LOG-IN
Displaying reports 1241-1260 of 3201.Go to page Start 59 60 61 62 63 64 65 66 67 End
R&D (FilterCavity)
Print this report.
YuhangZhao - 23:41, Thursday 19 November 2020 (2285)Get code to link to this report
Comparison of Thorlabs PSD (with amplification) and HAMAMATSU PSD (without amplification) for INPUT mirror oplev

To see the possibility of using HAMAMATSU PSD for INPUT mirror oplev, I did the comparison between Thorlabs PSD (with amplification) and HAMAMATSU PSD (without amplification) for INPUT mirror oplev.

The comparison result is shown in the attached figure.

Upper figure: (for INPUT pitch) REF0 is INPUT oplev spectrum with Thorlabs PSD. REF 2 Thorlabs PSD electronic noise after amplification. REF 4 INPUT oplev spectrum with HAMAMATSU 04 PSD. Red line: HAMAMATSU 04PSD electronic noise.

Lower figure: (for INPUT yaw)REF0 is INPUT oplev spectrum with Thorlabs PSD. REF 2 Thorlabs PSD electronic noise after amplification. REF 4 INPUT oplev spectrum with HAMAMATSU 04 PSD. Red line: HAMAMATSU 04PSD electronic noise.

It can be seen that Thorlabs PSD has a bit better SNR. This better SNR is proven to come from the amplification (proved in elog2114).

Images attached to this report
2285_20201119155941_inputcomp.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 23:26, Thursday 19 November 2020 (2284)Get code to link to this report
HAMAMATSU PSD electronic noise and DGS ADC noise

As reported in elog2281, the HAMAMATSU PSD electronic noise level is similar with DGS ADC noise. In this entry, I report a futher investigation of them.

1. I had a closer look into these two noise spectrum. As shown in the attached figure 1 (RED is elec noise, BLUE is ADC noise, GREEN/BROWN are integrated noise), elec noise is a bit higher than ADC noise but not a factor of  2. Therefore, to investigate further the electronic noise, we should amplify signal from PSD before it goes inside DGS system.

2. Before amplification of PSD individual signal, I checked the time series of individual signal from PSD. This is shown in the attached figures 2. We could see that it has very large noise. Therefore, it is very diffcult to amplify such large noisy signal. Indead, I found SR560 could only give a factor of 2 amplification for it before saturation. But the good thing is that, after the combination inside DGS, this noise is cancelled and resulted in a clean pitch/yaw electronic noise.

So if we really want to check better the electronic noise, we should combine PSD individual signal before they go inside DGS system.

Images attached to this report
2284_20201119155856_04eleadc.png 2284_20201119170903_wechatimg42.jpeg
R&D (FilterCavity)
Print this report.
YuhangZhao - 23:07, Thursday 19 November 2020 (2283)Get code to link to this report
Comparing 03PSD and 04PSD

By replacing 03PSD with 04PSD for BS oplev, I compared their performance.

The measured oplev spectrums are shown in the attached figure 1 (RED 04PSD, BROWN 03PSD), while the electronic noise comparison with ADC noise is shown in the attached figure 2 (RED is elec noise, BLUE is ADC noise, GREEN/BROWN are integrated noise).

We could conclude that both 03PSD and 04PSD have almost the same performance. The difference in position resolution is not affecting their performance now. So I guess this position resolution maybe related with beam size.

Images attached to this report
2283_20201119155559_0304.png 2283_20201119155802_04eleadc.png
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 18:18, Thursday 19 November 2020 (2282)Get code to link to this report
Q-mass does work

I tried to measure the residual gas molecules by a mass spectrometer.
The Q-mass could be operated by a front panel, and the pressure was 3.0*10-5 Pa, 7.8*10-7 Pa for H2o and N2 respectively.

Hoever, I could not connect to PC, and could not do degas.
This Q-mass needs degas but in order to do that, the connection to PC is needed...

I will ask the company.

R&D (FilterCavity)
Print this report.
YuhangZhao - 23:23, Wednesday 18 November 2020 (2281)Get code to link to this report
HAMAMATSU PSD C10443-03 test

Since we have problem of TAMA PSD, we considered to buy new PSD. Matteo asked two PSD from HAMAMATSU company for test. They are C10443-03 (we call it 03 later) and C10443-04 (we call it 04 later). The information of them can be found from this link.

