LOG-IN
Displaying reports 801-820 of 3273.Go to page Start 37 38 39 40 41 42 43 44 45 End
KAGRA MIR (Absorption)
Print this report.
MarcEisenmann - 19:36, Friday 11 March 2022 (2869)Get code to link to this report
Shinkosha7 absorption measurements restarted

Today I reinstalled the reference samples and checked the proper positions.

With the surface sample I got the crossing point at z = 35 mm (maximizes AC) and z_IU = 68 mm (maximizes AC/DC).

I measured R_surface = 16.91/W and R_bulk = 0.6212 cm/W.

I installed back SHINKOSHA 7 and did a long z scan.

In the attached figure you can see the comparison between various signals for the previous long z scan (red) and current situation (black) taken at the mirror X and Y centers.

Note that the z axis has been shifted for the red using the surfaces signal and then interpolated to the new measurement (step size of 0.05 mm instead of 0.1 mm).

As expected, the absorption is larger now and it seems that we have something like at least a factor 1.4 increase.

However, I was expecting to see a somehow constant increase but this is not the case..

I started a XY absorption measurement at the mirror center (ie X = 399.08mm, Y = 122.175 mm and Z = 71.625 mm)

Images attached to this report
2869_20220311113436_zscancomparison.jpg
Comments related to this report
MarcEisenmann - 13:45, Monday 14 March 2022 (2870)

We finished the first 3 measurements taken at the same positions as in Caltech ie at the mirror center, 10 mm after the first surface and 10 mm before the second one.

The results are attached to this entry and compatible with their measurements (and therefore also with Manuel's ones).

So we started absorption measurements in between these positions to get more data for the integrated map along z.

SimonZeidler - 10:50, Wednesday 16 March 2022 (2874)

Great achievement, Thank you!

Once the new data are ready, we can finish up the paper smiley

MarcEisenmann - 11:10, Wednesday 16 March 2022 (2875)

Thank you!

I also finished the last 2 measurements in between the 3 previous measurements that are attached to this entry.

KAGRA MIR (General)
Print this report.
MarcEisenmann - 11:16, Friday 11 March 2022 (2868)Get code to link to this report
PCI pump beam realigned

Katsuki, Marc

This is a summary of these past days activities.

As reported in entry 2863 we found out that the pump beam was larger than expected (48.5 um instead of 36 um).

In summary we had to act on the two lenses on the pump beam path to recover the good beam size and position following Jammt simulations.

These 2 lenses are now about 1 cm closer to the laser source.

During this realignment we also checked the the probe beam size to have a reference waist position.

We found out that it is not feasible to use the absorption DC photodiode together with the razor blade because there is scattering when we start to cut the beam that creates a spikes in the data and prevent a good fit of the data.

Furthermore, the probe beam is really large on this photodiode and is really astigmatic when setting up the imaging unit translation stage at z_IU = 0 mm (ie farthest from the translation stage).

In the end, we installed a power-meter in between the imaging unit lens and sphere and could get good data.

We also found out that a good step size for the translation stage is 20 um as it allows to get good enough resolution while not taking too long.

The attached figure reports the beam profiles of probe beam in vertical and pump beam in both vertical and horizontal directions.

The z axis is the same as Manuel's measurement (see elog 1089) ie the 0 mm is at 75mm from the breadboard.

We had to tweak a bit the tilt of the lenses to minimize the pump beam astigmatism.

We recovered the expected pump beam waist size and position so we will now switch to absorption measurement.

Note that probe beam power is 2.5 mW and pump beam power was about 150 mW for this measurement.

Images attached to this report
2868_20220311031343_beamprofile20220310.jpg
KAGRA MIF (OMC)
Print this report.
YoheiNishino - 19:44, Wednesday 09 March 2022 (2867)Get code to link to this report
Beam profile remeasurement

Nishino

I measured the beam profile between L2 and L3. I got better results than the previous one (see 2856).

 

  waist size waist position*
x 0.075 mm 134.5 mm
y 0.093 mm 125.3 mm

*starting position is 25 mm away from L2.

Images attached to this report
2867_20220309113130_omcbetl2l3xy.png 2867_20220309114122_inked1bfa9b40031343c597cd89b5ef0e289cli.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 16:15, Wednesday 09 March 2022 (2866)Get code to link to this report
Comment to Test of Minicircuits M3SWA-2-50DRB+ absorptive RF switch (Click here to view original report: 2865)

Taking the first 30ns of measurement data, I did a FFT analysis of the data and got a power spectrum density (PSD). Then the time span is shifted by 1ns several times to get the PSD evolution. In total, the 200ns data is shifted by 170 times to get the signal PSD change as a function of time. This is shown in attached figure one.

