LOG-IN
Displaying reports 3161-3180 of 3273.Go to page Start 155 156 157 158 159 160 161 162 163 End
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 02:40, Tuesday 07 July 2015 (123)Get code to link to this report
LMA sample 15033

Last friday (3rd July) I cleaned both surfaces of LMA coated sample #15033 (abs=12.8ppm) as shown in the first picture using the first contact polymeric optics cleaner.

I measured the absorption of the sample along Z axis. I made scans of 10mm in order to see both surfaces. The sample thickness is 6mm.

I made a scan (blue line in the second figure) and then I flipped the sample in order to see also the other surface (black line) and I notices a strange peak, then I moved the sample 0.1mm along Y axis and I repeated the scan (red line). The black line shows an absorption that has a not usual pattern. I guess it is because of some scrach on the surface, because just 0.1mm apart it doesn't show.

This monday (6th July) I made again the measurement of sample and of flipped sample, (third picture). It shows a variation among the two surfaces of a factor of 2. Maybe it is just a fluctuation of the absorption value because flipping the sample we never go back to the same X,Y position.

Images attached to this report
123_20150706192918_rimg0002.jpg 123_20150706192937_plotlma15033friday.jpg 123_20150706194202_plotlma15033monday.jpg
Comments related to this report
ManuelMarchio - 12:36, Saturday 18 July 2015 (135)

I wanted to understand the big difference between the absorption peak in the configuration where the pump is entering in the sample by the absorbing surface and the other configuration where the pump is entering the sample by the not absorbing surface (flipped sample). So I made again a calibration and also a calibration with the flipped surface reference sample and I found a calibration factor R=17 1/W (flipped) and R=13 1/W (not flipped). For the not flipped sample I made a calibration history in this post and I found a mean calibration factor R=11 1/W with fluctuatoins of 10%. I don't understand why R changes so much. 

If we calculate the LMA15033 absorption with those R factors we get 22ppm for the flipped surface and 14ppm for the not flipped surface.

R&D (Cryogenic)
Print this report.
SimonZeidler - 14:16, Friday 03 July 2015 (122)Get code to link to this report
Test of Vacuum System

We have finished the third cycle of pumping and observing the pressure after closing the gate valve on the cryostat in the ATC.

The results, compared with the other two cycles and the initial measurement, can be seen in the attached pictures.
Obviously, for the pumping, we can not get the vacuum that has been reached in the initial measurements but we could confirm the data from the second cycle with a final pressure of approx. 6*10^(-4) Pa after 60000s of pumping. The initial cycle reached approx. 4*10^(-4) Pa.
Again, the vacuum is by far better than in the first cycle after the cryostats repair.

The rising of the pressure after closing the gate valve follows quite exactly the development of the initial cycle. No important differences could be observed.

We also checked the connections for the temperature controller and are ready to braze the cabels on the respective connectors (will take only 30min, or so).
From our understanding, we are ready for doing the test of the cryogenic system and will start it on Monday, probably.

Non-image files attached to this report
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 11:24, Thursday 02 July 2015 (121)Get code to link to this report
Measurement of a low absorption sample of LMA

I put the surface calibration sample in the big sample holder and I found the position of the sample surface along the Z axis is about 8mm different with respect to the little reference sample holder. This just means the scan should start from Z0=12mm instead of Z0=20mm.

I put the sample of LMA  in the holder and I attached the holder to the translation stage, I acquired 1hour of noise and it turns out to be 18microV std with no spiky noise. This is means the presence of the sample holder doesn't increase the noise.

I acquired the scan of the coated sample they gave me at LMA (sample 15033, they measured abs=12.8ppm @1064). Pump power = 171mW (LD current = 1.2A)

The first image shows the scan along Z axis of the sample. We can see two peaks and this is not the coating absorption patter we see in the calibration. The reason is that we are looking at the surface which doesn't have a HR coating

The second image shows the scan along Z axis of the flipped sample, in order to see the right HR coated surface. It has a pattern very similar to the calibration scan, a high peak at the surface and the two smaller peaks of diffraction effects.

The absorption calculated from this measurements and the calibration factor with the reference sample we have (surf_abs = 22%) gives 107ppm, almost a factor of 8 higher than the nominal value (12.8ppm). We are understanding the meaning of this. We will check the calibration factor, how the software filtering works and how clean is the surface of the sample.

The formula is Abs=AC/(DC*Power*R) where R is the calibration factor, it has a value of 10+/-1 and it is calculated with the surface reference sample calibration (abs=22%).