From the datasheet, 03 and 04 PSD only have difference in position. I think the integrated noise spectrum can represent a position resolution. If so, the one with better position resolution should have a lower noise spectrum. Therefore, I firstly tested 03 which has a better position resolution. If it is better than the old PSD, we should use it in the future.

Together with Matteo, we made a customized circuit based on Mammoth connectors, bananna connectors and lemo connectors. In this way, the eight channels from HAMAMATSU PSD can be connected to power supply and ADC of DGS system. Four channels from PSD are used for power supply, which contain one plus(12V), one minus(-12V) ,and two grounds. Another four channels are called x1, x2, y1, and y2 separatly. The experimental set-up is shown in attached figure 1.

To convert four signal channels into pitch and yaw, I realized a matrix inside simulink file. This matrix is enclosed in a block called 'BS_test', as shown in the attached figure2.

Since old measurements have been saved in DGS system, I just plot new PSD data with old ones. As shown in the attached figure 3, there is comparison between old low gain TAMA PSD and 03PSD. We can see

1. REF0/1 (BS angular motion sensed by low gain TAMA PSD) is comparable with REF16/17 (BS angular motion sensed by 03PSD). This means that two PSD have the same gain.

2. REF8/9 (electronic noise of low gain TAMA PSD) is lower than REF24/25 (electronic noise of 03PSD).

3. Red lines in this figure shows the ADC noise. Actually this measurement is a bit strange. It shows that ADC noise is even higher than the electronic noise of low gain TAMA PSD. This can be correct if the ADC noise really becomes worse by itself.

4. Since the Red lines overlap with REF 24/25, it means the measured electronic noise maybe just ADC noise. So if we want to check better mirror angular motion, we need to amplify the signals coming from PSD.

I also compared the electronic noise between high gain TAMA PSD with 03PSD. The result is shown in the attached figure 4. Red lines are 03PSD electronic noise, which is lower than REF8/9 (high gain TAMA PSD electronic noise). 

We will also test 04 PSD to justify the relationship between PSD position resolution and noise spectrum.

Images attached to this report
2281_20201118152805_hamamatsupsd.jpeg 2281_20201118152850_img20201118105454.jpg 2281_20201118153026_untitled.png 2281_20201118153031_untitled2.png
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 15:34, Wednesday 18 November 2020 (2280)Get code to link to this report
Q-mass does not work

I pumped down the cryostat and turned on the Q-mass.
The LED lamp next to "POWER" turned on, though that of "Pa" did not.
It seems that this Q-mass may not be used anymore.

Furthermore, in order to degas the Q-mass, we need to pump down below 10-4 Pa, though it cannot reach below 6*10-4 Pa...
I gave up the measurement.

Images attached to this report
2280_20201118073419_qmass20201118.jpg
R&D (FilterCavity)
Print this report.
MatteoLeonardi - 15:46, Tuesday 17 November 2020 (2279)Get code to link to this report
Comment to System recovery after Mitaka power outage (Click here to view original report: 2278)
Some additional information regarding the recovery of the TAMA vacuum system after the power outage.

1) Two vacuum sensors were found not working. They are END station (arm side) and MID station (TAMA central side). The controller seems to work properly but it provide a "FAIL" message when reading the sensors.

2) The rotary pump in MID station has what it seems an oil leakage (see picture 1 and 2). I did not noticed this leakage when shutting it down on Friday but I did not payed particular attention to it, so I cannot tell if the problem was already present.

3) I could not find where to switch on the air dryer of South Arm. The West Arm air dryer are working.
Images attached to this comment
2279_20201117074132_img20201116113347.jpg 2279_20201117074141_img20201116113406.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 15:28, Tuesday 17 November 2020 (2278)Get code to link to this report
System recovery after Mitaka power outage

Eleonora, Matteo and Yuhang

All facilities and devices were recovered after the Mitaka power outrage.

The Mitaka power outage was done on 14th Nov., while the recovery work was done on 16th Nov.. Although everything was recovered, there were some issues we encountered during the recovery work.

