LOG-IN
Displaying reports 2721-2740 of 3273.Go to page Start 133 134 135 136 137 138 139 140 141 End
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 09:49, Tuesday 28 November 2017 (599)Get code to link to this report
Pump laser Power vs Current

I put a power meter right in front of the output fiber of the pump laser and characterized the laser power as a function of the current.

Non-image files attached to this report
R&D (FilterCavity)
Print this report.
YuhangZhao - 00:39, Tuesday 28 November 2017 (598)Get code to link to this report
Something in the temperature and humidity chamber
Please check the attached file.
Non-image files attached to this report
R&D (FilterCavity)
Print this report.
ZhaoYu-hang, TomuraAkihiro, - 19:02, Friday 24 November 2017 (597)Get code to link to this report
Beam Radius Measurement
This week we measured the beam radius after EOM. The measurement started from 100mm after EOM. We took 6 points and used Gaussian beam formula to fit. The result is shown as below.

The beam waist is 2.61667926e-05(m). We set the position of EOM as origin, the position of beam waist is 4.72853283e-02(m).

For details, please refer to the figure attached below.
Images attached to this report
597_20171124110225_beamfitting.png
Comments related to this report
AkihiroTomura - 13:32, Wednesday 29 November 2017 (601)

Vertical: w0 = 2.67e-5 m, z0 = 4.73e-2 m

Horizontal: w0 = 2.77e-5 m, z0 = 4.62e-2 m

where w0 is the spot size at a beam waist and z0 = a position of the beam waist.

R&D (FilterCavity)
Print this report.
AkihiroTomura,YuhangZhao,MatteoLeonardi - 11:46, Wednesday 22 November 2017 (596)Get code to link to this report
Filter cavity locking after FC pump shutdown

Three days after FC pump shutdown on last Saturday, we checked the filter cavity locking system. Soon the whole system recovered properly and the cavity was locked. Note that the transmission power while being locked was improved after we changed the demodulation phase. Now it is around 8.0V while it was 4.0V before.

 

Preparing for installation of coverage on the optical bench, we are changing positions of some optical components. We changed the main laser position slightly. Fortunately, it did not cause problems. One thing we want to notice is that we now have only two clamps fixing laser onto the bench because of lack of space. We do not recommend you to touch laser body.

R&D (FilterCavity)
Print this report.
MatteoLeonardi, ManuelMarchio - 10:21, Monday 20 November 2017 (595)Get code to link to this report
FC pump system recovery
After the planned electrical shutdown of 2017/11/18, the FC pump system has been switched on with success yesterday evening. Note that, due to the shutdown, all the general switches close to each pump station were switched off.
After two days without pumping the pressure in the tube close to the IM was approx. 5mbar, while close to EM it was approx 0.5mbar.
R&D (FilterCavity)
Print this report.
YuhangZhao, TomuraAkihiro, MatteoLeonardi - 22:35, Friday 17 November 2017 (594)Get code to link to this report
Measurement about locking and electronics
1. The measurement of non-filter output of Direct Digital Synthesizer(DDS)
We found on the electronic board of DDS that there is port for non-filter output. So we did some measurement. Although we can detect some signals, like -78dBm or -83dBm. It really means there is no effective output. The signal we detected is the signal leak from other channels.

2.Radio Frequency(RF) signal parameter
RF for Second Harmonic Generator(SHG): Frequency is 15.2MHz, Amplitude is 799mV(pp)
RF for Filter Cavity(FC) demodulation: Frequency is 78MHz, Phase is -0.08, Amplitude is 8.5V(pp)
RF for FC modulation: Frequency is 78MHz, Phase is 109.16, Amplitude is 1V(pp)
RF for Acoustic Optical Modulation(AOM): Frequency is 110MHz, Amplitude is 11dBm.
Amplifier for AOM: Mini Circuit ZHL-2(the gain according to data sheet is around 18dB)

3.Locking measurement
We accept the suggestion of yuefan to check the beam size. And we can see the peak value of transmitted power is around 1.2V as before. The error signal is around 4.7V peak to peak. After locking, the transmitted voltage is around 390mV.
Note here: The first and third value is from the oscilloscope around PC. The second value is from the oscilloscope on the rack. They have roughly 10 times difference for the same signal. We will check later where it is from, like oscilloscope setting or cable difference.

4.Before this week's locking, I found there is also drift. But it happened while the local control is off(see attached photo). And it is really similar with the photo I tool last time. I will check after this weekend where the beam will be next week.
Images attached to this report
594_20171117143506_1749502162.jpg
General (General)
Print this report.
ManuelMarchio, MatteoLeonardi - 18:06, Wednesday 15 November 2017 (593)Get code to link to this report
Tama clean booths map drawing

We made a drawing of the clean booth in Tama, in orange lines.