Images attached to this report
121_20150701164045_lma15033p171mwfilter.jpg 121_20150701164110_lma15033p171mwfilterswap.jpg 121_20150702101330_rimg0003768x1024.jpg
KAGRA VIS (OSEM)
Print this report.
TomotadaAkutsu - 15:01, Tuesday 30 June 2015 (120)Get code to link to this report
OSEM assembly procedure

I wrote an OSEM assembly procedure; please have a look!

http://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=3762

KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 17:14, Monday 29 June 2015 (119)Get code to link to this report
Comment to Strange AC noise (Click here to view original report: 117)

Today (Monday 29th) I acquired an hour of noise in the exactly same condition and I found 18microV of noise and no spiky noise at all. The only thing that changed since last time is that the system has been at rest for the weekend, about 48 hours.

KAGRA VIS (OSEM)
Print this report.
TomotadaAkutsu - 23:13, Friday 26 June 2015 (118)Get code to link to this report
Comment to OSEM LED screening preparation (Click here to view original report: 114)

[Abstract]

For measuring the LED intensity noise, the large DC voltage spoils the dynamic range of the FFT analyzer. The AC coupling could be usable to overcome the situation, so the difference between the AC and DC coupling inputs of 35670A's are measured. We should include the fit function for the calibration of measured data with the AC couping input.

[Measurements]

Measuring an input signal (or noise) with Ch1 and 2 of 35670A simultaneously and make it calculate the ratio Ch2/Ch1. A signal input is divided by a T-shape stuff and input into each channel.

  • (a) Freq. response of Ch2/CH1 are measured with a random noise input created by the 35670A itself.
  • (b) Swept sine mode to measure the same thing.

For the both cases, Ch2 is set to AC float and Ch1 to DC float. (I've tried other combination for just a confirmation.)

[Results]

The results of (a) and (b) are consistent, and  the fit transfer function is approximated by

f(x) = (i*x/x0)/(1+i*x/x0)

and

 x0 -> 0.485(5),

where I just assume f(x) -> 1 as x -> infinity.

You can use this function for the calbration when you use this 35670A with the AC coupling input (CH2).

Note: I'd like to avoid SR560s for the OSEM screening setup, as it appears to have large noise comparing to what I want to measure in the low frequency region, and each SR560 looks to have different noise level one by one... maybe our SR560s are getting old since they had been bought in 1998; one of them even sometimes has a jump in the DC offset during the measurement.

Images attached to this comment
118_20150626164136_bode20150626.png
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 20:14, Friday 26 June 2015 (117)Get code to link to this report
Strange AC noise

In order to check if the spiky noise decreases after placing the clean booth,  long time acquisition of AC noise was performed (1 hour).

I couldn't find spikes, but we see a strange beaviour: the noise increase with time to a very high level (1mV std). Then another acquisition of 5 minutes was performed and it can be seen the noise suddenly decrease alone at a certain point. I'm trying to find a reason to this strange behaviour. The booth fan are ON, I think they never stop alone. Maybe some fan of the air conditioning or other noisy devices, I don't know.

------------------------------------------------------------------------------------------------------------------------------------

I used a vacuum cleaner to clean the floor inside the clean booth and around it. The floor is not smooth, it's not easy to remove all the dust. 

Images attached to this report
117_20150626125842_probeon1hour.jpg 117_20150626125850_probeon5min.jpg
Comments related to this report
ManuelMarchio - 17:14, Monday 29 June 2015 (119)

Today (Monday 29th) I acquired an hour of noise in the exactly same condition and I found 18microV of noise and no spiky noise at all. The only thing that changed since last time is that the system has been at rest for the weekend, about 48 hours.

KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 18:03, Thursday 25 June 2015 (116)Get code to link to this report
Clean booth over the optical table

Members: Manuel, D.Tatsumi,

we clean up the optical table, remove unnecessary things, clean all cables, stick labels on cables, clean the floor.

We moved the clean booth from the upper floor to the optical table downstairs. And we moved the wheel desk from downstair to upstair.

Images attached to this report
116_20150625110041_img0130.jpg 116_20150625110050_img0133.jpg
KAGRA VIS (OSEM)
Print this report.
TomotadaAkutsu - 00:53, Thursday 25 June 2015 (114)Get code to link to this report
OSEM LED screening preparation

LEDs of OSEMs should be screened by measuring their intensity noise beforehand, so a setup for this business is prepared.

It is a quite simple setup holding a LED and PD closely with Thorlabs jigs (see pic).

For the driver for the LED and PD, today I just use the Okutomi-kun's circuit (made based on the actual OSEM satellite box design, but a little bit modified).

  • PD: Hamamatsu S1223-01
  • LED: OP232 (I'd like to change this to TSTS7100...)