The FFT has a bandwidth of 33MHz (since I used 30ns to make a FFT). Because the RF signal has a frequency of 110MHz, I took the frequency span of 99-132MHz to check the amplitude of RF switch output.

From this analysis, the fall time, which is the time that signal drops from 90% to 10%, is 10.6ns.

In addition, I also put a time-frequency-amplitude plot of this signal.

Images attached to this comment
2866_20220309081533_falltimefinal2.png 2866_20220309145605_timefamp.png
R&D (FilterCavity)
Print this report.
MichaelPage - 22:36, Tuesday 08 March 2022 (2865)Get code to link to this report
Test of Minicircuits M3SWA-2-50DRB+ absorptive RF switch

Pierre Prat (remote), Yuhang and Michael

We received the Minicircuits M3SWA-2-50DRB+ absorptive RF switch evaluation board. Nominally, it has a fall time of 4.6 ns, well within bounds of what we want (400 ns). In this case, the rise and fall times have been specified by the manufacturer as the time it takes to go from 10% to 90% of the peak voltage and vice versa. The circuit can accept high input power > 24 dBm at 100 MHz. It is powered by -5/+5 V supply. The switch is activated/deactiveated by a TTL (transistor-transistor logic) control signal. In short, voltages in a certain low threshold (0-0.8 V) are considered "OFF" and in a certain high threshold (2.1-5 V) are considered "ON". In this case, we can just use a square wave oscillating between 0 and 5 V, and then trigger the oscilloscope to follow the rise/fall of the RF signal.

Chip manual: https://www.minicircuits.com/pdfs/M3SWA-2-50DRB+.pdf
Evaluation board diagram: https://www.minicircuits.com/pcb/WTB-M3SWA250DRB+_P02.pdf
 

-- Test --

The RF switch was tested in the filter cavity clean room using the already present oscilloscope, function generator and RF amplifier(s). We brought a DC power supply to send -5V/+5V to power the RF switch, as well as a Tektronix AFG320 function generator to provide the control signal to the RF switch (0 to 5V square wave, checked at 1 Hz and 12 kHz). Both of these were tested first to make sure they give the required voltage and square wave signal.

A 500 MHz RF signal was sent from the filter cavity function generator to the switch -> RF amplifier -> oscilloscope. A 20 dB attenuator with 50 Ohm impedance was connected to the oscilloscope to prevent back reflection. Unfortunately, the first RF amplifier (Minicircuits ZHL2) we were using stopped outputting. We did take care to say the order in which you should make connections with the RF amplifier. I hope it is not permanently broken... 

 

-- Data --

The figure shows the fall time when a 1 Hz square wave is sent to the TTL port of the switch (rise time figure pending). The data is a bit low resolution. The lower half of the figure shows a zoom in of the timescale and indicates 10% of Vpk. This measurement doesn't seem very accurate, but regardless, the fall time is well below the target of 400 ns.

With these results, we moved the RF switch and the necessary electronics to the ATC cleanroom.

Images attached to this report
2865_20220308142949_falltime0308.jpg
Comments related to this report
YuhangZhao - 16:15, Wednesday 09 March 2022 (2866)

Taking the first 30ns of measurement data, I did a FFT analysis of the data and got a power spectrum density (PSD). Then the time span is shifted by 1ns several times to get the PSD evolution. In total, the 200ns data is shifted by 170 times to get the signal PSD change as a function of time. This is shown in attached figure one.

The FFT has a bandwidth of 33MHz (since I used 30ns to make a FFT). Because the RF signal has a frequency of 110MHz, I took the frequency span of 99-132MHz to check the amplitude of RF switch output.

From this analysis, the fall time, which is the time that signal drops from 90% to 10%, is 10.6ns.

In addition, I also put a time-frequency-amplitude plot of this signal.

R&D (FilterCavity)
Print this report.
NaokiAritomi - 16:49, Monday 07 March 2022 (2864)Get code to link to this report
IR locking accuracy with 1/f,1/f^4 filter, CCFC

This work is on 20220303. 

I measured CCFC error signal with green 1/f, 1/f^4 filters, and CCFC. The green transmission beam spot is upper side of camera (new beam spot). The green FC injection power was 23mW. The parameters for 1/f and 1/f^4 filters are as follows. Fig 1 shows the green OLTF.

filter input attenuator piezo gain UGF
1/f 0.8 8 11 kHz
1/f^4 0.2 8 13 kHz

The measured CCFC error signal with green 1/f, 1/f^4 filters, and CCFC are shown in Fig. 2. The CCFC amplitude was 142mVpp. CCFC filter gain is 1000 with 30Hz LPF.