1. The main power switch for the filter cavity arm lights were not found at the beginning. The reason is that this switch is located in the middle of filter cavity arm (I didn't know this at the beginning).

2. The sound issue of DDS. The solution is provided in elog1794.

3. To power on DAC, there is a DC voltage supply. At the beginning, I forgot to provide negative 18V. Then, while I increased the positive voltage, the maximum can reach only around 6.5V.

4. This time, none DDS channels flipped phase by 90deg.

5. The optimal temperature of SHG has been changed from 3.074 to 3.102. On the second day, the value changed back to 3.081.

6. The position of suspended mirror didn't change a lot. We don't need to change largly voltage offset sent to mirrors to recover FC flash.

7. PDH signal of FC lock has large 5kHz resonance even when FC is unlocked. It was figured out that this oscillation comes from SHG. And the problem is solved by reduce the gain of SHG to 2.1.

8. It was found that the input value of DGS doesn't change on medm. We checked medm sitemap/CDS. The situation is shown in the attached figure 1. It was figured out later that this is because I was using sine wave instead of square wave. After correcting to the good waveform, DGS worked again. The good sitemap/DGS is shown in the attached figure 2.

9. It was found that ratory pump in the middle arm has oil leakage.

Images attached to this report
2278_20201117075250_wechatimg40.jpeg 2278_20201117075315_good.png
Comments related to this report
MatteoLeonardi - 15:46, Tuesday 17 November 2020 (2279)
Some additional information regarding the recovery of the TAMA vacuum system after the power outage.

1) Two vacuum sensors were found not working. They are END station (arm side) and MID station (TAMA central side). The controller seems to work properly but it provide a "FAIL" message when reading the sensors.

2) The rotary pump in MID station has what it seems an oil leakage (see picture 1 and 2). I did not noticed this leakage when shutting it down on Friday but I did not payed particular attention to it, so I cannot tell if the problem was already present.

3) I could not find where to switch on the air dryer of South Arm. The West Arm air dryer are working.
RyutaroTakahashi - 12:57, Tuesday 24 November 2020 (2287)

Oil under the rotary pump was there from the old days. I replaced the rotary pump to new dry pump (ACP15). The TMP with the dry pump is working now at the mid point.

R&D (FilterCavity)
Print this report.
NaokiAritomi - 21:13, Monday 16 November 2020 (2277)Get code to link to this report
Comment to SHG cavity scan (Click here to view original report: 2269)

I measured SHG cavity scan again. This time, I put SHG temperature 2.8kOhm to avoid green conversion while nominal temperature is 3.1kOhm. Peak shape is still not completely symmetric possibly due to high IR injection power, but measured finesse is 70 which is reasonable value. 

Images attached to this comment
2277_20201116131337_shgcavityscan.png 2277_20201116131342_shgcavityscan2.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 17:26, Friday 13 November 2020 (2276)Get code to link to this report
Compare TF/Coherence w/wo excitation

To project PR/BS angular motion to AA signals, we should use TF measured with excitation. But what is the difference between with and without excitation is an interesting thing to check. So I did this check. The result is shown in the attached figure.

It can be seen that magnitude of TF can be similar or very different when the coherence becomes good.

Images attached to this report
2276_20201113092920_01.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 15:43, Friday 13 November 2020 (2275)Get code to link to this report
Modification of TAMA RF PD and use it to check CCFC

Matteo and Yuhang

In elog2236, before CCFC demodulated, its SNR was reported to be about 40dB with taking ~50% FC_ref . And the signal was ~-25dBm. So it was clear that SNR is large enough. Nevertheless, when we check the demodulated CCFC on the oscilloscope, the signal seems not to have an SNR of 40dB. So we thought the noise couples through the demodulation process, which could be solved by amplifying the CCFC signal.

The TAMA RF PD was used to acquire CCFC. And it was investigated when we used it to acquire CC1 from OPO reflection. That investigation proved that amplification was not effective to improve SNR. However, recently Matteo found that the resistor was chosen inappropriately during that time. In the first attached figure, the resistors used to amplify RF signals are marked. As shown in the attached figure 2, these resistors need to be 10Ohm and 100Ohm to have the best performance. However, it was chosen to be 1kOhm and 10kOhm when it was used for CC1. So we put the recommended 10Ohm and 100Ohm in TAMA RF PD.