When the LED is turned on, the DC ouput of the PD driver is 1.5 V; the transimpedance is 38.3k Ohm, so the photocurrent would be about 40uA. According to the specification sheet, the PD would have about 0.55 A/W at the wavelenth (OP232's light is 890nm), so the corresponding light flux collected by the PD would be around 72uW. Probably the current supply to the LED became too narrow in the Okutomi-kun's circuit (which was Akutsu's direction, sorry). The power supply of the driver is +/-14V, so even if the LED become brighter, the PD can deal with it.

Anyway, firstly I just setup the FFT analyzer (Agilent 35670A) so that it has DC coupling (the "range" was around 1.5V), but the dynamic range of the measurement system gets ruined due to the large DC voltage described above (see the graph attached).

Then today I simply set AC coupling and then the "range" can be reduced to around 12mV, and the measurements appear to be meaningful so far. As I'm not sure well what will happen in the lower frequency range when you select the AC coupling mode, so it would be better to use SR560 to cut DC signal because we can measure the transfer function and can be calbirated afterwards.

Images attached to this report
114_20150624173035_dcim1287.jpg 114_20150624173043_dcim1288.jpg 114_20150624175258_plot201506242.png 114_20150624175304_plot20150624ac.png
Comments related to this report
TomotadaAkutsu - 23:13, Friday 26 June 2015 (118)

[Abstract]

For measuring the LED intensity noise, the large DC voltage spoils the dynamic range of the FFT analyzer. The AC coupling could be usable to overcome the situation, so the difference between the AC and DC coupling inputs of 35670A's are measured. We should include the fit function for the calibration of measured data with the AC couping input.

[Measurements]

Measuring an input signal (or noise) with Ch1 and 2 of 35670A simultaneously and make it calculate the ratio Ch2/Ch1. A signal input is divided by a T-shape stuff and input into each channel.

  • (a) Freq. response of Ch2/CH1 are measured with a random noise input created by the 35670A itself.
  • (b) Swept sine mode to measure the same thing.

For the both cases, Ch2 is set to AC float and Ch1 to DC float. (I've tried other combination for just a confirmation.)

[Results]

The results of (a) and (b) are consistent, and  the fit transfer function is approximated by

f(x) = (i*x/x0)/(1+i*x/x0)

and

 x0 -> 0.485(5),

where I just assume f(x) -> 1 as x -> infinity.

You can use this function for the calbration when you use this 35670A with the AC coupling input (CH2).

Note: I'd like to avoid SR560s for the OSEM screening setup, as it appears to have large noise comparing to what I want to measure in the low frequency region, and each SR560 looks to have different noise level one by one... maybe our SR560s are getting old since they had been bought in 1998; one of them even sometimes has a jump in the DC offset during the measurement.

KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 19:03, Wednesday 24 June 2015 (113)Get code to link to this report
AC noise with Function Generator

I measured again the noise of the photodetector. The chopper is off and the lockin trigger comes from a function generator set at 430Hz. In order to check if the noise could come from the function generator, I measured the noise with the probe laser ON and with the probe laser OFF.

The AC noise std with the Probe ON is 18microV as usual.

The AC noise std with the Probe OFF is 0.7microV. Very similar to the case Chopper ON - Probe OFF measured before (0.8microV).

This means that the noise doesn't come from the function generator.

Other possible noise sources could be the HeNe power fluctuation or jitter position fluctuation due to some vibrations.

----------------------------------

During many noise acquisitions I noticed that sometimes (one time in 3 or 4 minutes) there is a strong peak of noise, about 2  order of magnitude higher than the usual noise. The system manual speaks about them and give this explanation: "There are various sources of spiky noise including dust particles crossing the probe beam path." 

General (Test)
Print this report.
TomotadaAkutsu - 10:35, Wednesday 24 June 2015 (112)Get code to link to this report
AEL: Satellite box preliminary test

A satellite box prototype was delivered to NAOJ last week, and was insitu revised/modified, having discussions with AEL group.

So the prototype circuit in NAOJ is not the final version; some capacitaces are removed due to their wrong assembly directions, D101 diodes are removed due to similar reason, and a Q101 transistor of Ch1 is removed during the investigation of the circuits.

Anyway, a basic measurement, the noise level of the output of the circuit is preliminarily measured. A PD (S1223-01) in a old design holder is connected to the circuit with a barrack conversion board, and the output's differential outputs are directly connected to a FFT analyzer (Agilent 35670A) with a front end setup of DC float.