I addition to the current clean booths, we drew a clean boot 2000x3000mm on the squeezing table and a 2000x2000mm on the PR tank.

Non-image files attached to this report
R&D (FilterCavity)
Print this report.
YuhangZhao - 22:38, Monday 13 November 2017 (592)Get code to link to this report
Measurement of frequency response of DDS
As we will use Direct Digital Synthesizer(DDS) to generate signal in the future. We need to know its feature. We take advantage the Agilent Technologies E8257C(signal generator) and the Hewlett Packard 8563E(spectrum analyzer) to check DDS's output for different frequency. We use E8257C as a clock for DDS, the frequency is 500MHz and the amplitude is 0dBm.

We can see from the figure attached that DDS has something like a low pass filter inside.
Images attached to this report
592_20171114050726_frequenyresponse.png
R&D (FilterCavity)
Print this report.
YuhangZhao, MatteoLeonardi - 22:33, Friday 10 November 2017 (591)Get code to link to this report
Locking refinement
After last time locking, we did a lot of things. This caused the unlocking of Filter Cavity.

I didn't know the drift problem before. I took the picture at laser injection window of PR chamber. You can see from Fig.1 the drift after around 12 hours. I remember that it will not happen when the local control is off.

After fixing them, we still cannot fix the problem. We discussed this during the meeting and Eleonora suggested to decrease the gain. So I decreased it from 100 to 20. This is really effective. We can easily lock the cavity then.

But the problem is the locking of TEM00 is not very good. For transmitted signal, we can see from the oscilloscope that the peak is around 1.2V. But after locking, the transmitted signal can be only 0.4V in the best case. Besides, we can lock with higher-order modes and they have almost the same level with TEM00. This should come from the bad alignment. But I don't know how to make alignment better. It's really difficult to find a good way to improve this. We can see almost no change on screen the modes, on oscilloscope by changing the local control offset of mirrors.

We also measure the noise spectrum of transmitted signal and error signal with different gains. Please refer to Fig.2 and Fig.3
Images attached to this report
591_20171110143157_drift.jpg 591_20171110143231_gain20.png 591_20171110143245_gain10.png
R&D (FilterCavity)
Print this report.
AkihiroTomura - 18:51, Thursday 09 November 2017 (590)Get code to link to this report
Angular variation of the end mirror

Members: Yuhang and Tomura

 

On 09 December, we found that the end mirror of the filter cavity had relatively larger variation in its pitch motion. As refer to in elog 428, the normal angular variation for the end mirror pitch with the loop closed was 3-4 urad. However, values we found there were 6-7 urad. As for the other mirrors (PR, BS, and IM), these values seemd decent or even better. We suppose this pitch noise on the end mirror can be one reason why we cannot lock the cavity although the beam seems to be well aligned. The reason of the noise itself is still unknown.

General (General)
Print this report.
ManuelMarchio - 12:53, Thursday 09 November 2017 (588)Get code to link to this report
Procedure to turn off tama300 facility before the power blackout

Takahashi-san showed us how to switch off the facility. Basically the vacuum system and the air compressor.
For each experiment, people working in tama must take care if turning off the instrumentation correctly.

Non-image files attached to this report
R&D (FilterCavity)
Print this report.
AkihiroTomura - 20:21, Tuesday 07 November 2017 (586)Get code to link to this report
PC re-booting and other problems

Members: Akihiro T., Matteo L. and Yuhang W.

Some remarks about the filter cavity locking

On 07 December, we tried to lock the filter cavity at first, and then we found some problems.

When we switched on the loop, the yaw of IM mirror went to auto-oscillation.

Trying to figure out where the problem came from, we measured the meachanical transfer function and the open loop transfer function of each mirror (pictures to be posted later, sorry).

As a result, the MTF and OLTF measured had similar form. We consulted Eleonora for this problem and she advised us that it was possible that the loop filters were not applied on the LABVIEW program which we usually use. This was because the PC was wrongly shutted down the other day. When the PC is shutted down, a program which offers us the loop filters is also shutted down. In this case, one need to boot all programs again correctly. Finally, after applying the loop filters, the whole system (the filter cavity) seemed decent. 

In the case of re-booting the PC, please boot the LABVIEW programs used to manipulate yaw and pitch of each mirror,and also a program named "my_Filter_Bank.vi" and load a file "filtrotot.txt" in it.

AOM miss-alignment

We also found that the first order diffraction of the green beam, which is inserted to the filter cavity, had less power. It is posssible that AOM is not alingned properly. We did rough measurement of beam power at some different points (shown below).