Then the modified TAMA PD was placed back to the reflection of FC with ~50% FC_ref going inside. We checked again amplifier noise, TAMA PD noise ,and CCFC sideband. The test result is shown in the attached figure 3. There is also a zoom-in of the CCFC sideband shown in the attached figure 4. We can see that the signal is amplified by ~10dB while noise is amplified by ~20dB. A factor of 10 of voltage amplification should corresponds to a factor 100 power amplification. So the 10dB signal amplification seems to be limited somehow.

But anyway, we demodulated this signal and checked it on oscilloscope. The signal didn't become much cleaner. 

Images attached to this report
2275_20201113074557_17.png 2275_20201113074602_58.png 2275_20201113074623_figure1.png 2275_20201113074628_figure2.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 13:35, Thursday 12 November 2020 (2274)Get code to link to this report
Comment to Projection of PR/BS angular motion to diagonalized AA signals (Click here to view original report: 2272)

The contribution of each DOF of PRBS angular motion is shown in the attached figures.

Images attached to this comment
2274_20201112053521_ff1.png 2274_20201112053526_ff2.png 2274_20201112053531_ff3.png 2274_20201112053537_ff4.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 19:22, Wednesday 11 November 2020 (2273)Get code to link to this report
Comment to Optimization of AA (check balance, demodulation phase, rotation angle and diagonalization matrix again) (Click here to view original report: 2271)

Today, I checked again the balance, demodulation phase, and rotation angle. The results are shown in the attached figure.

The balance between each channel is not optimized now. The demodulation phase is also not optimized. The pitch/yaw coupling becomes a bit better.

This result means the balance problem cannot be solved by a stable alignment.

Images attached to this comment
2273_20201111113626_wfs120201111.png 2273_20201111113640_wfs2balance20201111.png 2273_20201111113719_wfs1phase20201111.png 2273_20201111113730_wfs2phase20201111.png 2273_20201111113750_wfs1rotation20201111.png 2273_20201111113759_wfs2rotation20201111.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 17:30, Wednesday 11 November 2020 (2272)Get code to link to this report
Projection of PR/BS angular motion to diagonalized AA signals

To understand the noise contribution from PR/BS to the AA signal, we did several measurements and calculations. They are reported in elog2266 and elog2270. However, the sum of contribution was not done in the way of quadrature sum. So the results shown in elog2266 and elog2270 are not correct. Besides, elog 2270 used TFs which were measured without excitation. This should be replaced with TFs which are measured with excitation. This is due to that excitation can make the measurement of TF have more coherence. So we follow the same way of calculation with elog2270, except for the TFs and the noise sum method.

The measurement of TFs was done by exciting PR pitch, PR yaw, BS pitch, and BS yaw one by one. We call the measured TF with the style of TF_PRP_IP, which stands for the transfer function from PR pitch to INPUT pitch.

The quadrature sum of noise projection to INPUT pitch follows the equation:  IP = PRP*TF_PRP_IP + PRY*TF_PRY_IP +BSP*TF_BSP_IP +BSY*TF_BSY_IP. Here mirrors angular motion use name-style PRP to stand for PR pitch motion.

The results of noise projection and measurement is attached. Note that the data of PR/BS oplev signals and AA signals are directly from DGS. So the value itself doesn't correspond to a meaningful unit.

From this result, PR and BS angular motion are not the limiting noise of AA signals, except for INPUT pitch.

Images attached to this report
2272_20201111093221_1.png 2272_20201111093229_2.png 2272_20201111093235_3.png 2272_20201111093240_4.png
Comments related to this report
YuhangZhao - 13:35, Thursday 12 November 2020 (2274)

The contribution of each DOF of PRBS angular motion is shown in the attached figures.

R&D (FilterCavity)
Print this report.
YuhangZhao - 00:39, Wednesday 11 November 2020 (2271)Get code to link to this report
Optimization of AA (check balance, demodulation phase, rotation angle and diagonalization matrix again)

At the beginning of AA implementation, we checked balance (between each segment of QPD), demodulation phase (this is equavilent demodulation phase set in DGS, which can rotate the input I/Q signals by a matrix. In our case, we rotate signals to I phase.) and rotation angle (this is the quavilent rotation of QPD). All of these were done without the lock of AA loop. And we found some change from day by day.