Measurement of CCFC OLTF

CCFC OLTF can be measured by injecting a signal to sum port of CCFC filter (SR560) and measuring CCFC filter input/output. The relation between this measurement and CCFC OLTF is as follows.

(CCFC filter input/output)*CCFC filter = - G_CCFC/(1+G_green)

G_CCFC = - (CCFC filter input/output)*CCFC filter*(1+G_green)

Since the CCFC filter and G_green are known, G_CCFC can be obtained. Fig. 3 shows green, CCFC OLTF. The crossover frequency between green/CCFC is 1.6kHz.

Images attached to this report
2864_20220307084932_greenoltf20220303.png 2864_20220426023434_ccfclockingaccuracy20220303.png 2864_20220426023501_ccfcoltf20220303.png
KAGRA MIR (Absorption)
Print this report.
MarcEisenmann - 20:42, Friday 04 March 2022 (2863)Get code to link to this report
Comment to Shinkosha7 absorption (Click here to view original report: 2855)

We tried to investigate possible explanations for this discrepancy.

First we performed along z scan to be sure that we are able to see the 2 surfaces of the samples.

We could find S1 at 34.8 mm and S2 at 122 mm along z.

We can see the ac/dc signal decreasing with an increase of z (same as Manuel's measurement) but the signal is roughly half of what he got.

We have the same chopper frequency, we're injecting pure s polarization, but differences are that he was injecting about 10 W vs our 8.5 W, he set the DC to about 2.5 V vs 4V now and in his computation he is using 1.16 /cm instead of the 1.04 /cm later measured and I'm not sure how the transmission was taken into account.

For reference Manuel's measurements and analysis are in the KAGRA#7 folder.

One strong possibility is that we have a too large pump beam size. Indeed Manuel found out that it could cause some factor discrepancy when he upgraded the setup.

We characterized the beam size with the razor blade as reported in figure 1. The beam waist is 48.5 um instead of the expected 35 um.

Following Jammt simulation that indicates that the beam waist of 35 um by moving the last lens by ~5mm we started to realign but without clear improvement so we'll continue on Monday.

Images attached to this comment
2863_20220304124101_20220304pumpbeam.png
KAGRA MIR (Polarization)
Print this report.
HomareAbe - 17:37, Friday 04 March 2022 (2862)Get code to link to this report
Calibration of PBS
Abe

I tried to check the PBS transmission.

Without PBS, Intensity is 25.7 mW.

Minimum transmission is 114 uW.
QWP angle is 220 deg, HWP angle is 96 deg.

Maximum transmission is 24.0 mW.
QWP angle is 220 deg, HWP angle is 50 deg.
KAGRA MIF (OMC)
Print this report.
YoheiNishino - 16:41, Friday 04 March 2022 (2861)Get code to link to this report
Two issues found in the OMC components

Aso and Nishino did fit-check on Febrary 22nd.

There were 2 issues in the optical components

  1.  the sizes of the screw head for OBS1 were too large and they touched both sides of the OBS1 mirror.
  2.  the new mirror mount for OBS4 doesn't have a reference mark to set on the mirror

for the issue 1, Fukushima-san fixed it and Nishino did fit-check on March 4th.

Issue 2 is left on the date of this report.

KAGRA MIF (OMC)
Print this report.
YoheiNishino - 11:41, Friday 04 March 2022 (2860)Get code to link to this report
Comment to Beam profile measurement in the clean room in ATC (Click here to view original report: 2856)

I measured the beam profiles again on March 3rd. I took 10 samples for each point and used an average of them.

  width weist position*
x 0.1347+- 0.0006 mm -35.6 +-0.5mm
y 0.139+-0.004 mm 11.8+-2.3mm

* start position is two holes (~50 mm) distant from the center of the BS.

Images attached to this comment
2860_20220304033934_omcsheet56xy.png
R&D (FilterCavity)
Print this report.
NaokiAritomi - 21:20, Thursday 03 March 2022 (2859)Get code to link to this report
CCFC error signal with old/new green beam spot

I compared CCFC error signal with old/new green beam spot. The new green beam spot is upper side of camera as shown in elog2613. The old green beam spot is center of camera. 

The offset of BS pointing for old/new beam spot is as follows. 

  old  new
pitch 26 6
yaw 14 14

The attached figure shows CCFC error signal. The CCFC amplitude was 118mVpp. The new beam spot is better than old beam spot below 10Hz.