The measurement are done during the room lights are turned off as much as possible, and the PD is covered by an easy carton box. So far, for some reason (probably large 50Hz signal), the FFT's range should be set no less than 158 mV, and the most of the measured data are buried under the measurement system noise.

Anyway, the vertical axis is converted to input equivalent current noise density; divided by 2 due to its differential output, and divided by 38.3k Ohm of the transimpedance; see http://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=3499

Note that this is not a test for OSEM but just a basic measurement of the circuit itself. To say something about OSEM, the output should be calibrated to the unit of "m/rtHz".

I'm wondering why the 50Hz apperas such large in the circuit; comparing to the measurement result of an oplev driver so far, it appears a little bit too large...

Images attached to this report
112_20150624033500_figure1.png
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 22:50, Tuesday 23 June 2015 (111)Get code to link to this report
Calibration history

Members: Manuel, D.Tatsumi.

We made the calibration of  bulk and surface reference samples, in three different days.

For the bulk calibration, we found a difference of about 2% among the days.

For the surface calibration, we found a difference of about 10% among the days.

The plots show the AC/DC ratio during the scan along the z axis.

Images attached to this report
111_20150623154754_calibhistorybulk.jpg 111_20150623154802_calibhistorysurf.jpg
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 00:25, Tuesday 23 June 2015 (109)Get code to link to this report
AC noise
Members: Manuel, D.Tatsumi
 
We measured the reference signal from the chopper: it is a square waveform of 5V p-p at a frequency of 430Hz
We set a waveform generator at 430Hz but the knob was too sensitive to touch and the best we could do was 426Hz (the lockin time constant is 100ms, so we can be inside ±10Hz).
We aquired 3 minutes of AC signal. Noise standard deviation is 19microV, wich is comparable with the noise measured with the real chopper (18microV).
This means that the chopper vibration do not dominate the noise.
Images attached to this report
109_20150623042052_chopnochop.jpg
KAGRA VIS (General)
Print this report.
RyutaroTakahashi - 10:09, Thursday 18 June 2015 (108)Get code to link to this report
Adjustment of breadboard in MCE
Member: R. Takahashi, N. Ohishi, K. Doi

June 11
1) Clean the floor of clean booth.
2) Open the chamber.
3) Adjust the height of breadboard. The height was set to 490mm +/- 1mm from the base plate.
4) Measure the position of the breadboard by the straight edge and the pendulum. The position was deviated by -1.0mm, -1.75mm from the center of the chamber.

June 12
1) Measure the rotation of the breadboard by the straight edge and the pendulum. The position of support bolt was on the Y line within +/- 0.5mm.
2) Pick up some parts from the bottom of chamber. We found M10 long bolt, Al foil, I-bolt, and CCP base.
3) Close the chamber.
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 11:57, Wednesday 17 June 2015 (107)Get code to link to this report
AC noise -Lockin Time constant; Photodetector noise; Pump power needed

I acquired 3 minutes of the AC noise in the following conditions;

 - pump off

 - probe on

 - front door closed

  - various lock in integration time constants;

As shown in the plot , the higher the time constant, the lower the noise.

But we have to choose it depending on the total acquisition time, for example we cannot spend a week to acquire a surface map.

I report the values:

timeconstant (ms)

Noise std (microV)
10 47
30 14
100 18
300 9.5
1000 5.2

We remind the specification of the system says between 5 and 25 microV at time constant: 100ms; and chopper frequency: 380Hz.

-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Then we switch off the HeNe probe and measure the noise of the photodetector with 100ms time constant.

The photodetector noise standard deviation is 0.8 microV (see the histogram below) and it includes the detector noise and the lockin noise

-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------

For absorption of 22% we have about 350mV AC signal at 30mW pump power;

The signal should be at least 2 or 3 times the noise (18microV * 3);

For 12ppm we expect 19microV of AC signal at the same pump power (30mW)

This means that a pump power of 100mW might be sufficient to see a good signal for 12ppm. It follows that we need 10W to see absorptions of 0.1ppm

Images attached to this report
107_20150617035217_noisestdvstimeconst.jpg 107_20150617041842_probeoffhisto.jpg
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 10:36, Wednesday 17 June 2015 (106)Get code to link to this report
AC noise - Chopper frequency

I tuned the lockin gain to have a better resolution for little signals.

I acquired 3 minutes of noise for different modulation frequencies, in the following conditions:

 - Probe ON;

 - Pump OFF;

 - Front door closed;

 - Lockin time constant 100ms;

 

I made an histogram for every acquisition and I computed the standard deviation in order to find the best chopper frequency.

The lowet value of the noise standard deviation is at 430 Hz modulation frequency. The second best value is for 480 Hz, as used previously. The highest noises are at multiples of 50 Hz, as expected.