Since now we have AA locked. I checked them again and optimized accordingly. The old work gave already quite good gain/phase/angle values. For example, in the first two attached figures, we can see the balance situation of QPD1 and 2. The balance is already quite good. Since we could get exact values from figrue 1 and 2. I calculated the required gain to balance them. After applying these gain, the balance situation is shown in the attached figure 3 and 4. We can see that the balance becomes much better. But if this will change again from day by day should be examined later.

(Old gain is 1 for each segments.) New gain is shown in the following table

 

I1

Q1

I2

Q2

I3

Q3

I4

Q4

QPD1

1.000

1.000

0.802

1.661

0.673

1.059

1.204

1.557

QPD2

1.000

1.094

0.850

1.040

1.055

0.900

0.789

1.130

After that, I also optimized the demodulation phase. The result is shown in the attached figure 5 and 6.We could see that only negligible signals go to Q phase. This should also be checked again later.

New phase is shown in the following table

 

Segment1

Segment2

Segment3

Segment4

QPD1

97

100

97

99

QPD2

135

125

138

128

Then, the rotation angle was also optimized. The optimization result is shown in the attached figure 7 and 8. However, we could see that pitch to yaw coupling has become already around 15% in WFS1/2. This should also be checked later if they will change or not.

Old rotation angles are all zero. New angles are shown in the following table

QPD1

QPD2

13

5

In the end, a new matrix was developed to close AA loop. The loop filters/gains are the same with the old case. A comparison of diagonalized signals are shown in the attached figure 9. It can be seen that they don't have large difference.

pitch

WFS1

WFS2

Input

1

-0.5

End

1

2

yaw

WFS1

WFS2

Input

-1

0.8

End

-1

-3

Images attached to this report
2271_20201110164025_qpd1balance.png 2271_20201110164030_qpd2balance.png 2271_20201110164036_qpd1afterbalance.png 2271_20201110164044_qpd2afterbalance.png 2271_20201110164053_qpd1afterphasechange.png 2271_20201110164101_qpd2afterphasechange.png 2271_20201110164116_wfs1rotation.png 2271_20201110164124_wfs2rotation.png 2271_20201110164143_20201110.png
Comments related to this report
YuhangZhao - 19:22, Wednesday 11 November 2020 (2273)

Today, I checked again the balance, demodulation phase, and rotation angle. The results are shown in the attached figure.

The balance between each channel is not optimized now. The demodulation phase is also not optimized. The pitch/yaw coupling becomes a bit better.

This result means the balance problem cannot be solved by a stable alignment.

R&D (FilterCavity)
Print this report.
YuhangZhao - 23:53, Tuesday 10 November 2020 (2270)Get code to link to this report
AA diagonalized signal noise contribution from PR/BS

By measuring the transfer function from PR/BS p/y motion to Input/End diagonalized signals, we could calibrate PR/BS p/y motion to Input/End diagonalized signals. As a result, we could know how much PR and BS angular motion are contributing to the diagonalized AA signals (Input/End diagonalized signals).

Some measurement situation: The transfer function measurement was done without excitation. PR/BS local control was closed. 

Reconstruction method: Let's take reconstruction of Input_p as an example. Input_p = PR_p*TF + PR_y*TF + BS_p*TF + BS_y*TF. This means that we sum PR/BS p/y motion's contribution to AA diagonalized signal. (Here PR_p/y and BS_p/y signal is directly the oplev signal without any modification. The Input_p signal is also directly from DGS system)

The results are shown in attached figures. We could see that AA diagonalized signals are almost totally limited by PR/BS angular motions.

Images attached to this report
2270_20201110155538_figure1.png 2270_20201110155543_figure2.png 2270_20201110155548_figure3.png 2270_20201110155552_figure4.png
R&D (FilterCavity)
Print this report.
NaokiAritomi - 12:49, Tuesday 10 November 2020 (2269)Get code to link to this report
SHG cavity scan

I measured SHG cavity scan to check finesse of SHG (Pic. 1,2). The fitting result doesn't match the measured data very well possibly due to the non symmetric peak shape. From the fitting, I obtained SHG finesse of 43, but this is not consistent with design value of 72. 

