report
Improvement of 14MHz PD of coherent control

Chien-Ming, Yu-Hang,

We want to reduce the coherent noise coupled into the homodyne by reducing the intensity of CC beam entering the OPO. However, this action will cause the demodulation 14 MHz error signal (for coherent control of the pump beam phase) to become smaller. So, we tried to increase the gain of the photodetector at 14MHz by changing the capacitor CX and the amplifier resistors R1, R2, see Fig.1.

We let the capacitor CX become a plug-in base and test some combinations of capacitors. Currently we find the biggest signal at 14 MHz when using CX as 150pF+33pF. What is puzzling is that when we replace this combination with a single 180 pF capacitor, the signal is reduced by 8 dB on the spectrum analyzer. So, we are now using 150pF+33pF, and R1=500 ohms R2=5k ohms. The S/N now is improved 5 times more as shown in Fig.2.

Images attached to this report
1386_20190604181309_114mhzpd.png 1386_20190604181313_2snisimproved.png
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 19:20, Monday 03 June 2019 (1384)Get code to link to this report
Silicon mirror installation and rough alignment

I installed another silicon mirror inside the cryostat.
Then I did rough alignment of input beam with 2 STMs and the reflected beam reached to FI.

Tomorrow, I will scan the laser frequency and try to find transmitted HOMs.

Images attached to this report
1384_20190603121629_setup.jpg 1384_20190603121633_silicon.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 15:30, Monday 03 June 2019 (1383)Get code to link to this report
Some activity of coherent control PD

Yuhang and Chien-Ming

As we said before, we need a better PD for the coherent control1 lock. (PD set at OPO reflection) So we talked and tested several things.

1. Change the amplification factor of the op-amp. We tried two combinations of (500Om, 50kOm) and (500Om, 7.5kOm).  For the first combination, we tested the signal, but the amplitude and SNR are the same as before. While for the second combination, the amplitude is smaller while the SNR is a bit better.

We checked the spec of AD8057. The maximum gain we should give for this op-amp is 10. The further increase of the gain will degrade the bandwidth. And this is exactly what we see for increasing further the amplification factor.

It seems that we are also reaching the saturation current of the op-amp.

2. We also tried to use the RF amplifier, because maybe there is an optimal input level for the mixer to make it work better. But it seems RF amplifier increases both signal and noise.

3. We also tried to change the capacitor of the PD circuit. But we are still trying.