Images attached to this report
2859_20220303132006_ccfcbeamspot20220303.png
R&D (FilterCavity)
Print this report.
NaokiAritomi - 20:30, Thursday 03 March 2022 (2858)Get code to link to this report
FC GR lock investigation (5)

As reported in elog2850, FC is sometimes very stable, but not very stable most of the time. I noticed that the BS coil output was too large (~20000). After BS offload with picomotor, FC got more stable. Maybe BS was touching somewhere and that could cause the unlock. However, FC still sometimes unlocks. It might be better to open PR chamber and check PR suspension.

KAGRA MIF (OMC)
Print this report.
YoheiNishino - 12:02, Thursday 03 March 2022 (2857)Get code to link to this report
Optics alignment design for characterization of the new OMC

I designed an optics aligenment using g-trace (see fig 1).

Images attached to this report
2857_20220303035019_12473124631252212540125311247112519124831248820220303114947.png 2857_20220303040158_86672cca6d5c4557ac8906453e807c6a.jpeg
KAGRA MIF (OMC)
Print this report.
YoheiNishino - 11:18, Thursday 03 March 2022 (2856)Get code to link to this report
Beam profile measurement in the clean room in ATC

I measured the beam profile of the laser in the ATC clean booth. Yuhang and Michael are doing another experiment, so Aso-san and I will use the reflected light on the beam splitter(see figure 1).

Results:

  width weist position*
x 0.135 mm -39.4 mm
y 0.136 6.07 mm

* start position is two holes (~50 mm) distant from the center of the BS.

Images attached to this report
2856_20220303031611_cd9a0608bdcd4069ac1e3d1d2238322d.jpeg 2856_20220303031733_omcsheet4xy.png
Comments related to this report
YoheiNishino - 11:41, Friday 04 March 2022 (2860)

I measured the beam profiles again on March 3rd. I took 10 samples for each point and used an average of them.

  width weist position*
x 0.1347+- 0.0006 mm -35.6 +-0.5mm
y 0.139+-0.004 mm 11.8+-2.3mm

* start position is two holes (~50 mm) distant from the center of the BS.

KAGRA MIR (Absorption)
Print this report.
MarcEisenmann - 16:42, Tuesday 01 March 2022 (2855)Get code to link to this report
Shinkosha7 absorption

In order to get a more precise comparison between birefringence (integrated along z) and absorption (at a given z position) measurements, we performed absorption measurements at several z positions.

Note that in that case the shinkosha 7 orientation is still the same as the measurement done by Manuel ie arrow at the top and pointing towards the imaging unit.

All results are attached to this entry where I used same colorlimit as Manuel (0 to 200 ppm/cm) and similar colormap.

Similar patterns are visibles.

However, it seems that maximum absorption is quite lower than what was measured before...

One difference with the previous measurements is that we were using 0.5s waiting time and 70mm radius..

I'm now starting new measurements with differents lockin amplifier parameters to investigate this issue.

Note that the z values indicated here correspond directly to the translation stage values (therefore different than Manuel measurements where he corrected the z value to match the real position in the mirror)

Images attached to this report
2855_20220301084028_xz.png 2855_20220301084033_yz.png 2855_20220301084038_xyz40.png 2855_20220301084042_xyz50.png 2855_20220301084048_xyz70.png 2855_20220301084052_xyz90.png 2855_20220301084056_xyz100.png 2855_20220301084101_xyz110.png
Comments related to this report
MarcEisenmann - 20:42, Friday 04 March 2022 (2863)

We tried to investigate possible explanations for this discrepancy.

First we performed along z scan to be sure that we are able to see the 2 surfaces of the samples.

We could find S1 at 34.8 mm and S2 at 122 mm along z.

We can see the ac/dc signal decreasing with an increase of z (same as Manuel's measurement) but the signal is roughly half of what he got.

We have the same chopper frequency, we're injecting pure s polarization, but differences are that he was injecting about 10 W vs our 8.5 W, he set the DC to about 2.5 V vs 4V now and in his computation he is using 1.16 /cm instead of the 1.04 /cm later measured and I'm not sure how the transmission was taken into account.

For reference Manuel's measurements and analysis are in the KAGRA#7 folder.

One strong possibility is that we have a too large pump beam size. Indeed Manuel found out that it could cause some factor discrepancy when he upgraded the setup.

We characterized the beam size with the razor blade as reported in figure 1. The beam waist is 48.5 um instead of the expected 35 um.

Following Jammt simulation that indicates that the beam waist of 35 um by moving the last lens by ~5mm we started to realign but without clear improvement so we'll continue on Monday.