Images attached to this report
2269_20201110044918_shgcavityscan1.png 2269_20201110044924_shgcavityscan2.png
Comments related to this report
NaokiAritomi - 21:13, Monday 16 November 2020 (2277)

I measured SHG cavity scan again. This time, I put SHG temperature 2.8kOhm to avoid green conversion while nominal temperature is 3.1kOhm. Peak shape is still not completely symmetric possibly due to high IR injection power, but measured finesse is 70 which is reasonable value. 

R&D (Cryogenic)
Print this report.
SatoshiTanioka - 11:10, Monday 09 November 2020 (2268)Get code to link to this report
Q-mass installation

Takahashi-san, Tanioka

We installed a quadrupole mass spectrometer (Q-mass) to the cryostat as shown in the attached picture.
I will try to take data from next Monday.

Images attached to this report
2268_20201109030938_qmass.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 23:43, Friday 06 November 2020 (2267)Get code to link to this report
IR length locking noise when AA loop is closed or open (Unity gain frequency of AA is around 5Hz)

After the AA loop could be closed with high UGF, the locking noise was not checked again. So I did this check today.

The PR and BS local control was closed all the time. Then I measured locking noise when AA loop is closed or open. The result is shown in the attached figure.

The locking noise level is much higher than the one reported in elog2231, which has lower AA UGF.

If we compare the shape of the locking noise spectrum, it is very similar with the diagonalized AA_INPUT_PIT.

Images attached to this report
2267_20201106154328_figure1.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 18:37, Wednesday 04 November 2020 (2266)Get code to link to this report
Reconstruction of AA diagonalized signal with PR/BS mirror motion

The transfer function from PR/BS motion to AA diagonalized signal was measured and reported in elog2265. We were thinking to use this TF and PR/BS motion to reconstruct the AA diagonalized signal. Then we would like to compare this reconstructed signal with the measured AA diagonalized signal. We want to do this comparison because we found the AA diagonalized signal is higher than the corresponding INPUT/END oplev signal. This was reported in elog2245, which was strange for us because AA singal is usually less noisy.

Therefore, we measured the TF from PR/BS p/y to AA diagonalized signals when excitation was sent to PR/BS. As reported in elog2265, the coherence is only large enough between around 10 to 40Hz(the spectrum below 10Hz is not considered because the higher AA noise is mainly found above 10Hz). This was later figured out that, as shown in the attached figure 1, the excitation send to PR/BS make their motion higher than PSD noise level above 10Hz. In the usual case (no excitation), as shown in the attached figure 2, the spectrum above 10Hz is bascially PSD electronic noise.

Therefore, to calibrate PR/BS motion to AA diagonalized signal, I took one point above the PSD noise level while avoiding peaks or large deviation. After that, I assume the spectrum has 1/f2 slope because the measured AA diagonalized signal has also 1/f2 slope. And I got the attached figure 3 as the spectrum of PR/BS p/y motion above 10Hz.

To combine TF and PR/BS motion, I checked the coherence of TF. Since we have p/y coupling in AA, we found the PR/BS motion has the following coupling contribution for AA diangonalized signal.

 

PRP

PRY

BSP

BSY

AA_EP

1

 

1

 

AA_EY

 

1

 

1

AA_IP

1

 

1

 

AA_IY

1

1

1

1

Here '1' means PR/BS motion will contribute to AA. For example, AA_EY (reconstructed) = PRY*TF(PRY to AA_EY) + BSY*TF(BSY to AA_EY)

In this way, I got the reconstructed AA signal as shown in the attached last four figures. (measured AA signal is also shown for comparison). Note that the spectrum above 30Hz is not compared because there was no coherence.

The reconstruction fit well for END mirror. But the reconstruction is higher than the measurement for INPUT mirror.

Images attached to this report
2266_20201104103739_05.png 2266_20201104103758_49.png 2266_20201104103946_prbs.png 2266_20201104104249_aaendpitch.png 2266_20201104104253_aaendyaw.png 2266_20201104104258_aainputpitch.png 2266_20201104104304_aainputyaw.png