Right after SHG cavity: 57mW

Right after EOM: 49mW

After beam spriter before AOM: 9mW

Before iris in front of the chamber: 8mW

After the iris (the 1st order diffraction): 0.25mW

R&D (FilterCavity)
Print this report.
MatteoLeonardi - 14:16, Tuesday 07 November 2017 (585)Get code to link to this report
SHG length control modifications

Members: Matteo L., Yu-hang Wu

On monday 6th of November several measurements has been performed. Here a list of the measurements and the results with some discussion.

 

HV driver TF as function of the noise amplitude:

To understand if the low pass filter present in the HV driver transfer function (logentry 582) comes from a bad slew rate of the HV internal op amp, few measurements of the TF as function of the noise amplitude has been performed. The result is presented in Fig.1. If we compare the two measurement presented with the one done previously we see that the pole of the LP filter is unchanged with respect of the noise amplitude and it is always at 77Hz. Therefore the bad slew rate of the op amp does not seem the source of the low pass filter.

 

High pass filter and HPfilter+HVdriver characterization:

To try to remove the pole in the HV transfer function we added an active, first order high pass filter at the input of the HV driver. The schematic of the HP filter is presented in Fig.2 and Fig.3 shows its characterization. After inserting the HP filter at the input of the HV driver the transfer function of the system has been measured again and it's presented in Fig.4.

Fig.5 shows a recap of the  HVdriver TF before and after adding the HPfilter.

 

SHG characterization after the modification of the HVdriver:

After adding the HPfilter the SHG's open loop and cavity TFs has been measured (Fig.6 and Fig.7). In order to have a stable loop the parameters of the servo (SR560) were changed: Gain=200 (before =1000), f3dB=10Hz (unchanged), invert=ON (before =OFF). Inverting the error signal was necessary due to the 180deg phase delay introduced at low freq by the HPfilter (the op amp is in inverting configuration).

 

All the TFs presented are pieced together from traces with different frequencies span. Only in the SHG cavity TF (Fig.7) there seems to be some problems going from one span to the subsequent. I don't think this issue is cause by any physical effect but further investigation will be performed.

As a result of adding the HPfilter to the loop chain the SHG cavity TF is reasonably flat below the PZT resonance and the unitary gain frequency of the loop has been increased from approx. 1kHz to approx.4.5kHz. In the next days we plan to play with the servo parameters to improve the unitary gain freq. and the loop stability.

 

Huge dT guess:

After the previous work, the time strip-chart of the produced green and the SHG IR transmission, as seen by two phodotiode, has been recorded (Fig.8). As it's clear the low freq fluctuation is still present. To better understand what is the dominant factor in this low freq noise the thermal control of the LiNb crystal has been swithed off and the two signals has been recorded again (Fig.9). In this situation (Tcrystal = Troom) the SHG cavity does not produce any green but the transmitted IR si much more stable, therefore we suspect that the crystal thermal control while operating at the phase matching temperature introduces a dT noise that the cavity lenght control is not able to compensate.

We plan to investigate in the next days on the thermal control in order to measure the temperature stability and to improve the temperature stabilization.

Images attached to this report
585_20171107054641_hvdriver.png 585_20171107055244_highpassfilterschematic.png 585_20171107055259_hpfilter.png 585_20171107055458_hpfilterhvdriver.png 585_20171107055605_hpandhvrecap.png 585_20171107060024_shgoltf.png 585_20171107060031_shgcavitytf.png 585_20171107061544_shgoptttransmissiongreen.jpg 585_20171107061551_shgroomttransmissiongreen.jpg
R&D (FilterCavity)
Print this report.
MatteoLeonardi - 10:01, Monday 06 November 2017 (583)Get code to link to this report
Comment to Measurement of Transfer Function for High Voltage Driver (Click here to view original report: 582)
Additional information:

The High voltage driver (first picture) is from Matsusada Precision (PZJ-0.15P-LVS2) and the datasheet can be found at the following link: https://www.matsusada.com/pdf/pzj.pdf

The piezo used for the SHG cavity (first measurement) is from Piezomechanik GmbH (HPCh 150/15-8/3) and has a capacity of 790nF.

The piezo used for the IR mode cleaner (second measurement) is from PI (P-025.20H PICA) and has a capacity of 430nF.
Images attached to this comment
583_20171106101537_img20171106141433.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 00:04, Friday 03 November 2017 (582)Get code to link to this report
Measurement of Transfer Function for High Voltage Driver
We suspect the wired behavior of SHG's mechanical Transfer Function(TF) is caused by High Voltage Driver(HVD). So we decide to measure the TF of HVD.

We measured two HVDs. One is HVD used for SHG, the other is a new SHG. Each trace data is taken with two different frequency spans and overlapped. This is to make measurement precise.