Images attached to this report
106_20150617033133_histograms.jpg 106_20150617033153_stdnoisechopper.jpg
KAGRA AOS (General)
Print this report.
TomotadaAkutsu - 18:51, Tuesday 16 June 2015 (102)Get code to link to this report
Oplev's pillar test installation to the tunnel

Abstract: we did a test installation of an oplev's pillar to the tunnel.

Workers: Uchiyama and Akutsu

The test installation has been done around (1) PR3 chamber, (2) IFI chamber (for IMMT2 mirror), and (3) MCF chamber (for MCo mirror).

For all cases, the bolt holes on the floor are all dirty, and the M24x130 bolts are hard to fit to them without cleaning the bolt holes; even after the cleaning, it was so hard,

so ratchet socket wrenches are used to screw. Actually we didn't have M24 washers today, so the legs have not been fully fixed.

 (1) for PR3,

It is known that the pillar and the chamber's hinge will mechanically interfer, and the inconvenient fact was confirmed. We just slide the pillar a little from the planned location, and found a so-so place, but not know wether the optical table on the pillar can be installed or not. When the pillar stands, the top surface is almost horizontal without shims (shown by a whater leveler), and this means the floor surface here and the pillar's upper and lower surfaces are well manufactured.

(2) for IMMT2 mirror

Not much mechanical intereferences, but one concern is space between a barrelhead attached to a gate valve between MCF and IFI chambers. Still the bolts were hard. When the pillar stands, the top surface is almost horizontal without shims (shown by a whater leveler), and this means the floor surface here and the pillar's upper and lower surfaces are well manufactured. The height from the top surface to the beam line is measured to be 112.5 mm (the beam line height is just estimated from the chamber structure). Then the oplev beam height from the optical table on the pillar can be estimated 112.5 - 27 = 85.5 mm. The expected beam height was 85.85~85.9 mm, so it could be in the tolerant range.

(3) for MCo mirror

Not much mechanical intereferece, but one concern is space between a reducing flange and the pillar. Still the bolts were hard. Unfortunately, the floor surface for the pillar is not so good, and we need two pieces of shims to make the top surface horizontal. The height from the top surface to the beam line is measured to be 115 mm (the beam line height is just estimated from the chamber structure), which should be the same as the one arund IFI chamber, but now it is obvious to have 2.5 mm difference. This difference would be a little bit large if you take into consderaton about measurement errors. Then the oplev beam height from the optical table on the pillar can be estimated 115 - 27 = 88 mm. The expected beam height was 85.85~85.9 mm, so it could be a little bit larger than my expectation, but can be tuned.

Images attached to this report
102_20150616114556_dcim1209.jpg 102_20150616114613_dcim1211.jpg 102_20150616114703_dcim1215.jpg 102_20150616114722_dcim1217.jpg 102_20150616114807_dcim1222.jpg 102_20150616114825_dcim1224.jpg 102_20150616114953_dcim1226.jpg 102_20150616115049_dcim1227.jpg 102_20150616115107_dcim1237.jpg 102_20150616115116_dcim1235.jpg
KAGRA AOS (General)
Print this report.
SimonZeidler - 17:33, Tuesday 16 June 2015 (104)Get code to link to this report
PSD calculations

I got the two-dimensional map of the BS from Hirose-san.
I will try to implement these data into the "LightTools" software in the next days.

At the same time, I tried to fit the one-dimensional PSD with the K-correlation model with only limited success.
It seems that this particular model is not suitable for the data...or the data aren't enough.
From the map it should also be possible to calculate the two-dimensional PSD and to fit these data instead of the one-dimensional one.
I would like to try these calculations also during the next days to see how far I can use the PSD for the scattering Paper I am now writing.

KAGRA AOS (General)
Print this report.
SimonZeidler - 17:27, Tuesday 16 June 2015 (103)Get code to link to this report
Updated JASMINE scatterometer

The laser for the JASMINE scatterometer is now focussed on the sample.
I used a lens with f=200mm; the size of the focal point is now 0.25 - 0.3 mm.

Measurements can again be taken.

I will do some test measurements tomorrow on a Titanium sample.

KAGRA VIS (OSEM)
Print this report.
TomotadaAkutsu - 20:16, Monday 15 June 2015 (101)Get code to link to this report
Flexi circuits

Finally Tatsumi-san gets the (usable) flexi circuit prototypes for OSEMs, and he gave them to me. I firstly check whether the soldering can be done, and the results so far are good.

Images attached to this report
101_20150615131621_dcim1179.jpg 101_20150615131637_dcim1182.jpg