KAGRA MIR (General)
Print this report.
MarcEisenmann - 16:16, Tuesday 01 March 2022 (2854)Get code to link to this report
test setup for Soleil-Babinet compensator

Abe, Katsuki, Marc

We purchased a Soleil-babinet compensator that will be installed in PCI for future birefringence measurements.

We prepared a test setup on the optical table just in front of PCI clean room (the one where there is the reflectance measurement setup).

We will use the FC spare laser.

We installed 2 OD to have ~ 40 mW of power then 2 steering mirrors(Newport 5204)  to have proper alignment above a lign of holes.

We also brought the required componenents (1 PBS, 1 QWP and 4 HWPs).

R&D (FilterCavity)
Print this report.
NaokiAritomi - 23:10, Monday 28 February 2022 (2853)Get code to link to this report
Recovery of CCFC lock

As reported in elog2852, the CC detuning and CCFC demodulation phase should be adjusted. Since the CC detuning in elog2852 was 76Hz, first I changed CC PLL frequency by ~22Hz, but the shape of CCFC error signal was strange. Then I decided to change CC PLL frequency by 10Hz. The setting of CC PLL is as follows.

channel function frequency (MHz) binary number
CH0 CC PLL 20.99112421 1010 10111111 01011001 01000000
CH2 CC1/CCFC demod 13.99408281   111 00101010 00111011 10000000
CH3 CC2 demod   6.99704140     11 10010101 00011101 11000000

The setting of LEMO cables for demodulation is as follows.

Connection Color of LEMO cable
Between CCFC RF amplifier and mixer RF port green+yellow
Between DDS and mixer LO port green

Fig. 1 shows CCFC error signal. The CCFC calibration amplitude is 138mVpp. The mode matching is fixed to 0.9.

Today FC was quite stable and I could lock CCFC for the first time since last August! The CCFC filter gain is 1000 with 30Hz LPF. The Z correction, AA, BS pointing were engaged.

Fig. 2 shows the locking accuracy with CCFC.

Images attached to this report
2853_20220228150959_ccfcerrorsignal20220228t0081.png 2853_20220228185330_ccfclockingaccuracy20220228.png
R&D (FilterCavity)
Print this report.
NaokiAritomi - 22:20, Thursday 24 February 2022 (2852)Get code to link to this report
Recovery of CCFC error signal

First I aligned BAB to FC. The optimal p pol PLL frequency without green was 300MHz and BAB power before FC was 435uW. The maximum IR transmission of FC was 470.

Then I checked the CCFC error signal. I optimized the p pol PLL frequency to maximize the CCFC error signal. The optimal p pol PLL frequency with 20mW green was 240MHz and the CCFC error signal was 118mVpp. The setting of CC detuning is reported in elog2521. The setting of LEMO cables for demodulation is as follows.

Connection Color of LEMO cable
Between CCFC RF amplifier and mixer RF port brown
Between DDS to mixer LO port green

The attached figure shows CCFC error signal. The mode matching is fixed to 0.9. We need to adjust the demodulation phase and CC detuning.

Images attached to this report
2852_20220224142006_ccfc20220224.png
R&D (FilterCavity)
Print this report.
NaokiAritomi - 20:24, Thursday 24 February 2022 (2851)Get code to link to this report
CC PLL problem

I found that CC PLL could not be locked. I changed the phase detector polarity of ADF4002 from negative to positive. Then the fast loop of CC PLL could be locked, but slow loop could not be locked.

R&D (FilterCavity)
Print this report.
YuhangZhao - 23:18, Tuesday 22 February 2022 (2850)Get code to link to this report
FC GR lock investigation (4): Filter cavity lock can be quite easily acquired and kept for tens of minutes (AA/z_corr works as well)

Today I replaced the Qubig PD back to the one with DC output (simply called DC-Qubig later), whose change was done about one month ago (elog2801). Note that I just temporarily set up DC-Qubig, whose cables are still not deployed properly since it's hard to do by myself

After putting DC-Qubig back, to have a decent loop gain, I adjusted DDS2 channel CH1 amplitude from 1/4 to 1/2. After that, I took a measurement of open-loop transfer function. The unity gain frequency was around 13kHz. At the same time, the Rampeauto attenuation is 0, the Rampeauto gain is 8.

The setting for AA is shown in Fig.1. The setting for z_corr is shown in Fig.2. 

In the first minute of Fig.3, the filter cavity is controlled with PDH and z_corr. After using setting of Fig.1 for AA, the filter cavity transmission is stabilized. This figure shows about five minutes. But longer lock such as about 20min was observed tonight as well.

Images attached to this report
2850_20220222151836_18.png 2850_20220222151841_53.png 2850_20220222151846_35.png