As you can see from below, the TF of HVD truly has feature like low-pass.
Images attached to this report
582_20171102160251_tfforhvdshg.png 582_20171102160642_tfforhdvnew.png
Comments related to this report
MatteoLeonardi - 10:01, Monday 06 November 2017 (583)
Additional information:

The High voltage driver (first picture) is from Matsusada Precision (PZJ-0.15P-LVS2) and the datasheet can be found at the following link: https://www.matsusada.com/pdf/pzj.pdf

The piezo used for the SHG cavity (first measurement) is from Piezomechanik GmbH (HPCh 150/15-8/3) and has a capacity of 790nF.

The piezo used for the IR mode cleaner (second measurement) is from PI (P-025.20H PICA) and has a capacity of 430nF.
R&D (FilterCavity)
Print this report.
AkihiroTomura - 20:14, Thursday 02 November 2017 (581)Get code to link to this report
Searching an optimal setting of servo in SHG control

From the experiment we had done, we had got an open loop transfer function of the SHG control loop. To optimize this control loop, we changed the parameters of the servo (corner frequency and gain) monitoring the open loop transfer function, its unity gain frequency, and phase margin, using the network analyzer (Agilent 36540A).

I show several pairs of parameters below. On the 7th column we have 224Hz unity gain freq. and 30 degree phase margin, it seems somewhat better than the others.

With a gain greater than 2000, system became unstable.

Still, there need to be more investigation.

Corner freq.[Hz] Gain[dB] Unity gain freq.[Hz] Phase margin[degree]
10 1000 1000 1.5
10 200 512 11
10 100 352 14
10 50 224 23
3 1000 672 0.6
1 1000 384 15
0.3 1000 224 30
0.1 1000 96 42
0.03 1000 32 59
R&D (FilterCavity)
Print this report.
MatteoLeonardi - 17:13, Wednesday 01 November 2017 (580)Get code to link to this report
Comment to SHG transfer function (Click here to view original report: 579)
I'm not convinced on what we see in the mechanical TF of the SHG: at low frequency the behaviour of that TF should be flat, while in our case is not. It seems that there is a low pass included somewhere around 50Hz. If so, this lowpass might be the cause of the low phase margin at the unitary gain frequency (1kHz).
R&D (FilterCavity)
Print this report.
AkihiroTomura - 16:11, Wednesday 01 November 2017 (579)Get code to link to this report
SHG transfer function
On 10/31, Matteo L, Yuhang, and Tomura did measurements of SHG transfer function.
We used Agilent 35670A network analyzer.
All traces are pieced together from traces with different frequencies span.

Firstly, we investigated frequency response of an adder we used (1st figure).
We got a flat response in magnitude and phase.

Secondly, we took dark noise spectra of the adder and Agilent 35670A (2nd figure).

Thirdly, we measured a mechanical transfer function of SHG (3rd figure).
We guessed a peak around 25kHz indicated a resonance point of PZT.

Lastly, we measured an open loop transfer function of SHG (4th figure).
Unity gain frequency is around 1kHz.
Images attached to this report
579_20171101073907_addertf.png 579_20171101074103_darknoise.png 579_20171101074114_shgmechanicaltf.png 579_20171101074125_shgopenlooptf.png
Comments related to this report
MatteoLeonardi - 17:13, Wednesday 01 November 2017 (580)
I'm not convinced on what we see in the mechanical TF of the SHG: at low frequency the behaviour of that TF should be flat, while in our case is not. It seems that there is a low pass included somewhere around 50Hz. If so, this lowpass might be the cause of the low phase margin at the unitary gain frequency (1kHz).
R&D (FilterCavity)
Print this report.
MatteoLeonardi - 14:35, Wednesday 01 November 2017 (578)Get code to link to this report
telescope local control
This entry reports some features of the status of the telescope local control (PR and BS) as it was October, 30th 2017.
In the first picture the error signals and correction signals for PR.
In the second picture the error signals and correction signals for BS.
The resitual motion with the loop closed are:
  [urad]
PR yaw 0.78
PR pitch 1.40
BS yaw 2.21
BS pitch 4.31

The reference for the residual motion can be found in the previus entry n.428

Images attached to this report
578_20171101063442_prlocalcontrol.jpg 578_20171101063448_bslocalcontrol.jpg
R&D (FilterCavity)
Print this report.
MatteoLeonardi - 12:00, Tuesday 31 October 2017 (577)Get code to link to this report
ClearPulse photodetector mofication
I modified the resistor (R7) that sets the gain of the DC channel of the ClearPulse photodetector in transmission of the SHG.
The original resistor was 51 - 1W and the resistor now is 5k1 - 1/4W. The gain of that channel increased by a factor of 100 as expected.
Images attached to this report
577_20171031040002_aftersubstitution.jpg